Introduction

Lung cancer is the most common cause of death among oncological diseases. The treatment of choice is surgery, but it is indicated only in patients with a limited stage of the disease. Other therapeutic options are available today, in particular the three modality approach consisting of surgery, chemotherapy and radiotherapy in different scheduling schemes. In this scenario, radiotherapy has a crucial role in lung cancer treatment and the probability of local control is strictly related to the precise delineation of target volumes.

18F-Fluorodeoxyglucose (FDG) positron emission tomography (PET) has demonstrated significant value for radiotherapy of lung cancer. The incorporation of PET and PET/computed tomography (CT) data into the radiation treatment planning has recently gained widespread acceptance as it has been shown to improve gross tumour volume (GTV) definition [14].

A limitation of PET studies, performed during shallow breathing, is motion artefacts, which can reduce lesion contrast, underestimate the tracer uptake and overestimate the lesion size and thus the metabolic tumour volume (MTV) [5, 6]. Furthermore, if PET/CT scanners are used, the temporal mismatch between the short (in seconds) CT and the long (in minutes) PET scans may cause a spatial misalignment between CT and PET images with consequent errors in attenuation correction maps [710].

Respiratory motion can be managed either by motion suppression (deep inspiration breath hold or forced shallow breathing) or motion compensation [11, 12]; the last one includes a respiratory gating technique that typically bins the imaging acquisition to direct or indirect measurement of motion amplitude. In order to overcome the respiratory motion-related artefacts and the uncertainty in planning target volume (PTV) definition, respiratory-gated CT (RG-CT) has been widely evaluated, both from a methodological and a clinical point of view [1319]. Gated CT is strongly recommended today in PTV definition in lung cancer patients [20] as it can furnish individualized information on lesion motion and allows the use of smaller margins than standard CT [16, 21, 22]. Furthermore, a gated technique has been shown to reduce PTV and to be an effective imaging guidance strategy to personalize the lung cancer treatment [2325].

RG-PET/CT has been demonstrated as a feasible technique in routine clinical practice by improving image quality, quantification and diagnostic accuracy in the detection of lung lesions [2628]. Nevertheless, there is a lack of evidence of its usefulness in radiation treatment planning; to date only one paper evaluating its clinical utility in radiation treatment has been published showing that RG-PET/CT may better estimate the extent of tumour motion and, therefore, improve radiation treatment planning [29].

The purpose of the present study was to compare PTV defined on RG-PET/CT to the standard free-breathing CT-based PTV and to evaluate if RG-PET/CT can be useful to personalize PTV by tailoring the target volume to the lesion motion in lung cancer patients.

Materials and methods

Patients

From March 2010 to March 2012, we prospectively enrolled 13 patients (six men, mean age 70.0 years, range 45–88 years) with histologically proven lung cancer [1 small cell lung cancer (SCLC) and 12 non-small cell lung cancer (NSCLC)] who were candidates for radiation therapy. All patients gave their signed informed consent to the study. Clinical workup included 18F-FDG PET/CT and contrast-enhanced (ce) CT in all patients with the addition of mediastinoscopy (1/13 patients, 7.7 %) and bronchoscopy (4/13 patients, 30.8 %) completed within 1 month before the onset of radiation treatment. Clinical stage of the disease was IA–IVA. Of 13 patients, 9 (69.2 %) underwent chemotherapy concurrently (4/9 patients, 44.4 %) or prior to (5/9 patients, 55.6 %) radiation therapy. Radiation treatment was performed with radical intent in 12/13 patients (92.3 %) and as a part of neoadjuvant treatment in 1/13 patients (7.7 %). Of 13 patients, 1 (7.7 %) was referred to treatment for local relapse and the remaining 12/13 patients (92.3 %) for primary disease. In 7/13 cases (53.8 %) volumetric modulated arc therapy was used, whereas 6/13 patients (46.2 %) underwent 3-D conformal treatment. Clinical characteristics of the patients are reported in Table 1.

Table 1 Clinical characteristics of patients (n = 13)

Patient preparation and RG-PET/CT scanning protocol

The day before RG-PET/CT acquisition, patient compliance and set-up for gated scan was tested by positioning the patient into the gantry, on a flat table, in the same position and with the same immobilization devices for radiation treatment delivery. The respiratory curve was recorded for 12 min with a real-time position management (RPM) system (Varian Medical Systems, Palo Alto, CA, USA) to check the regularity of the breathing.

For PET/CT examination, patients were prepared according to common guidelines [30]. In particular, patients fasted at least 6 h before tracer administration and blood glucose was assessed immediately before tracer injection (cut-off level 170 mg/dl). Patients were intravenously injected with 4.0 MBq/kg of 18F-FDG and orally hydrated in the waiting time before scanning. Approximately 60 min after injection, RG-PET/CT studies were performed with a Discovery PET/CT 600 scanner (GE Healthcare, Milwaukee, WI, USA), integrated with an RPM system. For positioning, an internal laser on the scanner and skin markers were used for geometrical alignment of the patient during RG-PET/CT acquisition. One field of view (FOV) entirely encompassing the lesions was acquired in RG mode for both PET and CT. A free-breathing, ungated helical CT scan (16 slices, 40 mA, 120 kV, slice thickness 3.72 mm, rotation 0.8 s, pitch 1.375) was extended to the entire chest and upper abdomen for the contouring of the organs at risk in PTV. RG-CT was acquired in step-and-shoot mode, 40 mA, 120 kV, rotation time 0.5 s, revolution duration equal to patient’s breathing period +0.5 s, slice thickness 2.5 mm; RG-CT images were then retrospectively sorted into six phases of the respiratory cycle recorded during acquisition. RG-PET was acquired in 3-D list mode with 12 min duration; images were then retrospectively reconstructed into six breathing phases, 2 min/phase, with iterative reconstruction (ordered subset expectation maximization, OSEM), 256 matrix size, attenuation, random and scatter correction. Attenuation correction of RG-PET images was obtained from the corresponding breathing phases of RG-CT data (phase-matched attenuation correction).

PTV definition

Standard PTV definition

Free-breathing helical CT images, obtained during the PET/CT study, were transferred to the treatment planning system (TPS), an Oncentra MasterPlan 4.1 SP2 workstation (Nucletron BV, Veenendaal, The Netherlands), and standard clinical target volumes (CTV1) were manually contoured by the radiation oncologist. Standard PTV (PTV1) resulted from CTV1 with the addition of 1 cm of expansion in all directions in order to compensate for internal organ motion and set-up errors [12].

RG-PTV definition

RG-PET and RG-CT images were displayed on AW 4.6 workstation (GE Healthcare, Milwaukee, WI, USA), with preset window levels and colour scale according to a standardized institutional protocol for metabolic lesion contouring. The RG-MTV was initially obtained by an automatic segmentation algorithm commercially implemented on the workstation, with a fixed threshold of 42 % of the maximal activity of the target lesion [31]. Segmentation was performed on all images of each single phase of the patient’s respiratory cycle. The nuclear medicine physician (LG) visually evaluated the automatically generated volume and, if considered inappropriate, manually corrected. In particular, physiological uptake areas (cardiac wall, chest wall, vessel activity) were carefully excluded from contouring; in cases where portions of the pathological metabolic lesion were excluded by the automatic segmentation method, they were manually included. The RG-MTV was obtained by the union of the metabolic volumes contoured on each respiratory phase of gated PET data and represents the volume in which the metabolic lesion is moving during the whole breathing cycle (i.e. the metabolic internal target volume). Then, the RG-MTV was transferred to the TPS workstation in the Radiotherapy Department.

RG-CT-based CTV (CTV2) was defined by the radiation oncologist (SM, RN) by manually contouring each CT image in all six respiratory gates. The CTV2 resulted from the union of the volumes derived from each gate. RG-CT-based PTV (PTV2) was obtained from CTV2 by adding a margin expansion of 7–8 mm in head to feet direction and 5 mm in anterior to posterior and left to right direction to compensate for set-up errors [14, 21, 3234]. RG-PET/CT-based CTV (CTV3) was calculated by the radiation oncologist (SM, RN) by the union of CTV2 and RG-MTV. The RG-PET/CT-based PTV (PTV3) was then obtained with the addition of the same margin expansion previously mentioned for PTV2.

The lesion displacement was also assessed on the TPS workstation by calculating the movement of the geometric centre of the lesion in all phases of RG-CT. The displacement along the x, y, z axis and total displacement vector were assessed.

Critical organs (lung, heart, spinal cord) were contoured on ungated free-breathing helical CT images. The irradiation dose was delivered to each patient on the basis of PTV3 data.

Data analysis

The paired Student’s t test was used to assess statistical difference between CTV and PTV values. For each target lesion, the absolute and percentage differences between standard PTV1, PTV2 and PTV3 were calculated. The geometrical portion of PTV2 and PTV3 encompassed by PTV1 (intersection PTV1-PTV2 and intersection PTV1-PTV3, respectively) and the geometrical portion of PTV2 encompassed by PTV3 (intersection PTV2-PTV3) were automatically calculated on the TPS workstation as absolute and percentage values; similarly, the geometrical portion the PTV2 and PTV3 outside PTV1 (PTV2 out PTV1 and PTV3 out PTV1, respectively) and the geometrical portion of PTV2 outside PTV3 (PTV2 out PTV3) were assessed for each target lesion.

Results

Overall, 13 target lesions in 13 patients could be evaluated. Of 13 patients, 8 had nodal disease at PET/CT. In 3/13 patients (23 %) (patients 1, 6 and 13 in Table 2) nodal lesions were encompassed in target lesions as they were very close to them. In 5/13 patients (38.5 %) (patients 2, 3, 9, 11 and 12 in Table 2) nodal lesions were contoured apart and not included in CTV and PTV calculations. In Table 2 CTV and PTV values are reported for each lesion. In Table 3 the absolute and percentage differences of PTV, PTV intersection and the portion of PTVs out are reported for each lesion. The percentages of PTV2 out PTV1 and PTV3 out PTV1 were not lower than 10 % in 6/13 cases (46.2 %) and than 20 % in 3/13 cases (23.1 %). The percentage of PTV3 out PTV2 was not lower than 10 % in 3/13 lesions (23.1 %) and not lower than 20 % in only 1/13 lesions (7.7 %). In Table 4 the mean ± SD values of CTV, PTV, PTV intersection and PTVs out are reported. The CTV1 value was significantly smaller than CTV2 and CTV3, whereas no statistically significant difference was found between CTV2 and CTV3. Conversely, the PTV1 value was significantly larger than both PTV2 and PTV3, whereas PTV2 and PTV3 values did not show significant differences. No statistically significant difference was found among PTV intersections and PTVs out.

Table 2 CTV and PTV values for each lesion
Table 3 Absolute and percentage differences of PTV, PTV intersection and the portion of PTVs out for each lesion
Table 4 Mean ± SD values of CTV, PTV, PTV intersection and PTVs out

In Table 5, the lung lesion dimension on CT, lesion location, lesion displacement along orthogonal directions and vector displacement are reported for each lesion. The mean (median) ± SD lesion dimension was 39.4 (38.5) ± 16.5 mm. Six lesions were located in the inferior/medial lobe and seven in the superior lobe.

Table 5 Lung lesion dimension on CT, lesion location, lesion displacement along orthogonal directions and vector displacement for each lesion

Discussion

Respiratory motion can affect the quality of diagnostic PET/CT imaging. The effect of motion on volume definition is particularly relevant in radiation treatment planning, as the target lesion, although detectable, can be substantially altered in shape, borders and dimension. Information on the amount and nature of the organ motion is thus essential for the determination of internal margin size and technical solution for handling respiratory motion in PET/CT, which can be essential to personalizing target volume and optimizing radiation treatment strategy [23]. The aim of this paper was to compare PTV obtained from free-breathing helical CT, commonly used for radiation therapy planning, with PTV defined by using RG-PET/CT.

Our data showed that PTV1 (free-breathing helical CT-based) was significantly larger than both PTV2 (gated CT-based) and PTV3 (gated PET/CT-based) (p = 0.038 and p = 0.043, respectively). In a lesion by lesion analysis, PTV1 was larger than PTV2 and PTV3 in the same 11/13 (84.6 %) cases (Table 2). Compared to PTV2 and PTV3, PTV1 increase ranged from 3.0 % (Table 3, lesion 2) to 49.9 % (Table 3, lesion 11) and from 6.1 % (Table 3, lesion 2) to 53.5 % (Table 3, lesion 3), respectively. These results, in particular those regarding PTV2, are comparable to those previously published by other authors [15, 25] who demonstrated a significant reduction of PTV by using RG-CT for radiation therapy planning.

Data from this study indicate that an ungated PTV larger than RG-PTV does not imply that the former geometrically envelops the latter. In the present study, in patient 2 (Table 2) PTV1, PTV2 and PTV3 were very similar (22.9, 22.2 and 21.5 ml, respectively). The differences between PTV1/PTV2 and between PTV1/PTV3 were minimal with only 0.7 (3.0 %) and 1.4 ml (6.1 %), respectively (Table 3). Conversely, the PTV2 out and PTV3 out (volumes not encompassed in PTV1) were 5.6 (25.2 %) and 5.2 ml (24.1 %), respectively (Fig. 1). A similar condition could also be found in patient 4 (Tables 2 and 3). In these two cases, although similar PTVs values were found, a portion of the RG-PTV was outside the standard PTV due to unmatched geometrical position of the volumes. As a consequence, in these two cases the possibility of missing the lesion during treatment is high if an ungated imaging technique is used, leading to a greater probability of local relapse.

Fig. 1
figure 1

Patient 2. The yellow and red volumes represent PTV1 (helical CT-based, 22.9 ml) and PTV3 (RG-PET/CT-based, 21.5 ml), respectively. There is a portion of PTV3 (5.2 ml, 24.1 %) clearly outside of PTV1

In patient 3 (Table 2) PTV1 (106.3 ml) was consistently larger than both PTV2 and PTV3 (57.6 and 49.4 ml, respectively). PTV2 out and PTV3 out were only 1.5 and 0.2 %, respectively (Table 3). In this case ungated PTV is likely to envelop not only the lesion and its motion but also healthy tissue, with an increased probability of toxicity (Fig. 2). Finally, in patient 6, PTV1 (185.9 ml) was smaller than PTV2 (258.3 ml), and PTV3 (260.3 ml) with about 37 % of gated PTV outside PTV1. Again, the probability of missing the target lesion during standard PTV-based radiation treatment could increase.

Fig. 2
figure 2

Patient 3. The yellow and orange volumes represent PTV1 (helical CT-based, 106.3 ml) and PTV3 (RG-PET/CT-based, 49.4 ml), respectively. PTV1 is larger than RG-PTV and includes a portion of healthy tissues

As expected, our preliminary data showed that gated CTVs were significantly larger than standard CTV (Table 4). In fact, gated CTVs were derived from gated images and included motion of the lesion during breathing. Similar results were obtained by Aristophanous et al. [29] who evaluated the clinical utility of gated metabolic imaging in radiation treatment planning and demonstrated that RG-PET-based GTV in lung lesions was significantly larger than standard PET-based GTV.

Conversely, no significant differences were found between CTV2 and CTV3 (43.1 vs 44.8 ml, p = 0.161), between PTV2 and PTV3 (93.8 vs 97.0 ml, p = 0.161) and PTV out values. Similarly, no statistically significant difference was found among PTV intersections and PTV outside regions (Table 3). Although PTV2 out PTV1 and PTV3 out PTV1 values were very similar in most of cases, in 3/13 (23.1 %) lesions (Table 3, patients 1, 7 and 8), the percentage of PTV3 out was at least 5 % higher than that of PTV2 out. Furthermore, the portion of PTV3 out PTV2 was at least higher than 10 % in 3/13 (23.1 %) cases (Table 3, patients 1, 7 and 8). Currently these data cannot be used to draw conclusions and need to be confirmed in a larger population, but they could indicate that RG-PET/CT can add useful information to RG-CT for a better definition of the target volume in some cases.

In our study, lesion displacement did not correlate with PTV values. In particular, the lesions with the highest values of motion vector ΔS (patients 4, 5 and 10 in Table 5) showed different values of PTV2 out and PTV3 out, ranging from 19.4 % for lesion 4 to only 0.9 % (PTV3 out) for lesion 5. Furthermore, the lesion (patient 6) with the highest percentage of PTV3 out (37.5 %) showed a ΔS value of only 4.4 mm. This variability could be related to the lack of predictability of motion based on lesion location in our patients. The standard CTV expansion could thus be inappropriate to correctly define the internal target volume of the lesion.

In the present work, different margins of expansion have been applied to obtain standard and gated PTV. Consequently, it can be argued that the differences found in PTV values could be related to the different methodology rather than to the imaging techniques. However, when CTV is obtained from free-breathing CT, the lack of information on lesion motion forces the use of internal margins, in addition to set-up margins, to compensate for an uncertain and estimated motion and to reduce the risk of missing the lesion during treatment delivery. Conversely, when lesion motion is managed with a gated imaging technique, CTV can encompass the volume in which the lesion is moving during breathing (i.e. internal volume) and it is usually larger than standard CTV as found in our work; nevertheless, only set-up margins have to be added to CTV because internal margins have been already included [35]. Accordingly, the different margins of expansion used in the present work for PTV assessment can be considered strictly related to the imaging technique used for radiation treatment planning and reflects what is commonly performed in clinical practice. Although probably of interest from a merely investigational point of view, the use of the same margin expansion independently from the imaging technique cannot be considered a valid approach when translated into the clinical routine of radiation treatment planning [12].

Another issue in the present work is the use of anisotropic expansion in gated PTVs, with 5 mm in left to right and anterior to posterior directions and 7–8 mm in head to feet direction. The larger expansion in head to feet direction was not arbitrary and it was used because usually head to feet is the direction with both the larger motion [14, 21, 32, 33] and more consistent interfractional motion variability during radiotherapy courses [34]. Although a gating imaging technique is used, both factors can generate uncertainty in volume definition that can be compensated for by larger margin expansion in head to feet direction. In addition, a margin expansion of 7 and 8 mm was used for lesions located in the superior and medial-inferior lobe, respectively, as the lesion location can be correlated to different motion amplitude in head to feet direction [12, 14, 21, 34].

In the present work, the contouring of gated PET images was performed by the same nuclear medicine physician (LG) with an automated commercial software implemented on the workstation. PTVs were always manually defined by the same two radiation oncologists (SM, RN). Although the inter-observer variability in volume definition was not evaluated as it was not the aim of the present study, it is recognized that inter-physician GTV variation can have important dosimetric consequences, often more important than those of respiratory motion [12]. Notwithstanding, it has to be considered that all physicians involved in the processing of the data of this paper are well experienced in image contouring (more than 7 years) for radiation treatment planning and followed institutional standardized protocols.

In the present study, only one method was applied for contouring of PET images [31], as it has been the method commonly used in routine clinical practice for many years in our institution and commercially implemented in the workstation used for contouring. A comparison of different segmentation algorithms, although of interest, was not performed as it was not a matter of the investigation. Moreover, it has to be considered that to date there is no standard reference for metabolic target lesion contouring [36] and the manual procedure can also be considered a valid approach when methodology is standardized [37].

Compared to a standard scan, RG techniques are time consuming and require more time for scanner and patient set-up, acquisition and processing of data. Nevertheless, modern PET/CT scanners are commercially implemented with gating tools and powerful hardware and software capabilities that enable RG studies yet reduce the time for acquisition and processing data. This can be a positive impacting factor to further encourage the use of gating techniques in routine clinical practice. Furthermore, a single axial or helical planning CT scan acquired during free breathing is not an optimal study, because it may capture the random position of a tumour and/or generate artefacts of the patient’s anatomy. Planning on scans that are not representative of the tumour position during the time of the delivery of radiotherapy may result in systematic errors. Consequently, the use of a gating technique is strongly recommended for treatment planning in thoracic oncology, in spite of the need to deal with methodological issues [20].

A limitation of this study is the small sample size. However, while these are only preliminary data, they indicate the strength of a gating technique to optimize the radiation treatment planning in lung cancer patients. Secondly, another limitation of the study is the lack of comparison between PTV based on standard PET/CT and RG-PET/CT. However, the aim of our work was to compare a motion handling technique to the commonly used standard CT-based procedure for PTV definition. Finally, we cannot provide patient outcomes derived from the use of RG-PET/CT in radiation treatment planning as this was not a specific matter of investigation and also due to the limited number of patients analysed.

In conclusion, our preliminary data indicate that the use of RG-PET/CT in lung cancer can affect not only the volume of PTV but also its shape. When gated PET/CT information was added to gated CT, PTV changed in about 23 % of cases. The use of a gating technique is thus crucial for better delineating PTV by tailoring the target volume to the lesion motion.