Introduction

The quantitative assessment of renal blood flow (RBF) may help to understand the pathophysiological basis of kidney function and to evaluate pathophysiological events leading to vascular damage, such as renal arterial stenosis and chronic allograft nephropathy. The quantitative estimation of RBF by the use of H2 15O and dynamic PET has been developed and demonstrated by Nitzsche et al. [1]. The kinetic model of H2 15O is based on the assumptions that all activity is extracted by the parenchyma, extraction is very rapid, and tubular transport has not started or is insignificant at a level that does not influence the calculation of RBF [15]. With these assumptions, RBF has been estimated based on regions of interest (ROI) by the H2 15O dynamic PET approach [1, 3, 4]. Also, calculations to produce parametric images of RBF has been reported [5]. However, the quantitative computation of RBF has so far assumed that the blood/tissue partition coefficient of water (p, ml/g) is uniform for the whole region of renal tissue [3, 4], and/or that the contribution of radioactivity from the vascular space is negligible [57]. The influence on quantitative accuracy of these assumptions is unknown.

In previous studies RBF has been computed from the uptake rate (K 1, ml/min/g) [17]. Some studies also simultaneously computed the partition coefficient (p) [6, 7], and the apparent p values obtained ranged between 0.52 and 0.78 ml/g. From the published values of water content for tissue (76%) and blood (81%) [8], the p value can be physiologically determined as: p phys=0.94 ml/g [9]. The much smaller apparent p value might be due to the tissue mixture (or a partial volume effect) [10, 11] because of the composite structure of the kidney. The effects of the tissue mixture affect mostly K 1 and not clearance rate (k 2 min−1). Therefore the clearance rate of H2 15O (k 2 min−1) multiplied by p phys could be used for the calculation of blood flow rather than K 1 (ml/min/g) [11] when the effect of the tissue mixture is not negligible, although it is unknown how the glomerular filtration rate (GFR) additionally contribute to k 2. Thus, the influence of GFR on k 2 should be evaluated and allowed for in the computation of RBF.

The aim of this study was to develop a method to simultaneously calculate parametric images of K 1 and k 2 as well as the arterial blood volume (V A, ml/ml). The feasibility in terms of quantitative accuracy and image quality of calculated images was experimentally tested in healthy subjects. GFR was measured in each subject to investigate how much it contributes to the clearance rate (k 2, min−1). A simulation study was also performed to evaluate error sensitivities for possible error sources.

Materials and methods

Theory

The present formula was characterized by simultaneously estimating multiple parameters of uptake rate constant (K 1, ml/min/g) and clearance rate constant (k 2 ml/g) as well as activity concentration in the arterial vascular space (V A, ml/ml). The kinetic model for H2 15O was based on a single-tissue compartment model as follows:

$$Ci\left( t \right) = \left( {1 - V_A } \right) \cdot K_1 \cdot A_w \left( t \right) \otimes e^{ - k2 \cdot t} + V_A \cdot A_w \left( t \right)$$
(1)

where Ci(t) (Bq/ml) is radioactivity concentration in a voxel of PET image, A w(t) (Bq/ml) is the arterial input function, and ⊗ indicates the convolution integral.

In the present computation, we applied a basis function method (BFM) as introduced by Koeppe et al. [12] to compute the cerebral blood flow parametric image as well as the clearance rate constant simultaneously. Gunn et al. [13] applied this method to parametric imaging of both binding potential and the delivery of ligand relative to the reference region. The computation method has also been applied to myocardial blood flow studies to compute the uptake, clearance rates and blood volume [14, 15]. The BFM procedure for the present RBF computation is illustrated in Fig. 1. The BFM method enables parametric images to be computed by using linear least squares together with a discrete range of basis functions as the parameter value for k 2 incorporating the nonlinearity and covering the expected physiological range. The corresponding basis functions formed are:

$$F\left( {k_2 ,t} \right) = A_w \left( t \right) \otimes e^{ - k2 \cdot t} $$
(2)
Fig. 1
figure 1

Schematic diagram of the computation procedure by the BFM

For a physiologically reasonable range of k 2, i.e. 0 < k 2 < 15.0 ml/min/g, 1,500 discrete values for k 2 were found to be sufficient. Then Eq. 1 can be transformed for each basis function into a linear equation:

$$ \begin{array}{*{20}l} {{Ci{\left( t \right)} = \Theta \cdot F{\left( {k_{2} ,t} \right)} + \Psi \cdot A_{w} {\left( t \right)}} \hfill} \\ {{\Theta = {\left( {1 - V_{A} } \right)} \cdot K_{1} } \hfill} \\ {{\Psi = V_{A} } \hfill} \\ \end{array} $$
(3)

Hence for fixed values of k 2, the remaining two parameters Θ and Ψ can be estimated using the given basis function by standard linear least squares, and are represented as Θ k2 and Ψ k2. The value k 2 for which the residual sum of squares

$${\text{s}}\left( {{\text{k}}_{\text{2}} } \right)^2 = \sum\limits_t {\left( {Ci\left( t \right) - \Theta _{k2} \cdot F\left( {k_2 ,t} \right) - \Psi _{k2} \cdot A_w \left( t \right)} \right)^2 } $$
(4)

is minimized is determined by a direct search, and associated parameter values for this solution (K 1, k 2, V A) are obtained.

Subjects

Six healthy human subjects (the demographics are shown in Table 1) were studied under basal conditions and stimulation (after enalapril infusion) conditions. All subjects were nonsmokers and none of them was taking any medication. All subjects gave written informed consent. The study was approved by the Ethics Committee of the Hospital District of South-Western Finland, and was conducted in accordance with the Declaration of Helsinki as revised in 1966.

Table 1 Baseline characteristics of the six subjects studied

PET experiments

PET was carried out in 2-D mode using a GE Advance scanner (GE Medical Systems, Milwaukee, WI). After a 300-s transmission scan, two scans were undertaken with injection of H2 15O (1.0 to 1.5 GBq) into the cephalic vein of the right forearm. The first scan was under resting conditions and the other was under stimulated conditions, namely 20 min after infusion of 0.5 mg enalapril. The scan protocol consisted of 20 frames over a total of 240 s (15×4 s, and 5×10 s). During PET scanning, blood was withdrawn continuously through a catheter inserted into the left radial artery using a peristaltic pump (Scanditronix, Uppsala, Sweden). Radioactivity concentrations in the blood were measured with a BGO coincidence monitor system. The detectors had been cross-calibrated to the PET scanner via an ion chamber [16]. GFR was also measured in each subject [17]. To obtain the PET equivalent flow ratio for GFR, a kidney weight of 300 g and a cortex ratio of 70% were assumed [8].

Data processing

Dynamic sinogram data were corrected for dead time in each frame in addition to detector normalization. Tomographic images were reconstructed from corrected sinogram data by the OSEM method using a Hann filter with a cut-off frequency of 4.6 mm. Attenuation correction was applied with the transmission data. A reconstructed image consisted 128×128×35 matrix size with a pixel size of 4.3×4.3 mm and 4.2 mm with 20 frames. Measured arterial blood time–activity curves (TAC) were calibrated to the PET scanner and corrected for the dispersion (τ=5 and 2.5 s for intrinsic and extrinsic, respectively) [18] and delay [19]. The corrected blood TAC was used as the input function.

A set of K 1, k 2 and V A images was generated according to the BFM formula described above, using a set of dynamic reconstructed images and input function. Computations were programmed in C environment (gcc 3.2) on a Sun workstation (Solaris 10 Sun Fire 280R) with 4 GB of memory and two Sparcv9, 900-MHz CPUs.

Data analysis

A template ROI obtained by summing whole frames of a reconstructed dynamic image was drawn on an image of the whole region of each kidney (average ROI size for the all subjects was 153±43 cm3). Also, a ROI was drawn on a region of high tracer accumulation on the summed image as an assumed cortical region. Functional values of K 1, k 2 and V A were extracted from both ROIs, i.e. for the whole region and the cortical region, respectively. Data re shown individually or as means±SD. Student’s paired t test was used for comparisons between the physiological states and p values <0.05 were considered significant.

The ROI for the whole region was divided plane-by-plane into subregions of ten pixels each. The subregions were created by extracting pixels first from the horizontal direction and then from the vertical direction inside the whole ROI in each slice. Each subregion consisted of a single area with the same number of pixels. Functional values of K 1, k 2 and V A were extracted from each subregion. Tissue TACs were also obtained for each subregion from corresponding dynamic images. The three parameters K 1, k 2 and V A were estimated using the Eq. 1 and the input function fitted to the tissue TACs by the nonlinear least-squares fitting method (NLF, Gauss-Newton method). Functional values of K 1, k 2 and V A from corresponding subregions were then compared between the methods. Regression analysis was performed.

The model relevancy introducing p and/or V A into the computation was tested using the Akaike Information Criterion (AIC) [20]. The most appropriate model provides the smallest AIC. The tissue TACs from the subregions were fitted and AICs were computed for models with the three parameters K 1, k 2 and V A, fixing p (=K 1/k 2) at 0.35 ml/g (mean value obtained in the present subjects), fixing V A at 0 ml/ml, and fixing V A at 0.15 ml/ml (mean value obtained in the present subjects).

Error analysis in the simulation

Error propagation from errors in the input function for the present BFM formula was analysed for two factors: delay and dispersion in arterial TAC. It is known that the measured arterial TAC is delayed and more dispersed relative to the true input TAC in the kidney because of the time for transit of blood through the peripheral artery and the catheter tube before reaching the detector [18, 19]. Calculations of RBF so far have employed a fixed partition coefficient (p, =K 1/k 2, ml/g) and/or assumed the blood volume (V A, ml/ml) as negligible throughout the whole renal region and do not estimate it regionally. BFM formulae with a fixed value of p (BFM-pfix) and blood volume V A (BFM-vfix) in addition to the present BFM formula, and the error in these formulae, were analysed.

A typical arterial input function obtained from the present PET study was used in the present simulation as the true input function. Applying this input function to the water kinetic model in Eq. 1, a tissue TAC was created assuming values for normal kidney tissue (K 1=2.0 ml/min/g, V A=0.14 ml/g [5], and p=0.4 ml/g, corresponding to the estimated means in cortical region in all subjects in this study).

Time in the input function was shifted from −4 to 4 s to simulate the error sensitivity due to the error in the time delay, where a positive error represents an over-correction of the time delay. The input function was convoluted or deconvoluted with a simple exponential [18] by shifting the time constant from −4 to 4 s to simulate the error sensitivity due to error in dispersion correction, where a negative error represents under-correction, as described previously [18, 21]. Values of K 1 and k 2 were calculated using simulated input functions and the tissue TACs based on the BFM formula. Errors in these calculated K 1 and k 2 values are presented as percentage differences from the assumed values. Then, the value of p was varied from 0.3 to 0.5 ml/g and the tissue TAC was generated as above to simulate the error from the value of p in BFM-pfix formula. Also, the V A value was varied from 0.0 to 0.4 ml/ml and the tissue TAC was generated to simulating the error from V A in BFM-vfix formula. Then, K 1 and k 2 were calculated using the true input function and the created tissue TACs, assuming p=0.4 ml/g and V A=0.0 ml/ml in the BFM-pfix and BFM-vfix formulae, respectively. Error in K 1 and k 2 values due to fixing p is presented as the percentage difference in K 1 and k 2 as a function of p. Error in K 1 and k 2 values due to neglecting V A is presented as the percentage difference in K 1 and k 2 as a function of V A. Also, K 1 and k 2 were computed with V A fixed at 0.14 ml/ml in the BFM-vfix formula from the set of the tissue TACs, in which K 1 and p were fixed at 2.0 ml/min/g and 0.4 ml/g, respectively, and VA was varied. The percentage difference in K 1 and k 2 between the two conditions, i.e. the initial (K 1=2.0 ml/min/g and V A=0.14 ml/ml) and changed conditions (presented as ΔK 1 and Δk 2, respectively) is presented as a function of the percentage difference in the assumed V A from 0.14 ml/ml (ΔV) to investigate the extents to which the change in K 1 and k 2 were estimated when K 1 and k 2 were computed in the BFM-vfix formula.

Results

Experiments

The relationships of the regional ROI values of K 1, k 2 and V A between NLF and BFM are shown in Fig. 2. The regression lines obtained were K 1,BFM=0.93K 1,NLF−0.11 ml/min/g (r=0.80, p<0.001), k 2,BFM=0.96k 2,NLF−0.13 ml/min/g (r=0.77, p<0.001), and V A,BFM=0.92V A,NLF−0.00 ml/ml (r=0.97, p<0.001), where the subscripts show the methods used for calculating the parametric values; the slopes were not significantly different from unity.

Fig. 2
figure 2

Relationships of (a) K 1, (b) k 2 and (c) V A between the ROI-based NLF method and pixel-based BFM. The regression lines were K 1,BFM=0.93K 1,NLF−0.11 ml/min/g (r=0.80, p<0.001), k 2,BFM=0.96k 2,NLF−0.13 ml/min/g (r=0.77, p<0.001), and V A,BFM=0.92V A,NLF−0.00 ml/ml (r=0.97, p<0.001)

The fitted curve by the present model estimating K 1, k 2 and V A fitted better than the other two models fixing p (=K 1/k 2) or V A. An example of fitted curves is shown in Fig. 3. Also, the AIC values from three parameter fitting were the smallest for all subjects except two values for two parameter fitting fixing V A in patient 2 and fixing p in patient 3, although some AIC values were similar (Table 2). These results show that the present method with three parameter fitting is feasible for computing RBF.

Fig. 3
figure 3

Curves fitted to the measured tissue TAC from the different computation methods. Three parameters: K 1, k 2 and V A were computed. p-fixed: K 1 and V A were computed with p (=K 1/k 2) fixed at 0.35 ml/g. V A-fixed: K 1 and k 2 were computed with V A fixed at 0.15 ml/g. V A-ignored: K 1 and k 2 were computed without taking into account V A

Table 2 AIC values for the models

Values of K 1, k 2·p phys and V A were obtained for the whole renal region and cortical region (Table 3). The K 1 values were smaller than k 2·p phys values and the ratio between them ranged from 0.35 to 0.45, suggesting that K 1 values underestimated RBF due to the partial volume effect. Both K 1 and k 2·p phys were not significantly different between the resting and stimulated conditions for the whole renal region and the cortical region, respectively, although the value of V A was higher under the stimulated conditions than under the basal conditions. The GFR obtained was 78±4 ml/min, corresponding to a clearance rate of 0.37±0.02 ml/min/g and to 9.6% of the k 2 obtained for the cortical region under the normal conditions.

Table 3 Values of K 1, k 2·p phys and V A (n=6) in the whole renal region and the cortical region calculated by the present method for the baseline conditions and the stimulated conditions

Representative K 1 and k 2·p phys images generated by the present method are shown in Fig. 4. The quality of the image is acceptable. The K 1 and k 2·p phys values ranged from 1.5 to 2.0 ml/min/g and 3.0 to 5.0 around cortical region, respectively, and some parts showed higher values than these. The average time required to compute the parametric images was 2 min 23 s.

Fig. 4
figure 4

Representative parametric images of K 1 (left) and k 2·p phys (right) for a subject under baseline conditions. Coronal (upper) and transverse (lower) views are shown

Error analysis

The sizes of the errors introduced in both K 1 and k 2 were less than 20% for estimation of delay and the dispersion time constant up to 2 s (Fig. 5). The error sensitivity in K 1 and k 2 was 40% when the partition coefficient was 0.35 (Fig. 6). The magnitude of the error was markedly enhanced when the blood volume was ignored (Fig. 7a), and if the arterial blood volume increased by 25%, K 1 and k 2 were overestimated by 20% (Fig. 7a).

Fig. 5
figure 5

Error propagation from the error in input time (a) and dispersion time constant (b) to K 1 and k 2 (the two lines were identical). Positive and negative values of error indicate over- and under-correction of delay time and dispersion time, respectively

Fig. 6
figure 6

Error propagation from the partition coefficient (p, ml/g) to K 1 and k 2. When the true p was varied between 0.6 and 0.8 ml/g, the size of the error in RBF was simulated assuming p=0.7 ml/g

Fig. 7
figure 7

a Error propagation from the arterial blood volume (V A, ml/ml) to K 1 and k 2 (the two lines were identical). When the true V A changed from 0.0 to 0.4 ml/ml, the size of the error in K 1 and k 2 calculated assuming V A=0.0 ml/ml was simulated. b Error propagation from the change in arterial blood volume from 0.14 ml/ml (ΔV A) to the change in K 1 and k 2 from the initial conditions (ΔK 1 and Δk 2, ml/min/g) (the two lines were identical)

Discussion

We have presented an approach to generating quantitative K 1, k 2 and V A images using H2 15O and PET applying the BFM computation method. The validity of this approach in healthy human subjects under resting and stimulated conditions is described. The rate constant values of K 1 and k 2·p phys obtained from the parametric images were consistent against NFL and the quality of the K 1 and k 2·p phys images obtained was acceptable. The smaller K 1 against k 2·p phys values suggested that the K 1 values underestimated the absolute RBF value due to the partial volume effect. The simulation showed that the delay time and dispersion time constant should be estimated within an accuracy of 2 s, and V A and p cannot be ignored/fixed to estimate the rate constants of K 1 and/or k 2. Also V A cannot be ignored, even when only relative rate constant values are needed. These findings suggest that the present k 2 obtained BFM technique provides an RBF image with reasonable accuracy and quality.

In the present study the rate constants of K 1 and k 2 were experimentally computed, and the ratios obtained ranged from 0.35 to 0.45 ml/g, which corresponds to the apparent kidney–blood partition coefficient. The much smaller apparent p value might be due to a partial volume effect, as has been demonstrated in a previous brain and cardiac study [10, 11], because of the composite structure of the kidney, the spatial resolution of the reconstructed image and breathing movement of the patient during the scan. When the rate constant K 1 is underestimated due to the partial volume effect, k 2·p phys could be applied for RBF rather than K 1. The present study showed that the contribution of GFR to the clearance rate was only 10%, and that k 2·p phys is more appropriate for RBF assessment, although further study of how the GFR changes under stimulated conditions is required. The k 2·p phys value in the cortical region obtained in the present study was 3.64±2.15 ml/min/g under normal condition, a value within the normal range of 4 to 5 ml/min/g reported in the literature [22]. Middlekauff et al. [2325] applied the ROI base analysis, and showed similar RBF values around 4 ml/min/g. These findings also support the use of k 2·p phys for the calculation of RBF. The different values of RBF between the present study and the previous studies [35] might be due to differences in the approaches.

The present computation of RBF by the BFM has two main advantages over the NLF. One is the ability to produce a voxel-by-voxel quantitative parametric map, and the other is faster computing speed. In fact, the parametric images were obtained within a reasonable time, i.e. 2.5 min with an image size of 128×128 pixels with 35 slices and 22 frames. The time could be further reduced by applying a threshold to omit pixels with lower values. From a clinical standpoint, voxel-by-voxel analysis is preferred to ROI-based analysis because the operator can independently define ROIs to improve reproducibility, and faster computations are important for analysing very large datasets.

Kinetic parameters estimated by the NLF agreed well with those estimated by the BFM as shown in Fig. 2. The disagreement in some rate constant values between the voxel-based (BFM) and ROI-based computation methods might have been due to the composite structure between the cortical region and its surroundings, or to image noise. Although superior to the NLF in terms of computing speed and ability to generate parametric maps, the BFM shares the same source of errors as the NLF because they use the same model and assumption. Delay and dispersion in input function, motion of the patient during a study [2628], and flow heterogeneity [29] are sources of error for parameters estimated by both the NLF and BFM. Selection of a specific range of k 2 and the number of basis function can affect the accuracy and precision of the estimated parameters in neuroreceptor studies [30, 31]. However, the range was 0 to 15 ml/min/g in the present computation with H2 15O, and the limits of this range would be acceptable for the present computation. In practice, selection of a wider range and/or a large number of discrete values of the basis function is slow and inefficient against the required accuracy and precision.

The present simulation study showed that if V A is neglected or fixed, not only the absolute rate constants, i.e. RBF value, are overestimated, but estimated changes in RBF between two physiological states could be over- or underestimated. These findings suggest that V A should be included to obtain either absolute or relative values of RBF. For p, the present simulation revealed that the error sensitivity in RBF for that value was significant. The values of p for the whole and cortical regions were 0.35 and 0.42 ml/g, respectively. If the value was fixed at 0.4 ml/g, a 40% overestimation in RBF for regions with a p of 0.35 occurred. Thus, regional difference in p introduce error in quantitative RBF values. Also the AIC analysis showed that introducing the extra parameters of p and V A did not increase the AIC value against the others. These findings suggest that both p and V A need to be estimated simultaneously with quantitative RBF, especially when changes under different conditions are assessed.

Knowledge of RBF is mostly needed in determining the severity of renovascular disease. Although the degree of renal artery stenosis is easily diagnosed, its actual effect on RBF remains difficult to quantify. In clinical work, estimates of GFR have not shown very good accuracy in relation to possible interventional treatment. Also, there is no good clinical method to easily measure single-kidney or regional RBF. We can obtain the effective renal plasma flow (ERPF) by infusing p-aminohippuric acid and measuring the urine and plasma concentrations, but this method only gives the total ERPF for both kidneys. An alternative is a magnetic resonance (MR) based method, which is problematic in patients with chronic kidney disease, because the contrast agent gadolinium is contraindicated in these subjects [32]. The present PET-related methodology may provide quantitative estimate of regional RBF, and be clinically applicable under conditions such as chronic allograft nephropathy and acute kidney insufficiency. The procedure – as presented here – still involves a small degree of invasiveness because of blood sampling. However, many noninvasive methods for estimating input functions have been proposed [35, 2325, 33, 34], and their implementation will allow RBF to be determined in a fully noninvasive fashion, particularly for clinical purposes.

In conclusion, although some issues remain to be investigated, this study shows the feasibility of measurement of RBF using PET with H2 15O.