Introduction

Iodine-124 positron emission tomography (124I PET) lately has emerged as an attractive modality for lesion dosimetry prior to (131I) radioiodine therapy of differentiated thyroid cancer (DTC) [16]. For accurate dosimetry, the absolute activity concentration in the lesion must be measured reliably. Correction must be made for the partial recovery effect, namely, every PET scanner’s systematic underestimation of activity in lesions smaller than two to three times the scanner resolution [79]. To accomplish such correction, one must determine both in two-dimensional (2D) and three-dimensional (3D) modes the absolute recovery coefficient of 124I. This value is calculated by dividing the activity concentration of (test) objects of various sizes as determined by the scanner by the activity concentration of the same objects measured by, for instance, a well counter.

To optimize the accuracy of PET with 124I, the complex decay scheme of this radionuclide also must be taken into account. This decay scheme includes a cascade gamma radiation coincidence, which, for instance, increases spurious activity and decreases image contrast of 124I PET relative to PET using the pure positron emitter, fluorine-18 (18F) [10, 11]. Only a limited number of phantom studies [1114] on different PET scanners were performed to establish 124I quantitative imaging. To our knowledge, none of these studies has reported absolute rather than relative recovery coefficients and none has used 18F as a gold standard for accuracy.

We therefore performed the present prospective study to determine the absolute recovery coefficients of the University of Duisburg-Essen Nuclear Medicine Clinic’s stand-alone PET and combined PET/computed tomography (CT) systems using 124I. To delineate the effect of cascade gamma rays, we also scanned our phantoms using 18F as a reference. Moreover, we validated the resultant recovery correction coefficient under typical conditions observed in DTC patients by performing three consecutive scans under our clinical whole-body protocol. The prospective study unexpectedly observed a significant underestimation of absolute 18F activity in objects far larger than the threshold size for the partial recovery effect. To assess whether this error was institution- or system-specific, we performed an additional exploratory study with the same model of scanner in another institution as well as a scanner from another manufacturer. In the present publication, we report the results of both the prospective and exploratory studies.

Materials and methods

Figure 1 illustrates the design of our prospective and exploratory studies. Methodological details are given below.

Fig. 1
figure 1

Illustration of the study design. The PET systems examined were located at the universities in Essen (E), Bad Oeynhausen (BO), and Tübingen (T), Germany. The study comprised prospective and explanatory parts using phantoms with different objects filled with 124I or 18F as reference

Radionuclides

18F was obtained from a Cyclone 18/9 (IBA, Louvain-la-Neuve, Belgium) according to standard methodology. 124I was produced in the CV 28 cyclotron (Cyclotron Corporation, Berkeley, CA, USA) and purified under methods described in detail elsewhere [15, 16].

Calibrated activity measurement—dose calibrator and well counter

Using the dose calibrator, high activities of 124I or 18F were measured using a CRC®-15R dose calibrator (Capintec Inc., Ramsey, NJ, USA). The samples were placed in a P6 vial (Amersham International, Buckinghamshire, UK) containing 1 ml of radioactive solution. For 124I, the dose calibrator was calibrated by cross-validation of a sample activity with the measurements of the Physikalisch-Technische Bundesanstalt (PTB), the institution in Germany responsible for national radiation measurement standards. Additionally, both the 124I and 18F activities were checked via gamma spectroscopy (Genie 2000 software, Canberra, Australia).

The well counter (1480 Wallac Wizard; Perkin-Elmer, Wellesley, MA, USA) was calibrated based on the activity measurements displayed by the CRC®-15R dose calibrator. The well counter’s linearity (count rate vs. activity) was checked for the activity ranges to be used in the phantom measurements. Under conditions identical with those of the calibration measurements, the well counter was used to measure small activities, such as aliquots obtained from the spheres placed inside the phantoms or from phantom cavities. These samples were placed in a 2-ml vial containing 1 ml of radioactive solution.

PET tomographs

For the prospective study (Fig. 1), PET imaging was performed using two (ECAT) EXACT HR+ PET scanners at the University of Duisburg-Essen Nuclear Medicine Clinic: one a stand-alone system (CTI/Siemens, Erlangen, Germany), which images in both 2D and 3D modes [17], and the other the PET component of a combined PET/CT scanner, the BIOGRAPH EMOTION DUO (Siemens Medical Solutions, Hoffman Estates, IL, USA), which lacks septa and therefore images in 3D only [18].

The exploratory study (Fig. 1) used a second stand-alone EXACT HR+ with both 2D and 3D imaging capabilities from the Institute of Radiology, Nuclear Medicine and Molecular Imaging of the Heart and Diabetes Center of Nordrhein-Westphalia, Bad Oeynhausen, Germany. Additionally, the exploratory study employed a stand-alone PET scanner of another model and manufacturer (GE ADVANCE; General Electric Medical Systems, Milwaukee, WI, USA) from the University of Tübingen Nuclear Medicine Clinic, which images in both 2D and 3D modes [19, 20].

The EXACT HR+ PET system consisted of 32 rings (63 image planes with a plane separation of 2.43 mm). The energy window ranged from 350 to 650 keV (2D and 3D mode). The GE ADVANE PET system consisted of 18 rings (35 image planes with a plane separation of 4.25 mm). The energy window ranged from 300 to 650 keV in 2D and 375 to 650 keV in 3D mode. The axial field-of-view (FOV) was 15.2 cm for both the EXACT HR+ and GE ADVANCE systems.

CT tomograph and scanning parameters

The CT component of the BIOGRAPH combined PET/CT scanner was a dual-slice spiral CT scanner. All CT scans were performed using a 130 kVp tube voltage, 130 mAs, a 3 mm per rotation table speed, and a 3-mm slice width, except that an 8 mm per rotation table speed and a 5-mm slice width were used in the scans performed to validate our recovery correction methods under typical clinical conditions.

PET tomograph calibration

Because of the unexpected result in the prospective study, the PET calibration will be described thoroughly. Before any phantom measurements were performed, the PET scanners underwent a complete normalization including gantry setup to equalize the detector photomultiplier tube gains and energy thresholds, adjustment for relative efficiency variations, and calibration to enable quantitative measurements. The branching ratios for the PET tracers used were 0.97 (18F) and 0.23 (124I) [21]. Cross-calibrations between each PET system and the dose calibrator or well counter were performed using 18F. PET systems were calibrated in 2D or 3D mode using a standard cylindrical reference phantom (20 cm axial length, 20-cm outside diameter, 6.265 l inside volume measured by fluid displacement). This phantom was filled with 18F activities according to the manufacturer’s recommendations. The activity concentration was determined using the dose calibrator and aliquots of the phantom contents in the well counter. The phantom was placed on the dedicated holder provided by the manufacturer and located centrally within the FOV.

A “cold” transmission scan (acquisition time 120 min) was acquired on the stand-alone PET systems for attenuation correction using cold water (no radioactivity) in the phantom during transmission measurements. The attenuation correction of the BIOGRAPH PET system was based on the CT images; therefore, the measurements started with a spiral CT scan of the cylinder when this phantom already was filled with the 18F activity. The emission scans were acquired for 120 min in 2D mode and 60 min in 3D mode. The emission and transmission images were reconstructed with standard filtered backprojection (FBP) with a 5-mm Gaussian filter. The measured attenuation map after segmentation was used for attenuation correction. The analysis was based on circular regions of interest (ROIs) drawn centrally over the transversal plane with a 17-cm diameter. Only a distance of about ±6.5 cm in the axial direction from the middle of the phantom was considered in the determination of the average activity concentration. The PET calibration factors were adjusted appropriately.

There were two items worth mentioning. First, selecting a circular ROI that equalled the actual diameter of the cylinder phantom (20 cm) resulted in 5% increase of the calibration factor, which was caused by edge effects (intensity drop at the edges). Second, it would seem suitable to perform cross-calibration using the non-pure positron emitter 124I instead of the pure positron emitter 18F. The resulting calibration factors would be specific for the radionuclide used. Use of 124I for cross-calibration probably gives rise to a significant amount of spurious activity introduced by cascade gamma rays that is not removed in the standard software packages of PET systems suppliers, leading to a seemingly increased PET activity in the cylinder phantom, particularly in 3D mode. In the present study, measurements on the PET systems calibrated with 18F corroborated this suggestion. Filling the cylinder phantom with known prepared activity concentration of 124I resulted in an overestimation of the measured 124I activity concentration of 8% in 3D mode; however, the measured activity was accurate in 2D mode (1%).

Phantoms and their preparation

Resolution phantom

For the phantom for measuring the transversal spatial resolution, transversal (in-plane) spatial resolution (x, y) in water was measured in two directions (radially and tangentially) using a line source orthogonal to the transverse plane that was embedded in the cylinder phantom (20-cm axial length, 20-cm outside diameter) filled with cold water. The line source, consisting of polyethylene tubing (0.8-mm inner diameter, 1.6-mm outer diameter), was looped back through the phantom to provide distances of 1 cm and 7 cm from the central axis of the FOV. The transversal spatial resolution was measured at four positions each at the center (z=0) and one-fourth of the axial FOV (z=±1/4 FOV), i.e., (x=1 cm, y=0), (x=7 cm, y=0), (x=0, y=1 cm) and (x=0, y=7 cm). The line source was filled with either 124I or 18F. These measurements were performed on the Essen EXACT HR+ and BIOGRAPH PET systems only.

Recovery phantoms

The National Electrical Manufacturers Association (NEMA)/International Electrotechnical Commission (IEC) 2001 body phantom (elliptically shaped torso phantom) without a lung insert [22, 23] and a cylinder phantom (20 cm in axial length, 20-cm outside diameter) were used as recovery phantoms. For the recovery measurements, both these phantoms contained six refillable spheres of different sizes. The inner diameters (volumes) of the spheres were: 9.7 mm (0.47 ml), 12.6 mm (1.05 ml), 17.1 mm (2.60 ml), 22.2 mm (5.75 ml), 28.0 mm (11.44 ml), and 37.0 mm (26.53 ml).

The spheres in the body phantom were filled from a stock solution of 124I or 18F to measure the “hot” spot recovery coefficients (radioactivity-containing spheres in a cold background) for the Essen EXACT HR+ and BIOGRAPH PET systems. In the present paper, the activity concentration measured using the dose calibrator is referred to as the “prepared” activity concentration, whereas the activity concentration obtained from the PET image is termed the “measured” activity concentration. The prepared 18F sphere activity concentration was about 80 kBq/ml in 2D and 40 kBq/ml in 3D mode; the prepared 124I sphere activity concentration was about 60 kBq/ml in both modes. The phantom cavities were filled with water only.

The spheres in the cylinder phantom were filled with 18F only to measure the hot spot recovery coefficients for not only the Essen EXACT HR+ and BIOGRAPH scanners (prospective study), but for another EXACT HR+ located at a different institution in Bad Oeynhausen (BO) and for a PET system from another manufacturer, namely the GE ADVANCE (exploratory study) located in Tübingen (T). The prepared 18F sphere activity concentration was about 80 kBq/ml in 2D mode and 40 kBq/ml in 3D mode.

Validation phantom

To validate the recovery correction method that was derived in the prospective study, the body phantom was used with 124I-containing spheres only to mimic in vivo conditions seen in DTC patients. Validation was performed with the Essen EXACT HR+ and BIOGRAPH PET systems only. Based on our standard administered 124I activity of about 30 MBq [6] to patients and the 70 kg weight of a standard adult human, the torso activity concentration ought to be about 0.43 kBq/ml, assuming homogeneous activity distribution and a tissue density of 1 g/ml. For the spheres a concentration of about 50 kBq/ml [6] was used that falls within the concentration range of 124I-accumulating lesions at 24 h. This concentration coincided with the NEMA protocol for 18F. In the phantom measurement, the prepared activity concentrations were 43.12 kBq/ml for the sphere and 0.44 kBq/ml for the torso. The resulting prepared source-to-background ratio was therefore 98 as directly checked with the well counter using aliquots obtained from spheres and the phantom cavity, i.e., nearly pure hot spot imaging.

PET(/CT) acquisition protocol

For the acquisition for transversal spatial resolution determination, the phantoms were placed centrally within the FOV. For the Essen EXACT HR+ stand-alone PET, a cold 60-min transmission scan was acquired before the line source were filled with radioactivity. BIOGRAPH measurements started with a spiral CT scan with the activity already filled in the line source. The emission scans lasted 2 h each in 2D and 3D mode.

For the acquisition for recovery coefficient determination, the acquisitions were performed with high statistical precision. The acquisition parameters for the body and cylinder phantoms were identical. For the three stand-alone PETs (Essen and Bad Oeynhausen EXACT HR+ scanners, GE ADVANCE scanner), a cold 180-min transmission scan was acquired before the spheres were filled with radioactivity. The total 68Ge/68Ga rod source activity was 240 MBq (EXACT HR+) or 550 MBq (GE ADVANCE) at the time of measurement. Two emission scans each were acquired successively in 2D and 3D modes. The 18F emission scans lasted 2 h in 2D and 1 h in 3D mode. The 124I emission scans lasted 3 h in both modes. BIOGRAPH measurements started with a spiral CT scan with the activity already filled into the spheres. For the EXACT HR+ component of the BIOGRAPH, 18F emission scans lasted 2 h and 124I emission scans lasted 3 h. In addition, the PET calibration factors on the scanners were checked with subsequent measurements using a standard reference cylinder provided by the manufacturer that was homogeneously filled with 18F.

The phantom acquisition for validation of the recovery correction method was identical with our typical whole-body PET(/CT) patient measurement. On the Essen EXACT HR+ PET system, the durations of the emission (transmission) scans were 480 s (240 s) in 2D mode and 281 s (138 s) in 3D mode. BIOGRAPH measurements started with a whole-body spiral CT scan. After completion of this scan, the bed was advanced automatically to the PET and one-bed emission began. The emission scan lasted 300 s. Three consecutive scans were performed.

Image reconstruction

The phantom images were reconstructed using the typical patient’s iterative image reconstruction algorithm. The image reconstruction was performed after Fourier rebinning with attenuation-weighted ordered-subset expectation maximization (AW-OSEM) using two iterations, eight subsets, and a 5-mm Gaussian filter. The data were corrected for attenuation by measured 68Ge/68Ga transmission including image segmentation in PET, while PET/CT used the CT-based attenuation correction approach. Standard scatter correction in 2D and 3D modes was implemented in the PET(/CT) system. The resulting PET images of the body and cylinder phantoms had 128×128 voxels, and those of the transversal spatial resolution phantom had 512×512 voxels. The voxel size of the images of the body and cylinder phantom was 1.72×1.72×2.43 mm3 (EXACT HR+ and BIOGRAPH) or 2.34×2.34×4.25 mm3 (GE ADVANCE). The voxel size of the images of the transversal spatial resolution phantom was 0.43×0.43×2.43 mm3 for the Essen EXACT HR+ and BIOGRAPH PET systems.

The above OSEM 2×8 reconstruction with image segmentation was used as the default. To evaluate the specific degree to which differences in the recovery coefficients and transversal spatial resolution were related to image reconstruction, additional analysis was performed. Images were reconstructed using OSEM reconstructions with various numbers of iterations and subsets (OSEM 1×32 and OSEM 4×16) and standard FBP. The images were smoothed with a 5-mm Gaussian filter. For recovery curves, this additional analysis was performed for the BIOGRAPH PET system only. For the transversal spatial resolution, the analysis was performed on both the Essen EXACT HR+ and BIOGRAPH PET systems.

Images were reconstructed with the software packages ECAT 7.2.1 (Essen and Bad Oeynhausen EXACT HR+), Syngo Somaris/5, VA40C (BIOGRAPH), and ADVANCE release 6.0 (GE ADVANCE).

Data analysis

Transversal spatial resolution

The image spatial resolution was expressed as full width at half maximum (FWHM). The width of the reconstructed line spread function in the tangential and radial directions was measured for each line source position. These line profiles were fitted with a Gaussian function. The average value (n=12) in the tangential and radial directions obtained at transversal (1 cm and 7 cm) and axial positions (center and ±1/4 FOV) are given. Images and data were analyzed with PMOD 2.5 software (PMOD Technologies Ltd., Zurich, Switzerland) and PeakFit 4.0 (Jandel Scientific, San Rafael, CA, USA).

Activity concentration

The activity concentration was measured using circular ROIs with an 8-mm diameter (FWHM-sized ROIs) as suggested by others [9, 23]. These ROIs were positioned at the apparent center of the sphere, and the mean of the voxels in the FWHM-sized ROI was calculated. The transversal planes were used for analysis. For spheres with an inner diameter >17.1 mm, the average ROI value was calculated from adjacent planes (±1). The concentrations of spheres with an inner diameter ≤17.1 mm were obtained using a line profile (the mean of the FWHM-sized ROIs vs. axial distance) around the maximum. The line profile was fitted with a Gaussian function and the Gaussian amplitude was taken as the activity concentration. Images were analyzed with PMOD 2.5 software.

Recovery coefficient and error analysis

The absolute measured recovery coefficients were directly calculated using the measured activity concentration of the FWHM-sized ROI of 8 mm (see above) divided by the prepared activity concentration as determined by the well counter. The resulting recovery curves were parameterized using a three-parameter sigmoid function: \( RC{\left( d \right)} = \alpha {\left[ {1 + {\left( {\gamma \mathord{\left/ {\vphantom {\gamma d}} \right. \kern-\nulldelimiterspace} d} \right)}^{\beta } } \right]}^{{ - 1}} \). The symbol RC indicates the size-dependent recovery coefficient, d is the diameter of the sphere, and α, β, γ are fitting parameters.

Different sources of error occurred in the determination of the recovery coefficients. First, the estimated relative standard deviation (RSD), RSD ROI, of the mean value in the FWHM-sized ROI (each containing n=15 voxels) including ROI placement was larger for the smallest spheres than for the largest sphere (≈7% vs. ≈1%). Second, the uncertainty of cross calibration was another source of error. PET calibration factors were checked after each recovery measurement. The error of the cross-calibration was estimated by nine calibration measurements. The mean relative deviation between the measured and prepared activity concentrations in the cylinder was 0.8% and the corresponding RSD Cross was ±2.2%. Third, the last error was associated with the activity concentration measurement of the aliquots of the phantom spheres in the well counter. The mean relative deviation between the dose calibrator and well counter estimated by repeated measurements (n=9) was 0.9% and the corresponding RSD Counter was ±1.0%. The total error estimate, ΔRC, according to Gaussian error propagation law was \( \Delta RC = RC \cdot {\sqrt {RSD^{2}_{{ROI}} + RSD^{2}_{{Cross}} + RSD^{2}_{{Counter}} } } \). In the figures, the uncertainties are displayed as vertical error bars around each data point.

Recovery correction method

For a large source-to-background ratio, i.e., nearly pure hot spot imaging, the corrected activity concentration, C cor, can be estimated with C cor = C msd /RC [24]. The measured activity concentration, C msd , was obtained using circular ROIs with a diameter of 8 mm (FWHM-sized ROI). The procedure was identical with the determination of the recovery coefficient as described above. The torso activity concentration (background) was determined using circular ROIs with a 16-cm diameter drawn on planes on which the spheres were not visible, but excluding the first and last five planes and using additional circular ROIs with a 4-cm diameter located in the center of the torso phantom cavity containing the spheres. The mean of the 53 ROIs was taken as the torso activity concentration. For each whole-body scan, the corrected activity concentration of each sphere was determined using the recovery coefficients obtained from the parameterized recovery curve. The individual percentage deviations between the recovery-corrected and prepared activity concentrations (physical decay corrected to the first scan) were calculated for each sphere.

Statistical analysis

The errors of the measured quantities were assumed to be normally distributed. The two-sided Z value for a difference was used to compare recovery coefficients, FWHM values, and source-to-background ratios. For all statistical analyses, a P value <0.05 was considered statistically significant (i.e., a Z value equal to 1.96 or 95% confidence interval).

Results

Transversal spatial resolution

Table 1 lists the results of the transversal spatial resolution measurements, which represent the transversal image resolution in clinical applications. No significant differences were noted between results obtained with the two acquisition modes or image reconstruction methods. The mean ± standard deviation (SD) FWHM was 8.3 ± 0.2 mm for 124I and 7.4 ± 0.2 mm for 18F; these values differed significantly (P < 0.002). A small degree of resolution degradation of 1 mm was noted for 124I vs. 18F.

Table 1 Average transversal spatial resolution (FWHM* in mm, SD  ±0.2 mm) in water under different image reconstruction methods and acquisition modes for 124I (18F)

Recovery coefficients

The absolute hot spot recovery coefficients of the body phantom are shown in Figs. 2, 3, and 4 (Essen EXACT HR+ and BIOGRAPH). Table 2 summarizes the recovery coefficients of the 37-mm spheres as an example of a large object.

Fig. 2
figure 2

Absolute recovery coefficients of 124I and 18F for EXACT HR+ PET system located in Essen (E) measured with a body phantom containing radionuclide-filled spheres, acquired in 2D (a) or 3D (b) modes. Analysis was based on 8-mm FWHM-sized ROIs. Image reconstruction was performed after Fourier rebinning (AW-OSEM at two iterations and eight subsets with a 5-mm Gaussian filter). The sizes of the error bars indicate the uncertainty (see text). The data points were fitted using a three-parameter sigmoid function (solid line)

Fig. 3
figure 3

Absolute recovery coefficients of 124I for the BIOGRAPH PET system measured with a body phantom containing radionuclide-filled spheres, acquired in 3D mode: (a) OSEM 2×8, (b) OSEM 1×32, (c) OSEM 4×16, (d) FBP. Methodology is described in the caption of Fig. 2

Fig. 4
figure 4

Absolute recovery coefficients of 18F for the BIOGRAPH PET system measured with a body phantom containing radionuclide-filled spheres, acquired in 3D mode: (a) OSEM 2×8, (b) OSEM 1×32, (c) OSEM 4×16, (d) FBP. Methodology is described in the caption of Fig. 2

Table 2 Absolute hot spot recovery coefficients (in %) for 18F (124I) for an object (37-mm sphere) far larger than the threshold size for partial recovery effect*

Recovery was lower with 124I than with 18F. Based on the individual values of the spheres, the mean ± SD decreases relative to 18F were −12.7±4.6% in 2D, −12.4±4.8% in 3D (Essen EXACT HR+), and −13.8±5.6% in 3D (BIOGRAPH). The measured 124I recovery coefficients for the 37-mm sphere were 81% in 2D and 75% in 3D mode (Essen EXACT HR+) and 77% in 3D mode (BIOGRAPH). Hence, the 124I recovery for large objects was higher in 2D than in 3D mode; this difference was statistically significant (P=0.03).

The recovery curves obtained from images reconstructed with various OSEM parameters and FBP are depicted in Fig. 3 for 124I and Fig. 4 for 18F. Recovery coefficients were larger for the OSEM- than for the FBP-reconstructed images, except in the case of the largest (37-mm) sphere, for which the recovery coefficients were identical using both algorithms (maximum deviation from FBP: ±1.5%). Based on the individual values for the spheres with diameters <37 mm, the mean ± SD observed recovery increase for OSEM relative to FBP was 11.4±5.6% for 18F and 7.5±4.2% for 124I. For most spheres, recovery values obtained within OSEM were similar regardless of the iteration and subset parameters. However, for the 22.2-mm and 28.0-mm spheres, recovery values appeared be about 5% larger for OSEM 2×8 than for OSEM 1×32 or OSEM 4×16.

The observed 18F recovery coefficients of the body phantom for large objects were 90% in 2D and 87% in 3D mode (Essen EXACT HR+) and 89% in 3D mode (BIOGRAPH). The differences in 2D and 3D mode were not significant, but the measured 18F recovery coefficient for the 37-mm sphere significantly deviated from full recovery (P<0.001) regardless of the image reconstruction method used. Because of this unexpected finding, the scatter geometry of the phantom was changed through the substitution of a cylinder phantom for the body phantom [9]. Identical spheres were used with either phantom. Figure 5 shows the resulting 18F recovery coefficients. These recovery coefficients did not significantly differ from those obtained using the body phantom (Table 2). With the cylinder phantom, full recovery again was not reached (P < 0.001).

Fig. 5
figure 5

Absolute recovery coefficients of 18F for the EXACT HR+ in 2D (a) and 3D modes (b) and BIOGRAPH PET systems (c) both located in Essen (E) measured with a cylinder phantom. Methodology is described in the caption of Fig. 2

We therefore performed an exploratory study measuring 18F recovery coefficients using the cylinder phantom and another EXACT HR+ PET system located in Bad Oeynhausen, Germany, or a PET system from a different manufacturer (GE ADVANCE) operated in Tübingen, Germany. The default OSEM reconstruction was used for the EXACT HR+ PET system. However, the data from the GE ADVANCE were reconstructed using OSEM 4×16 because the regions reconstructed with OSEM 2×8 using GE ADVANCE software had not reached the convergence compared with its FBP-reconstructed images. This effect was attributed to the manufacturer-specific implementation of the OSEM algorithm. The recovery curves from the exploratory study are shown in Figs. 6 and 7, and the measured recovery coefficients for the 37-mm sphere are listed in Table 2. Almost full recovery was reached for the GE ADVANCE, but again not for the EXACT HR+ PET system (P<0.004).

Fig. 6
figure 6

Absolute recovery coefficients of 18F for another EXACT HR+ located in Bad Oeynhausen (BO) measured with a cylinder phantom, acquired in 2D (a) and 3D modes (b). AW-OSEM 2×8 image reconstruction was used. Further methodological description is given in the caption of Fig. 2

Fig. 7
figure 7

Absolute recovery coefficients of 18F for a GE ADVANCE PET system measured with a cylinder phantom, acquired in 2D (a) and 3D modes (b). OSEM 4×16 image reconstruction was used (see text). Further methodological description is given in the caption of Fig. 2

Validation of recovery correction method

The individual percentage deviations between the recovery-corrected and the prepared activity concentrations are given in Table 3. Using all spheres, the mean relative deviation ± SD between recovery-corrected and prepared sphere concentrations were −0.5%±6.0% in 2D and −3.5%±5.3% in 3D mode for Essen EXACT HR+ and −3.2%±11.0% in 3D mode for the BIOGRAPH PET system. The results of repeatability measurements including recovery correction indicated that, except for the smallest spheres, the range of uncertainty of the mean recovery-corrected activity concentrations was ±10%. The reproducibility of the smallest sphere was inferior. In particular with BIOGRAPH PET imaging, the mean relative deviation was −21%.

Table 3 Validating the recovery correction method using the body phantom prepared under typical conditions observed in lesion 124I PET dosimetry in DTC*

The measured activity concentrations at start time derived from the images using the 37-mm sphere were 35.5 kBq/ml in 2D and 32.9 kBq/ml (Essen EXACT HR+) and 32.1 kBq/ml (BIOGRAPH) in 3D mode. The measured mean torso activity concentrations at start time ± SD was 0.62±0.37 kBq/ml in 2D and 0.74±0.33 kBq/ml in 3D mode for the Essen EXACT HR+ and 0.69±0.20 kBq/ml in 3D mode for the BIOGRAPH PET system. The corresponding measured source-to-background ratios were therefore 57 (EXACT HR+, 2D), 45 (EXACT HR+, 3D), and 47 (BIOGRAPH, 3D). The measured source-to-background ratios significantly deviated from the prepared source-to-background ratio in 3D (P<0.01), but did not in 2D mode. The background increase can be mainly attributed to the presence of spurious activities due to gamma-decay cascades of 124I resulting in decrease of image contrast [11].

Discussion

For an accurate representation of the tracer concentration, the voxel counts in images must be related to the actual concentration via a constant PET calibration factor. Partial volume effects cause this calibration factor to depend on the size and shape of the imaged object: the measured activity concentration is underestimated for an object (i.e., lesion) with a diameter smaller than two to three times the scanner resolution [79]. Thus, along with image contrast (and therefore tracer specificity), spatial resolution is a major limiting factor for accurate quantification. Nonetheless, partial volume effects can be corrected for using empirically obtained recovery coefficients.

These recovery coefficients are generally derived from experiments with physical objects. These empirical methods mimic the object imaged in clinical practice by using simple geometrical shapes such as cylinder or spheres. In PET tumor imaging such as 18F-fluorodeoxyglucose or 124I PET, it is generally accepted that, in a first approximation, the shape of the lesions can be described by a sphere model and the activity distribution is assumed to be homogenous. Under these assumptions, recovery correction requires one measurement using a phantom with spheres of known diameters and activity concentrations (lesion and background) that closely mimic in vivo conditions [24].

The absolute recovery curves that we obtained for 124I and 18F are shown in Figs. 2, 3, 4, 5, 6, 7. The spatial resolution in PET depends on the positron energy of the isotope of interest [25]. The recovery of the non-pure positron emitter 124I (maximum positron energy: 2.14 MeV) is expected to be inferior to the recovery of the pure positron emitter 18F (maximum positron energy: 0.63 MeV) [14]. In fact, the recovery for EXACT HR+ and BIOGRAPH PET systems was indeed lower for 124I than for 18F. The measured line-spread functions (a common measure of the spatial resolution) were, however, similar for both isotopes. The observed degradation in transversal resolution was only 1 mm (Table 1), which agrees with published data [14]. This finding suggests that the observed degradation of the 124I recovery coefficient is not only associated with the longer range of the more energetic 124I positron relative to that of the 18F positron, but mainly with the complex decay scheme of 124I. 124I emits gamma rays that are in cascade with the emitted positron, a phenomenon that often results in spurious coincident photon pairs. The impact of these spurious coincidences is expected to create greater inaccuracy in 3D than in 2D mode if no correction method is applied [10]. Using the standard PET reconstruction software, measurements on the EXACT HR+ and BIOGRAPH PET systems confirmed this expectation: the 124I recovery coefficients were slightly inferior in 3D vs. 2D mode (Fig. 2).

The recovery coefficient of the pure positron emitter 18F should converge to 100% for objects larger than two to three times the scanner resolution and decrease significantly for objects smaller than this [79]. As shown in Figs. 2, 3, 4, 5, 6, 7, the asymptotic behavior of the recovery curves is clearly visible and the recovery coefficients significantly decrease for spheres with diameters ≤22.2 mm (about two to three times the scanner resolution) for both 18F and 124I. The curvature of the recovery curves is therefore in good agreement with theoretical considerations and experiments [79]. However, the absolute value of the 18F recovery coefficients for the 37-mm sphere as an example of a large object (four to five times the scanner resolution) significantly deviated from 100% for both the body and cylinder phantoms on EXACT HR+ operated at two different institutions and BIOGRAPH PET/CT system (Table 2). This unexpected result for 18F was still observed after re-calculating the recovery coefficient using the rescaled 18F activity concentration of the reference cylinder (a reference for a large object) as recommended by other authors [9, 24]. Our results cannot be compared with those of other working groups [17, 28], as these investigators defined the relative recovery coefficients of the 37-mm sphere to have 100% recovery.

A possible explanation of the observed bias in 18F for large objects in the present study might relate to the number of (effective) iterations in the image reconstruction. However, calculated 18F recovery coefficients of the 37-mm sphere were independent of the image reconstruction method. The various image reconstruction methods resulted in identical absolute recovery values for the largest sphere. In agreement with the literature [26, 27], the recovery coefficient decreased faster with decreasing sphere diameter for FBP-reconstructed images (Figs. 3 and 4) because the measured activity concentration is generally underestimated compared to the concentration obtained with OSEM-reconstructed images.

This finding is indicative of a system-inherent bias. 18F recovery measurement was performed on a different PET scanner. The GE ADVANCE PET system also contained bismuth germanate detectors. As shown in Fig. 7, the absolute recovery value for the largest sphere did not significantly deviate from 100%, that is, full recovery was reached, suggesting that the discrepancy indeed was system-specific. Since the GE ADVANCE PET system did not show the bias in 18F, one has to conclude that the described procedure has to be done on every PET system with which one intends to do quantitative imaging.

The observed system-inherent bias in 18F (deviation of ≈10%) may be associated with the observation of Lubberink and colleagues [28] that the EXACT HR+ PET scanner is not linear. These investigators measured the hot spot recovery coefficient (HSRC) and cold spot recovery coefficient (CSRC) and used the theoretical relationship, HSRC =1-CSRC, for a system that responds linearly [24]. Lubberink and colleagues [28] showed with 18F that the measured HSRC and CSRC deviated from the theoretical relationship. The authors of the present study also performed this test of system linearity using images of the body phantom that were reconstructed via FBP algorithm (linear operation). We substantiated the non-linear behavior of both EXACT HR+ scanners and of the BIOGRAPH PET system used in the present study (data not shown). Thus, this non-linear behavior is an indication that corrections such as those for attenuation and, especially, scatter do not perform accurately. Further investigations are necessary to elucidate the main cause of the system non-linearity of response and to clarify the source of errors for the observed bias in 18F recovery.

Another issue in the present study was to validate the accuracy of the recovery correction method including reproducibility measurements under clinical imaging conditions typically found in 124I PET of DTC patients. Using a GE ADVANCE PET scanner, Eschmann and colleagues [2] reported for 124I phantom measurements in 2D mode an inaccuracy of about 10%. Similar inaccuracies were found in the present study. Specifically, the maximum uncertainty range of about ±10% was found for spheres ≥12.6 mm in diameter for the EXACT HR+ and BIOGRAPH PET systems (Table 3). Obviously, the limitation of the recovery correction method applied under clinical conditions was reached for the smallest sphere (9.7 mm) that is in agreement with published data [9]. More specifically, even if the volume of the lesion can be determined very accurately, the applicability of the recovery correction method is limited for lesions <12.6 mm (≈1.5 times the scanner resolution).

Partial volume correction based on measured recovery coefficients is possible for simple geometric objects like spheres and for diameters larger than 1.5 times the scanner resolution. Such correction is, however, non-trivial to perform for objects that, for example, significantly deviate from spherical shape or possess inhomogeneous activity distribution. However, recovery correction as currently applied [24] is nonetheless a reasonable way to a more accurate quantification compared to that obtained with gamma-camera based dosimetry [29] and is in most cases acceptable to estimate the absorbed dose of subsequent 131I therapy using serial 124I PET(/CT) imaging such as in thyroid cancer.

Conclusion

In 124I PET dosimetry, recovery correction must be applied even to large lesions, but is a reasonable way to a more accurate estimation of the absorbed dose of 131I radioiodine therapy to metastases of thyroid cancer than that obtained using gamma cameras. Recovery correction is effective in lesions as small as approximately 1.5 times the FWHM (in our case, ≥12.6 mm). Using the EXACT HR+ PET scanner, an unidentified source of bias in 18F quantification was demonstrated and appeared to be system-specific. Therefore, to ensure accurate quantification, it is strongly recommended to determine the absolute recovery coefficients of each radionuclide and PET system used.