Introduction

The pediatric obesity epidemic and rising rates of metabolic syndrome [1] are critical health issues. Metabolic syndrome is a collection of conditions [2,3,4], including insulin resistance, dyslipidemia and cardiovascular disease, and it is associated with non-alcoholic fatty liver disease (NAFLD) [2,3,4,5,6,7,8], which is characterized by hepatic steatosis [7,8,9]. Analysis of body composition, especially the amount, distribution and content of adiposity, is important for studying metabolic syndrome and associated diseases. Fat can be characterized as either white adipose tissue (WAT) or brown adipose tissue (BAT). WAT stores triglycerides while BAT metabolizes fat to produce heat [10]. In the abdomen, WAT is categorized as either visceral adipose tissue (VAT), located inside the abdominal cavity, or subcutaneous adipose tissue (SAT), located outside the abdominal cavity [10]. BAT is generally located in the axillary, supraclavicular, cervical, paravertebral, perirenal and periadrenal regions [10]. Tissue fat content is related to the proportion of different types of fat; higher fat content may indicate “WAT-like” tissue and lower fat content may be associated with “BAT-like” tissue [11,12,13]. Excess VAT can increase the risk for obesity, metabolic syndrome and NAFLD [10, 14, 15], whereas BAT might protect against obesity [10, 12, 13, 16]. In addition, ectopic organ fat, such as hepatic steatosis, is associated with an increased risk for metabolic syndrome [3, 5, 6].

Evidence shows that events during fetal life and early childhood play a central role in the pathogenesis of obesity and metabolic syndrome [17,18,19]. Therefore, knowledge about body composition during infancy is important for understanding the early origins of obesity and metabolic syndrome. Infants born to obese mothers with and without gestational diabetes have increased hepatic fat, a potential biomarker for future metabolic diseases [20, 21]. However, current techniques for analyzing body composition and hepatic fat in infants have limitations. The gold standard for body composition analysis is measurement in cadavers [22, 23]. These studies are time-consuming and complex and do not support longitudinal in vivo monitoring [22]. Computed tomography (CT) and dual-energy X-ray absorptiometry are noninvasive imaging techniques for body composition analysis [14] but involve radiation [22, 24]. While ultrasound is a noninvasive and non-ionizing imaging tool that has been used to measure body composition and hepatic fat, it is operator-dependent, has poor resolution, reproducibility and accuracy [22, 25], and becomes unreliable in deeper regions of the body [22]. The gold standard for quantifying hepatic fat is a liver biopsy [7, 9, 26]; however, biopsies are invasive, limited by sampling bias, and have complications [26].

Magnetic resonance techniques, such as magnetic resonance imaging (MRI) and magnetic resonance spectroscopy (MRS), are noninvasive and can be used for longitudinal studies. MRI and MRS are well suited for children because they do not involve ionizing radiation [22, 27]. MRS is regarded as the noninvasive reference standard for quantifying hepatic fat content in children and adults [28,29,30]. However, MRS techniques have limitations because of partial volume effects, spatial sampling bias, and motion. MRS techniques only measure fat content in a specified voxel and do not provide information about fat distribution. In contrast, MRI has excellent soft-tissue contrast and can obtain high-resolution 3-D images [22, 27]. Intensity-based MRI techniques can provide spatially resolved information about fat distribution and volume in infants [15, 20, 21, 31, 32], but they do not provide information about fat content. Chemical-shift-encoded MRI techniques can provide spatially resolved information about fat distribution, volume and content [11, 32] by quantifying proton-density fat fraction (PDFF), a standardized MRI biomarker [33].

Conventional MRI scans are performed using Cartesian sampling [11,12,13, 15, 16, 20, 21, 31, 32, 34,35,36], which is limited by sensitivity to motion-induced coherent aliasing artifacts that can obstruct the anatomy of interest, degrade image quality in organs affected by motion, and hinder quantification. Therefore, scans using Cartesian trajectories are typically performed while breath-holding. However, infants cannot breath-hold, breathe heavily and irregularly, and might cry during the MRI exam. This leads to motion artifacts and non-diagnostic images [37,38,39,40]. To improve image quality for clinical infant scans, sedation or anesthesia can be prescribed. However, sedation or anesthesia is not feasible in a research setting or for body composition analysis [37,38,39,40]. Because of the lack of age-appropriate technology, a limited number of MRI studies have investigated VAT, SAT, BAT and hepatic fat in infants [11, 12, 15, 16, 20, 21, 31, 34,35,36].

Compared to Cartesian MRI, non-Cartesian MRI based on 3-D stack-of-radial sampling might provide greater robustness to motion because of the dispersed distribution of motion artifacts [41, 42]. The 3-D stack-of-radial trajectory allows for considerably less obtrusive motion artifacts that do not obscure the anatomy of interest, allowing the MRI scan to be performed during free-breathing [42, 43]. A recently developed free-breathing 3-D stack-of-radial MRI fat quantification technique (free-breathing radial) demonstrated agreement with conventional breath-held Cartesian MRI and reference breath-held MRS in a fat phantom and in the abdomen of healthy adults and older children [42, 43]. In our pediatric study, free-breathing radial MRI produced high-quality images and achieved accurate and repeatable hepatic fat quantification compared to conventional breath-held Cartesian MRI and reference breath-held MRS [43]. We believe the capabilities of free-breathing radial MRI can be extended to quantify body composition and hepatic fat in infants. Hence, this study’s purpose was to (1) compare the image quality of free-breathing radial MRI [42] to that of conventional Cartesian MRI in the liver and (2) evaluate the feasibility of the free-breathing radial MRI technique [42] to quantify body composition and hepatic fat in infants.

Materials and methods

Infant study population

This study was approved by our institutional review board. All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and national research committees and with the 1964 Helsinki declaration and its later amendments, or comparable ethical standards.

Parents/legal guardians provided informed consent. Healthy infants ages 1–12 months were eligible for participation. Exclusion criteria included known liver diseases, infections that affect the liver (e.g., toxoplasmosis, rubella, cytomegalovirus, human immunodeficiency virus or herpes simplex virus), major congenital anomalies or diseases that involve the liver, inborn errors of metabolism, and contraindications for MRI.

Infant preparation and magnetic resonance imaging experiments

Preparation and scanning procedures were adapted from previous recommendations [37,38,39,40]. See the Appendix for a detailed description of the preparation and scanning procedures. Preparation time for the MRI scan began when the infant entered the MRI suite and included feeding, swaddling and rocking the infant until he or she became sleepy. Earplugs and earmuffs (MiniMuffs; Natus Medical, Pleasanton, CA) were placed on the infant’s ears for hearing protection (Fig. 1). Next, infants were placed on a vacuum immobilizer (Med-Vac; CFI Medical, Fenton, MI) and clips were fastened on the immobilizer (Fig. 1). Air was not removed from the vacuum immobilizer. The infant was placed on the posterior portion of the head array coil at this time (Fig. 1). Finally, the anterior portion of the head coil was attached (Fig. 1). If the infant cried, the anterior portion of the head coil was removed. After all of the coils were positioned and the infant was calm, the MRI table and infant were advanced into the MRI bore and the total preparation time was recorded. An example of the infant preparation procedures is shown in Fig. 1.

Fig. 1
figure 1

Infant MRI procedures. a–c Initial preparation for the infant scan. a First, feed and swaddle the infant and apply hearing protection. b Next, place the infant on the vacuum immobilizer. c Finally, fasten the vacuum immobilizer clips. Air was not removed from the vacuum immobilizer. d–e Preparation on the MRI table. d Place the infant on the table with the posterior head coil. e Next, add blankets on either side of the infant to prop up the body matrix array. Finally, position the body matrix array and the anterior head coil

After advancing the table and infant into the MRI bore, the start time for the scan was recorded. If the infant cried during the first MRI scan and did not stop crying or the parent/legal guardian requested to stop the scan, the infant was removed from the scanner and comforted in the immobilizer. If the infant settled down, the infant was placed on the MRI table again and set-up procedures on the MRI table were repeated (Fig. 1). The total scan time was calculated after the MRI scan protocol was completed.

The infant MRI experiments were performed on a 3-tesla (T) scanner (Magnetom Skyra or Prisma; Siemens Healthineers, Erlangen, Germany) using a 20-channel head array coil, 32-channel spine array coil and 18-channel body matrix array. A free-breathing radiofrequency-spoiled bipolar multi-echo gradient-echo 3-D stack-of-radial sequence with golden angle ordering [41] and gradient error and eddy current correction [42] (free-breathing radial) was used to acquire abdominal and head and chest scans. The 3-D slabs for the head and chest scan were prescribed to overlap slightly with the abdominal scan for each infant, to cover the body contiguously. Free-breathing radial abdominal scans were repeated back-to-back to assess intra-session repeatability (scan 1 and scan 2). We acquired a commercially available radiofrequency-spoiled bipolar multi-echo gradient-echo 3-D Cartesian sequence [44] with four-fold undersampling and controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) [45] reconstruction (qDixon; The Liver Lab, software version VE11, Siemens Healthineers, Erlangen, Germany). All scans were performed without breath-holding or sedation. Imaging parameters for the free-breathing radial and Cartesian scans are reported in Table 1.

Table 1 Imaging parameters for infant abdominal, and head and chest MRI scans

Magnetic resonance image reconstruction and proton-density fat fraction calculation

Free-breathing radial images and PDFF maps were generated offline in MATLAB version R2017b (MathWorks, Natick, MA). Free-breathing radial images were reconstructed fully sampled (according to the Nyquist criteria) using gradient and eddy current correction [42], linear density compensation, gridding and adaptive coil combination. PDFF maps were determined by fitting the multiple-echo-time images to the gradient-echo signal model with multiple fat peaks [46] to solve for the field map, single effective R2*, water-only, and fat-only maps with complex fitting using a graph cut algorithm [47,48,49]. Magnitude discrimination was used to minimize noise bias to calculate PDFF maps from the water- and fat-only images [50]. Cartesian MR images and PDFF maps were calculated on the scanner using a prototype reconstruction method with the same signal model (prototype version 963, qDixon; The Liver Lab, Software version VE11, Siemens Healthineers, Erlangen, Germany).

Liver image quality analysis

All reconstructed images and PDFF maps were converted to Digital Imaging and Communications in Medicine (DICOM) format for analysis in OsiriX software version 6.0 (Pixmeo SARL, Bernex, Switzerland). Image quality was evaluated by an experienced pediatric radiologist (S.G., >10 years of experience), who was blinded to the trajectory (Cartesian or radial) on a scale of 1–3 for motion artifacts. The criteria for a score of 1 was non-diagnostic images with significant artifacts in the liver (i.e. bad image quality); 2 indicated minor motion artifacts in the liver; 3 indicated no motion artifacts in the liver (i.e. good image quality). We determined the percentage of images that fell into each score category.

Body composition analysis

Fat composition was defined as VAT, SAT and BAT volumes (cm3). Fat content was defined as the PDFF (0–100%). Body composition was measured and analyzed in Horos software version 3 (The Horos Project, horosproject.org). The VAT, SAT and BAT were segmented using the 2-D region growing segmentation and brush tools on the free-breathing radial PDFF maps (all slices in the 3-D volume). VAT was defined as adipose tissue in the intra-abdominal area [27, 32]. SAT was defined as adipose tissue above the muscle fascia and below the skin in the abdomen [27]. VAT and SAT were subjectively delineated on all abdominal axial slices avoiding blood vessels, bones, intramuscular fat and organs [27, 32]. BAT was analyzed on three consecutive reformatted coronal slices of the supraclavicular region [12, 16, 36]. Areas of BAT were defined as any accumulation of adiposity in the supraclavicular area found in the lower neck and the axillary regions [12, 16, 36]. The number of slices and time to contour all the slices to determine VAT, SAT and BAT for each subject was recorded and was reported as median ± interquartile range. All of the segmented areas of VAT, SAT and BAT on the PDFF maps were summed per pixel count to compute the volume and mean PDFF in each subject. All body composition measurements were performed by the same study coordinator (K.V.L.) for consistency. All slices and segmentations were reviewed and confirmed by a pediatric radiologist (S.G., >10 years of experience).

Hepatic fat quantification

Hepatic fat was measured on a single 15mm × 15mm × 15mm region of interest encompassing five 3-mm slices on free-breathing radial PDFF maps. These regions of interest were placed to avoid large blood vessels and bile ducts and in anatomically corresponding regions in the liver to reduce variations in position caused by motion. A pediatric radiologist (S.G., >10 years of experience) confirmed the placement of the regions of interest. The hepatic PDFF determined from each infant was reported individually for free-breathing radial scan 1 and scan 2, and as an average of free-breathing radial scan 1 and scan 2.

Statistical analysis

Descriptive statistics, preparation time and scan time were reported as median ± interquartile range and range (minimum, maximum) or percentage (number). We compared differences in the distribution of liver image-quality scores between free-breathing radial scan 1, free-breathing radial scan 2, and Cartesian scans using McNemar-Bowker statistical tests for dependent categorical variables in STATA software version 12 (StataCorp, College Station, TX). A P-value <0.05 was considered significant. We assessed repeatability of free-breathing radial hepatic PDFF in MATLAB by calculating the mean difference, absolute mean difference, within-subject standard deviation, and coefficient of repeatability [51, 52] between back-to-back repeated scans (scan 1 and scan 2). Free-breathing radial VAT, SAT, BAT (volume and PDFF) and hepatic PDFF were calculated for all subjects and reported as median ± interquartile range and range (minimum, maximum) among subjects.

Results

Infant study population

Ten infants were enrolled in the study, and nine completed the entire study (90% completion rate). One infant completed only one free-breathing radial abdominal MRI scan because of crying. Characteristics for the study population are reported in Table 2. For ten infants, the median ± interquartile range and range (minimum, maximum) preparation time was 32±7 min (15 min, 81 min) and scan time was 24±11 min (15 min, 34 min), respectively (Table 3). Excluding the infant with the incomplete scan, the scan time was 24±11 min (20 min, 34 min). Of the ten infants, four were scanned without the anterior head coil attached.

Table 2 Subject characteristics
Table 3 Infant characteristics and procedure times reported as median ± interquartile range and range (minimum, maximum) (n=10)

Liver image quality

Representative axial and coronal reformat images of the liver with image-quality scores ranging 1–3 are shown in Fig. 2. Free-breathing radial images of the liver demonstrated higher image-quality scores than Cartesian abdominal images (P=0.011). Compared to free-breathing radial scan 1, free-breathing radial scan 2 showed a higher proportion of image-quality scores of 3 (89% vs. 30%, P=0.025). A summary of the proportion of free-breathing radial (scan 1 and scan 2) and Cartesian images receiving each score is shown in Table 4. One hundred percent of Cartesian images received non-diagnostic liver image-quality scores of 1 because of severe motion aliasing artifacts. In contrast, 0% of free-breathing radial (scan 1 and scan 2) abdominal images received liver image-quality scores of 1. Free-breathing radial images had minor motion artifacts that affected the liver less (image-quality score of 2) or no motion artifacts that affected the liver (image-quality score of 3).

Fig. 2
figure 2

Representative Cartesian and free-breathing radial infant abdominal MR images from subject 3, a 6.4-month-old girl. These examples have image-quality scores of 1–3. A score of 1 represents non-diagnostic images (i.e. bad image quality), a score of 2 represents diagnostic images with minor motion artifacts in the liver, and a score of 3 represents diagnostic images with no motion artifacts in the liver. Solid arrows indicate severe motion artifacts (score=1). Dashed arrows indicate small structures with minor motion blurring (score=2). Open arrows indicate small structures that appear sharp (score=3). FB free-breathing

Table 4 Summary of the image-quality scores

Body composition analysis and hepatic PDFF quantification

VAT, SAT and BAT (volume and PDFF) and hepatic PDFF for all infants are listed in Table 5. The median ± interquartile range (minimum, maximum) number of slices used to contour SAT and VAT was 40±0 (32, 44), and in all subjects three slices were used to contour BAT. The median ± interquartile range (minimum, maximum) times to contour VAT, SAT and BAT were 92±57 min (42, 146), 20±8 min (10, 30) and 38±12 min (32, 70), respectively. PDFF maps from the infant with the lowest VAT, VAT-PDFF, SAT, SAT-PDFF, BAT and BAT-PDFF are shown in Fig. 3. This was a 2.5-month-old preterm boy born with intrauterine growth restriction. The hepatic PDFF for this boy (3.8%) was close to the median hepatic PDFF (3.4%). In contrast, PDFF maps from the infant with the highest hepatic PDFF (6.0%) are shown in Fig. 4. This 3.2-month-old infant had a family history notable for non-alcoholic fatty liver disease. Hepatic PDFF quantification using free-breathing radial abdominal scans demonstrated repeatability with mean difference = 0.21%, absolute mean difference = 0.56%, within-technique standard deviation = 0.69%, and coefficient of repeatability = 1.90%.

Table 5 Infant hepatic fat and body composition measurements from free-breathing radial scans for the ten infant subjects
Fig. 3
figure 3

Representative infant abdominal, and head and chest proton-density fat fraction (PDFF) maps (range: 0–100%) from subject 7, a 2.5-month-old boy. Abdominal PDFF maps are shown in axial and coronal reformat orientations. The head and chest PDFF map is shown in the coronal reformatted orientation. The coronal reformatted image was combined from the slabs acquired during the abdominal scan and the head and chest scan. Hepatic PDFF (HPDFF) in the liver is shown in a representative region of interest. Visceral adipose tissue (VAT) and subcutaneous adipose tissue (SAT) compartments are contoured on axial abdominal images. The brown adipose tissue (BAT) compartment is contoured on head and chest coronal reformatted images. In addition, 3-D rendered views of the abdominal PDFF map are displayed, with the orientations denoted as right (R), left (L), anterior-superior (AS), anterior-inferior (AI), posterior-superior (PS) and posterior-inferior (PI). This boy was born preterm with intrauterine growth restriction. He had the lowest VAT volume (17.0 cm3), VAT-PDFF (34.2%), SAT volume (148.4 cm3), SAT-PDFF (77.8%), BAT volume (0.7 cm3) and BAT-PDFF (17.3%) in the study cohort. FB free-breathing

Fig. 4
figure 4

Representative infant abdominal, and head and chest proton-density fat fraction (PDFF) maps (range: 0–100%) from subject 2, a 3.2-month-old boy. Abdominal PDFF maps are shown in axial and coronal reformat orientations. The head and chest PDFF map is shown in the coronal reformatted orientation. The coronal reformatted image was combined from the slabs acquired during the abdominal scan and the head and chest scan. Hepatic PDFF (HPDFF) in the liver is shown in a representative region of interest. Visceral adipose tissue (VAT) and subcutaneous adipose tissue (SAT) compartments are contoured on axial abdominal images. The brown adipose tissue (BAT) compartment is contoured on head and chest coronal reformatted images. In addition, 3-D rendered views of the abdominal PDFF map are displayed, with the orientations denoted as right (R), left (L), anterior-superior (AS), anterior-inferior (AI), posterior-superior (PS) and posterior-inferior (PI). This boy was born full term and had a family history of non-alcoholic fatty liver disease. He had the highest HPDFF (6.0%), second highest VAT volume (63.5 cm3), second highest BAT volume (2.2 cm3), and third highest BAT-PDFF (36.4%) in the study cohort. FB free-breathing

Discussion

In this study, quantification of body composition and hepatic fat in infants was feasible using free-breathing radial MRI. Moreover, free-breathing radial MRI had higher image quality in the liver compared to Cartesian MRI. Free-breathing radial quantification of hepatic PDFF demonstrated repeatability with coefficient of repeatability <2%. Our technique allows for longitudinal studies and could be used to investigate the pathogenesis of and test interventions for metabolic diseases.

For all infants in our study population, the preparation time was approximately 30 min and scan time was approximately 25 min. The scan times in this study included the acquisition of two free-breathing radial abdominal scans, a free-breathing radial head and chest scan, and an additional Cartesian abdominal scan. For the infant with an incomplete MRI, only one free-breathing radial abdominal scan was acquired and the scan time was 15 min. In this study, all of the free-breathing radial abdominal, and head and chest scans were acquired with oversampling to ensure enough data were collected during the MRI exam. Therefore, the scan time reported in this study was longer than it would have been if the data had been acquired prospectively as fully sampled. If only a single fully sampled free-breathing radial abdominal scan is acquired, the scan time for the MRI exam can be reduced to approximately 15 min. Free-breathing radial imaging did not require sedation or anesthesia, thereby eliminating additional cost and time and potential complications.

Among the ten infants, the median ± interquartile range of hepatic PDFF was 3.4±1.1%. The results using our new free-breathing radial technique are consistent with previously reported hepatic PDFF values in infants [20, 21]. The infant with the highest hepatic PDFF in our cohort (6.0%) also had VAT and BAT volumes and BAT-PDFF that were among the highest compared to other infants. This infant’s family history was notable for non-alcoholic fatty liver disease, and there was a maternal history of insulin resistance. Infants of diabetic mothers are at increased risk for metabolic syndrome [53, 54]. Previous studies reported significant differences in hepatic PDFF between infants born to mothers with gestational diabetes versus infants born to mothers without gestational diabetes. In these studies, hepatic PDFF positively correlated with maternal body mass index [20, 21]. It remains controversial whether NAFLD is an independent risk factor for metabolic syndrome [55]. However, large studies demonstrate an association between NAFLD and metabolic syndrome in children [56]. Further research in a larger cohort of infants will be required to investigate hepatic PDFF in infancy and the relationship between hepatic PDFF, NAFLD and future metabolic syndrome.

The characterization of VAT and BAT are important for improving the understanding and management of metabolic syndrome [10,11,12,13, 15, 20, 36]. Excessive WAT in infancy is linked to later onset of metabolic syndrome [10, 14, 15, 20]. Beyond the newborn period, larger volumes of BAT in infants, children and young adults are associated with a decreased risk of metabolic syndrome [10, 12, 13, 16]. Furthermore, it has been speculated that lower BAT fat content (PDFF) in infants and children might be associated with a decreased risk for metabolic syndrome later in life [11,12,13]. Interestingly, in our study, an infant who was born preterm with intrauterine growth restriction had the lowest VAT, SAT and BAT (volume and PDFF) compared to the other infants. The clinical significance of these findings remains unknown. Moreover, in this study we did not measure lean body mass, a key component of body composition. Larger increases in lean body mass have been associated with smaller decreases in BAT [36]. Infants born preterm, small for gestational age, large for gestational age, and with intrauterine growth restriction have increased insulin production and decreased insulin sensitivity and lean body mass [17, 57,58,59]. When these infants experience rapid weight gain in early childhood, there is an increased risk for later metabolic syndrome [17, 60, 61]. Additional work is needed to decipher the relationship between maternal, fetal and neonatal body composition (volume and content) and hepatic PDFF with metabolic syndrome.

Previous studies in newborns with a mean age <2 weeks observed abdominal VAT ranging from 11 cm3 to 44 cm3, and SAT ranging from 50 cm3 to 159 cm3 [15, 20, 31]. Compared to our results, VAT and SAT volumes observed in these newborn studies were lower than the volumes we observed in our study (median VAT=52.0 cm3, median SAT=267.7 cm3). This most likely reflects age differences. Other studies in infants with a mean age of approximately 2 months reported total body VAT and SAT measurements. Hence VAT and SAT volumes observed in these studies were higher than the values we obtained because these volumes were measured from total body fat rather than abdominal fat alone [62].

Two-echo and six-echo Cartesian chemical-shift-encoded MRI techniques have been used to quantify fat content [11, 12, 16, 32, 35]. Two-echo techniques do not correct for confounding factors, such as the multi-peak spectrum of fat, and might underestimate PDFF [62]. On the other hand, six-echo techniques correct for confounding factors, including the multi-peak spectrum of fat [62]. BAT-PDFF and SAT-PDFF reported in our study were similar to those in studies of infants with a mean age of 1–2 months [11, 12, 16]. Of these studies, two used a two-echo chemical-shift-encoded MRI technique [12, 16] and did not correct for confounding factors, which might underestimate PDFF [62]. Our study, along with another study [11], used a six-echo chemical-shift-encoded MRI technique that corrects for confounding factors. Determining fat content for body composition by accurate quantification of PDFF might be clinically relevant. Evidence suggests that a lower BAT-PDFF during infancy is associated with a lower risk of metabolic syndrome [11, 13]. Lower PDFF might indicate increased metabolism of fat to produce heat, a process known as non-shivering thermogenesis that is important for preventing hypothermia [10,11,12,13]. Fat with a higher PDFF might function more like WAT, while fat with a lower PDFF might function more like BAT [11,12,13]. Our PDFF results are consistent with these studies; in our study median VAT-PDFF was higher than median BAT-PDFF (42.2% versus 26.1%).

Compared to free-breathing radial scan 1, free-breathing radial scan 2 showed a higher proportion of liver image-quality scores of 3 (89% vs. 30%, P=0.025). These differences could be a result of more motion, irregular breathing or heavy breathing during free-breathing radial scan 1 compared to scan 2. Free-breathing radial imaging was typically performed first during the MRI exam. At the beginning of the scan, we observed that infants were often startled and cried. We believe this is most likely secondary to the noises generated by the MRI. Soon after, infants usually stopped crying, fell asleep, and moved less as the exam progressed. Hence during free-breathing radial scan 2, infants might have fallen into a deeper sleep and moved less compared to free-breathing radial scan 1. In this study, no self-navigation, respiratory gating or motion compensation was performed. Free-breathing radial imaging can be used to generate a self-navigation signal, which could be incorporated in future work to further improve image quality [63,64,65].

The practicality of infant MRI scanning should be considered. First, MRI is expensive compared to other imaging modalities. This limits access to infant MRI for research and clinical purposes. Second, MRI is loud because of the rapidly changing gradients [66]. To reduce noise and avoid hearing damage during the MRI exam, earplugs and earmuffs are required. Third, MRI can produce heating in the subject, characterized in terms of the specific absorption rate [66]. To reduce the specific absorption rate, we used a low flip angle of 5° for the free-breathing radial and Cartesian acquisitions to produce a low whole-body specific absorption rate of 0.08–0.10 W/kg and B1+rms of 0.6 μT. Finally, abdominal body composition analysis is time-consuming and requires training. In this study, the median time for contouring VAT, SAT and BAT was approximately 90 min, 20 min and 40 min, respectively.

This study has limitations. First, the sample size was small and the range of hepatic PDFF in this population was narrow. Larger sample sizes with a wider range of hepatic PDFF and longitudinal studies are needed to elucidate associations between hepatic PDFF, VAT, SAT and BAT (volume and PDFF), and demographics, clinical data and risk factors for metabolic syndrome. Second, BAT can be present in other regions besides the supraclavicular and axillary regions, such as the cervical, paravertebral, perirenal and periadrenal regions. Because supraclavicular fat is considered the largest depot of brown fat in infants [67], we chose to measure the supraclavicular along with the axillary region for BAT. We were unable to validate that all fat in the supraclavicular region was BAT. To do this, a biopsy or positron emission tomography (PET), which uses radiation [22, 67], would be required. Because this technique was evaluated in a research setting, these confirmatory procedures were not possible. In previous work, autopsies have confirmed that voxels with intermediate quantitative MRI PDFF values reflect BAT [67]. Third, we did not compare our hepatic fat results to the gold standard, liver biopsy, or a breath-held Cartesian scan. Performing a liver biopsy for research purposes in healthy infants is unethical. Breath-held MRI scans were not possible in this study because infants cannot voluntarily perform a breath-hold and sedation was not used. However, we assessed the repeatability of free-breathing radial imaging for hepatic fat quantification and our technique demonstrated repeatability with coefficient of repeatability <2%. Furthermore, previous studies in a fat phantom, adults and children have demonstrated agreement in quantitative PDFF between free-breathing radial MRI, breath-holding Cartesian MRI and breath-holding MRS [42, 43]. Fourth, we did not analyze repeatability for abdominal body composition because of the lengthy analysis time. Head and chest scans were not repeated because of scan time considerations. However all of our measurements were consistently performed by one coordinator and verified by a radiologist. Fifth, preparation for infant MRI can be time-consuming and requires staff training and parental education. We demonstrated that with training the median preparation time and scan time for an infant MRI exam were approximately 32 min and 24 min, respectively. Finally, this study required the infant to be asleep or calm during the MRI exam. If the infant began to cry during the MRI exam, the exam could be incomplete. While our new free-breathing radial technique is robust to breathing motion, additional corrections might be needed for severe motion. To improve the motion robustness of our free-breathing radial technique, motion detection using a self-navigation signal might be used to characterize and compensate for severe motion [63,64,65], which could be explored in future studies.

Conclusion

Free-breathing 3-D quantification of infant body composition and hepatic fat is feasible using a free-breathing radial MRI technique within a median scan time of 24 min and might be reduced to within 15 min if only a single free-breathing radial abdominal scan is required. Free-breathing radial MRI demonstrated improved liver image quality compared to conventional Cartesian MRI and quantified hepatic PDFF with repeatability of mean difference <0.25% and coefficient of repeatability <2%. The new free-breathing radial MRI technique might help unravel the early origins of metabolic diseases by providing accurate and detailed information about body composition and hepatic fat in infants.