Introduction

Congenital heart defects are the most frequently encountered congenital disorders. [1] The incidence of congenital heart defects reported in literature varies between 4 and 50 cases per 1,000 live births depending on the definition used [2]. Natural survival to adulthood for moderate to severe congenital heart defects – i.e. conduits, transposition of the great arteries, Tetralogy of Fallot or univentricular hearts – is less than 50%. Due to advances in cardiac surgery, survival has increased considerably during the past decades and >96% of patients with congenital heart defects – surviving their first year – will be alive at 16 years of age [2, 3]. Despite these advances, residual lesions requiring reoperation and complications like ventricular dysfunction and arrhythmia are prevalent in this population and require follow-up over time. Another important group in pediatric cardiology are patients with ventricular dysfunction due to a cardiomyopathy – classified as dilated, restrictive, hypertrophic, arrythomogenic or unclassified cardiomyopathy – with a cumulative incidence of about 1.24 per 100,000 children younger than 10 years of age [4, 5]. The increasing number of childhood cancer survivors, estimated to be one in every 540 adults in the United States, adds to this group due to anthracycline-induced cardiotoxicity leading to left ventricular dysfunction and – in a subgroup – to dilated or restrictive cardiomyopathy [6].

In patients with congenital heart defects and cardiomyopathies, cardiac imaging is important to diagnose the presence and extent of cardiac and extracardiac pathology, to determine the optimal treatment strategy and to provide the referring physician with a detailed overview of pre-procedural intra- and extracardiac anatomy. Although echocardiography remains the first-line imaging modality in pediatric patients, cardiac magnetic resonance (MR) imaging has become the most important subsequent imaging modality for follow-up of children with cardiac disease. The major advantages of cardiac magnetic resonance imaging over other imaging modalities such as cardiac catherization and CT are that cardiac magnetic resonance imaging is noninvasive, it avoids application of ionizing radiation and can provide physiological, functional and anatomical information, all within a single examination.

Challenges of cardiac magnetic resonance imaging in patients with congenital heart disease

Obtaining high-quality cardiac MR imaging examinations in children has inherent challenges. The anatomical structures that have to be visualized are smaller, heart rates are higher and the ability to perform long and repeated breath-holds is limited. This requires the cardiac MR imaging acquisitions to be of both higher spatial and temporal resolution than in adults, with short imaging times and preferably without sequential breath-holds. As spatial resolution, acquisition time and signal-to-noise ratio are inherently dependent on each other, higher spatial resolutions and shorter imaging times will lead to decreased signal-to-noise ratio (Fig. 1). Fortunately, advances in cardiac MR imaging hardware, acquisition techniques and image reconstruction have mitigated this trade-off enabling high-quality imaging at short acquisition times. This review aims to provide an overview of the advances in cardiac MR imaging hardware and acquisition techniques that are important for imaging pediatric and adult patients with congenital heart disease and discusses relevant advances in cardiac MR imaging techniques used to assess function, anatomy, flow and tissue characterization (Table 1).

Fig. 1
figure 1

MR imaging constraints. With MR imaging there is an inherent trade-off in each imaging protocol between spatial and temporal resolution, acquisition time and signal-to-noise ratio. For example, increasing spatial and temporal resolution will always lead to decreased signal-to-noise ratio or longer acquisition times when all other sequence parameters are kept equal. Improving one factor will therefore always negatively influence the other unless additional adaptations are made

Table 1 Summary of new sequences in cardiac magnetic resonance imaging

Technical advances

Hardware

The most obvious change in MR hardware has been the introduction of higher field strengths greater than 1.5 T. Many facilities worldwide have installed 3-T field-strength magnets for use in cardiovascular imaging as all major vendors now offer large field-of-view capabilities with advanced transmit and receive coil arrays. Imaging at 3 T has the advantage of an intrinsically higher signal-to-noise ratio, which ameliorates the trade-off between higher spatial resolution, faster acquisitions and high signal-to-noise ratio [7]. However, imaging at higher field strengths is also associated with increased image artifacts due to inhomogeneties in the main magnetic field (B0) and nonuniform radiofrequency excitation (B1). Advanced radiofrequency shimming algorithms as well as adiabatic excitations have been developed to address these problems [8].

Parallel to the introduction of high field imaging, several other developments in MR hardware have greatly impacted cardiac magnetic resonance imaging in the past decades. Improved gradient performance has enabled much shorter repetition times (TR) and image acquisition duration [9, 10]. Further improvements in gradient performance, however, are limited by physiological and safety constraints such as radiofrequency-induced heating and peripheral nerve stimulation [11, 12].

The introduction of multi-array, highly sensitive receiver coils is another strategy that enables accelerated acquisition, while maintaining adequate signal-to-noise ratio, by using parallel imaging techniques. With parallel imaging, a reduced dataset in the phase encoding directions of k-space is acquired by using multiple coils elements simultaneously for signal reception [13, 14]. Sensitivity encoding (SENSE) and generalized autocalibrating partially parallel acquisition are the parallel imaging techniques most frequently used. These techniques are, however, fundamentally different. Generalized autocalibrating partially parallel acquisition acquires reduced data in the frequency domain – before Fourier transformation – and sensitivity encoding in the imaging domain – i.e. after Fourier transformation. The factor by which acquisition duration is reduced is roughly equal to the parallel imaging factor, depending on which parallel imaging algorithm is used. Because undersampling is inherent in parallel imaging strategies it is important to obtain a sensitivity profile of each coil element in reference to a sensitivity map of the body coil, to avoid reconstruction artifacts. In the generalized autocalibrating partially parallel acquisition algorithm, this calibration step is integrated into the acquisition by acquiring several additional k-space lines, whereas in the sensitivity encoding algorithm the calibration is performed in a separate reference acquisition.

A final advance in MR hardware has been the introduction of small microchips that enable conversion of the analog signal picked up by the receiver coil to a digital signal in the receiver coil itself. The major advantage of this technology is the preservation of signal-to-noise, which is otherwise reduced by having to travel through receiver cables [15].

Undersampled acquisition

At present, image acceleration due to advances in hardware is limited to approximately a factor of about 10, but advances in pulse sequence design have enabled further gains in imaging speed [16]. Undersampling is the most frequently applied strategy; the basic principle is that k-space is undersampled using different sampling strategies or data sharing – also called view-sharing – exploiting spatial redundancy, temporal redundancy or both. By acquiring only part of the k-space the acquisition is accelerated. Variable-density techniques are commonly used for undersampling k space. The most often used strategy is sampling the center more frequently than the peripheral k-space. As only part of k-space is filled, missing data have to be reconstructed using specific, optimized algorithms to obtain accurate and artifact-free images. Some of the most frequently used strategies include keyhole-4-D TRAK, BRISK, CURE, TRICKS, TWIST and others (Table 2). For a detailed description of these techniques, we refer to the excellent overview by Tsao and Kozerke [16].

Table 2 Cardiac magnetic resonance imaging acquisition techniques: vendor cardiovascular MR acronyms

A recently introduced method for k-space undersampling known as compressed sensing deserves specific mention. In compressed sensing, k-space is sampled in a near-random fashion, which results in aliasing artifacts presenting as noise rather than discrete “image-like” foldover artifacts. In compressed sensing, iterative reconstruction techniques and various a priori constraints are used to suppress the noise due to aliasing, thus enabling faster and/or higher-resolution MR imaging [17]. In various studies, investigators have demonstrated the synergistic benefits of parallel imaging in combination with compressed sensing in children [18, 19].

Motion compensation

Of particular interest in pediatric patients are techniques to compensate for patient motion. With regard to cardiac imaging, three types of motion can be distinguished: 1) cardiac motion, 2) respiratory motion and 3) bulk motion (e.g., due to the inability of the subject to lie still in one position).

For cardiac imaging, motion-free periods suitable for anatomical imaging are end systole (rest period of approximately 50 ms) or mid-diastole (variable rest period, mainly dependent on heart rate). Advanced MR imaging platforms have the ability to target both of these rest periods selectively for imaging. Note that acquisition duration can be calibrated carefully to the exact rest period duration by varying the number of k-space lines acquired (in increments of 1 TR or several ms). Thus, in contrast to CT imaging where the heart rate often needs to be adapted to the rotation time of the CT scanner by using beta blockers, MR imaging specifically allows for tailoring the acquisition to the patients’ heart rate.

The easiest way to compensate for respiratory motion is breath-holding performed by the patient. In cases where this is not possible, monitoring diaphragmatic motion with dedicated sequences offers a reliable alternative strategy, albeit at the expense of scanning efficiency [20]. Recently, several groups have proposed the idea of self-navigation, an approach whereby information about respiratory motion-induced displacement of the heart is obtained by oversampling the center of k-space relative to the peripheral part [2123]. The main advantage of self-navigation schemes is the much better scan efficiency (up to 100% vs. approximately half this value for conventional respiratory navigator gating). The same principles employed for cardiac and respiratory motion compensation can be applied to correct for bulk motion as well. For detailed information about these techniques, the reader is referred to the excellent review of motion compensation strategies in MRI by van Heeswijk et al. [24].

Anatomy

Black-blood imaging

High-contrast black-blood images are very well suited for morphological assessment of cardiac structures. In many institutions, black-blood imaging with a standard (double inversion recovery) spin-echo pulse sequence is part of the routine imaging protocols. However, conventional black-blood imaging uses single-slice breath-holds, which is time-consuming and less favorable in children. To overcome this, various single-shot techniques, either turbo or fast spin-echo using halfscan (undersampling) with sensitivity encoding or stimulated echo acquisition mode, have been developed and validated against conventional single breath-hold black-blood techniques [25, 26]. The single-shot black-blood techniques have imaging times as short as 180 ms, enabling respiratory navigator gating or single breath-hold multislice black-blood acquisitions, while maintaining high spatial resolutions (Fig. 2) [25].

Fig. 2
figure 2

Black-blood imaging. Black-blood images at the level of the pulmonary bifurcation of a 22-year-old man after arterial switch operation for transposition of the great arteries. Multiple breath-hold (a) and single-shot (b) black-blood images obtained at the level of the ascending aorta both depict a stenosis of the left branch of the pulmonary artery. Multiple breath-hold (a) used a multishot turbo spin-echo sequence with scan parameters: TR/TE: 857/10 ms, flip angle 90°, matrix 212×132, acquired voxel size 1.51×2.52 mm, slice thickness 8 mm, 18 slices and an acquisition duration of 120 s. For the single-shot black-blood image, a single-shot turbo spin-echo sequence was used with the following parameters: TE/TR: 857/50 ms, flip angle 90°, matrix 228×176, acquired voxel size 1.4×1.81 mm, slice thickness 8 mm, 12 slices and an acquisition duration of 11.4 s

Contrast-enhanced MR angiography

Three-dimensional contrast-enhanced MR angiography (MRA) can be considered the state-of-the-art technique for noninvasive assessment of the vasculature. The first report on feasibility of contrast-enhanced MRA in infant-like conditions was by Hernandez et al. [27] in 2001, who simulated infant imaging conditions in small baboons. Multiphase 3-D dynamic MRA was performed using a spoiled gradient recalled-echo sequence during quiet breathing, achieving excellent image quality for the great vessels [27]. However, the imaging times were still long (average of 23 s per phase) and the respiratory movement caused blurring of the images. Shorter acquisition times were achieved using the above discussed view-sharing techniques in combination with various interpolation techniques (i.e. keyhole-4-D TRAK, TWIST and TRICKS), which are very well suited for contrast-enhanced imaging. This permitted higher resolutions and time resolved – but high spatial resolution – 4-D MRA (Table 1), with almost equal signal-to-noise ratio, contrast-to-noise ratio and vessel sharpness for the great vessels as conventional single-phase 3-D MRA (Figs. 3 and 4) [28]. The main advantages over conventional 3-D MRA are that time-resolved MRA resolves the problem of contrast timing and adds diagnostic information such as visualization of intra- and extracardiac shunts in approximately 60% of the patients [2830]. However, some smaller vascular structures might be depicted more clearly with conventional 3-D single-phase MRA [30]. Therefore, both techniques are often used in clinical practice to supplement each other. As breath-hold times limit the spatial resolution that can be obtained for MRA, and motion – either cardiac or respiratory – causes blurring of the MRA images, cardiac and respiratory-gated MRA was developed as described above. In patients with congenital heart disease, ECG- and respiratory-gated high-resolution MRA resulted in significantly improved delineation of (intra) cardiac anatomical structures and image quality, and vessel sharpness was equal to first-pass MRA [18, 30, 31]. However, to achieve high contrast-to-noise ratio during these relatively long acquisitions, two out of three studies used a blood pool contrast agent and the third study – Dabir et al. [31] – used a higher field strength [18, 31, 32].

Fig. 3
figure 3

Four-dimensional MRA of a 21-year-old women with chronic thrombo-embolic pulmonary hypertension shows successive full volume maximum intensity projections of a time-resolved gadolinium-enhanced MRA acquisition using a non-ECG-triggered spoiled gradient-echo sequence with SENSE (TR/TE 3.1/1.14 ms, flip angle 30°, matrix 268×267, acquired voxel size 1.49×1.49×4.6 slices, 60 slices, 12 dynamics, 3.4 s/per dynamic). There is clearly less pulmonary vascular enhancement in the right lung compared to the left lung (arrows)

Fig. 4
figure 4

Gadolinium-enhanced vs. non-contrast-enhanced MRA of a 16-year-old boy after arterial switch operation with a moderately severe left pulmonary artery stenosis. a Left anterior oblique maximum intensity projection of a dynamic gadolinium-enhanced MRA using a non-ECG triggered spoiled gradient-echo sequence (TR/TE 5.2/1.49 ms, flip angle 12°, matrix 192×171, acquired voxel size 0.94 × 0.94 × 2 mm, 60 slices, 2.8 s/per dynamic). b Corresponding non-contrast-enhanced respiratory navigator-gated balanced steady-state free precession MRA (TR/TE 4.9/2.4 ms, flip angle 90°, matrix 280×176, acquired voxel size 1.25×1.25×1.50 mm and 75 slices). Trigger delay was set to the end-diastolic resting period in this patient and acquisition duration at 80 ms per cardiac cycle

Non-contrast-enhanced MRA

Volume-targeted or 3-D whole-heart non-contrast-enhanced (NCE) MRA techniques are well-suited to evaluate coronary anatomy and obstruction [33, 34] or to evaluate the thoracic vasculature in infants and pediatric patients (Figs. 4 and 5). Free-breathing 3-D radial steady-state free precession (SSFP) using respiratory movement correction is currently the method of choice to perform NCE-MRA [22, 33, 35]. To achieve maximum image contrast, a T2-preperation pulse and frequency-selective fat-suppression pulse are used. Recently, Bornert et al. [36] evaluated dual-echo Dixon fat suppression, which further improved image quality by delivering higher image contrast and diminishing ghosting and motion blurring. For optimal image quality, precise cardiac and respiratory gating (±2 mm) tailored to the individual patient is important. This can be hindered by heart rate variability, which decreases image quality and is especially a problem in children [37]. Generally, the imaging window is set to mid-diastolic resting period of the heart, in children – with high heart rates – this period is very short, resulting in motion blurring of the images. To avoid this, individual adaptations of the image window should be considered, in children and infants the end systolic resting period might provide better image quality than the mid-diastolic resting period (Fig. 5).

Fig. 5
figure 5

Three-dimensional whole-heart non-contrast-enhanced (NCE) MRA in a 6-month-old boy with Kawasaki disease and grossly aneurysmatic coronary arteries. a Transverse source image at the level of the aortic root and left main coronary artery shows dilated right, left main, proximal left anterior descending and left circumflex coronary arteries (arrows). b Double oblique multiplanar reformation in left oblique sagittal orientation shows grossly dilated RCA. Note that the proximal right coronary artery has a diameter nearly equal to the aortic root (arrows; TR/TE 4.9/2.4 ms, flip angle 90°, TFE factor 16, matrix 280×176, acquired voxel size 1.25×1.25×1.50 mm and 75 slices). Because of the high heart rate the trigger delay was set to the endsystolic resting period. The acquisition duration was set to 50 ms per cardiac cycle

Function

B-SSFP cine imaging

Assessment of cardiac function using balanced SSFP (b-SSFP) cine imaging is one of the cornerstones of cardiovascular MR. Conventionally, single breath-hold multislice b-SSFP cine images are acquired in short-axis or axial orientation to cover the entire heart (online supplementary video), which is time-consuming and cumbersome. Starting the arrival of different acceleration techniques, single or dual breath-hold 4-D b-SSFP cine sequences have been developed using either highly parallel imaging, sensitivity encoding undersampling or compressed sensing (Fig. 6) [3840]. Unfortunately, 3-D b-SSFP suffers from decreased blood-myocardial contrast and endocardial blurring, leading to underestimation of left and right ventricular volumes compared to conventional multislice techniques. To solve this, the use of intravascular contrast agents for 4-D b-SSFP was introduced, increasing image quality, blood-myocardial contrast and accuracy of volumetric measurements [39, 41]. In small children, the temporal and spatial resolution that can be achieved using 4-D b-SSFP might not be sufficient. To prevent multiple breath-holds in these patients, free-breathing respiratory gated multislice cine b-SSFP – using short echo and repetition time with multiple signal averages – can be used, which proved equally accurate as multislice acquisition for volumetric measurements [42].

Fig. 6
figure 6

Dual breath-hold whole-heart cine. Conventional single breath-hold per slice 2-D cine images (a) and reconstructed dual breath-hold 3-D whole-heart cine images (b) in the short-axis orientation, obtained in a 24-year-old man with repaired Tetralogy of Fallot. Three-dimensional whole-heart cine images (b) can be retrospectively reconstructed in every desired imaging plane. However, due to the isotropic acquisition with much thinner slices compared to conventional cine imaging, there is lower in-plane spatial resolution and lower blood-to-myocardium contrast. Conventional cine images were acquired with a balanced steady-state free precession sequence (a), scan parameters: TE/TR 1.7/3.4 ms, flip angle 60°, matrix 192×183, voxel size 1.25×1.25 mm, slice thickness 8 mm, 30 cardiac phases. For the dual breath-hold whole-heart cine images (b), a balanced steady-state free precession sequence was employed with the following scan parameters: TE/TR 1.4/2.8 ms, flip angle 50°, matrix 144x144x25, voxel size 2.43×2.43×2.5 mm, 30 cardiac phases

Myocardial perfusion cardiovascular magnetic resonance

In adult cardiology, ischemia detection for coronary artery disease is one of the most important diagnostic questions. Myocardial perfusion MR imaging, using first-pass contrast enhancement of the myocardium during stress (with adenosine) and rest, has an excellent sensitivity and specificity to detect coronary artery stenosis (Fig. 7) [4345]. Directly following contrast, infusion images are acquired dynamically using either gradient echo, SSFP with preperation (inversion or saturation) or hybrid echo-planar imaging (EPI)-gradient echo with preparation pulse. Undersampling and parallel imaging have improved ventricular coverage and imaging times; however, specificity at 1.5 T is still hampered by low signal-to-noise ratio, contrast-to-noise ratio and dark-rim artifacts [43, 46]. To improve myocardial perfusion cardiovascular MR, higher field strengths (3 T) might be beneficial, as they improve signal-to-noise ratio and contrast-to-noise ratio and decrease dark-rim artifacts, improving diagnostic accuracy [44, 46]. Indications to perform myocardial perfusion investigations are less widespread in the pediatric population; however, it is of supplementary value in selected patient groups (i.e. aberrant coronary arteries or after reimplantation of the coronary arteries). The few studies performed in children demonstrate that myocardial perfusion imaging is feasible in this patient group and reported no serious adverse events related to adenosine infusion [47, 48].

Fig. 7
figure 7

Myocardial perfusion in a 19-year-old woman with an anomalous left circumflex coronary artery from the pulmonary artery, which was reimplanted in the aortic root at 13 years of age. A routine follow-up nuclear myocardial perfusion scan was inconclusive and an adenosine stress myocardial perfusion cardiac magnetic resonance imaging was performed. Stress perfusion images (a) show a subendocardial perfusion defect in the mid-inferior wall, which is not seen in the matching rest perfusion images (arrows) (b). The late gadolinium enhancement image (c) shows limited subendocardial hyperenhancement in the same area (arrow), suggesting a mid-inferior region with subendocardial ischemia and a limited irreversible component. For myocardial perfusion imaging, a single-shot balanced-fast field echo sequence with 100 ms saturate prepulse was performed directly after administering contrast both at rest and during adenosine stress. Images were taken in short axis (3 slices), 4-chamber and 2-chamber orientation with the following scan parameters: TR/TE 2.6/1.31 ms, flip angle 50°, matrix 128×141, voxel size 2.73×2.86 mm, slice thickness 10 mm (with a gap of 9.05 mm for short-axis slices) and 60 dynamic phases

Flow

Flow-sensitive imaging techniques allow for quantification of flow over time and are therefore useful in quantification of valve insufficiencies or shunt fractions (Qp:Qs) [49, 50]. Most frequently, a velocity-encoded, phase-contrast sequence is used, which is based on the principle that the phase of flowing relative to stationary spins along a magnetic gradient is proportionate to flow velocity. Velocity encoding can determine through-plane flow using phase-encoding or in-plane velocity, using frequency encoding. In case of valvular stenosis, both techniques might be used to obtain the best assessment possible. A major disadvantage of 2-D flow measurements is the inability to take into account the valve displacement during the cardiac cycle; direct measurements will likely underestimate the true regurgitate volumes of atrioventricular valves [51]. Furthermore, flow is measured in a single plane, underestimating eccentric regurgitations. Newer 4-D velocity-encoded phase contrast acquisitions allow for more accurate quantification of flow and regurgitant volumes and are able to visualize 3-D flow patterns (Fig. 8) [52, 53]. Furthermore, 4-D velocity-encoded MR is a free-breathing technique and, using compressed sensing and parallel imaging, high spatial resolutions can be obtained (near isotropic 3 mm3) [52, 53]. The high spatial resolutions even proved sufficient to accurately quantify ventricular volumes in the same 4-D dataset [19]. Furthermore, the desired slice orientation can be determined offline after acquisition. These advantages make 4-D phase contrast acquisitions very valuable in pediatric patients with congenital heart diseases, in whom complex anatomy can make accurate slice orientation difficult. Disadvantages remain the relatively long acquisition times of 5–15 min and the longer postprocessing time.

Fig. 8
figure 8

Four-dimensional velocity-encoded imaging. Visualization of the systolic flow patterns in the left ventricular outflow tract and ascending aorta (a), and of neo-aortic valve regurgitation (b) in a 16-year-old boy after arterial switch operation for transposition of the great arteries. A color scale is used to visualize the flow velocities recorded using 4-D velocity-encoded imaging. Images were acquired using a 3-D gradient echo with Cartesian acquisition with the following scan parameters: TR/TE 7.7/4.2 ms, flip angle 10°, matrix 108×60, acquired voxel size 3.43×3.65×3.50 mm, 18 slices using 30 phases and three directions of flow encoding

Tissue characterization

Late gadolinium enhancement

Imaging of focal fibrosis using late gadolinium enhancement is incorporated into standard imaging protocols of both ischemic and nonischemic cardiomyopathies in many institutions. Currently, 2-D late gadolinium enhancement with (phase-sensitive) inversion-recovery (PSIR) gradient echo sequence still is the reference standard to assess viability of myocardium on cardiac MR imaging (Fig. 9) [5456]. In pediatric patients, 2-D late gadolinium enhancement can be performed using respiratory gated free-breathing, achieving equal imaging quality as repeated breath-holds [57]. However, 3-D acquisition seems preferable as signal-to-noise ratio is improved for late gadolinium enhancement imaging, facilitating increased spatial resolution [58]. Without acceleration, 3-D imaging is time-consuming and artifacts may occur due to changes in inversion time and heart rate during the acquisition, both negatively influencing myocardial nulling. Compressed sensing is one of the techniques proposed to facilitate 3-D late gadolinium enhancement imaging with high isotropic spatial resolution in short acquisition time (Fig. 10). The increased resolution is not only of importance for the pediatric population but was also shown to increase the success of radiofrequency ablation procedures in patients with recurrent ventricular arrhythmia. Adaptations to conventional compressed sensing have been proposed, such as low-dimensional-structure self-learning and thresholding, which reduces blurring artifacts and can achieve high isotropic resolution [59]. To prevent imperfect myocardial nulling due to changes in inversion time, PSIR can be used [55] or acquisition times can be shortened by improving the efficiency of respiratory gating with for example continuously adaptive windowing strategy [60].

Fig. 9
figure 9

Conventional late gadolinium enhancement imaging in a 45-year-old patient after repair for Tetralogy of Fallot with suspected cardiac sarcoidosis. Late gadolinium enhancement imaging was performed in short axis (a) and right ventricular outflow tract view (b) 15 min after administration of 0.2 mmol/kg gadolinium contrast agent. Delayed hyperenhancement can be seen at the sight of the homograft implantation (arrows in a and b). This is a typical finding in patients after homograft implantation and is not suspect for cardiac involvement of sarcoidosis. An inversion-recovery gradient echo was performed using the following imaging parameters: TR/TE 3.5/1.15 ms, flip angle25°, matrix 160×131, acquired voxel size 2x2.44x10 mm, 10 slices

Fig. 10
figure 10

Near-isotropic free-breathing 3-D late gadolinium enhancement sequence. Assessment of myocardial late gadolinium enhancement and presence of thrombus in a 50-year-old uncooperative male stroke patient with severe cardiac arrhythmia. Image acquisition was done during free-breathing and 3 signals averaged to reduce motion artifacts. A 3-D inversion-recovery gradient echo sequence was employed 15 min after administration of 0.2 mmol/kg gadolinium contrast agent. Imaging parameters used were: TR/TE 3.5/1.2 ms, flip angle 25°, matrix 93×87, acquired voxel size 3.43×3.65×5.00 mm, 32 slices. In the 4-chamber orientation (a), a large thrombus is present in the left ventricular apex (marked with T), surrounded by transmurally infarcted myocardium (arrows). Because of the nearly isotropic spatial resolution reformations in the left 2-chamber (b) and short-axis orientations could be made after completion of the acquisition

Myocardial edema

Imaging of acute myocardial edema using cardiac magnetic resonance imaging was described as early as 1991 by Gagliardi et al. [9], demonstrating altered signal intensity using T2-weighted spin-echo sequences in the presence of myocarditis. Short-tau-inversion recovery (STIR) black-blood technique proved superior to T2-weighted spin-echo for musculoskeletal and abdominal MR and was introduced for cardiac imaging 1996 by Simonetta et al. [61]. They demonstrated that STIR can generate myocardial images which are highly sensitive to changes in tissue relaxation times, with minimum flow and motion artifacts (Fig. 11). Although introduced in the 1990s, it was not until 2004 that STIR was validated in a larger group of patients to establish its sensitivity to distinguish acute (myocardial edema) from chronic (fibrosis) myocardial infarction, with success [10]. The same research group tested the diagnostic value of a comprehensive cardiac MR imaging protocol to diagnose myocarditis, using presence of early relative enhancement (gRE), presence of late gadolinium enhancement and high signal intensity – relative to skeletal muscle – on T2-weighted black-blood STIR, as diagnostic criteria [62]. The highest diagnostic accuracy was achieved if all three were combined and ≥2 features were positive, which was later adapted to the Lake Louise consensus criteria for suspected myocarditis on cardiac MR imaging [63, 64]. However, black-blood STIR has some important limitations such as signal drop-out, slow-flow blood artifacts, image quality impairment in tachycardia (or normal high heart rate in children) and lower signal-to-noise ratio. To achieve optimal image contrast and quality STIR sequence, parameters should be chosen carefully with special regard for timing and design of the pulse sequence, quality of the TSE readout pulses, bandwidth and slice thickness. To improve the robustness and clinical applicability of edema imaging on cardiac MR imaging, noncontrast sequences have been developed over the past years. Aletras et al. [65] propose Acquisition for Cardiac Unified T2 Edema (ACUT2E), a combined TSE-SFFP approach to achieve optimal CNR, signal-to-noise ratio and image quality to depict acute myocardial infarction with a reduction of artifacts, although the absolute T2 values are systematically overestimated with this method [65]. Furthermore, noncontrast T1 mapping (also see below) was compared to both conventional black-blood STIR and ACUT2E TSE-SSFP and yielded a higher diagnostic accuracy to detect acute myocardial infarction, even though more regions were affected by artifacts. The higher diagnostic accuracy is likely the result of eliminating the need to compare myocardium to a remote reference region of interest, which is mandatory for both black-blood STIR and ACUT2E [66]. Other advantages of T1 mapping include that it is robust to tachycardia and breath-holds are considerably shorter, making it advantageous for children.

Fig. 11
figure 11

Myocardial edema in myocarditis in a 22-year-old man with suspected perimyocarditis. A T2-weighted short-T1 inversion (STIR) recovery sequence in short-axis orientation reveals subtle diffusely increased signal of the myocardium, which is most clearly seen when myocardial signal intensity is compared to that of skeletal muscle in the same image. The diffusely increased signal intensity indicates that there is diffuse myocardial edema, which is in line with a diagnosis of myocarditis. The following scan parameters were used: TR/TE: 2,182/100 ms, flip angle 90°, matrix 236×175, acquired voxel size 1.36×1.83 mm, slice thickness 8 mm, 15 slices

T1 mapping

After the success of late gadolinium enhancement for imaging of focal fibrosis, much attention was directed to noninvasive quantification of diffuse myocardial fibrosis in the past decade. This process is presumed to be accelerated in many congenital and non-congenital cardiac diseases [67]. Multiple cardiac MR imaging methodologies have been developed to quantify diffuse myocardial fibrosis, using T1 mapping with or without contrast administration. First, a new imaging sequence was developed by Messroghli and colleagues to overcome the limitations of the conventional look-locker approach, the modified look-locker inversion recovery (MOLLI), which was later modified into shortened-MOLLI, both facilitating high-resolution T1 maps of the human myocardium within a single breath-hold [68, 69]. The first clinically applicable cardiac MR imaging method was validated against myocardial biopsy in 2010 by Flett and colleagues [70], using equilibrium-contrast cardiac MR imaging with a standard MOLLI acquired before and after contrast equilibrium. The theory behind this method is that at contrast equilibrium, the extracellular gadolinium concentration is equal in blood and myocardium and the extracellular volume fraction of the myocardium can be determined, which correlates well to collagen volume fraction on biopsy. However, although accurate and highly repeatable, time to contrast equilibrium is relatively long and multiple breath-holds are required making this a complex and time-consuming technique, especially if used in children [70]. Several solutions were developed in the past years to shorten and simplify acquisition. First, to simplify the contrast protocol a bolus-only technique – assuming dynamic equilibrium – was designed. Both contrast protocols – bolus vs. continuous infusion – were compared to each other and to histological results by White and colleagues [71]. The bolus-only proved equally accurate in estimating extracellular volume fraction, provided the total fraction was less than 0.4 [71]. Extent of diffuse myocardial fibrosis, measured by bolus or equilibrium-contrast T1 mapping, has since been linked to decreased functional status and outcome in several pathologies, including congenital heart disease and childhood cancer survivors [7274]. To further simplify the acquisition, a third contrast protocol was investigated by Miller and colleagues [75], using isolated postcontrast myocardial T1 measurement. In their study, isolated post bolus-contrast T1 measurement and extracellular volume fraction estimation using pre- and post-bolus-contrast T1 mapping was compared to histology, they conclude that isolated postcontrast T1 measurement is insufficient to reliably estimate extracellular volume fraction [75]. Finally, a protocol not requiring contrast administration, non-contrast T1 mapping, was used to image diffuse myocardial fibrosis. It was compared to histology in patients with aortic stenosis and, although it only moderately correlated to collagen content measured on histology, T1 mapping was able to differentiate healthy from diseased and was related to the severity of aortic stenosis [76]. Following the rapid innovations in the area of quantification of extracellular volume fraction and T1 mapping, the cardiac MR imaging working group of the European Society of Cardiology and the Society of Cardiovascular Magnetic Resonance recently released a consensus statement. In accordance with the above stated, they observe that extracellular volume fraction measurement appears to be more accurate than isolated postcontrast T1 measurement (Fig. 12). Furthermore, they state that a bolus-only approach with at least 15-min delay is sufficient for most myocardial extracellular volume fraction applications.[77]

Fig. 12
figure 12

T1 mapping and extracellular volume measurement. Examples illustrate extracellular volume maps in a patient with no abnormalities (a) and a patient with dilated cardiomyopathy (b). Note the larger extracellular volume fraction in the patient with dilated cardiomyopathy. Extracellular volume maps were obtained by acquiring two separate T1 maps with the modified look-locker 3–5 sequence: a single-shot balanced FFE readout (TR/TE 3.2/1.17 ms, flip angle 35°, matrix 288×288, acquired voxel 2×2×8 mm, 3 slices, SENSE = 1.5). One map was obtained before contrast injection, and the second was acquired 15 min after bolus injection of 0.2 ml/kg contrast agent

Imaging of stents and intracardiac devices

In patients with congenital heart disease, interventional devices and stents are increasingly used to decrease the need for surgery and increase convenience of the patient. Khan et al. [78] researched the disruption caused by the interventional devices and the influence of type of pulse sequence and field strength. Although signal void in the immediate vicinity of the device was common, significant disruption of surrounding tissue was only seen for the stainless steel Flipper detachable embolization coil. Field strength and pulse sequence type did not significantly influence image disruption or signal void [78]. Nordmeyer et al. [79] investigated the diagnostic accuracy of CT angiography and different cardiac magnetic resonance imaging sequences compared to conventional angiography to assess patency and re-stenosis of nitinol, platinum-iridium and stainless steel stents. Combining high flip angle gradient recalled echo, SSFP and high flip angle MRA resulted in high diagnostic accuracy for nitinol and platinum-iridium stents. For stainless steel stents the diagnostic accuracy compared to angiography was slightly lower due to large penumbra of signal loss [79].

Hybrid cardiac catheterizations, combining radiographic and MR imaging to accurately address clinical questions, are being performed for both diagnostic and interventional purposes. An obvious advantage of MR guiding is decreased radiation exposure, which predisposes to development of cancer, especially when radiation is imposed on smaller infants [80]. Other advantages include improved visualization of soft tissue and cardiac structures and the ability to combine invasive pressure measurements with flow quantification performed on cardiac MR imaging, ensuring accurate assessment of the hemodynamics and pulmonary vascular resistance [81]. The development of catheters is ongoing, currently passive guidewires – not actively conveying signals to the cardiac magnetic resonance imaging scanner – are used for such guided catheterizations, as active wires carry a risk of heating. Recently, a compatible and safe guidewire for MR imaging was developed and the first studies in children are currently being conducted [81].

Conclusion

Patients with congenital heart disease increasingly survive into adulthood and are in need of repeated high-quality cardiac MR imaging throughout life. The technical advances as outlined above enable high-fidelity assessment of anatomical as well as functional parameters and tissue characterization in short imaging times and underscore the preferential role of cardiac MR imaging as comprehensive cardiac imaging technique of first choice.