Introduction

Laser-induced breakdown spectroscopy (LIBS) is a spectroscopic technique that uses a focused, pulsed laser to form a plasma in air as well as in or on liquid and solid surfaces [19]. By analyzing the emission spectra from the LIBS plasma, the atomic composition of an analyte can be determined. No sample preparation is necessary for LIBS, making it a very popular technique in many applications that include environmental monitoring [1012], material analysis [13], forensics [14], characterization of fossils [15], biological identification [12, 16], analysis of works of art [17], and space exploration [18, 19]. Though the applications of LIBS are vast, LIBS suffers a disadvantage in sensitivity when compared to other spectroscopy techniques, such as inductively coupled plasma atomic emission spectroscopy (ICP-AES) [2024].

Double-pulse LIBS (DP-LIBS) is typically used to improve sensitivity compared to single-pulse LIBS (SP-LIBS) [25]. DP-LIBS involves the use of two laser pulses, either collinear or orthogonal, that are separated temporally on the order of nano- or microseconds. An orthogonal configuration uses a non-ablative pulse parallel to the surface after the ablative pulse is fired into the sample [1, 2636]. Using an orthogonal DP-LIBS configuration with a non-ablative pulse fired above the surface followed by an analytic/ablation pulse, Stratis et al. found a 20-fold improvement in LIBS spectral intensity compared to SP-LIBS on glass [34]. Scaffidi et al. used an orthogonal DP-LIBS method with a pre-pulse fired into the sample first [33]. The first laser pulse was generated from a femtosecond pulse-width laser and was followed by a nanosecond pulse-width laser (the analytical pulse) fired above the surface through the ablation plume created from the first pulse. This femtosecond–nanosecond method resulted in significant 30-fold increase in spectral intensity. In a collinear configuration, both laser pulses are fired along the same path, and the LIBS measurement is made from the plasma generated by the second laser pulse. Sattmann et al. used this method, which resulted in twofold increase in spectral intensity and a sevenfold increase in ablation [37]. Both orthogonal and collinear DP-LIBS can improve spectral intensity and signal-to-noise (S/N); however, depending on the available equipment and application, one or the other maybe preferred.

It has been noted in the literature that the mechanisms that lead to signal enhancement in DP-LIBS are not clear [38, 39], and some proposed mechanisms are often contradictory [40]. While the DP-LIBS process is highly complex, there are three factors that appear to be of primary importance in producing signal enhancements: (1) pulse-plasma coupling energetics between first laser-induced plasma (LIP) and the second laser pulse and LIP, (2) sample heating in the context of the plasma produced by the first laser pulse, and (3) local atmospheric pressure and number density effects [39, 40]. A simplified description of the collinear DP-LIBS process starts when the first laser pulse interacts with the sample matrix beginning the sample heating and ablation process. The first LIP then freely expands into the cold environment of the buffer gas [40]. This expansion is similar to that postulated for SP-LIBS that would involve a low-pressure region with forward particle flow, in the direction of the ablating laser pulse that is usually normal to the sample surface [41]. The first stage is followed by a transition regime with momentum transferred to the buffer gas forming a shock wave that would be followed by a diffusion stage at higher gas pressure. The result would be a confined, sphere-like structure for the first LIP or SP-LIP [42]. During the expansion, the first LIP loses some energy in the process of ionizing the buffer gas. The second laser pulse in DP-LIBS interacts with the heated environment of the first LIP and sample created by the first laser pulse. The preheated target surface produced by the first laser pulse and LIP reduces the breakdown threshold as well as the background gas density behind the shock wave of the first LIP [43], which reduces the laser shielding for the second laser pulse [26]. The rarefied gas and higher temperature produced by the first LIP creates an environment that leads to a faster and greater expansion of the plume for DP-LIBS compared to SP-LIBS [36, 40]. Expansion of the second LIP in the rarefied gas created by the first LIP saves some of its energy, leading to an increase of the emission signal. DP-LIBS signal enhancements include increased detection time and emission volume that result from a combination of effects [25, 39].

There also appears to be an agreement that the DP-LIBS mechanisms are dependent on the inter-pulse delay. For example, more efficient ablation is observed for inter-pulse delays less than a few hundred nanoseconds. This is rationalized because of the pre-heated sample and rarefied atmosphere in front of the target created by the first laser mentioned above that reduce the breakdown threshold [38, 40]. However, with longer delays, the sample has time to equilibrate with the environment produced by the first laser pulse, and there is less increase in mass ablation. Galbacs et al. [43] have suggested that this mechanism may last through inter-pulse delays as long as 10 μs. Another plasma condition that varies with inter-pulse delay is the plasma temperature, which is higher with DP-LIBS and leads to longer lifetimes; however, the effect gets averaged out with integration over the whole plasma volume [40] and with longer inter-pulse delays [38]. Thus, longer inter-pulse delay times between development of the first LIP and the second laser pulse tend to negate effects observed with short, nanosecond inter-pulse delays.

In addition to DP-LIBS, atmospheric conditions (i.e., pressure and gas composition) also impact the observed spectral intensity [44, 45]. Gas composition also influences LIBS spectra and can be used to increase signal intensity. Results reported by Iida [45] revealed that that He and Ar atmospheres were both effective for increasing LIBS spectral intensity compared to air at one atmosphere (∼760 Torr). At higher pressures, there are more collisions with the buffer gas, which leads to decreased plasma lifetime, but an enhancement of energy transfer from buffer gas to analyte [46]. Heavier buffer gases lead to slow plume expansion, resulting in shorter plasma length and better plasma confinement [47, 48]. The mass ablation can also be affected by the gas composition. While the ablation rate dropped in the order He > Ne > air > Ar, Sdorra and Niemax [49] reported the signal intensities were found to decrease in the order of Ar > Ne > He. These observations were attributed to stronger plasma excitation in argon. Similar to DP-LIBS, expansion of a laser produced plume in a background, or buffer gas, is also complicated and difficult to model thoroughly [41, 50]. This can be further complicated when changes in pressure are included [46]. Reduced pressures have been used to increase spectral resolution and S/N, but often at the expense of signal intensity [44, 51]. However, Iida [45] found that as pressure was decreased to 100 Torr, LIBS spectral intensity increased for both Ar and air buffer gases. If pressure is decreased too far, then the LIBS intensity tends to erode because of the lack of plasma confinement. At reduced pressures, the plasma expansion is more elongated [52]. The spectral intensity increase at low pressures is likely due to reduced plasma shielding, resulting in more ablation as well as reduced Stark broadening [44]. Similar to DP-LIBS, the mechanisms responsible for the observations related to gas composition and pressure are not fully understood, partly because of the variations in experimental conditions [44].

In this study, a collinear DP-LIBS scheme was used along with manipulation of the atmospheric conditions to investigate the potential for further improvements in sensitivity. Since the second laser pulse in the orthogonal DP-LIBS method is parallel to the surface, it would become progressively more difficult for the second laser pulse to intersect the laser-induce plasma (LIP) created by the first laser pulse because as the atmospheric pressure is reduced, the plume elongates along the path of the first ablative laser pulse [52]; thus, a collinear arrangement was employed in this study. The effect of DP-LIBS was evaluated in Ar, He, and air at pressures ranging from atmospheric pressure to 1 Torr. For evaluation purposes, the DP-LIBS results were compared with those for SP-LIBS experiments under the same atmospheric conditions.

Experimental

The collinear DP-LIBS experimental setup with atmospheric chamber is illustrated in Fig. 1. Two laser pulses, both at 1,064-nm, were combined into the same beam path with a beam splitter. A Continuum Powerlite II Nd:YAG (Santa Clara, CA, USA) supplied the first pulse (45-mJ), and a Continuum Surelite I Nd:YAG (Santa Clara, CA, USA) supplied the second pulse (45-mJ). The timing between the two pulses was controlled with a delay generator and monitored with an oscilloscope from the laser Q-switch. A single-pulse is assumed when the inter-pulse delay is 0-s, which would provide the equivalent total laser energy produced by DP-LIBS (90-mJ). The laser pulses were focused onto a brass sample housed in a controlled atmospheric/vacuum chamber, which was purged with air, Ar, or He. The pressure of the bath gases was varied from atmospheric (630 Torr at an elevation of 1,430 m) to 1 Torr, which was measured using a pressure gauge. To minimize changes related to laser ablation depth, the sample was moved between experiments using an x,y-translation stage.

Fig. 1
figure 1

Experimental setup for collinear DP-LIBS with variable atmospheres

LIBS measurements were taken perpendicularly to the plasma expansion along the direction of the incident laser beam (i.e., perpendicular to the sample surface). The optical collection geometry was chosen for efficient plume coverage, with a field of view (FOV) of up to 1 in. (∼25-mm) from the target surface. To obtain the most representative light collection as possible, the sample desorption location was set at the focal length of the lens to minimize divergence. The collection angle of the fiber optic bundle with a numerical aperture (NA) of 0.22 is 25.4°. A fused silica lens with 2 in. diameter with an 85-mm focal length and NA of 0.299 was used that translates into a total collection angle of 34.8°, which is greater than the acceptance angle of the fiber optic. Therefore, the sample location was set at approximately the mid-point of the lens. This geometry was used to ensure that light from the sample was within the acceptance angle of the fiber optic bundle (lens half-angle 17.4°). Fiber optic was located at approximately the focal length of the lens (85-mm). Because the quartz vacuum window (thickness of 0.105-mm, refractive index 1.4585) bends the light in slightly, the fiber optic location was adjusted to optimize the signal.

The fiber optic was coupled to an Acton SpectraPro® 500i Czerny-Turner spectrometer (Acton, MA, USA) equipped with a Princeton Instruments PI-max 512 × 512 pixel ICCD camera (Trenton, NJ, USA). The spectrograph used an 1,800-g/mm grating for measurements of the 481-nm zinc line and a 150-g/mm grating for broadband measurements. The inter-pulse delay of the DP-LIBS experiment was varied from 0 to 120 μs. The ICCD was operated with a 20-μs gate width and a delay of 1 μs after the firing of the second laser pulse. SP-LIBS experiments were conducted by setting the inter-pulse delay to 0-s so that the energy equivalent to the two laser pulses was preserved [26]. Winspec/32®, provided by Princeton Instruments, was used to control the spectrograph and camera to collect spectra. Unless specified otherwise, reported spectra are accumulations of eight shots. Spectra were exported in ASCII and plotted in MatLab® from MathWorks® (Natick, MA, USA). The peak area was calculated and a baseline was subtracted in MatLab®.

For excitation temperature measurements, the 150-g/mm grating was used to measure the lines 510.6, 515.3, and 521.8 nm from Cu emission. The peak areas from these lines were used to create a Boltzmann plot for each condition using the equation Ii = giAi/λ exp[−Ei/(kTe)]. In the Boltzmann equation, Ii is the line intensity, k is the Boltzmann constant, and gi, Ai, Ei, and λ are spectroscopic constants corresponding to the statistical weight, transitions probability, energy of the excited state, and wavelength, respectively. Spectral constants found in Hafez et al. [46] were used. After rearranging and taking the natural log of the Boltzmann equation, the line intensity and constants were used to create an x,y-plot, from which the excitation temperature, Te, was determined from the slope.

Results and discussion

The complexity of DP-LIBS and variations of LIBS results with gas composition and pressure may partly be due to the experimental configurations and parameters [43, 44]. Bogaerts et al. [38] noted that the trend in plasma temperature for DP-LIBS is generally reported to increase, although some reports show negligible change or even lower temperatures. These types of variation may depend on the experimental configuration as well as the observation times. Because of these complications, this study was conducted to survey the effects that combing DP-LIBS with various gas compositions and pressures would have on signal intensity with the current experimental setup.

A collinear DP-LIBS configuration was used in this experiment due to the simplicity of the experiment and its compatibility with LIBS plume geometry at reduced pressures. As the first laser plasma expands toward the laser source, the second laser pulse will intersect a larger portion of the first laser plasma than if an orthogonal arrangement were used. If the ablative pulse is fired first and the analytic pulse is fired second in an orthogonal configuration, the ablative plume will vary wildly in size and density similar to SP-LIBS work reported by Yalcin et al. [52]. Thus, firing a pre-ablative pulse in an orthogonal DP-LIBS configuration above a surface at reduced pressures may result in no plasma formation because plasma intensity of gases decrease with decreasing pressure in low pressure regimes (less than 400 Torr) [53]. In orthogonal DP-LIBS, the location at which the analytic pulse is fired in the plume would be another variable, making the study difficult to extract meaningful information. Therefore, a collinear arrangement was selected for this study.

SP-LIBS serves as the benchmark against which DP-LIBS results are compared. In the literature, not all DP-LIBS experiments compare SP to DP in the same manner [40]. For an appropriate comparison of collinear DP-LIBS with SP-LIBS, it is important that the laser energy deposition for the two experiments be as similar as possible. When a single laser is used that can provide two pulses separated by a variable delay, the SP-LIBS has been performed either with a single laser pulse equivalent in energy to the two laser pulses [37] or with an inter-pulse delay of 0 s between the pulses [26]. In the present case, two different lasers are used to provide the two pulses for DP-LIBS; therefore, SP-LIBS experiments were performed with an inter-pulse delay of 0 s so that any variation in laser beam quality between the two lasers would be preserved and replicated for SP-LIBS. However, it should be noted that SP-LIBS is not always defined in the same manner in the literature. Sometimes, it is stated that the first laser pulse is equivalent to SP-LIBS or that energy of the DP-LIBS is twice that of SP-LIBS [38, 40]. Therefore, care must be taken to clarify or know what comparison is being made when interpreting DP-LIBS versus SP-LIBS results.

An atmosphere of Ar seems to result in the most intense spectra as seen in Fig. 2 that compares a DP-LIBS spectrum with an 80-μs inter-pulse delay in 630 Torr Ar with a SP-LIBS spectrum in 630 Torr air for a brass sample. The increase in intensity is related to an increase in sensitivity. For example, the ratio between the S/N for the spectrum acquired in 630 Torr Ar with a 80-μs inter-pulse delay and the S/N for 630 Torr air with 0-μs inter-pulse delay is about ∼70. In addition to an increase in intensity of the Cu lines at 510, 515, and 521 nm, the Zn lines also increased. Because Zn is the minor component of the brass sample (∼5%), the changes in the Zn line at 481-nm were used to monitor effects of atmospheric changes, and the experiments were repeated to observe changes in the Cu lines for temperature measurements.

Fig. 2
figure 2

DP-LIBS with a 80-μs inter-pulse delay in 630 Torr Ar compared with SP-LIBS in 630 Torr air

In Fig. 3a, DP-LIBS in 300 Torr air with a 30-μs inter-pulse delay has approximately a fivefold increase in intensity when compared to SP-LIBS in air for the 481-nm Zn line. Figure 3b plots the peak area of the 481-nm Zn line versus inter-pulse delay at different pressures of air. It can be seen in Fig. 3b that the maximum intensity is seen in 300 Torr with an inter-pulse delay of 30-μs. In general, 100, 300, and 630 Torr DP-LIBS is rather effective at increasing the spectral intensity of Zn 481-nm compared to SP-LIBS at these pressures; however, a slight decrease in intensity is seen with increasing inter-pulse at 1 and 10 Torr air. In Fig. 3c, the excitation temperature is plotted versus inter-pulse delay for different pressures with air as the bath gas. It is seen here that the inter-pulse delay and the air pressure have little effect on the excitation temperature, with all conditions varying between 7,500 and 6,300 K.

Fig. 3
figure 3

a DP-LIBS in 300 Torr air with a 30-μs inter-pulse delay compared to SP-LIBS in 630 Torr air. b Effect of DP-LIBS on spectral intensity with increasing inter-pulse delay and at varying air pressures. c Effect of DP-LIBS on excitation temperature with increasing inter-pulse delay and at varying air pressures (SP-LIBS occurs with 0-μs inter-pulse delay)

DP-LIBS in Ar has a similar effect as DP-LIBS in air. In Fig. 4a, DP-LIBS at 630 Torr Ar is compared with SP-LIBS in 630 Torr. DP-LIBS in 630 Torr Ar with an inter-pulse delay of 80 μs is ∼2.5-fold greater than SP-LIBS in 630 Torr Ar. Figure 4b plots the peak area from the 481-nm Zn line versus inter-pulse delay at varying Ar pressures. DP-LIBS has an effect on increasing the spectral intensity in 630, 300, and 100 Torr Ar when compared to SP-LIBS. At low Ar pressures of 1 and 10 Torr, the intensity changes very little as the inter-pulse delay is increased. DP-LIBS in 300 and 630 Torr Ar is nearly as effective and seems to provide the most intense spectra when compared to LIBS for all other gases studied. The excitation temperature plotted versus inter-pulse delay at varying Ar pressures (Fig. 4c) shows little change with increase inter-pulse delay and only a slight change depending on the pressure, ∼9,500 K for 100 Torr Ar and ∼11,000 K for 1 Torr Ar.

Fig. 4
figure 4

a DP-LIBS at 630 Torr Ar with a 80-μs inter-pulse delay compared to SP-LIBS also at 630 Torr Ar. b Effect of DP-LIBS on spectral intensity with increasing inter-pulse delay at varying Ar pressures. c Effect of DP-LIBS on excitation temperature with increasing inter-pulse delay and at varying Ar pressures (SP-LIBS occurs with 0-μs inter-pulse delay)

The results indicate that DP-LIBS on brass in an atmosphere of 630 Torr Ar gives the most intense spectral intensity for the Zn line at 481 nm when compared to spectra acquired in air. At low pressures of 1 and 10 Torr, DP-LIBS resulted in no increase in spectral intensity for either Ar or air compared to SP-LIBS. These results are consistent with those reported by Hafez et al. [46], who looked at SP-LIBS in lower pressures of Ar (1–7 Torr) and observed that signal increases with increasing Ar pressures. While higher pressures result in more collisions with the buffer gas, which decreases plasma lifetime, the observed enhancement was attributed to energy transfer from the buffer gas to analyte. Another reason that Ar provides stronger signal intensities than air may be increased plasma confinement [47], and it is more inert than air [54]. The plasma temperatures for air and Ar are within the range of reported values [25, 26, 37, 55]. The lack of variation with inter-pulse delay is likely due to the long delay times used in the current study.

While the Ar and air data are rather well behaved and consistent with literature reports, the results with a helium buffer gas are somewhat erratic. Results of DP-LIBS in He are different than those observed for DP-LIBS in Ar or air in that DP-LIBS in He did not result in the most intense spectra. Therefore, Fig. 5 compares SP-LIBS in 100 Torr He (which is the most intense in He) with SP-LIBS in 630 Torr He. SP-LIBS in 100 Torr results in a ∼7-fold increase in 481-nm Zn intensity when compared to SP-LIBS at 630 Torr He. As can be seen in Fig. 5b, DP-LIBS seems to have no positive effect on spectral intensity as the inter-pulse delay is increased, except for DP-LIBS at 300 and 630 Torr that showed only a modest increase. DP-LIBS with a 30-μs inter-pulse delay in 100 Torr He results in a significant decrease in intensity and could be the result of the first pulse perturbing the second (analytical) pulse. In a similar result as the other gases, DP-LIBS at low pressure of He (1 and 10 Torr) has nearly no effect on the spectral intensity. Figure 5c shows the excitation temperature measurements versus inter-pulse delay in varying He pressures. The excitation temperature in 1 and 10 Torr He showed very little change as the inter-pulse delay was increased and ranged between 9,800 K and 11,000 K. At 100 Torr He, the temperate was relatively constant, varying only slightly around 17,600 K. At 630 Torr He, the maximum excitation temperature of ∼25,000 K was seen with an inter-pulse delay of 10-μs, while a minimum of temperature of ∼18,500 K was observed with 0-μs inter-pulse delay. At 300 Torr He, the excitation temperature varied wildly and had no clear trend.

Fig. 5
figure 5

a SP-LIBS at 100 Torr compared to SP-LIBS in 630 Torr in He. b Effect of DP-LIBS with increasing inter-pulse delay with varying He pressure. c Effect of DP-LIBS on excitation temperature with increasing inter-pulse delay and at varying He pressures (SP-LIBS occurs with 0-μs inter-pulse delay)

One possible explanation for the He observations would be if the plume expanded out of the FOV, which is important as the pressure is decreased [56]. Because plume expansion at reduced pressures likely becomes less spherical and more ellipsoidal [52], the experimental light collecting arrangement may be more critical to homogeneously capturing the plasma emission. For a pressure of 0.1 mbar (0.075 Torr), Herrera et al. [57] found a radius of 0.24-cm and an axial expansion of 0.29-cm with a delay time of 50-ns. Based on results from Yalcin et al. [52], the plume may expand axially to ∼5-mm; however, they did not use delays as long as those used in the current study. Even though the same experimental setup was used for all three buffer gases, less plume confinement is expected for the lighter weight He buffer gas. This is also important because the temperature in the plume is not homogeneous and varies with time and pressure [48]. Thus, if the FOV were inadequate, then at lower pressures and longer delay times, the plume might extend beyond the FOV such that only the back end of the plume would have been observed and the results would not be representative of the whole plasma. Based on simulations and experiments, Wineford and co-workers [58, 59] suggest that plasma asymmetry can have an effect on the observed emission spectra depending on the light collecting geometry or angle of view. Plume asymmetry (i.e., non-spherical) is expected at reduced pressures [52], but can also occur with DP-LIBS [60]. In the current experimental setup, a lens coupled to a fiber optic was used to collect more of the plume, which can be better than only using a fiber optic [59], to give a FOV of ∼25-mm along the axis of the plasma (i.e., normal to the sample surface). It is conceivable that increased expansion in the radial direction was not adequately captured in the helium studies. Future experimental designs should consider spatially resolved data collection [60] or optical arrangements to collect the light emission more efficiently and homogeneously from all dimensions [61].

Analyte signal intensities in the presence of helium have been reported to be erratic. Henry et al. [62] added helium to air and monitored carbon signal intensities. They observed that the carbon signal initially decreased as helium was added until it reached a minimum with a 75% He concentration. Further addition of He resulted in an increase in signal intensity until a maximum was observed at 100%. However, plots of the peak-to-base (P/B) and S/N versus He concentration revealed monotomic increase for both P/B and S/N. Henry et al. suggest that the presence of He perturbs the plasma conditions, especially decreasing the absorption of laser energy and the electron density. At high concentrations of He, such as those used in the present study, Henry et al. also suggest that metastable helium states could transfer energy to the analyte directly. Therefore, studies using helium as the buffer gas should be explored with additional monitoring for the behavior of He lines.

The variation in ionization potentials between the buffer gases may also contribute to the observations, especially those in helium. Chen et al. [48] noted that plasma expansion is faster with ionization than without. Argon has an ionization potential of 15.76 eV, while He has the highest ionization potential of all the buffer gases at 24.58 eV [47]. For air, ionization potentials of the primary constituents nitrogen (N2) and oxygen (O2) are 15.58 and 12.06 eV [63], respectively. Besides the ionization potential, the molecular dissociation energy should also be considered when comparing SP-LIBS and DP-LIBS in air [40]. Oxygen further complicates issues because of the potential chemical reactivity with the analyte [64]. While the first LIP loses some of its energy to ionizing the buffer gas, the second LIP expands in the rarefied gas of the first LIP, allowing it to conserve its energy and enabling an increase in emission intensity. When the ionization energy of the buffer gas is really high, the plasma cannot ionize the gas: thus, SP- and DP-LIBS spectra have similar intensity, as in the case of helium.

Conclusion

By manipulating the gas composition and pressure along with the use of DP-LIBS, the spectral intensity can be increased. In this study, DP-LIBS in an Ar buffer gas provided the most intense spectrum for the 481-nm line of Zn, which is a minor component in brass, than background gases of air or helium. The results also suggest that the use of a DP-LIBS method in a He environment would provide no benefit; however, it should be noted that the calculated plasma temperatures in He were less consistent than those in Ar or air. In addition, DP-LIBS at reduced pressures less than ∼100 Torr generally had no positive effect with the inter-pulse delays explored in this study. The experimental configuration of these studies involved collecting plasma radiation orthogonally to the expansion of the plume, similar to previous DP-LIBS experiments at atmospheric pressures. However, this configuration may not be the most effective for uniformly collecting light from the plume. Future work should include comparison of the results in this study with a similar study that collects plasma radiation more efficiently and homogeneously.