Abstract
This critical review discusses the results published between 2000 and 2005 on the development of analytical systems based on the luminol chemiluminescent and electrochemiluminescent reactions. An increasing number of non-specific detection systems based on the enhancing, inhibiting or catalysing effect of a large range of compounds have been published. Possible detected compounds and their concomitant presence in samples are discussed. Chemiluminescent and electrochemiluminescent reactions were also found to merge in biochip and microarray development as a possible substitute to the well-established but hardly quantitative fluorescent detections.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Luminescent transitions of excited molecules or atoms to a state of lower energy are characterized by electromagnetic radiation dissipated as photons in the ultraviolet, visible or near-infrared region. These luminescent reactions are classified according to the energy source involved during the excitation step: thus, most classical light emission reactions are referred to as bioluminescence (from in vivo systems), chemiluminescence (from a chemical reaction), electroluminescence (from an electrochemical reaction) and photoluminescence (from UV, visible or near-IR radiations). More trivial reactions are also described as pyroluminescence (from flame-excited metal atoms), radioluminescence (from irradiation by X-rays or γ-rays), sonoluminescence (from ultrasonication of dissolved substance) and thermoluminescence (from solids subjected to mild heating).
Chemiluminescence reactions are generally oxido-reduction processes and the excited compound, that is the reaction product, has a different chemical structure from the initial reactant. Several hundreds of organic and inorganic compounds are at the origin of chemiluminescence reactions which can occur in liquid or solid phases, or at solid–liquid or solid–gas interfaces [1–3].
This critical review will mainly focus on liquid-phase light-emitting reactions based on the oxidation of 5-amino-2,3-dihydrophthalazine-1,4-dione (luminol). Scheme 1 shows the overall reactions in aqueous medium. Luminol oxidation leads to the formation of an aminophthalate ion in an excited state, which emits light when returning to the ground state. The quantum yield of the reaction is low (≈0.01) and the emission spectrum shows a maximum at 425 nm [4].
As suggested above, this reaction can be triggered through a wide range of catalysts, more or less specific for a particular oxidizing species and with varying efficiencies. Peroxidase enzymes, and particularly the one extracted from horseradish, are usually considered as the most effective catalyst, working at pH relatively close to neutral (8–10) and having a quite high specificity for a particular co-oxidant, the hydrogen peroxide. This chemiluminescence production system is the basis of biochemistry-based analytical systems such as biosensors, immunoassays, immunosensors and microarrays. The reaction is then used to reveal any interaction with a peroxidase-labelled entity, or to detect hydrogen peroxide first produced by a specific reaction.
The electrochemical oxidation of luminol is usually considered as the second most efficient way of triggering the reaction. The scheme is similar to the peroxidase-catalysed one with hydrogen peroxide as co-oxidant and a large range of working pH.
Finally, transition metal cations (Co2+, Cu2+, Cr2+, Fe2++, Fe3+, Hg2+, Mn4+, Ni2+) and their complexed forms (e.g. ferrocene, ferricyanide) can be used as catalyst with mitigated performances which are linked to the relatively low signal to noise ratio obtained and the required elevated pH of the reaction.
Applications of these three types of chemiluminescent reactions during the last five years will be presented and discussed. Particular attention will be given to: (i) the detection of compounds through their effect on the chemiluminescence reaction rate, (ii) the use of the chemiluminescent reaction as a label in separation systems, (iii) the coupling of this reaction to other enzymatic reactions and (iv) the on-chip uses on these reactions.
Reaction triggering
Bio-catalysis and chemi-catalysis
As suggested above, the bio-catalysis of the luminol chemiluminescence is considered as the most powerful way of triggering the light emission reaction. The heme-containing proteins, particularly horseradish peroxidases (EC 1.11.1.7), are able to catalyse the chemiluminescent reaction of luminol in the presence of hydrogen peroxide. The use of this enzyme has the advantage over other catalysts that the chemiluminescent reaction can proceed at near-neutral pH values (8–8.5).
In spite of numerous works [5–9], the complex mechanism of the peroxidase-catalysed reaction and the stoichiometry remain hypothetical. Roughly speaking, the reaction sequence leading to light generation can be divided into two main processes (Fig. 1a). First, in the course of a series of enzymatic steps and in the presence of luminol (LH−) and hydrogen peroxide, horseradish peroxidase (HRP) is successively converted into intermediary complexes (complex I and complex II) before being regenerated to free peroxidase. These enzymatic steps produce luminol radicals (L−., LH.) which then enter a complex chemical pathway to finally generate luminol hydroperoxide (LO2H−), the precursor of the light emitter (excited 3-aminophthalate ion). This second part of the mechanism can be slightly modified, in terms of the species involved and their relative concentrations, when the enzyme to hydrogen peroxide molar ratio changes: when this ratio is high, a superoxide pathway is involved; for low ratios the predominant pathway is the diazaquinone route.
This reaction can be used for detecting peroxidase-labelled molecules, hydrogen peroxide generating labelled enzymes and hydrogen peroxide enzymatic precursors, i.e. substrates for oxidase enzymes generating hydrogen peroxide.
When transition metal cations are used to catalyse the reaction, these are involved in the production of luminol radicals, in the same way as HRP is, but under harsher pH conditions. The analytical applications using such catalysts will therefore focus more on the detection of molecules affecting the rate or efficiency of the reaction and the detection of hydrogen peroxide enzymatic precursors.
Electro-catalysis
An original and unusual way to obtain highly sensitive hydrogen peroxide detection is the electrogenerated chemiluminescence or electrochemiluminescence of luminol (ECL). In a mechanistic study of this ECL reaction, Sakura et al. [10] had proposed that luminol was first oxidised at the electrode surface and then reacted, mole to mole, with hydrogen peroxide (Fig. 1b). The theoretical ratio (photon produced)/(H2O2 consumed) is then 1, whereas it is only 0.5 for the peroxidase-catalysed reaction. Moreover, avoiding the use of fragile enzymes for the catalysis of the chemiluminescent reaction could lead to more stable and reproducible sensors. Consequently, regarding the sensitivity of hydrogen peroxide detection, the electrogenerated chemiluminescence of luminol will be more efficient than the peroxidase-catalysed reaction [11]. Because of this high sensitivity for hydrogen peroxide most of the applications of ECL are dedicated to the detection of hydrogen peroxide generating label enzymes and hydrogen peroxide enzymatic precursors. Less often, this electrogenerated reaction is used for the detection of luminol-labelled molecules.
The main competitors to luminol label in the field of analytical chemistry are the ruthenium complexes [12]. Such molecules, which can be regenerated after having emitted their photon(s), appeared to be more appealing electroluminescent labels.
Analytical applications
Direct chemiluminescent detection of compounds in flow injection analysis (FIA)
The last five years have witnessed a drastic increase of the number of publications focussed on the use of chemiluminescence for the flow injection analysis of compounds [13] as diverse as metal ions, pesticides, amino acids and more generally pollutants and toxicants. This trend, mainly owing to the simplicity of development and use of chemiluminescent FIA systems, has led to the description of numerous case studies whose shear volume reveals their lack of specificity [14–32].
Indeed, a large number and type of compounds were tested, according either to their inhibiting, enhancing or catalytic effect on the chemiluminescent reaction. An overall look on this literature evidences quite interesting detection limits, with absolute detected quantities down to the sub-femtomolar level, but also obvious potential cross reactivity problems (Table 1) [33]. The presence of different toxicants for example in human urine will lead to the measurement of a mean signal corresponding to both inhibited and enhanced chemiluminescent reaction. The same remark could be made from water samples that are, as the human fluids, quite complex media obviously containing more than one chemiluminescently active component. Nevertheless, in particular situations—e.g. in process analytics—one and only one particular analyte is expected and needs to be quantified. In these cases, such sensitive analytical systems could then be really helpful.
A solution to overcome these interferences and cross reactivity problems should be found in the use of a separation step such as liquid chromatography [34], high-performance liquid chromatography [35–37] or capillary electrophoresis [34, 38–40] before the detection. The characterization of the retention time for each species present in the samples then leads to more accurate results. Two different uses of these coupled systems are (i) separation and detection through the effect of each separated compound on the CL reaction [36, 37, 39, 40], (ii) labelling of the sample components with luminol before the separation and detection through their emitted chemiluminescence [34, 35]. For the latter purpose, a large range of luminol derivatives (Fig. 2) can be used for the labelling of aminoacids and peptides, ibuprofen, amphetamine and methamphetamine, histamine and much more. The detection limits obtained with these chemiluminescent separative systems are usually in the fmol–pmol range.
An extension of such separation systems is the use of molecularly imprinted polymer recognition elements jointly with chemiluminescent detection [41, 42]. Polymers are synthesised by using a process of template-induced formation of specific molecular recognition sites (binding or catalytic) in a material where the template directs the positioning and orientation of the material’s structural components by a self-assembling mechanism. The binding by the polymer of the target molecule, which has chemiluminescent activity (usually catalytic), will then help to increase the specificity of the analytical systems.
Bio-specific chemiluminescent detection of compounds
Interesting approaches to the specific detection of particular compounds in a complex mixture are based on the sequential use of highly specific enzymatic reactions and chemiluminescent detection. Chemiluminescent systems requiring hydrogen peroxide for the light emission (Fig. 1) are of particular interest because in addition to the determination of H2O2, several other compounds can be analysed by coupling with H2O2-generating enzymatic reactions [43–49]. This makes it possible to detect specific compounds, including e.g. glucose, lactate, urate, cholesterol, xanthine, choline, in complex media, taking advantage of the specificity of the oxidase enzyme used.
For this type of biosensor, FIA fibre optic sensors associated with membrane- or polymer-based immobilized enzymes (Fig. 3a) have been extensively used; however, these appear to be declining in popularity (according to the number of publications) since the introduction of miniaturisation. Nevertheless, the immobilization of both the oxidase enzyme and the peroxidase in millimetre-sized sensing layers, in contact with a fibre optic, has led to the achievement of sensitive, specific and stable analytical systems [3, 50, 51]. Interesting results were also obtained based on ferricyanide or electrochemically catalysed chemiluminescence (Fig. 3b). The typical detection performances of such macrosystems are presented in Table 2. They exhibit satisfactory detection limits and ranges and are in most cases validated in complex samples such as human sera. These systems benefit from the high sensitivity of the light measurement systems such as photomultipler tubes. More applications of these coupled reactions are detailed in the following section, which presents the up-to-date miniaturized chemiluminescent microarrays and biochips systems.
When the compound of interest is not a substrate for oxidase enzymes or could not be converted into such a substrate, the chemiluminescence reaction can still be used in a biospecific system through the use of immunoreactions with peroxidase- or luminol-labelled antibodies. In such cases the specifically detected target can be pesticides [52], toxins [53], explosives [54], detergent [55], proteins [56, 57] or living cells [58] (Table 3).
Peroxidase-labelled antibodies are the most widely used chemiluminescent tools in the bio-analytical field. Indeed, these protein complexes are used for immunoassay on standard 96-well microplates, for immunosensor detection and for on-chip immunodetection (see below).
Antibodies labelled with luminol or luminol derivatives [59] have exclusively been used as a label for electrochemiluminescent systems. Only a few works with such a label have been [59, 60], mainly because of the difficulties encountered when attempting to achieve its attachment through standard chemical reactions compatible with retaining the protein integrity. Indeed, the luminol molecule has an aromatic amine as the only available functional group and this is not easily covalently linked to biomolecules. Luminol derivatives with more reactive functional groups, such as N-(4-aminobutyl)-N-ethylisoluminol (ABEI), have also been tested as labels [61] but these have shown lower light emission properties when grafted directly to proteins.
Bio-specific on-chip chemiluminescent detection of compounds
One of the major advances in chemiluminescent analytical tools during the last five years has been the use of these light-emitting reactions in miniaturized systems such as biochips and microarrays. Thus, the sensing layers' multiplexing has led to the achievement of multi-parameter systems with interesting quantitative possibilities when compared to the equivalent fluorescent systems.
Passive systems for immunodetection (Table 4), i.e. where the solid phase only provides an immobilization support, have been described for the quantitative detection of multiple toxins [62], antibiotics [63] or allergen-specific antibodies [64]. An example of such systems is presented in Fig. 4. A homogeneous immobilization chemistry was performed on the entire surface of the support (microscope glass slide) and different proteins spotted to generate the array [63]. The low light background of the charge coupled device (CCD) system and the bright light emission of the peroxidase labelling system (see Fig. 4 micrographs) enabled the quantitative detection of the different molecules (11 antibiotics) concomitantly.
More elaborate systems involving heterogeneous immobilization chemistry based on latex bead deposition [65, 66] have also been described. These systems use a chemiluminescent detection and exhibited interesting analytical performances for the detection of HIV-1-specific antibodies or p53 gene sequence mutations. In these cases, specific immobilization chemistry could be performed according to the molecule to be immobilized and furthermore the low background signal of the CCD imaging system led to sensitive quantification.
Similar work with passive supports but for the multiparametric detection of DNA sequences were also published recently [67]. Analysis of influenza A viruses (H1N1, H3N2, H1N2, and H5N1) and influenza B viruses showed that chemiluminescent microarray-based methods were capable of rapid and unambiguous identification of all types and subtypes of viruses by use of random PCR products. DNA sequences were then labelled with peroxidase through biotin–streptavidin adducts.
Interesting results were also obtained, for DNA sequence detection, by directly using the CCD sensor as an immobilization substrate [68]. Thus, the photons emitted by the peroxidase-labelled DNA were detected with high sensitivity through the CCD array (Fig. 5) and led to really interesting performances with low-pM detection limits for target sequences.
Micro-chemiluminescent biosensors and flow injection biosensors (μFIA) were also described based on the HRP-catalysed reaction [69] or electrochemically triggered reactions (Fig. 2b) [70–72]. Figure 6 presents one of these ECL micro-biosensors based on the use of screen-printed electrode arrays. This active support acts as immobilization substrate and ECL triggering material, and enabled the detection of different compounds (glucose, lactate and choline) concomitantly and with good sensitivities (Table 4). The micrographs in Fig. 6 present the typical electrochemiluminescent images obtained, which are characterised by good spatial resolution and no interference problem between the different sensing elements.
Conclusions
Luminol chemiluminescence reactions are very attractive, mainly owing to their ease of use, high sensitivity and simplicity of the instrumentation used for the technique. During the last five years, a large number of studies have been reported on the use of non-specific chemiluminescent reactions for the detection of an unexpectedly large number of compounds. As has been discussed, such a large number of potential activators, inhibitors and catalysers evidences obvious lacks of specificity of the developed direct detection systems. More specific results were obtained by coupling these non-specific reactions with separation steps such as chromatography or electrophoresis systems. Major advances were also obtained through the miniaturization and multiplexing of chemiluminescent analytical systems. Thus, micro-biosensors and arrayed immunoassays have achieved performances similar to those of the macro systems developed during the late 1990s.
Chemiluminescent reactions therefore appeared to be good candidates for biochip developments albeit with some potential restrictions. The chemiluminescent image presented in Fig. 7 evidences a possible problem of the chemiluminescent detection on microarrays. Indeed, for a constant spot diameter of 150 μm, increasing the density of peroxidase within the spot generates an increasing emission area. Such phenomenon, probably due to the lateral diffusion of luminol radical or excited 3-aminophthalate, will therefore impose a limit on the achievable spot density that could be detected through peroxidase-catalysed chemiluminescence [65]. A more complete understanding of such phenomenon is therefore required to fully explore the on-chip chemiluminescent detection possibilities.
References
Isacsson U, Wettermark G (1974) Anal Chim Acta 68(2):339
Stedman DH, Fraser ME (1985) Analytical applications of gas phase chemiluminescence. In: Burr JG (ed) Chemi- and bioluminescence. Dekker, New York, p 439
Blum LJ (1997) Bio-,chemi-luminescent sensors. World Scientific, Singapore
Roswell DF, White EH (1978) The chemiluminescence of luminol and related hydrazides. In: Fleischer S, Fleischer B (eds) Methods in enzymology. Academic Press, London, p 409
Lundin A, Hallander LOB (1987) Mechanisms of horseradish peroxidase catalysed luminol reaction in presence of various enhancers. In: Schölmerich J et al (eds) Bioluminescence and chemiluminescence: new perspectives. Wiley, Chichester, p 555
Merényi G, Lind J, Eriksen TE (1990) J Biolumin Chemilumin 5(1):53
Nakamura M, Nakamura S (1998) Free Radic Biol Med 24(4):537
Hiner ANP, Raven EL, Thorneley RNF, García-Canovas F, Rodríguez-Lopez JN (2002) J Inorg Biochem 91(1):27
Rodríguez-Lopez JN, Lowe DJ, Hernández-Ruiz J, Hiner ANP, García-Canovas F, Thorneley RNF (2001) J Am Chem Soc 123(48):11838
Sakura S (1992) Anal Chim Acta 262(1):49
Marquette CA, Blum LJ (1999) Anal Chim Acta 381:1
Lee W-Y (1997) Microchimica Acta 127(1–2):19
Fuster Mestre Y, Lahuerta Zamora L, Martínez Calatayud J (2001) Luminescence 16(3):213
Huertas-Pérez JF, García-Campan?a AM, González-Casado A, Gámiz-Gracia L, Martínez Vidal JL (2005) Talanta 65(4):980
Xu H, Duan CF, Zhang ZF, Chen JY, Lai CZ, Lian M, Liu LJ, Cui H (2005) Water Res 39(2–3):396
Song Z, Yue Q, Wang C (2005) Food Chem 94(3):457
Worsfold PJ, Achterberg EP, Bowie AR, Cannizzaro V, Charles S, Costa JM, Dubois F, Pereiro R, San Vicente B, Sanz-Medel A, Vandeloise R, Vander Donckt E, Wollast P, Yunus S (2002) J Autom Methods Manag Chem 24(2):41
Li B, Wang D, Xu C, Zhang Z (2005) Microchimica Acta 149(3–4):205
Li Y, Zhu C, Wang L (2005) Microchimica Acta 150(1):95
Song Z, Wang L (2001) J Agric Food Chem 49(12):5697
Song Z, Hou S (2002) Talanta 57(1):59
Huang Y, Liu W (2005) J Pharm Biomed Anal 38(3):537
Yao D, Vlessidis AG, Evmiridis NP (2004) Microchimica Acta 147(1–2):1
Taha ZH (2003) Talanta 61(1):3
Tsukada Y, Yasutake M, Kusama Y, Kishida H, Takano T, Jia D, Tsukada S (2003) Life Sci 72(9):989
Liu X, Du J, Lu J (2003) Luminescence 18(5):245
Zhang S, Li H (2001) Anal Chim Acta 444(2):287
Song Z, Zhao T, Wang L, Xiao Z (2001) Bioorg Med Chem 9(7):1701
Liu H, Zhang L, Zhou J, Hao Y, He P, Fang Y (2005) Anal Chim Acta 541(1–2):125
Safavi A, Hormozi Nezhad MR, Karimi MA (2005) Luminescence 20(3):170
Ratanawimarnwong N, Amornthammarong N, Choengchan N, Chaisuwan P, Amatatongchai M, Wilairat P, Nacapricha D, McKelvie ID (2005) Talanta 65(3):756
Satienperakul S, Cardwell TJ, Kolev SD, Lenehan CE, Barnett NW (2005) Anal Chim Acta 554(1–2):25
Ussher SJ, Achterberg EP, Worsfold PJ, Yaqoob M, Nabi A (2005) Anal Chem 77(7):1971
Yamaguchi M, Yoshida H, Nohta H (2002) J Chromatogr A 950(1–2):1
Vázquez BI, Feás X, Lolo M, Fente CA, Franco CM, Cepeda A (2005) Luminescence 20(3):197
Zhou J, Cui H, Wan G, Xu H, Pang Y, Duan C (2004) Food Chem 88(4):613
Pérez-Ruiz T, Martínez-Lozano C, Tomás V, Martín J (2004) J Chromatogr A 1026(1-2):57
Tsukagoshi K, Fujii T, Nakajima R (2004) Chem Lett 33(8):1000
Jiang HL, He YZ, Zhao HZ, Hu YY (2004) Anal Chim Acta 512(1):111
Tsukagoshi K, Nakahama K, Nakajima R (2003) Chem Lett 32(7):634
Du J, Shen L, Lu J (2003) Anal Chim Acta 489(2):183
Henry OYF, Cullen DC, Piletsky SA (2005) Anal Bioanal Chem 382(4):947
Wu F, Huang Y, Huang C (2005) Biosens Bioelectron 21(3):518
Wu F, Huang Y, Li Q (2005) Anal Chim Acta 536(1–2):107
Tsafack VC, Marquette CA, Pizzolato F, Blum LJ (2000) Biosens Bioelectron 15(3–4):125
Panoutsou P, Economou A (2005) Talanta 67(3):603
Li YX, Zhu LD, Zhu GY, Zhao C (2002) Chem Res Chin Univ 18(1):12
Kiba N, Miwa T, Tachibana M, Tani K, Koizumi H (2002) Anal Chem 74(6):1269
Kiba N, Ito S, Tachibana M, Tani K, Koizumi H (2001) Anal Sci 17(8):929
Marquette CA, Blum LJ (2001) Recent developments in luminol-H2O2 chemiluminescence-based flow injection assays. In: Recent research development in pure applied analytical chemistry. Transworld Research Network, Kerala (India), p 9
Monk DJ, Walt DR (2004) Anal Bioanal Chem 379(7–8):931
Marquette CA, Blum LJ (2000) Talanta 51:395
Marquette CA, Coulet PR, Blum LJ (1999) Anal Chim Acta 398:173
Wilson R, Clavering C, Hutchinson A (2003) Anal Chem 75:4244
Zhang R, Hirakawa K, Seto D, Soh N, Nakano K, Masadome T, Nagata K, Sakamoto K, Imato T (2005) Talanta 68(2):231
Marks RS, Margalit A, Bychenko A, Bassis E, Porat N, Dagan P (2000) Appl Biochem Biotechnol 89:117
Luo J-X, Yang X-C (2003) Anal Chim Acta 485(1):57
Varshney M, Li Y, Griffis CL, Nanapanneni R, Johnson MG (2003) J Rapid Methods Autom Microbiol 11(2):111
Marquette CA, Blum LJ (1998) Sens Actuators B Chem 51:100
Hersh LS, Vann WP, Wilheim SA (1979) Anal Biochem 93:267
Calvo-Munoz ML, Dupont-Filliard A, Billon M, Guillerez S, Bidan G, Marquette C, Blum L (2005) Bioelectrochemistry 66(1–2):139
Rubina AY, Dyukova VI, Dementieva EI, Stomakhin AA, Zasedatelev AS, Nesmeyanov VA, Grishin EV (2005) Anal Biochem 340(2):317
Knecht BG, Niessner R, Weller MG, Strasser A, Dietrich R, Märtlbauer E (2004) Anal Chem 76(3):646
Fall BI, Niessner R, Weller MG, Eberlein-König B, Ring J, Behrendt H (2003) Anal Chem 75(3):556
Marquette CA, Blum LJ (2004) Anal Chim Acta 506(2):127
Marquette CA, Degiuli A, Imbert-Laurenceau E, Mallet F, Chaix C, Mandrand B, Blum LJ (2005) Anal Bioanal Chem 381(5):1019
Kessler N, Ferraris O, Palmer K, Marsh W, Steel A (2004) J Clin Microbiol 42(5):2173
Mallard F, Marchand G, Ginot F, Campagnolo R (2005) Biosens Bioelectron 20(9):1813
Davidsson R, Emnéus J, Johansson B, Passoth V, Bengtsson M, Laurell T (2004) Lab Chip 4(5):488
Corgier BP, Marquette CA, Blum LJ (2005) Anal Chim Acta 538(1–2):1
Marquette CA, Blum LJ (2003) Sens Actuators B Chem 90(1–3):112
Marquette CA, Degiuli A, Blum LJ (2003) Biosens Bioelectron 19(5):433
Tsafack VC, Marquette CA, Leca B, Blum LJ, Tsafack VC (2000) Analyst 125(1):151
Marquette CA, Leca BD, Blum LJ (2001) Luminescence 16(2):159
Marquette CA, Ravaud S, Blum LJ (2000) Anal Lett 33(9):1779
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Marquette, C.A., Blum, L.J. Applications of the luminol chemiluminescent reaction in analytical chemistry. Anal Bioanal Chem 385, 546–554 (2006). https://doi.org/10.1007/s00216-006-0439-9
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-006-0439-9