Introduction
Most biosensors are dedicated to specific tasks, such as glucose measurement, and are therefore not transferable between applications. In addition, some of them lack the specificity to be useful without any confirmation analysis. This leads us to the current situation, where, with an increasing number of parameters and the need for quantification, other analytical techniques, such as chromatography, are more attractive. To overcome these limitations, multianalyte biosensors have been developed. Most of them are based on the parallelization of conventional biosensors. However, the increasing technical expenditure with increasing number of parameters limits this approach to only a few channels. Despite the tremendous success of microarray technology, which is based on pioneering work performed in the 1980s [1, 2, 3] and at the beginning of the 1990s, the application of microarrays in biosensing is a quite recent development.
Comparison of microarray biosensors
In 1999, one of the first biosensors based on microarray technology [4–6] was presented in the field of environmental analysis (Table 1). The system was designated a “parallel affinity sensor array (PASA)”, a designation which emphasizes both the spatial array pattern and the fully parallel readout, which differentiated it from older systems which achieved multianalyte detection by sequential collection of data and subsequent evaluation. The PASA systems consist of a series of precision syringe pumps, a flow cell and a cooled (high-sensitivity) CCD camera for parallel measurement. The assays are analogous to immunoassays performed in microtitration plates; however, chemiluminescence detection is used instead of a chromogenic substrate. Peroxidase-catalyzed chemiluminescence (luminol/hydrogen peroxide) shows very good sensitivity, and, more relevant in this context, it leads to a very simple and efficient optical system, which needs neither wavelength-selective elements (like filters or spectrometers) nor light sources (like lasers) for excitation, in contrast to fluorescence. The PASA system has been extended in several steps and applied in different fields, including allergy diagnosis (specific and total IgE detection in human blood serum [8]) and the high-speed detection of antibiotics in milk [9]. In the latter instrument, an assay time of less than 5 min was achieved for ten analytes. Other systems [10, 11, 12, 13] were based on total internal reflection fluorescence (TIRF). Mobile systems designed for field deployment were applied in the field of biological warfare agents and explosives. Up to nine analytes could be determined simultaneously [12]. A second system based on TIRF was demonstrated for pesticide analysis in water [14]. Recently, the parallel detection of three analytes was presented [15]. In Table 1 some systems are compared with their key specifications. Multianalyte capillary systems often rely on several capillaries, which are coated with different reagents [16, 17, 18]. In 2002, however, a biosensor with four different reagent zones in one capillary was demonstrated [19].
Multivariate and univariate microarrays
Single analyte immunoassays always show some degree of cross-reactivity and so are not able to deliver analyte concentrations. Only analyte equivalents are available. Without any further information, equivalents cannot be converted into concentrations. That is one of the main reasons why many immunoassays are used for screening purposes only, and need some confirmation analysis (for instance, by HPLC). Even if “generic antibodies” (group-selective antibodies) are used, only semi-quantitative results can be expected, because most cross-reactivity patterns are not “flat” enough. Microarrays and other highly parallel techniques offer a powerful approach for overcoming this problem. By using overlapping cross-reactivities, even complex mixtures may be resolved and quantified. This chemometric approach could be termed a “multivariate microarray” [20]. For this purpose a larger set of antibodies (or other reagents) directed against one analyte group is needed. Today the production of antibodies is still the limiting step and so this approach is rarely used. It is much easier to set up a “univariate microarray” with many antibodies with essentially non-overlapping cross-reactivities. The design of such a system is relatively straightforward; only some unexpected cross-reactivities have to be dealt with, for example by the exchange of an antibody. In “univariate microarrays” all of the assays can be evaluated independently. There is, in fact, no difference to simple immunoassays. Unfortunately, the advantages of “multivariate microarrays”, such as substance identification and quantification, are lost in “univariate” systems. However, they can be used for screening purposes, similar to their microtitration plate counterparts. Their main benefit is their ability to screen against many compounds in one run, and so they are perfectly suited for economic non-target analysis.
High-speed microarrays
Microarrays are acknowledged to be powerful research tools. Unfortunately, many microarray experiments are very slow and typically need several hours or sometimes even days to be completed. Consequently, the application of microarrays to fast sensing applications was mostly disregarded. However, taking into account that many immunoassays are performed in a kinetic mode anyway, it is obvious that microarrays can be favorably used for fast applications, too. Currently, assay durations of less than 5 min can be achieved [9]. Even shorter assay times may be conceivable. Alternatively, encoded microsphere assays based on flow-cytometer technology might be used for the development of fast multianalyte biosensors [21]. However, their on-line potential is still largely unexplored.
Outlook
Recent work in the field of microarray biosensors shows that there is much potential for novel multianalyte applications. However, techniques for the fast and economic production of high-quality antibodies are still needed to fulfill the needs of the growing field of protein microarrays. Unfortunately, recombinant techniques have not lead to the expected breakthrough, as yet. In addition, problems with intellectual property may have delayed the widespread and economic use of existing antibodies in microarrays. Furthermore, fierce patent battles have surrounded microarray technology, some of which are not completely settled, and these may have deterred many companies from entering the market. Another problem to be solved is the complexity involved in manufacturing protein microarrays.
However, the additional dimensions in the designs of such arrays in relation to standard immunoassays provide almost infinite opportunities for researchers to solve their analytical problems. The ability to overcome cross-reactivity problems, hence the potential for quantitative results, the excellent scalability, and the high speed of such microarray biosensors makes them a powerful analytical technique.
References
Gordon J, Hawkes R, Towbin H, Niday E (1982) British Patent Application GB 2 099 578 A
Ekins R, Chu F, Biggart E (1989) Anal Chim Acta 227:73–96
Gallo D, Riggs JL, Schachter J, Emmons RW (1981) J Clin Microbiol 13:631–636
Schuetz AJ, Winklmair M, Weller MG, Niessner R (1999) Multianalyte detection with an affinity sensor array. In: Roda A, Pazzagli M, Kricka LJ, Stanley PE (eds) Bioluminescence and chemiluminescence – perspectives for the 21st century (Proc 10th Int Symp, Bologna, Italy). Wiley, Chichester, UK, pp 67–70
Winklmair M, Schuetz AJ, Weller MG, Niessner R (1999) Fresen J Anal Chem 363:731–737
Weller MG, Schuetz AJ, Winklmair M, Niessner, R (1999) Anal Chim Acta 393:29–41
Weller MG (2004) Protein microarrays for biosensor applications. In: Schena M (ed) Protein microarrays. Jones and Bartlett, Sudbury, MA
Fall BI, Eberlein-König B, Behrendt H, Niessner R, Ring J, Weller MG (2003) Anal Chem 75:556–562
Knecht BG, Strasser A, Dietrich R, Märtlbauer E, Niessner R, Weller MG (2004) Anal Chem 76:646–654
Rowe CA, Tender LM, Feldstein MJ, Golden JP, Scruggs SB, MacCraith BD, Cras JJ, Ligler FS (1999) Anal Chem 71:3846–3852
Rowe-Taitt CA, Hazzard JW, Hoffman KE, Cras JJ, Golden JP, Ligler FS (2000) Biosens Bioelectron 15:579–589
Rowe-Taitt C, Anderson GP, Lingerfelt BM, Feldstein MJ, Ligler FS (2002) Anal Chem 74:6114–6120
Sapsford KE, Charles PT, Patterson CH Jr, Ligler FS (2002) Anal Chem 74:1061–1068
Barzen C, Brecht A, Gauglitz G (2002) Biosens Bioelectron 17:289–295
Rodriguez-Mozaz S, Reder S, Lopez de Alda M, Gauglitz G, Barceló D (2004) Biosens Bioelectron 19:633–640
Narang U, Gauger PR, Kusterbeck AW, Ligler FS (1998) Anal Biochem 255:13–19
Yacoub-George E, Meixner L, Scheithauer W, Koppi A, Drost S, Wolf H, Danapel C, Feller KA (2002) Anal Chim Acta 457:3–12
Huth SP, Warholic PS, Devou JM, Chaney LK, Clark GH (2002) J AOAC Int 85:355–364
Mastichiadis C, Kakabakos SE, Christofidis I, Koupparis MA, Willetts C, Misiakos K (2002) Anal Chem 74:6064–6072
Weller MG (2003) Anal Bioanal Chem 375:15–17
Nolan JP, Sklar LA (2002) Trends Biotechnol 20:9–12
Acknowledgements
The valuable contributions of B. G. Knecht, A. J. Schütz, M. Winklmair, B. I. Fall, B. Eberlein-König, H. Behrendt, J. Ring, A. Strasser, R. Dietrich, E. Märtlbauer, R. Niessner, U. Panne, H. Tao, C. Zander, I. Gregor, and C. Gerlach are gratefully acknowledged. I also thank the German Ministry of Education and Research (BMBF, project 02-WU9573), FEI (Forschungskreis der Ernährungsindustrie e.V., Bonn), the AiF, the German Ministry of Economics and Labour (Project 32 ZN), and the Technische Universität München for their financial support.
Author information
Authors and Affiliations
Corresponding author
Additional information
This article is based on a presentation from the 2nd International Workshop on Multianalyte Biosensing Devices, Tarragona, Spain, 2004
Rights and permissions
About this article
Cite this article
Weller, M.G. Optical microarray biosensors. Anal Bioanal Chem 381, 41–43 (2005). https://doi.org/10.1007/s00216-004-2688-9
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-004-2688-9