The complications of 648 consecutively inserted Universal AO pedicle screws (140 in the thoracic spine and 508 in the lumbar spine) performed by one surgical team to treat 91 patients with spinal problems, were reviewed. The spinal pathology consisted of: scoliosis (34 patients), degenerative lower lumbar spinal disease (25 patients), neoplastic spinal disease (11 patients), thoracic kyphosis (8 patients), spinal fractures (7 patients), lumbo-sacral spondylolisthesis (3 patients), and osteomyelitis (3 patients). Intraoperative complications were: screw misplacement (n = 3), nerve root impingement (n = 1), cerebrospinal fluid leak (n = 2) and pedicle fracture (n = 2). Postoperative complications were; deep wound infection (n = 4), screw loosening (n = 2) and rod-screw disconnection (n = 1). The conclusion was that pedicle screw fixation has an acceptable complication rate and neurological injury during this procedure is unlikely.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Blumenthal S, Gill K (1993) Complications of the Wiltse pedicle screw fixation system. Spine 18:1867–1871
Frymoyer J (1991) The adult spine, principles and practice, vol 2. Raven Press, p 2002
Louis R (1986) Fusion of the lumbar and sacral spine by internal fixation with screw plate. Clin Orthop 203:1933
Matsuzaki H, Tokuhashi Y, Matsumoto E, et al (1990) Problems and solutions of pedicle screw plate fixation of the lumbar spine. Spine 15:1159–1163
McGuire RA, Amundson GM (1993) The use of internal fixation in spondylolisthesis. Spine 18: 1662–1672
Mulholland RC (1994) Pedicle screw fixation in the spine (editorial). J Bone Joint Surg Br 76:517–519
Ohlin A, Karlsson M, Duppe H, et al (1994) Complications after transpedicular stabilisation of the spine. Spine 19: 2774–2779
Roy-Camille R et al (1986) Internal fixation of lumbar spine with pedicle screw plating. Clin Orthop 203:7–17
Sanford HD, Myers LM (1992) Complications of lumbar spinal fusion with transpedicular instrumentation. Spine 17:183–189
Gertzbein SD, Robbins SE (1990) Accuracy of pedicular screw placement in vivo. Spine 15:11–14
Weinstein SL (1994) The paediatric spine, principles and practice, vol 2. Raven Press, p 1723
Weinstein JN, Zeilke K, Bohm H (1993) Anatomic and technical consideration of pedicle screw fixation. Clin Orthop 284:34–36
West JL, Oglivie JW, Bradford DW (1991) Complications of the variable plate pedicle screw fixation. Spine 16: 576–579
Zendrick M (1987) Analysis of the morphometric characteristics of the thoracic and lumbar pedicle. Spine 12: 160–166
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Faraj, A.A., Webb, J.K. Early complications of spinal pedicle screw. Eur Spine J 6, 324–326 (1997). https://doi.org/10.1007/BF01142678
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01142678