Abstract
Purpose
To evaluate the performance of the Panomat® P-10 micro-volumetric infusion pump for its use in drug administration at minimal flow rates (μL·hr−1; e.g., intrathecal application).
Methods
Fluid delivery at steady state conditions, and after vertical displacement of the syringe pump by −50 cm was determined gravimetrically. The Panomat® P-10 infusion pump was evaluated at 4, 10, 20, 50 and 100 μL·hr−1, and compared to a conventional syringe pump assembly at 100, 200, 500 and 1000 μL·hr−1. Measurements were repeated twice with two different devices of each syringe pump system, and with two syringes. Data are reported as mean ± SD.
Results
Steady state fluid delivery of the Panomat® P-10 infusion pump revealed less than 5% deviation to set flow rate at 10, 20, 50 and 100 μL·hr−1, and 12% deviation at 4 μL·hr−1. Mean zero-drug delivery time (ZDDT) after lowering the pump by 50 cm at 4 μL·hr−1 flow rate was 38.4 ± 7.3 min. At 100 μL·hr−1 and with original infusion line ZDDT was almost 20 times shorter when compared to the conventional syringe pump assembly (1.5 ± 0.5 min vs 28.5 ± 5.0 min).
Conclusion
The tested Panomat® P-10 micro-volumetric pump shows an acceptable flow accuracy as well as a low susceptibility to vertical displacement, and is therefore suitable for continuous drug administration at minimal flow rates. The technology used in this pump carries potential implications for a new generation of syringe pumps.
Résumé
Objectif
Evaluer la performance de la pompe à perfusion microvolumétrique Panomat® P-10 utilisée pour l’administration de médicaments à débit minimal (μL·hr−1; l’application intrathécale, par ex.). Méthode : L’administration de liquide à l’état d’équilibre, et après le déplacement vertical de la pompe à perfusion de −50 cm, a été déterminée par gravimétrie. La pompe à perfusion Panomat® P-10 a été évaluée à 4, 10, 20, 50 et 100 μL·hr−1 et comparée à une pompe à perfusion classique à 100, 200, 500 et 1000 μL·hr−1. Les mesures ont été faites deux fois avec deux appareils différents pour chaque système de pompe à perfusion et avec deux seringues. Les données sont décrites par la moyenne ± l’écart type.
Résultats
L’administration de liquide à l’état d’équilibre, de la pompe à perfusion Panomat® P-10, présentait moins de 5% de déviation du débit fixé à 10, 20, 50 et 100 μL·hr−1, et 12% de déviation à 4 μL·hr−1. La durée moyenne de perfusion nulle (DPN), après avoir abaissé la pompe de 50 cm, à 4 μL·hr−1 a été de 38,4 ± 7,3 min. À 100 μL·hr−1, et avec les tubulures de perfusion d’origine, la DPN a été presque 20 fois plus courte qu’avec le système classique de pompe à perfusion (1,5 ± 0,5 min vs 28,5 ± 5,0 min).
Conclusion
La pompe microvolumétrique Panomat® P-10 démontre une fiabilité du débit acceptable ainsi qu’une faible susceptibilité au déplacement vertical. Elle convient donc à l’administration continue de médicament à débit minimal. La technologie utilisée pour cette pompe comporte des implications potentielles pour une nouvelle génération de pompes à perfusion.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Cook RI. Syringe pump assemblies and the natural history of clinical technology (Editorial). Can J Anesth 2000; 47: 929–35.
Albright AL, Barron WB, Fasick MP, Polinko P, Janosky J. Continuous intrathecal baclofen infusion for spasticity of cerebral origin. JAMA 1993; 270: 2475–7.
Gilmartin R, Bruce D, Storrs BB, et al. Intrathecal baclofen for management of spastic cerebral palsy: multicenter trial. J Child Neurol 2000; 15: 71–7.
Lonnqvist PA. How continuous are continuous drug infusions? (Editorial). Intensive Care Med 2000; 26: 660–1.
Kern H, Kuring A, Redlich U, et al. Downward movement of syringe pumps reduces syringe output. Br J Anaesth 2001; 86: 828–31.
Weiss M, Hug MI, Neff T, Fischer J. Syringe size and flow rate affect drug delivery from syringe pumps. Can J Anesth 2000; 47: 1031–5.
Neff TA, Fischer JE, Schulz G, Baenziger O, Weiss M. Infusion pump performance with vertical displacement: effect of syringe pump and assembly type. Intensive Care Med 2001; 27: 287–91.
Weiss M, Fischer J, Neff T, Baenziger O. The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps. Anaesthesia 2000; 55: 1094–8.
Weiss M, Banziger O, Neff T, Fanconi S. Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med 2000; 26: 776–9.
Lonnqvist PA, Lofqvist B. Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectors. Intensive Care Med 1997; 23: 998–1001.
Author information
Authors and Affiliations
Corresponding author
Additional information
Disclosure statement: The manufacturer (Disetronic Medical Systems AG, Burgdorf, Switzerland) provided two Panomat®-P-10 microvolumetric infusion pumps and infusion lines for the study without charge. The authors do not hold any agreement on the device and did not receive any financial support from the manufacturer for the study.
Rights and permissions
About this article
Cite this article
Neff, T.A., Fellmann, C., Fuechslin, R.M. et al. The Panomat® P-10 micro-volumetric infusion pump is suitable for continuous drug administration at minimal flow rates. Can J Anesth 49, 1048–1052 (2002). https://doi.org/10.1007/BF03017900
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF03017900