Abstract
A multicentcr study used 756 samples from 251 patients in 12 institutions to compare arterial (PaO2, PaCO2) with transcutaneous (PsO2, PsCO2) oxygen and carbon dioxide tensions, measured usually at 44°C. Of these samples, 336 were obtained from 116 neonates, 27 from 25 children with cystic fibrosis, and 140 from 40 patients under general anesthesia. Ninety-one patients were between 4 weeks and 18 years of age, 32 were between 18 and 60 years, and 12 were over 60. The ratio of transcutaneous to arterial P(s/a)CO2 was 1.01 ±0.11 with PaCO2 less than 30 mm Hg, increasing to 1.04 ±0.08 at PaCO2 greater than 40 mm Hg. Mean bias and its standard deviation (PsCO2 — PaCO2) were + 1.3 ± 3.9 mm Hg in the entire group, + 1.8 ± 4.2 mm Hg in neonates (NS). Bias was +0.2 ± 2.7 mm Hg when PaCO2 was less than 30 mm Hg (N = 175, NS), 1.0 ± 3.4 with 30 < PaCO2 < 40 (n = 329,p < 0.001), and +2.04 ± 4.00 mm Hg with 40 < PaCO2 < 70 (n = 229,p < 0.001). These data suggest that, using transcutaneous PCO2 monitors with inbuilt temperature correction of 4.5%/‡C, the skin metabolic offset should be set to 6 mm Hg. The linear regression was PsCO2 =1.052(PaCO2)-0.56, Sy·x = 3.92, R = 0.929 (n = 756); and PsCO2 = 1.09(PaCO2)-1.57, Sy·x = 4.17, R = 0.928 in neonates (n = 336). The use of vasopressors and vasodilators had no significant effect on bias or its standard deviation or on regression slope and intercept (n = 78). In cystic fibrosis patients, bias and standard deviation were 0.0 ± 1.7 mm Hg (n = 27). Under anesthesia, PsCO2 = 1.07PaCO2-1.58, with bias and standard deviation = 0.6 ± 3.5 (n = 140). For oxygen, at PaO2 ≤ 80 the ratio P(s/a)O2 = 1.05 ± 0.16 in nconates and 0.93 ± 0.21 in older patients, but when PaO2 > 80, P(s/a)O2 fell to 0.88 ± 0.18 in neonates and 0.74 ± 0.21 in older patients. The errors were significantly greater (p < 0.001) in older patients than in neonates above but not below 80 mm Hg, and within both groups errors were significantly greater above than below 80 mm Hg.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Tremper KK. Transcutaneous PO2 measurement. Can Anaesth Soc J 1984;31:664–677
Tremper KK, Barker SJ. Transcutaneous oxygen measurement: experimental studies and adult applications. Int Anesthesiol Clin 1987;25(3):67–96
Phan CQ, Tremper KK, Lee SE, Barker SJ. Noninvasive monitoring of carbon dioxide: a comparison of the partial pressure of transcutaneous and end-tidal carbon dioxide with the partial pressure of arterial carbon dioxide. J Clin Monit 1987;3:149–154
McEvedy BAB, McLeod ME, Mulera M, et al. End-tidal, transcutaneous and arterial PCO2 measurements in critically ill neonates: a comparative study. Anethesiology 1988;69:112–116
Severinghaus JW. A combined transcutaneous PO2-PCO2 electrode with electrochemical HCO3 stabilization. J Appl Physiol 1981;51:1027–1032
Thunstrom AM, Stafford MJ, Severinghaus JW. A two temperature, two PO2 method of estimating the determinants of tcPO2. In: Huch A, Huch R, and Lucey J, eds. Continuous transcutaneous blood gas monitoring. Birth Defects, Vol XV, No. 4. New York: Liss, 1979:167–182
Hazinski TA, Severinghaus JW. Transcutaneous analysis of arterial PCO2. Med Instr 1982, 16:150–153
Glenski JA, Cucchiara RF. Transcutaneous O2 and CO2 monitoring of ncurosurgical patients: detection of air embolism. Anesthesiology 1986;64:546–550
Mindt W, Eberhard P, SchÄfer R. Methodologic factors affecting the relationship between transcutaneous and arterial PCO2. In: Huch R, Huch A, eds. Continuous transcutaneous blood gas monitoring. New York: Marcel Dekker, 1983:199–215
Cassady G. Transcutaneous monitoring in the newborn infant. J Pediatr 1983;103:837–848
Epstein MF, Cohen AR, Feldman HA. Estimation of PaCC2 by two noninvasive methods in the critically ill newborn infant. J Pediatr 1985;106:282–286
Naifeh K, Severinghaus JW. Validation of a maskless CO2 response test for sleep and infant studies. J Appl Physiol 1988;64:391–396
Mclberg SG, Vesterager P. Temperature effects in transcutaneous carbon dioxide monitoring. In: Huch R, Huch A, eds. Continuous transcutaneous blood gas monitoring. New York: Marcel Dekker, 1983:217–231
Naifeh K, Severinghaus JW, Kamiya J, Kraft M. Effect of aging on estimates of hypercapnic ventilatory response during sleep. J Appl Physiol 1989;66:1956–1964
Palmisano BW, Severinghaus JW. Clinical accuracy of a combined transcutaneous PO2-PCO2 electrode. Crit Care Med 1984;12:276
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Palmisano, B.W., Severinghaus, J.W. Transcutaneous Pco2 and Po2: A multicenter study of accuracy. J Clin Monitor Comput 6, 189–195 (1990). https://doi.org/10.1007/BF02832145
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02832145