Abstract
Statistical decision theory indicates that the best way to locate a signal of known waveshape is to use a matched filter. The degree of improvement actually achieved is investigated using actual signals of varying signal-to-noise ratio in which the maternal complex has been automatically eliminated by computer processing. It is shown that a simplified 3-point version of the matched filter provides almost as much improvement with the added advantage that prior knowledge of the fetal waveshape is not required. It is possible to improve the performance further by using a two-stage process that exploits the almost-periodic characteristic of the fetal ECG.
Sommaire
Le calcul statistique montre que la meilleure façon de localiser un signal de forme d’onde connue est d’utiliser un filtre adapté. Le degré de résolution réellement atteint est obtenu enutilisant des signaux réels dont le rapport signal sur bruit est variable, et dans lequel le complexe maternel a été automatiquement éliminé par calculateur. On montre qu’une version simplifiée due filtre adapté offre une aussi bonne résolution, avec l’avantage qu’il n’est pas nécessaire de connaître à l’avance la forme de l’onde foetale. Il est possible d’améliorer encore la performance en utilisant un processus èn deux étapes qui exploite la caractéristique quasi-périodique de l’ECG foetal.
Zusammenfassung
Aus der Entscheidungstheorie der Statistik geht hervor, daß ein Signal bekannter Wellenform am besten mit Hilfe eines angepaßten Filters lokalisiert wird. Der gegenwärtig erreichte Stand wird mit Signalen verschiedenen Signal/Rausch-Verhältnisses untersucht dabei wird der mütterliche Komplex durch Datenverarbeitung automatisch eliminiert. Es wird gezeigt, daß eine vereinfachte 3-Punkte-Version des angepaßten Filters fast diesen Stand erreicht. Ein zusätzlicher Vorteil ist, daß die fetale Wellenform nicht im voraus bekannt sein muß. Eine weitere Verbesserung ist durch Benutzung eines Zweistufenverfahrens möglich, bei dem die fast periodische Charakteristik des fetalen EKG ausgenutzt wird.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Favret, A. G. andMarchetti, A. A. (1966) Fetal electrocardiograph wave forms from abdominal wall recordings.Obstet. Gynec. 27, 355–362.
Goddard, B. A., Newell, J. A., Edwards, R. L. andFarr, R. F. (1966) A clinical foetal electrocardiograph.Med. Electron. biol. Engng 4, 159–67.
Hon, E. H. andHess, O. W. (1957) Instrumentation of fetal ECG.Science 125, 533–54.
Hon, E. H. andHuang, H. S. (1962) The electronic evaluation of fetal heart rate-VII: Premature and missed beats.Obstet. Gynec. 20, 81–90.
Hon, E. H. andLee, S. T. (1963) Noise reduction in fetal electrocardiograph—II: Averaging techniques.Am. J. Obstet. Gynec. 87, 1086–1096.
Maio, R. A. (1967) Detection of almost periodic signals using an adaptive optimum filter. Master’s Thesis, Catholic University of America, Washington, D.C.
Smyth, C. M. (1953) Experimental electrocardiography of the fetus.Lancet i, 1124–26.
Steinberg, C. A., Abraham, S. andCaceres, C. A. (1962) Pattern recognition in clinical electrocardiogram.I.E.E.E. Trans. Bio-Med. Electron. BME-9, 23–30.
Sureau, C. andTrocellier, R. (1963) Etude de quelques problems techniques en electrocardiographie foetale.Med. Electron. biol. Engng 1, 181–88.
Turin, G. L. (1960) An introduction to matched filters.I. R. E. Trans. Inform. Theory IT-6, 311–329.
Van Bemmel, J. H. andVan Der Weide, H. (1966) Detection procedure to represent the foetal heart rate and electrocardiogram.I.E.E.E. Trans. Bio.-Med. Electron. BME-9, 175–182.
Van Bemmel, J. H. (1967) Private communication.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Favret, A.G. Computer matched filter location of fetalR-waves. Med. & biol. Engng. 6, 467–475 (1968). https://doi.org/10.1007/BF02474285
Received:
Revised:
Issue Date:
DOI: https://doi.org/10.1007/BF02474285