Abstract
Dual x-ray absorptiometry (DXA) provides a convenient, non-invasive method of assessing skeletal bone mineral which is widely used for clinical studies. This report describes a study to estimate the effective dose of radiation (ICRP-60(1990)) to a typical female patient from scans performed on three DXA scanners: the Hologic QDR-1000, QDR-1000/W and QDR-2000. The scans modes studied were: total body; anteroposterior (AP) lumbar spine; lateral lumbar spine; proximal femur; distal forearm. An ionization chamber and tissue-equivalent phantom were used to determine entrance surface dose and percentage depth-dose curves for each scan mode. Anatomical data from ICRP-23 (Reference Man) and a body section atlas were used to estimate the absorbed dose to each organ in the scan fields. Effective dose was estimated using the ICRP-60 tissue weighting factors and the fraction of each organ in the scan field. Results are summarized below. Figures for the effective dose are given both excluding and (in brackets) including the ovaries to cover the cases of postmenopausal and premenopausal women respectively.
Article PDF
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
References
Tothill P. Methods of bone mineral measurement. Phys Med Biol 1989;34:543–72.
Wahner HW, Dunn WL, Brown ML, Morin RL, Riggs BL. Comparison of dual-energy X-ray absorptiometry and dual photon absorptiometry for bone mineral measurements of the lumbar spine. Mayo Clin Proc 1988;63:1075–84.
Mazess R, Collick B, Trempe J, Barden H, Hanson J. Performance evaluation of a dual-energy x-ray bone densitometer. Calcif Tissue Int 1989;44:228–32.
Cullim ID, Ell PJ, Ryder JP. X-ray dual-photon absorptiometry: a new method for the measurement of bone density. Br J Radiol 1989;62:587–92.
Broerse JJ, Carmichael JHE. Quantity criteria for diagnostic radiographic images. CEC working document. 2nd ed. 1990.
Kalender WA. Effective dose values in bone mineral measurements by photon absorptiometry and computed tomography. Osteoporosis Int 1992;2:82–7.
ICRP Publication 60. 1990 recommendations of the International Commission on Radiological Protection. Annals of the ICRP 1991;21:no. 1–3.
Pye DW, Hannan WJ, Hesp R. Effective dose equivalent in dual X-ray absorptiometry. Br J Radiol 1990;63:149.
Rawlings DJ, Faulkner K, Chappie CL. The influence of scan time on patient dose and precision in bone mineral densitometry. In: EFJ Ring, editor. Current reseach in osteoporosis and bone mineral measurement II. London: British Institute of Radiology, 1992:23–4.
Laskey MA, Flaxman ME, Barber RW, Trafford S, Hayball MP, Lyttle KD, Crisp AJ, Compston JE. Comparative performance in vitro and in vivo of Lunar DPX and Hologic QDR-1000 dual energy X-ray absorptiometers. Br J Radiol 1991;64:1023–9.
Purdie D. Screening for osteoporosis: the Humberside experiment. In: EFJ Ring, editor. Current research in osteoporosis and bone mineral measurement II. London: British Institute of Radiology, 1992:84–5.
Reid DM. Screening for osteoporosis: which site and at what cost? In: EFJ Ring editor. Current research in osteoporosis and bone mineral measurement II. London: British Institute of Radiology, 1992:85.
Kanis JA, Geusens P, Christiansen C. Guidelines for clinical trials in osteoporosis. Osteoporosis Int 1991;1:182–8.
Steiger P, von Stetten E, Weiss H, Stein JA. Paired AP and lateral supine dual x-ray absorptiometry of the spine: initial results with a 32 detector system [abstract]. Osteoporosis Int 1991;1:190.
Herd RJM, Blake GM, Parker JC, Ryan PJ, Fogelman I. Total body studies in normal British women using dual energy x-ray absorptiometry. Br J Radiol 1993;66:303–8.
ICRP Publication 23. Report of the task group on reference man. Oxford: Pergamon Press, 1975.
Eycleshymer AC, Schoemaker DM. A cross-sectional anatomy. New York: Appleton and Century, 1911.
Hughes JS, Shaw KB, O'Riordan MC. Radiation exposure of the UK population: 1988 review. National Radiation Protection Board Report NRPB-R227. London: HMSO, 1989.
Shrimpton PC, Wall BF, Jones DG, Fisher ES, Hillier MC, Kendall GM, Harrison RM. A national survey of doses to patients undergoing a selection of routine x-ray examinations in English hospitals. National Radiological Protection Board NRPB-R200. London: HMSO, 1986.
Conway BJ, McCrohan JL, Antonsen RG, Reuter FG, Slayton RJ, Suleiman OH. Average radiation dose in standard CT examination of the head: results of the 1990 NEXT survey. Radiology 1992;184:135–40.
ICRP Publication 26. Recommendations of the International Commission on Radiological Protection. Annals of the ICRP 1977;1:no. 3.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Lewis, M.K., Blake, G.M. & Fogelman, I. Patient dose in dual x-ray absorptiometry. Osteoporosis Int 4, 11–15 (1994). https://doi.org/10.1007/BF02352255
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02352255