Abstract
The paper reports on the histological effects of chronic implantation of self-sizing spiral cuff nerve electrodes on the cat sciatic nerve. The implantation period is about 4.4 months. Four different experimental conditions are evaluated: control, sham, bare cuff (cuffs without contacts and leads) and full cuff. The total number of axons in the nerves of the control group is compared with the three other groups. The surface occupied by collagen fibres in the nerve section, perineurium thickness, fibre diameter and myelin thickness are also measured. The average number of axons in the control nerves is found to be 16416 (±1509) and does not differ significantly from the three other groups (p>0.1). Collagen measurements show an extrafascicular epineurial fibrosis in the two implanted groups that is found to be significantly different (p<0.05). No differences are encountered in the perineurium thickness analysis. Fibre diameter distributions show a regular bimodal pattern for all groups. Centrality (mean and Pm) and dispersion statistics (P25 and P75) extracted from fibre diameter distributions do not reveal significant differences. Myelin thickness distributions are also similar for all groups, as well as centrality and dispersion statistics. The present morphometrical results suggest that the effects produced by a chronic spiral cuff implant on this animal model are negligible.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Agnew, W., McCreery, D., Yuen, T., andBullara, L. (1988): ‘Development of safe techniques for selective activation of neurons’. Technical Report, Contract N01-NS-62397, NIH-NINCDS
Agnew, W., McCreery, D., Yuen, T., andBullara, L. (1989): ‘Histological and physiological evaluation of electrically stimulated peripheral nerve: considerations for the selection of parameters’,Ann. Biomed. Eng.,17, pp. 39–60
Agnew, W., andMcCreery, D. (1990): ‘Considerations for safety with chronically implanted nerve electrodes’,Epilepsia,31, (Suppl. 2), pp. S27-S32
Aziz, W., Firrell, J., Ogden, L., andBreidenbach, W. (1999): ‘Blood flow in a chronic entrapment neuropathy model in the rabbit sciatic nerve’,J. Reconstr: Microsurg.,15, pp. 47–53
Baer, G., Talonen, P., Shneerson, J., Markkula, H., Exner, G., andWells, F. (1990): ‘Phrenic nerve stimulation for central ventilatory failure with bipolar and four-pole electrode systems’,Pacing Clin. Electrophys.,13, pp. 1061–1072
Bosnjak, R., Dolenc, V., andKralj, A. (1999): ‘Biomechanical response in the ankle stimulation of lumbosacral nerve roots with spiral cuff multielectrode: preliminary study’,Neurol. Med. Chirur.,39, pp. 659–666
Brown, W., Ferguson, G. G., Jones, M., andYates, S. (1976): ‘The location of conduction abnormalities in human entrapment neuropathies’,Can. J. Neurol. Sci.,3, pp. 111–1118
Chervin, R., andGuilleminault, C. (1997): ‘Diaphragm pacing for respiratory insufficiency’,J. Clin. Neurophysiol.,14, pp. 369–377
Cornefjord, M., Sato, K., Olmarker, K., Rydevik, B., andNordborg, C. (1997): ‘A model for chronic root compression studies: Presentation of a porcine model for controlled, slow-onset compression with analyses of anatomic aspects, compression onset rate, and morphologic and neurophysiologic effects’,Spine,22, pp. 946–957
Creasey, G., Elefteriades, J., DiMarco, A., Talonen, P., Bijak, M., Girsch, W., andCantor, C. (1996): ‘Electrical stimulation to restore respiration’,J. Rehab. Res. Dev.,33, pp. 123–132
Cruz-Orive, L., andWeibel, E. (1981): ‘Sampling designs for stereology’,J. Microscop. 122, pp. 235–257
Dahlin, L., andKanje, M. (1992): ‘Conditioning effect induced by chronic nerve compression, an experimental study of the sciatic and tibial nerves of rats’,Seand. J. Reconstr. Hand Surg. 26, pp. 37–41
Dahlin, L., Rydevick, B., McLean, W., andSjöstrand, J. (1984): ‘Changes in fast axonal transport during experimental nerve compression at low pressure’,Exp. Neurol.,84, pp. 29–36
Dyck, P. (1969): ‘Experimental hypertrophy neuropathy’,Arch. Neurol.,21, pp. 73–82
Dyck, P., Thomas, P., Lambert, E., andBunge, R. (1984): ‘Peripheral neuropathy’,in W.B. Saunders Company, pp. 769, 802
Fern, R., andHarrison, P. (1994): ‘The contribution of ischaemia and deformation to the conduction block generated by compression of the cat sciatic nerve’,Exp. Physiol.,79(4), pp. 583–592
Glenn, W., andPhelps, M. (1985): ‘Diaphragm pacing by electrical stimulation of the phrenic nerve’,Neurosurgery,17, pp. 974–984
Grill, W., andMortimer, J. (1994): ‘Electrical properties of implant encapsulation tissue’,Ann. Biomed. Eng.,22, pp. 22–33
Gundersen, H. (1977): ‘Notes on the estimation of the numerical density of arbitrary profiles: the edge effects’,J. Microsc.,111, pp. 219–223
Heijmans, H. (1999): ‘Connected morphological operators for binary images’,Comput. Vis. Image Underst.,73(1), pp. 99–120
Kim, J., Manuelidis, E., William, W., Glenn, W., Fukuda, Y., Cole, D., andHogan, J. (1983): ‘Light and electron microscopic studies of phrenic nerves after long-term electrical stimulation’,J. Neurosurg.,58, pp. 84–91
Larsen, J. (1998): ‘Stereology of nerve cross sections’,J. Neurosci. Methods,85, pp. 107–118
Larsen, J., Thomsen, M., Haugland, M., andSinkjaer, T. (1998): ‘Degeneration and regeneration in rabbit peripheral nerve with long-term nerve cuff electrode implant: a stereological study of myelinated and unmyelinated axons’,Acta Neuropathol.,96, pp. 365–378
Lequesne, P., andCasey, E. (1974): ‘Recovery of conduction velocity distal to a compressive lesion’,J. Neurol. Neurisurg. Psychiatry,37, pp. 1346–1351
Lloyd, D., andChang, H. (1948): ‘Afferent fibers in muscle nerves’,J. Neurophysiol.,11, pp. 199
Lundborg, G. (1970): ‘Experimental studies on intraneural microvascular pathophysiology and nerve function in a limb subjected to a temporary circulatory arrest’,Scand. J. Plast. Reconstr. Surg.,6(Suppl.), pp. 287–294
Lundborg, G. (1975): ‘Structure and function of the intraneural microvessels as related to trauma, edema formation and nerve function’,J. Bone Joint Surg.,57A, pp. 938–948
Mackinnon, S., Dellon, A., andHudson, A. (1984): ‘Chronic nerve compresion: an experimental model in the rat’,Ann. Plast. Surg.,13, pp. 112–120
Mackinnon, S., Dellon, A., andHudson, A. (1985): ‘A primate model for chronic nerve compression’,J Reconstr. Microsurg.,1, pp. 185–195
Mayhew, T., andSharma, A. (1984a): ‘Sampling schemes for estimating nerve fiber size. i. methods for nerve trunks of mixed fascicularity’,J. Anat.,139, pp. 45–58
Mayhew, T., andSharma, A. (1984b): ‘Sampling schemes for estimating nerve fiber size. ii. methods for unifascicular nerve trunks’,J. Anat.,139, pp. 59–66
McCreery, D., Agnew, W., Yuen, T., andBullara, L. (1992): ‘Damage in peripheral nerve from continuous electrical stimulation: comparison of two stimulus waveforms’,Med. Biol. Eng. Comput.,30, pp. 109–114
McCreery, D., Yuen, T., Agnew, W., andBullara, L. (1994): ‘Stimulus parameters affecting tissue injury during microstimulation in the cochlear nucleus of the cat’,Hear. Res.,77(1–2), pp. 105–115
McNeal, D., Waters, R., andReswick, J. (1977): ‘Experience with implanted electrodes’,Neurosurg.,1, pp. 228–229
Mortimer, J. (1990): ‘Electrode for functional neuromuscular stimulation’. Technical Report, Contract N01-NS-7-2396, final report, NIH-NINCDS Neural prostheses program
Naples, G., Mortimer, J., Schneider, A., andSweeney, J. (1988): ‘A spiral nerve cuff electrode for peripheral nerve stimulation’,IEEE Trans.,BME-35, pp. 905–916
Nielson, K., Watts, C., andClark, W. (1976): ‘Peripheral nerve injury from implantation of chronic stimulating electrodes for pain control’,Surg. Neurol.,5, pp. 51–53
Ochoa, J., Danta, G., Fowler, G., andGilliatt, R. (1971): ‘Nature of the nerve lesion caused by pneumatic torniquet’,Nature,233, pp. 265–271
Romero, E., Cuisenaire, O., Denef, J., Delbeke, J., Macq, B., andVeraart, C. (2000): ‘Automatic morphometry of nerve histological sections’,J. Neurosci. Methods,97, pp. 111–122
Rydevik, B., andLundborg, G. (1977): ‘Permeability of intraneural microvessels and perineurium following acute, graded experimental nerve compression’,Scand. J. Plast. Reconstr: Surg.,11, pp. 179–187
Rydevik, B., Lundborg, G., andBagge, U. (1981): ‘Effects of graded compression on intraneural blood flow’,J. Hand Surg.,6, pp. 3–12
Sauter, J., Berthord, H., andJeanrenaud, B. (1983): ‘Simple electrode for intact nerve stimulation and/or recording in semichronic rats’,Pflugers Arch.,397, pp. 68–69
Segal, R., Stacey, B., Rudy, T., Baser, S., andMarkham, J. (1998): ‘Spinal cord stimulation revisited’,Neurol. Res.,20, pp. 391–396
Strauss, W., andSprague, J. (1944): ‘The dinnervation of the interosseous muscles and the mechanism of the toe spreading reflex of the hind-limb in the cat’,Am. J. Physiol.,142, pp. 391–396
Suderland, S. (1968): ‘Nerves and nerve injuries’ in E. and S. Livingstone
Upshaw, B., andSinkjaer, T. (1997): ‘Natural versus artifitial sensors applied in peroneal nerve stimulation’,J. Appl. Physiol.,82, pp. 227–231
Van Kerrebroeck, E., Der Aa, H.V., Bosch, J., Koldewijn, E., andVorsteveld, J. (1997): ‘Sacial rhizotomies and electrical bladder stimulation in spinal cord injury. Part i: Clinical and urodynamic analysis. dutch study group on sacral anterior root stimulation’,Eur. Urol.,31, pp. 263–271
Veraart, C., Grill, W., andMortimer, J. (1993): ‘Selective control of muscle activation with a multipolar nerve cuff electrode’,IEEE Trans.,BME-40, pp. 640–653
Weese-Mayer, D., Morrow, A., Brouillette, R., Ilbawi, M., andHunt, C. (1989): ‘Diphragm pacing in infants and children. a lifetable analysis of implanted components’,Am. Rev. Resp. Dis.,139, pp. 974–979
Weisl, H., andOsborne, G. (1964): ‘The pathological changes in rats' nerves subject to moderate compression’,J. Bone Joint Surg.,46B, pp. 297–306
Yuen, T., Agnew, W. andBullara, L. (1984): ‘Histopathological evaluation of dog sacral nerve after chronic electric stimulation for micturition’,Neurosurg.,14, pp. 449–455
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Romero, E., Denef, J.F., Delbeke, J. et al. Neural morphological effects of long-term implantation of the self-sizing spiral cuff nerve electrode. Med. Biol. Eng. Comput. 39, 90–100 (2001). https://doi.org/10.1007/BF02345271
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02345271