Abstract
Objective: The present study was intended to evaluate the “in vivo” endotracheal (ET) tube resistance and respiratory mechanics in mechanically ventilated patients with respiratory failure by using fiber optic catheters.Design: Two fiber optic catheters, consisting of a thin probe with a pressure transducer on the tip, were used. The first was placed at the proximal side of the ET tube and the second was positioned distally beyond the end. A low compliant air-filled catheter connected to a traditional pressure transducer was placed close to the proximal fiber optic device to compare the pressure values obtained with both systems.Setting: The study was performed in the General Intensive Care Unit of Rome “La Sapienza”, University Hospital.Patients and participants: Seven patients admitted for the management of acute respiratory failure of different etiologies were included in the protocol. All the patients were intubated and mechanically ventilated for a least 48h prior to the investigation.Measurements and results: The endotracheal tube resistance was obtained both by the end-inspiratory occlusion method and measuring pressure proximally and distally to the ET tube. The measurement of respiratory mechanics was obtained proximally and distally to the ET tube. Different flows and tidal volume changes were performed. The results showed that the fiber optic device gives an adequate evaluation of airway pressure and the possibility for an easy detection of obstructions and/or deformations of the ET tube. The area described by inspiratory and expiratory pressure recorded at both sides of the ET tube showed a positive relationship between the surface and flows while no surface changes were shown when the tidal volumes were modified. Thoraco-pulmonary compliance measured proximally and distally to the ET tube gave rise to a small and statistically insignificant differenceConclusion: This study confirms that 48h after the positioning of ET tubes the airflow resistance is significantly higher than might be expected from the “in vitro” data. The presence of the endotracheal tube can interfere with the evaluation of thoraco-pulmonary mechanics, particularly in dynamic conditions. The fiber optic system represents an interesting and simple tool for the evaluation of ET tube resistance and pulmonary mechanics in patients undergoing mechanical ventilation.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Lavietes MH, Rochester DF (1981) Assessment of airway function during assisted ventilation. Lung, 159:219
Butler JP, Leith DE, Jackson AC (1986) Principles of measurement: application to pressure, volume and flow. Handbook of Physiology. The Respiratory System, III:15
Mead J, Whittenberger JL (1954) Evaluation of interruption technique as a method of measuring pulmonary air-flow resistance. J Appl Physiol, 6:408
Bates JHT, Rossi A, Milic-Emili J (1985) Analysis of the behaviour of the respiratory system with constant inspiratory flow. J Appl Physiol 58:1840
Rossi A, Cottfried SB, Zocchi L, Higgs BD, Lennox S, Carverly PMA, Begin P, Grassino A, Milic-Emili J (1985) Measurement of static compliance, of the total respiratory system in patients with acute respiratory failure during mechanical ventilation: the effect of intrinsic PEEP. Am Rev. Respir Dis 131:672
Bolder PM, Healy TEJ, Bolder AR, Beatty PCW, Kay B (1986) The extra work of breathing through adult endotracheal tubes. Anesth Analg 65:853
Demers RR, Sullivan MJ, Paliotta J (1977) Airflow resistance of endotracheal tubes. JAMA 237:1362
Wright PE, Marini JJ, Bernard GR (1989) In vitro versus in vivo comparison of endotracheal tube airflow resistance. Am Rev Respir Dis 140:10
Johanningman JA, Branson RD, Hurst JM, Davis K Jr, Jonson DJ (1989) Change in oxygen consumption while breathing through endotracheal tubes. Am Rev Respir Dis 17:S27
Sullivan M, Paliotta J, Saklad M (1976) Endotracheal tube as a factor in measurement of respiratory mechanics. J Appl Physiol 41:590
Behrakis PK, Higgs BD, Baydur A, Zin WA, Milic-Emili J (1983) Respiratory mechanics during halothane anesthesia and anesthesia-paralysis in humans. J Appl Physiol 55:1085
Gottfried SD, Rossi A, Higgs BD, Calverley PM, Zocchi L, Bozic C, Milic-Emili J (1985) Noninvasive determination of respiratory system mechanics during mechanical ventilation for acute respiratory failure. Am Rev Respir Dis 131:114
Loring SH, Elliot EA, Drazen JM (1979) Kinetic energy loss and convective acceleration in respiratory resistance measurement. Lung 156:37
Menon AS, Webr ME, Chaqng HK (1984) Model study of flow dynamics in human central airways. Part III: oscillatory velocity profiles. Respir Physiol 55:255
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
De Blasi, R.A., Conti, G., Antonelli, M. et al. A fibre optics system for the evaluation of airway pressure in mechanically ventilated patients. Intensive Care Med 18, 405–409 (1992). https://doi.org/10.1007/BF01694342
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01694342