Abstract
The aim of this study was to evaluate the precision and accuracy of dual-energy X-ray absorptiometry (DXA) for measuring bone mineral content at different sites of the skeleton in rats. In vitro the reproducibility error was very small (<1%), but in vivo the intra-observer variability ranged from 0.9% to 6.0%. Several factors have been shown to affect in vivo reproducibility: the reproducibility was better when the results were expressed as bone mineral density (BMD) rather than bone mineral content (BMC), intra-observer variability was better than the inter-observer variability, and a higher error was observed for the tibia compared with that for vertebrae and femur. The accuracy of measurement at the femur and tibia was assessed by comparing the values with ash weight and with biochemically determined calcium content. The correlation coefficients (R) between the in vitro BMC and the dry weight or the calcium content were higher than 0.99 for both the femur and the tibia. SEE ranged between 0.0 g (ash weight) and 2.0 mg (Ca content). Using in vitro BMC, ash weight could be estimated with an accuracy error close to 0 and calcium content with an error ranging between 0.82% and 6.80%. TheR values obtained between the in vivo and in vitro BMC were 0.98 and 0.97 respectively for femur and tibia, with SEE of 0.04 and 0.02 g respectively. In conclusion, the in vivo precision of the technique was found to be too low. To be of practical use it is important in the design of experimentation to try to reduce the measurement error. This can be achieved by performing measurements in the same position, by repeating measurements several times and by using the mean values of several BMD calculations performed by the same observer on each BMD measurement. Furthermore, better reproducibility can be obtained on the vertebra or the femur than on the tibia.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Kanis JA, Geusens P, Christiansen C. Guidelines for clinical trials in osteoporosis: a position paper of the European Foundation for Osteoporosis and Bone Disease. Osteoporosis Int 1991;1:182–8.
Hassager C, Christiansen C. Current techniques for bone mass measurement. Baillieres Clin Obstet Gynaecol 1991;5:807–15.
Kalu DN, Hardin RR, Cockerham R. Evaluation of the pathogenesis of skeletal changes in ovariectomized rats. Endocrinology 1984;115:507–12.
Wronski TJ, Cintron M, Doherty AL, Dann LM. Estrogen treatment prevents osteopenia and depresses bone turnover in ovariectomized rats. Endocrinology 1988;123:681–6.
Kimmel DB, Wronski TJ. Nondestructive measurement of bone mineral in femurs from ovariectomized rats. Calcif Tissue Int 1990;46:101–10.
Vanderschueren D, Van Herck E, Suiker AMH, Visser WJ, Schot LPC, Bouillon R. Bone and mineral metabolism in aged male rats: short and long term effects of androgen deficiency. Endocrinology 1992;130:2906–16.
Shen V, Dempster DW, Birchman R, Xu R, Lindsay R. Loss of cancellous bone mass and connectivity in ovariectomized rats can be restored by combined treatment with parathyroid hormone and estradiol. J Clin Invest 1993;91:2479–87.
Mosekilde LI, Danielsen CC, Knudsen UB. The effect of aging and ovariectomy on the vertebral bone mass and biochemical properties of mature rats. Bone 1993;14:1–6.
Ammann P, Rizzoli R, Slosman D, Bonjour JP. Sequential and precise in vivo measurements of bone mineral density in rats using dual-energy x-ray absorptiometry. J Bone Miner Res 1992;7:311–6.
Juhn A, Weiss A, Mendes D, Silbermann M. Non-invasive assessment of bone mineral density during maturation and aging of Wistar female rats. Cells Materials 1991;Supp 1:19–24.
Ongphiphadhanakul B, Alex S, Braverman LE, Baran DT. Excessive 1-thyroxine therapy decreases femoral bone mineral densities in the male rat: effect of hypogonadism and calcitonin. J Bone Miner Res 1992;7:1227–31.
Griffin MG, Kimble R, Hopfer W, Pacifici R. Dual-energy X-ray absorptiometry of the rat: accuracy, precision, and measurement of bone loss. J Bone Miner Res 1993;8:795–800.
Rozenberg S, Peretz A, Caufriez C, Robyn, Ham H. The validity of bone mineral measurements at the vertebral site is age related. In: Christiansen C, Johansen JS, Rijs BJ, editors. Osteoporosis. Viborg, Denmark: Norhaven, 1987:72–3.
LeBlanc AD, Evans HJ, Marsh C, Schneider V, Johnson PC, Jhingran SG. Precision of dual photon absorptiometry measurements. J Nucl Med 1986;27:1362–5.
Johnson J, Dawson-Hughes B. Precision and stability of dual-energy x-ray absorptiometry measurements. Calcif Tissue Int 1991;49:174–8.
Pouilles JM, Tremollieres F, Todorovsky N, Ribot C. Precision and sensitivity of dual-energy x-ray absorptiometry in spinal osteoporosis. J Bone Miner Res 1991;6:997–1002.
Gotfredsen A, Pødenphant J, Nørgaard H, Nilas L, Herss Nielsen VA, Christiansen C. Accuracy of lumbar spine bone mineral content by dual photon absorptiometry. J Nucl Med 1988;29:248–54.
Ho CP, Kim RW, Schaffler MB, Sartoris DJ. Accuracy of dual-energy radiographic absorptiometry of the lumbar spine: cadaver study. Radiology 1990;176:171–3.
Louis O, Van den Winkel P, Covens P, Schoutens A, Osteux M. Dual-energy x-ray absorptiometry of lumbar vertebrae: relative contribution of body and posterior elements and accuracy in relation with neutron activation analysis. Bone 1992;13:317–20.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Rozenberg, S., Vandromme, J., Neve, J. et al. Precision and accuracy of in vivo bone mineral measurement in rats using dual-energy X-ray absorptiometry. Osteoporosis Int 5, 47–53 (1995). https://doi.org/10.1007/BF01623658
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01623658