Abstract
Using global constraints and dynamic programming, a new model-based segmentation algorithm was developed to determine myocardial borders and basal plane. The segmented image is transformed to a countrate polar map and the infarct size (I.S.) is determined by comparison with a reference polar map. In order to evaluate our method the algorithm was applied to heart phantoms, to software simulations and to animal studies. In the last experiments, Tc-99m Sestamibi was used as a perfusion agent.
The total myocardial volume and infarct size of a Jasczack phantom were overestimated, especially when I.S. was expressed in absolute rather than relative values. It was proven by software simulations of cardiac Spect studies that those errors were mainly due to finite resolution effects causing a clear overestimation of myocardial thickness. Implementation of a constant thickness in the algorithm resulted in a much better correlation with actual values.
In a dog experiment the size of total myocardial volume of the area at risk during occlusion and of the final infarct size after thrombolysis was correlated with the histologic values obtained by planimetry after TTC staining. In 8 studies, an excellent correlation between the histologic area at risk versus the estimated perfusion defects was obtained (r=0.97).
The automatic delineation of myocardial borders and valve plane was excellent even when perfusion defects were present. Manual intervention was only necessary in certain slices where a clear overlap between liver and myocardium was present in the dog studies. Segmental polar maps expressing count rate and volume information provided a visual and quantitative tool to evaluate the influence of thrombolysis in acute ligation experiments.
It is concluded that the new algorithm is ready to be used in a clinical environment for the quantitative evaluation of perfusion defects after acute myocardial infarction and for the follow-up of the therapeutic strategy.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Nuyts J, Mortelmans L, Suetens P et al. Model-Based quantification of myocardial perfusion images. J Nucl Med 1989; 30: 1992–2001.
Geltman E, Abenschein D, Devries S. Assessment of coronary thrombolysis. Card Clin 1987; 5: 55–66.
Eisner R, Nowak D, Pettigrew R, Fajman W. Fundamentals of 180 degrees acquisiton and reconstruction in SPECT imaging. J Nucl Med 1986; 27: 1717–28.
Nakajima K, Shuke N, Taki J, Ichihara T, Motomura N, Bunko H, Hisada K. A simulation of dynamic SPECT using radiopharmaceuticals with rapid clearance. J Nucl Med 1992; 33: 1200–6.
Eckner F, Brown B, Davidson D. Dimensions of normal human hearts. After standard fixation by controlled pressure coronary perfusion. Arch Path 1969; 88: 497–507.
Goldstein R, Mullani N, Wong W et al. Positron imaging of myocardial infarction with rubidium-82. J Nucl Med 1986; 27: 1824–9.
Chang W, Henkin R, Buddemeyer E. The sources of overestimation in the quantification by SPECT of uptakes in a myocardial phantom: Concise communication. J Nucl Med 1984; 25: 788–91.
Koral K, Wang X, Rogers W. SPECT compton-scattering correction by analysis of energy spectra. J Nucl Med 1988; 29: 195–202.
Hutton B, Bailey D, Fulton R. Estimates of left ventricular volumes by equilibrium radionuclide angiography: Importance of attenuation correction. J Nucl Med 1985; 26: 317–8.
Hosoba M, Wani H, Toyama H et al. Automated body contour detection in SPECT: Effect on quantitative studies. J Nucl Med 1986; 27: 1184–91.
Manglos S, Jaszczak R, Floyd C et al. Nonisotropic attenuation in SPECT: Phantom tests of quantitative effects and compensation techniques. J Nucl Med 1987; 28: 1584–91.
Eisner R, Noever T, Nowak D. Use of cross-correlation function to detect patient motion during SPECT imaging. J Nucl Med 1987; 28: 97–101.
Eisner R, Churchwell A, Noever T et al. Quantitative analysis of the tomographic thallium-201 myocardial bulls-eye display: Critical role of correcting for patient motion. J Nucl Med 1988; 29: 91–7.
Eisner R, Tamas M, Cedarholm J et al. Estimation of left ventricular mass from SPECT Tl-201: A new algorithm based on the ‘bullseye’ display validated in animal studies (abstract). J Nucl Med 1988; 29: 945.
Greckle W, Frank T, Links J et al. Correction for patient and organ movement in SPECT: Application to exercise thallium-201 cardiac imaging. J Nucl Med 1988; 29: 441–50.
Machac J, Howard L. Balk E et al. Computer modeling of planar myocardial perfusion imaging: Effect of heart rate and ejection fraction on wall thickness and chamber size. J Nucl Med 1986; 27: 653–9.
Barat J, Brendel A, Colle J et al. Quantitative analysis of left-ventricular function using gated single photon emission tomography. J Nucl Med 1984; 25: 1167–74.
Wackers F, Gibbons R, Verani M et al. Serial quantitative planar technetium-99m isonitrile imaging in acute myocardial infarction: Efficacy for non invasive assessment of thrombolytic therapy. JACC 1989; 14: 861–73.
Verani M, Jeroudi M, Mahmarian J et al. Quantification of myocardial infarction during coronary occlusion and myocardial salvage after reperfusion using cardiac imaging with technetium-99m hexakis 2-methyoxyisobutyl isonitrile. JACC 1989; 12: 1573–81.
Caldwell JH, Williams DL, Harp GD et al. Quantitation of size of relative myocardial perfusion defect by single-photon emission computed tomography. Circulation 1984; 70: 1048–56.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Mortelmans, L., Nuyts, J., Vanhaecke, J. et al. Experimental validation of a new quantitative method for the analysis of infarct size by cardiac perfusion tomography (SPECT). Int J Cardiac Imag 9, 201–212 (1993). https://doi.org/10.1007/BF01145322
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01145322