Abstract
Cast Co-Cr-Mo was heat treated to determine whether a significant improvement in the alloy's mechanical properties could be obtained without loss of corrosion resistance. Tensile, corrosion fatigue, and corrosion tests were carried out using individually cast test pieces. The effect on the tensile properties of solution treatment at 1240° C and of ageing at 720° C was determined for a large range of heat treatment times. In addition the effect of other heat treatments described in the literature was determined. Partial solution treatment gave the greatest improvement in corrosion fatigue behaviour. None of the heat treatments significantly affected the corrosion resistance. In spite of the improvements obtained, it was concluded that for orthopaedic implant applications requiring high corrosion fatigue strength, cast Co-Cr-Mo was less suitable than other currently available products.
Article PDF
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
References
D. G. Taylor,Proc. Roy. Soc. B192 (1976) 145.
H. S. Dobbs,Eng. Med. 7 (1978) 107.
N. J. Grant,Trans. Amer. Soc. Met. 39 (1947) 281.
G. A. Hoffman andC. F. Robards, National Advisory Committee for Aeronautics TN (1951) 2513.
J. R. Lane andN. J. Grant,Trans. Amer. Soc. Met. 44 (1952) 113.
F. J. Clauss, F. B. Garrett andJ. W. Weeton, National Advisory Committee for Aeronautics TN (1955) 3512.
J. W. Weeton andR. A. Signorelli,Trans. Amer. Soc. Met. 47 (1955) 815.
R. Hollander andJ. Wulff,Eng. Med. 3 (1974) 8.
F. G. Hodge andT. S. Lee,Corros. 31 (1975) 111.
T. M. Devine andJ. Wulff,J. Biomed. Mater. Res. 9 (1975) 151.
J. Cohen, R. M. Rose andJ. Wulff,ibid. 12 (1978) 935.
M. Lorentz, M. Semlitsch, B. Panic, H. Weber andH. G. Willert,Eng. Med. 7 (1978) 241.
D. O. Cox, PhD dissertation, University of California (1977).
A. M. Hall andF. R. Morral,Mater. Design Eng. 63 (1966) 76.
A. N. Hughes, B. A. Jordan andS. Orman,Eng. Med. 7 (1978) 135.
A. T. Kuhn,Biomater. 2 (1981) 68.
A. Acharya, E. Freise andE. H. Greener,Cobalt 47 (1970) 75.
B. C. Syrett andS. S. Wing,Corros. 34 (1978) 138.
T. M. Devine andJ. Wulff,J. Electrochem. Soc. 123 (1976) 1433.
P. Süry andM. Semlitsch,J. Biomed. Mater. Res. 12 (1978) 723.
T. P. Hoar andD. C. Mears,Proc. Roy. Soc. A294 (1966) 486.
H. J. Mueller andE. H. Greener,J. Biomed. Mater. Res. 4 (1970) 29.
J. R. Cahoon, R. Bandyopaohya andL. Tennese,ibid. 9 (1975) 259.
J. Brettle andA. N. Hughes,Eng. Med. 7 (1978) 142.
A. Cornet, D. Muster andJ. H. Jaeger,Biomater. Med. Dev. Art. Arg. 7 (1979) 155.
J. Kruger, ASTM STP684, (1979) 107.
L. G. Johnson, “The Statistical Treatment of Fatigue Experiments” (Elsevier, Amsterdam, 1964).
C. P. Sullivan, J. D. Varin andM. J. Donachie,Met. Eng. Quart, ASM (1969) 16.
J. Vander Sande, R. J. Coke andJ. Wulff,J. Met. Trans. 7A (1976) 389.
F. R. Morral,J. Met. 20 (1968) 52.
S. Ahier, PhD thesis, University of Surrey (1969).
A. J. T. Clemow andB. L. Daniell,J. Biomed. Mater. Res. 13 (1979) 265.
R. W. Revie andN. D. Greene,Corros. Sci. 9 (1969) 755.
H. S. Dobbs andM. J. Minskl,Biomater. 1 (1980) 193.
H. S. Dobbs andJ. L. M. Robertson,Eng. Med. 11 (1982) 175.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Dobbs, H.S., Robertson, J.L.M. Heat treatment of cast Co-Cr-Mo for orthopaedic implant use. J Mater Sci 18, 391–401 (1983). https://doi.org/10.1007/BF00560627