Summary
Reproducibility of lateral spine dual energy X-ray absorptiometry (LAT DEXA) scans using a Lunar DPX-L scanner was assessed in a cadaveric phantom and in patients. One hundred phantom measurements over 7 months demonstrated a longitudinal stability of 1.7% (coefficient of variation, CV). Additional scans were performed with the phantom rotated by up to 20° in each of the three orthogonal planes to assess the effects of variable patient positioning. Horizontal and vertical rotation of the spine had little effect on the estimated bone mineral density (BMD), however, axial rotation of greater than 8° led to errors in the BMD measurement. One hundred consecutive patients had two lateral scans performed within 1 month. BMD (range 0.10–1.6 g/cm2) was determined for each scan by one operator. Significant overlap from ribs and pelvis was often seen with L2 and L4 vertebrae but one vertebra (L3) could be measured in every case. Intraoperator and interoperator variability was assessed by three experienced operators, each analyzing 10 patients' scans on five separate occasions, and was found to be less than 1.1% for a single vertebra. BMD estimation of vertebral bodies and midslices by lateral DEXA scans (CV% of 3.8% and 4.6%) have a 95% confidence interval of 0.074 g/cm2 and 0.096 g/cm2, respectively for two vertebrae. This variability is due mainly to axial rotation, with operator variability, horizontal rotation, and vertical rotation having little effect on BMD estimation.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Valkema R, Verheij LF, Blokland JAK, Zwinderman AH, Bijvoet OLM, Papapoulos SE, Pauwels EKJ (1990) Limited precision of lumbar spine dual-photon absorptiometry by variations in the soft-tissue background. J Nucl Med 31(11):1774–1781
Genant HK (1990) Methods of bone mineral density. In: Ring EFJ (ed). Current research in osteoporosis and bone mineral measurement. p 15
Uebelhart D, Duboef F, Meunier PJ, Delmas PD (1990) Lateral dual-photon absorptiometry: a new technique to measure the bone mineral density at the lumbar spine. J Bone Miner Res 5(5):525–531
Pouilles J-M, Tremollieres F, Todorovsky N, Ribot C (1991) Precision and sensitivity of dual-energy x-ray absorptiometry in spinal osteoporosis. J Bone Min Res 6:997–1001
Rupich R, Pacifici R, Griffin M, Vered I, Susman N, Acioli LV (1990) Lateral dual energy radiography: a new method for measuring vertebral bone density. J Clin Endocrinol Metab 70(2): 1768–1770
Pocock NA, Eberl S, Eisman JA, Yeates MG, Sambrock PN, Freund J, Duncan A (1987) Dual-photon bone densitometry in normal Australian women: a cross-section study. Med J Australia 146:293–297
Duboeuf F, Meunier PJ, Delmas PD (1991) Lateral measurement of the bone mineral density (BMD) of the lumbar spine measured with a Lunar DPX L: preliminary results. Calcif Tissue Int 48(suppl):A76
Lunar DPX-L Technical Manual: Documentation Version 7/91. Lunar Corporation, Madison, Wisconsin
DPX bone densitometer operator's manual: documentation version 8/90 (1990) Lunar Corporation, Madison, Wisconsin
Nilas L, Hassager C, Christiansen C (1988) Long-term precision of dual photon absorptiometry in the lumbar spine in clinical settings. J Bone Miner Res 3(4):305–315
Kraenzlin ME, Kraenzlin CA, Haas HG (1991) Dual energy absorptiometry (anterior-posterior and lateral) versus quantitative computer tomography in the assessment of spinal bone mass. Calcif Tissue Int 48(suppl): A77
Eberl S, Johnson S, Yeates MG, Pocock NA, Ramsay S, Hille NA, Kelly PJ, Sambrock PN, Eisman JA (1990) Reproducibility of X-ray bone mineral density analyser. Aust N Z J Med 20(3)(suppl):445
Denton MD, Eberle RW, Scott JC (1990) Bone mineral density in lateral projection using DEXA. J Bone Miner Res (suppl) 5: 2:S179
von Stetton EC, Steiger P, Kelly TL, Stein JA (1991) Paired AP and lateral spine dual x-ray absorptiometry improves precision of repeat lateral scans. Bone Mineral Conference, Bad Reichenhall, Germany
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Larnach, T.A., Boyd, S.J., Smart, R.C. et al. Reproducibility of lateral spine scans using dual energy X-ray absorptiometry. Calcif Tissue Int 51, 255–258 (1992). https://doi.org/10.1007/BF00334484
Received:
Revised:
Issue Date:
DOI: https://doi.org/10.1007/BF00334484