Summary
Detailed studies for evaluating the optimum nickel sulphide button size in the estimation of platinum group elements (PGE) by inductively coupled plasma mass spectrometry (ICPMS) were carried out. PGE losses in various stages of the process have been studied. The results indicate that among the three button sizes investigated, i.e., 2.5, 5 and 8 g, the 2.5 g button was found to be adequate for the full recovery of the PGE. Losses of PGE in the fusion slag and in the pulp at the second filtration stage have been found to be negligible. Tellurium coprecipitation was found to minimize the losses during the dissolution of the nickel sulphide button.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Robert RVD, Van Wyk E, Palmer R (1971) Natl Inst Metall, Johannesburg, Rep No 1371, 1–5
Date AR, Davis AE, Cheung YY (1987) Analyst 112:1217–1222
Asif M, Parry SJ (1989) Analyst 114:1057–1059
Jackson SE, Fryer BJ, Gosse W, Healey DC, Longerich HP, Strong DF (1990) Chemical Geology 83:119–132
Steele TW, Levin J, Copelowitz T (1975) Natl Inst Metall, Johannesburg, Rep No 1696, 1–50
Bowditch DC (1973) Bull Austral Mining Develop Labs (AMDEL) 15:71–87
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Reddi, G.S., Rao, C.R.M., Rao, T.A.S. et al. Nickel sulphide fire assay — ICPMS method for the determination of platinum group elements: a detailed study on the recovery and losses at different stages. Fresenius J Anal Chem 348, 350–352 (1994). https://doi.org/10.1007/BF00323133
Received:
Revised:
Issue Date:
DOI: https://doi.org/10.1007/BF00323133