Summary
Designs for flexible ureteropyeloscopes should be based on the configuration of the intrarenal collecting system. This study measured retrospectively the ureteroinfundibular angle and infundibular lengths of 30 patients treated for calculus disease. Recommendations are given for the design of flexible ureteropyeloscopes that could visualize the entire normal intrarenal collecting system.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Aso Y, Ohtawara Y, Suzuki K, Tajima A, Fujita K (1984) Usefulness of fiberoptic pyeloureteroscope in the diagnosis of the upper urinary tract lesions. Urol Int 39: 355–357
Bagley DH (1987) Ureteral endoscopy with passively deflectable, irrigating flexible ureteroscopes. Urology 29: 170–173
Bagley DH, Huffman JL, Lyon ES (1987) Flexible ureteropyeloscopy: diagnosis and treatment in the upper urinary tract. J Urol (in press)
Marshall VF (1964) Fiberoptics in urology. J Urol 91: 110–114
Takagi T, Go T, Takayasu H, Aso Y (1971) Fiberoptic pyeloureteroscope. Surgery 70: 661–666
Takayasu H, Aso Y (1974) Recent development for pyeloureteroscopy: guide tube method for the introduction into the ureter. J Urol 112: 176–178
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Bagley, D.H., Rittenberg, M.H. Intrarenal dimensions. Surg Endosc 1, 119–121 (1987). https://doi.org/10.1007/BF00312699
Issue Date:
DOI: https://doi.org/10.1007/BF00312699