Abstract
There had been attempts to utilize spectral information for tissue characterization soon after the invention of Computed Tomography, but only recently Dual Energy CT has achieved a significant role in clinical radiology.
To perform Dual Energy CT, it is necessary to generate x-rays with different energies, mostly as polychromatic spectra. On the other hand, the detector has to be capable to differentiate x-ray quanta of different energies. There are four technical approaches to meet these requirements, of which the Dual Source CT, the rapid voltage switching and the layer detector technology are available or being implemented.
To obtain relevant diagnostic information, there have to be substances with spectral properties which reflect the pathology by their presence or distribution. Most important is the photoelectric effect of elements like uric acid, iron, calcium, iodine or xenon gas, which are present in pathological structures or can be administered as contrast material. The identification and quantification of these elements can be used to diagnose several diseases.
Access provided by Autonomous University of Puebla. Download chapter PDF
Similar content being viewed by others
Keywords
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.
1 History
First attempts to use spectral information in computed tomography date back to the late 1970s (Millner et al. 1979; Avrin et al. 1978; Chiro et al. 1979; Genant and Boyd 1977). At that time, two separate scans were acquired and either projection data or reconstructed data were postprocessed. However, the lacking stability of the CT density values, the long scan times which often caused patient motion between scans, the limited spatial resolution, and the difficulty of postprocessing were the main reasons why the method never achieved broad clinical acceptance (Kelcz et al. 1979). With the necessity to acquire both scans separately, the use of contrast material and its differentiation by dual-energy or spectral analysis was impossible. Since 2006, dual-energy CT has experienced a revival and success in clinical application. The main reason was the introduction of dual-source CT that made it possible to acquire two spiral scans simultaneously with different X-ray spectra by running the two X-ray sources at different tube voltages (Flohr et al. 2006; Johnson et al. 2007a). Today, there are several technical approaches to dual-energy CT.
2 X-ray Spectra
Generally, there are three requirements for spectral CT imaging: First, X-ray sources providing X-ray quanta with different energies are necessary. Obviously, this can be achieved optimally when using two separate X-ray sources emitting different photon energies. However, this criterion is basically also met by single X-ray sources, because they generally provide polychromatic spectra. The X-ray spectrum of a normal tube with a tungsten anode consists of a continuous part, the Bremsstrahlung, and discrete peaks according to the energy levels of the electrons in the shell of the tungsten atoms. Figure 1 shows the X-ray spectra that are obtained from an X-ray tube operated at 140 and 80 kV. Obviously, the area under the curve for equivalent tube currents differs by a factor of about 4–5, and the tube current needs to be adapted to obtain a similar output of quanta from both tubes. The higher energy spectrum is dominated by the characteristic lines of the tungsten anode, while the lower energy spectrum mainly consists of Bremsstrahlung. In this case, the mean photon energies are 71 and 53 keV, respectively, for the primary spectra. It is desirable to have as little overlap as possible between the spectra; therefore, the lowest and highest potentials offered by the respective CT scanner are always used for dual-energy acquisitions. A tube voltage lower than 80 kV is not useful because too much of the quanta would be absorbed by the human body; values higher than 140 kV are generally not available and would result in very low soft tissue contrast that likely would contribute very little to tissue differentiation. Obviously, it is not possible to obtain monochromatic X-rays with today’s tube technology; monochromatic X-ray sources so far do not provide a sufficient output of quanta for clinical applications. An exception is monochromatic synchrotron radiation that can be used for experimental setups. As synchrotrons are far too large for use in a rotating gantry, the object has to be rotated between the source and the detector, i.e., it is not possible to use them for clinical imaging.
3 Detector Technology
The second requirement is that the detector has to be able to differentiate quanta of different energies. With technologies in use in clinical CT today, this is not directly possible with a single detector. The detector integrates the fluorescent light intensities of all photons detected during a single readout interval, but does not give account of their energy. Current approaches either rely on entirely separate X-ray sources and corresponding separate detectors, on reading out the projection data at different time points, or on using a two-layer or “sandwich” detector with different spectral sensitivities. In the near future, cadmium-based materials such as CdZnTe may serve as semiconductors for photon-counting detectors that resolve the energy of each individual photon. However, this detector technology cannot yet cope with the high photon flux required for clinical CT.
These first two requirements define the spectra of the X-ray sources and the corresponding detectors, representing the sensitivity of the systems for photons, detectors of different energies. The more these spectra differ, the stronger the contrast-to-noise ratio of the resulting spectral information, as long as both sufficient transmission and attenuation can be achieved in the human body at these photon energies.
4 Tissue Properties
The third requirement is a sufficient difference in spectral properties of the materials under investigation. The attenuation that is quantified in CT to characterize different body tissues is caused by three physical processes. Compton scatter is the largest component of attenuation. It is related to the electron density and not to the protons in the atomic core that would allow a material differentiation (McCullough 1975). Rayleigh scatter also is related to the electrons and not to the atomic core, but contributes only a negligible amount. The third process, the photoelectric effect, is strongly related to the atomic number, Z, of the material, i.e., to the number of protons of the atomic core (Fig. 2). Therefore, only elements with a considerable difference in Z values will be distinguishable by their spectral properties. This difference can be characterized with the so-called dual-energy index (DEI):
where x80 is the CT value in HU at 80 kV tube potential and x140 the value of the respective voxel at 140 kV (Johnson et al. 2007a).
According to Alvarez and Macovski (1976), the photoelectric interaction with the K shell is proportional to the third power of the atomic number (Z). Therefore, high values apply for Z values of 53 (iodine) or 54 (xenon). The elements of which the human body consists, i.e., hydrogen (Z = 1), oxygen (Z = 8), carbon (Z = 6) and nitrogen (7), have low Z numbers and hence do not show a sufficient photo effect and spectral behavior that would allow a differentiation (Michael 1992). Bone with its high content in calcium (Z = 20) and fat, which only consists of hydrogen and carbon, represent tissues that differ from other body tissues significantly, which also explains their good differentiation in standard CT. This feature has also been employed in approaches for the quantification of obesity or for the identification of calcifications in pulmonary nodules (Cann et al. 1982; Svendsen et al. 1993). The most clinically useful application of dual-energy CT can be expected for the differentiation of iodine (Kruger et al. 1977; Riederer and Mistretta 1977; Nakayama et al. 2005), which is generally used in CT as a contrast agent and the distribution of which can be masked by the underlying tissue.
5 Dual-Source CT
At present, there are three different approaches to clinical dual-energy CT commercially available or under investigation. The most straightforward approach is dual-source CT with two X-ray sources running at different voltages with two corresponding detectors (Fig. 2). Additionally, a filter can be used to further harden the higher energy spectrum, i.e., to rid the spectrum of low-energy quanta. This setup has several advantages. First, established X-ray sources and detector materials can be used. Also, the image reconstruction can rely on established methods. But the most important benefit of this technology is that the tube voltage and current can be adjusted freely to obtain the largest possible difference in photon energies with similar total amounts of quanta from both tubes. A disadvantage of this approach is the considerable hardware effort, making the system expensive. Also, the space in the CT gantry is only sufficient for a smaller second detector so that the field of view of the dual-energy scans is restricted. The acquired projection data primarily have to be reconstructed by standard filtered backprojection, separately for the two simultaneously acquired spiral datasets. The fact that the acquired projection data have an offset of 90° at equal z-axis positions means that a primary postprocessing of projection data is impossible because there are no equivalent projections. Still, data at equivalent z-positions are sampled simultaneously, so the angular offset does not imply a temporal offset between the acquisitions. A problem is posed by the cross-scatter that is oriented perpendicular to the primary course of beam, thus hitting the second detector and contaminating its data. However, comparative Monte Carlo simulations show a good spectral separation with dual-source CT (Fig. 3) (Kappler et al. 2009).
6 Rapid Voltage Switching
Another approach that requires less hardware effort is rapid voltage switching. With this method, the tube voltage follows a pulsed curve, and projection data is collected twice for every projection, one at high and one at low tube voltage (Fig. 4). Optimally, the tube current should be modulated at least inversely to the voltage so that the difference in the number of photons is kept as small as possible (Grasruck et al. 2009). The major advantage of this approach is that it can be realized at considerably lower cost, because no major additional hardware is required. Limitations include the slow acquisition, because the rotation time has to be reduced to less than half to allow the collection of the additional projections. Also, the course of the tube current and voltage retain a non-rectangular, curved shape so that the resulting spectral difference does not correspond to the nominal tube voltages. The adaptation of the current also has limitations so that there is generally a far lower signal at low energy than at high energy. Figure 5 shows the resulting spectra with a rather large difference in size.
7 Layer Detector
A third approach is not to generate different photon spectra but to work with two detector layers that have their maximum sensitivity for different photon energies (Fig. 6). The sensitivity is determined by the scintillator material, e.g., consisting of ZnSe or CsI in the top and Gd2O2S in the bottom layer. This setup has the advantage that only one standard tube is required. Disadvantages include the hardware effort for the layer detector and the lower dose efficiency of such a setup. Also, the obtainable spectral difference is rather low so that the contrast of the spectral information is limited or requires additional dose. Figure 7 shows the sensitivity profile of a layer detector with ZnSe and Gd2O2S scintillators.
8 Sequential Acquisition
Another approach that is not being implemented currently is sequential filtering or voltage switching, i.e., two subsequent rotations are acquired in a sequence mode at the same table position with different tube voltages and optionally an additional filter. This would imply a rather minor hardware effort and may be a viable option especially for systems with broad detectors and a rather high number of simultaneously acquired slices. An obvious disadvantage is the rather long delay between both acquisitions, which is too long to preclude artifacts from cardiac or respiratory motion or changes in contrast material opacification.
9 Radiation Exposure
Regarding radiation exposure, dual-energy CT based on dual-source CT acquisitions does not require a higher patient dose than a routine CT scan of the same body region. It is possible to tailor the tube current such that the dose from both tubes matches that of a routine single-source CT protocol (Johnson et al. 2007a). Recent comparative phantom studies with external validation have proven that dual-energy acquisitions can provide similar or even improved contrast-to-noise ratios at equivalent dose (Schenzle et al. 2010). This is in contrast to investigations by Ho et al. 2009 who observed two to three times higher doses for dual-energy CT. However, their setup was based on a single-source system using rapid voltage switching and contained neither a normalization of image noise nor of dose, so the lower energy spectrum was obtained with the same tube current time product as the single energy scan. In another study with normalization to equivalent low-contrast detectability (Li et al. 2010), Li et al. observed an additional dose of 14% in the body and 22% in the head with rapid kV switching. Specific clinical studies comparing the dose efficiency of different dual-energy CT systems are lacking, but it is evident that dual-energy CT does not necessarily imply an increased dose.
10 Clinical Applications
Meanwhile there are numerous clinical applications of dual-energy CT. Among these are bone removal from the carotid (Morhard et al. 2009) or peripheral runoff (Sommer et al. 2009) CT angiography datasets, the reconstruction of virtual noncontrast images or quantification of iodine enhancement in lesions of solid organs (Graser et al. 2009), the depiction of iodine (Thieme et al. 2008; 2009) or xenon gas (Chae et al. 2008) distribution in the lung and the differentiation of kidney stones (Graser et al. 2008), or the identification of gout tophi (Johnson et al. 2007b). Also, image quality and display can be optimized, e.g., by generating monoenergetic images (Voit et al. 2009), or images with optimized contrast (Holmes et al. 2008). As this more specific or functional additional information can be obtained without an additional dose, it should be exploited whenever diagnostically useful.
11 Summary
In summary, dual-energy CT offers the possibility to exploit spectral information for diagnostic purposes. There are different technical approaches that all have inherent advantages and disadvantages. Some systems are commercially available, and there is already quite a number of well-established clinical applications.
References
Alvarez RE, Macovski A (1976) Energy-selective reconstructions in X-ray computerized tomography. Phys Med Biol 21:733–744
Avrin DE, Macovski A, Zatz LE (1978) Clinical application of Compton and photo-electric reconstruction in computed tomography: preliminary results. Invest Radiol 13:217–222
Cann CE, Gamsu G, Birnberg FA, Webb WR (1982) Quantification of calcium in solitary pulmonary nodules using single- and dual-energy CT. Radiology 145:493–496
Chae EJ, Seo JB, Goo HW et al (2008) Xenon ventilation CT with a dual-energy technique of dual-source CT: initial experience. Radiology 248:615–624
Chiro GD, Brooks RA, Kessler RM et al (1979) Tissue signatures with dual-energy computed tomography. Radiology 131:521–523
Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268
Genant HK, Boyd D (1977) Quantitative bone mineral analysis using dual energy computed tomography. Invest Radiol 12:545–551
Graser A, Johnson TR, Bader M et al (2008) Dual energy CT characterization of urinary calculi: initial in vitro and clinical experience. Invest Radiol 43:112–119
Graser A, Johnson TR, Hecht EM et al (2009) Dual-energy CT in patients suspected of having renal masses: can virtual nonenhanced images replace true nonenhanced images? Radiology 252:433–440
Grasruck M, Kappler S, Reinwand M, Stierstorfer K (2009) Dual energy with dual source CT and kVp switching with single source CT: a comparison of dual energy performance. Proc SPIE 7258:72583R
Ho LM, Yoshizumi TT, Hurwitz LM et al (2009) Dual energy versus single energy MDCT: measurement of radiation dose using adult abdominal imaging protocols. Acad Radiol 16:1400–1407
Holmes DR III, Fletcher JG, Apel A et al (2008) Evaluation of non-linear blending in dual-energy computed tomography. Eur J Radiol 68:409–413
Johnson TR, Krauss B, Sedlmair M et al (2007a) Material differentiation by dual energy CT: initial experience. Eur Radiol 17:1510–1517
Johnson TR, Weckbach S, Kellner H, Reiser MF, Becker CR (2007b) Clinical image: dual-energy computed tomographic molecular imaging of gout. Arthritis Rheum 56:2809
Kappler S, Grasruck M, Niederloehner D, Strassburg M, Wirth S (2009) Dual-energy performance of dual kVp in comparison to dual-layer and quantum-counting CT system concepts. Proc SPIE 7258:725842
Kelcz F, Joseph PM, Hilal SK (1979) Noise considerations in dual energy CT scanning. Med Phys 6:418–425
Kruger RA, Riederer SJ, Mistretta CA (1977) Relative properties of tomography, K-edge imaging, and K-edge tomography. Med Phys 4:244–249
Li B, Yadava G, Hsieh J (2010) Head and body CTDIw of dual energy x-ray CT with fast-kVp switching. In: SPIE Medical Imaging, San Diego, CA, paper 7622–7669
McCullough EC (1975) Photon attenuation in computed tomography. Med Phys 2:307–320
Michael GJ (1992) Tissue analysis using dual energy CT. Australas Phys Eng Sci Med 15:75–87
Millner MR, McDavid WD, Waggener RG, Dennis MJ, Payne WH, Sank VJ (1979) Extraction of information from CT scans at different energies. Med Phys 6:70–71
Morhard D, Fink C, Graser A, Reiser MF, Becker C, Johnson TR (2009) Cervical and cranial computed tomographic angiography with automated bone removal: dual energy computed tomography versus standard computed tomography. Invest Radiol 44:293–297
Nakayama Y, Awai K, Funama Y et al (2005) Abdominal CT with low tube voltage: preliminary observations about radiation dose, contrast enhancement, image quality, and noise. Radiology 237:945–951
Riederer SJ, Mistretta CA (1977) Selective iodine imaging using K-edge energies in computerized x-ray tomography. Med Phys 4:474–481
Schenzle JC, Sommer WH, Neumaier K et al (2010) Dual energy CT of the chest: how about the dose? Invest Radiol 45:347–353
Sommer WH, Johnson TR, Becker CR et al (2009) The value of dual-energy bone removal in maximum intensity projections of lower extremity computed tomography angiography. Invest Radiol 44:285–292
Svendsen OL, Hassager C, Bergmann I, Christiansen C (1993) Measurement of abdominal and intra-abdominal fat in postmenopausal women by dual energy X-ray absorptiometry and anthropometry: comparison with computerized tomography. Int J Obes Relat Metab Disord 17:45–51
Thieme SF, Becker CR, Hacker M, Nikolaou K, Reiser MF, Johnson TR (2008) Dual energy CT for the assessment of lung perfusion–correlation to scintigraphy. Eur J Radiol 68:369–374
Thieme SF, Johnson TR, Lee C et al (2009) Dual-energy CT for the assessment of contrast material distribution in the pulmonary parenchyma. AJR Am J Roentgenol 193:144–149
Voit H, Krauss B, Heinrich MC et al (2009) Dual-source CT: in vitro characterization of gallstones using dual energy analysis. Rofo 181:367–373
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2011 Springer-Verlag Berlin Heidelberg
About this chapter
Cite this chapter
Johnson, T.R.C., Kalender, W.A. (2011). Physical Background. In: Johnson, T., Fink, C., Schönberg, S., Reiser, M. (eds) Dual Energy CT in Clinical Practice. Medical Radiology(). Springer, Berlin, Heidelberg. https://doi.org/10.1007/174_2010_43
Download citation
DOI: https://doi.org/10.1007/174_2010_43
Published:
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-01739-1
Online ISBN: 978-3-642-01740-7
eBook Packages: MedicineMedicine (R0)