Abstract
Retrospective analysis of neonatal cranial ultrasonic movies has been effectively realized by developing a novel movie analysis system. Despite the recognition of the diagnostic potential of neonatal cranial ultrasonic movies, several issues still persist in handling a large-scale archive of approximately 1000 cases. These issues include extensive, redundant computations and inconsistent data. The proposed system was designed to address these issues through a systematic organization of processes. It includes a primary layer for fully-automated preprocessing such as frequency spectrum analysis and several consequent layers for semi-automated retrospective analysis such as detection of probe-stabilized scene and assessment of pulsatile tissues. Four test cases were examined by the conventional and novel systems to evaluate the performance. The results demonstrated that the novel system performed essential tasks three to five times faster, while also providing new capabilities that the conventional system lacks. Overall, the proposed system confirmed a significant improvement of efficiency in retrospective analysis compared with the conventional process flow and the retrospective analysis has been realized to the entire archive for the first time, revealing its clinical applicability for further analysis of ischemic diseases.
Access provided by Autonomous University of Puebla. Download conference paper PDF
Similar content being viewed by others
Keywords
1 Introduction
Ultrasonic (US) movie is widely used by pediatricians at neonatal intensive care unit (NICU) to observe the brain tissues through the anterior fontanelle of neonates with potential risks of ischemic and related diseases such as Hypoxic-Ischemic Encephalopathy (HIE) [1] and Intraventricular Hemorrhage (IVH) [2, 3]. Since early detection of these diseases is essential for timely intervention, pediatricians have diagnosed such abnormalities from the shape and motion of brain tissues in brightness mode images and movies while taking care of some factors such as choice of suitable equipment, optimal settings, appropriate scanning protocols, and scanning experience to ensure its image quality [4].
One of cardinal viewpoints in the diagnosis of ischemic diseases is to observe a pulsatile motion of brain tissues due to blood flow of neighboring arteries. In order to assist such pediatric diagnosis, we have conducted several studies in detecting and visualizing the pulsatile tissues from US movies with time-frequency analysis [5,6,7,8]. It was based on a heartbeat-frequency component I(fHB) in the frequency spectrum obtained from a time-variant echo intensity of each pixel over 64 frames in a movie fragment extracted from a scene without any motion of the US probe. I(fHB) was provided to assist pediatric diagnosis by visualizing as a superimposed image while overlaying a color gradation proportional to I(fHB) onto the original US image [8]. It was also given by statistically analyzing as a pulsatile tissue measures. These procedures are shown in Fig. 1. An extensive archive of neonatal cranial US movies was also prepared for a variety of studies, and its retrospective analysis was examined by using a conventional analysis system. However, it was problematic to handle the entire archive due to several issues in the conventional analysis system such as extensive, redundant computations and inconsistent data.
In this study, we introduce an approach for handling the entire archive by developing a novel analysis system dedicated to neonatal cranial US movies of ischemic diseases while mitigating potential challenges associated with our conventional system. Additionally, several test cases are formulated to compare the performance of two systems.
The reminder of this paper is as follows. In Sect. 2, the archive associated with the conventional image analysis system and its issues are presented. Sect. 3 proposes a novel image analysis system to alleviate such issues. Next, Sect. 4 illustrates some experiments to evaluate the novel system performance. Finally, achievements of this paper are summarized in Sect. 5.
2 Movie Archive and Issues in Its Retrospective Analysis Using the Conventional System
The movie archive utilized in this study consists of an immense volume of cranial US movies with approximately 1000 cases recorded at NICUs in the previous affiliated institutions (Himeji Red Cross Hospital and Saiseikai Hyogo-ken Hospital) of the author (Kitsunezuka, Y.) and provided anonymously under the approval of the affiliated department, or the ethics committee after its establishment. It consists of around 213 h of movie and encompassing around 1000 cases. Table 1. shows the statistics of the examined archive. Since the US apparatus was different between the groups of different recording periods, the resolution and the fan-shaped field of view (FOV) should be considered for each group.
The process flow in the previous studies using the conventional analysis system is illustrated in Fig. 2. In this figure, rounded and angular rectangles indicate the processes and their resultant data, respectively. The gray background indicates a process that requires manual intervention for each case or movie fragment, while the white one means what can be performed for multiple cases or fragments automatically at once.
In that conventional process flow, four types of manual process were required for each scene prior to the time-frequency analysis in order to select appropriate scenes for detecting pulsatile tissues. They are: 1) the anatomical identification of tomographic section and diagnosis annotation, 2) probe-stabilized scene detection to eliminate the visual incoherence due to the sway of probe, 3) fHB estimation according to the heartbeat frequency for each neonate, and 4) cranial ROI detection to focus on the area of brain tissues. All of them are performed by visual observation of the movie.
The case selection process was repeatedly invoked each time a certain purpose of study was considered for corresponding analysis and visualization. Therefore, it was difficult to investigate the relationship between pulsatile tissues and pathologic conditions. The problem of this process flow revolved around the uncertainty of the success or failure of all the manual processes until the visualization of the time-frequency analysis were observed. For example, the probe-stabilized scene was not valid until the I(fHB) superimposed image was observed to confirm that it did not contain motion artifacts, as shown in Fig. 3. If it contained motion artifacts, we had to re-extract the scene and perform the time-frequency analysis again. In most cases, that process required some iterations with the feedback from the results of the time-frequency analysis. Since each of the other manual processes also required its own verification, a complete manual process included complex iterations. Therefore, the analysis of the entire archive was challenging because as more scenes were processed, various problems with iterative processes arose such as extensive, redundant computations and inconsistent data.
3 A Proposal of the Novel Image Analysis System
To tackle the issues in the conventional process flow, a novel image analysis system has been proposed. Figure 4 shows the systematic process flow enabled by that system. The shape and the background color of the rectangle in Fig. 4 indicate the same process type as those shown in Fig. 2. The novel system adopts a multilayered architecture to arrange processes into distinct layers including: 1) primary layer, which prepares extensive amount of intermediate data, 2) refinement layer, which filters or transforms intermediate data into more lightweight and concentrated data, and 3) exploratory layer, which discovers data patterns that are associated with the pathologic conditions.
One of the important design features of this system is that all the processes are performed on each movie fragment rather than on each movie scene. In addition, prior to any manual process, time-frequency analysis and its visualization are performed on the entire archive, regardless of the appropriateness for detecting pulsatile tissues, and the results are recorded as an intermediate data. An element of the intermediate data consists of a fragment spectrum that includes the whole spectrum I(f) corresponding to a movie fragment, and an I(fHB) superimposed image made from the initial frame of the fragment. Example of spectrum I(f) and corresponding I(fHB) superimposed image was reflected in Fig. 1. Since a huge number of intermediate data was generated to cover the entire archive, it can be directly used for subsequent processes without regeneration, which is a clear advantage over the conventional system. Therefore, this design feature allowed all the processes of intermediate data generation to be positioned as the primary, feedback-free layer in the process flow.
Another important feature of this system is to efficiently perform the manual processes by using both intermediate data and the US movies. Among the manual processes, the section identification and ROI detection should be performed by visual observation of US movies since it is based on anatomical knowledge. However, the probe-stabilized scene can be effectively detected, by visual observation of a series of I(fHB) superimposed images to ensure the absence of motion artifacts [9]. Furthermore, fHB can be effectively estimated from the peak frequency of the fragment spectrum. Since iteration was no longer required for all the manual processes, they were positioned as a refinement layer in the process flow to generate pulsatile tissue measures from the intermediate data. In other words, this layer refines the large amount of data from the primary layer into more focused data for further analysis.
The final output of this flow is the result of correlation analysis between the diagnosis data and the pulsatile tissue measures to assist pediatric diagnosis. Since it is expected to apply several pediatric diagnoses such as prognosis of HIE and risk assessment of IVH, this process was positioned as the exploratory layer in the process flow and designed to be performed for a certain subset of the archive.
In our system, a fragment was constructed from 64 frames (2 s), and Blackman was selected as the windowing function, by considering the appropriateness to handle the actual fHB of neonates (2 to 3 Hz) with the standard video rate (29.97 fps).
4 Evaluation of System Performance
In order to compare the performance between the conventional and the proposed systems, four test cases were designed and conducted as follows.
-
1.
Time to extract the valid probe-stabilized scenes: This test aims to evaluate the time required to extract and validate all the probe-stabilized scenes from a 20-min US movie of a certain neonate.
-
2.
Time to calculate the pulsatile tissue measures: This test aims to evaluate the time required to calculate the pulsatile tissue measures from the 1200 probe-stabilized scenes for 100 neonates with a common fan-shape ROI.
-
3.
Applicability to the prospective studies in NICU: This test aims to examine the system applicability for a kind of prospective study in the NICU to compare an ongoing disease with the previous cases. We gave a new US movie of an ongoing case and specified 3 previous cases in the archive, and tested the ability to generate and visualize the pulsatile tissue images of them within 1 h, assuming an acceptable latency in NICU. The pulsatile tissue image means a special I(fHB) superimposed image generated from the validated probe-stabilized scenes.
-
4.
Applicability to the retrospective (case-control) studies: This test aims to examine the system applicability for retrospective studies to compare two case groups in the archive. We specified two case groups (HIE positive and negative) including around 10 patients for each in the archive and tested the ability to extract a certain pulsatile tissue measure and to visualize its sectional or time dependence.
Table 2. shows the results of four test cases. The extraction of valid probe-stabilized scenes was five times faster in the novel system. The primary reason for this result is the ability of the pre-generated I(fHB) superimposed images to eliminate the iteration of manual extraction of probe-stabilized scenes. Likewise, the calculation of pulsatile tissue measures was three times faster in the novel system because it utilized the pre-generated fragment spectrum and the refinement data.
The novel system succeeded in instantly generating the pulsatile-tissue images for both ongoing and multiple previous cases, while the conventional system failed to do so due to time-consuming iterative processes. From this result, it was found that the applicability to prospective studies in the NICU was realized exclusively in the novel system.
The novel system also succeeded in extracting the pulsatile tissue measures and visualizing their sectional or time variation in specified two case groups, while it was impossible for the conventional system because the valid probe-stabilized scene was unknown. Figure 5 shows examples of pulsatile tissue images and corresponding measures obtained in HIE negative and positive case groups retrospectively. A normalized pulsatile area in a fan-shape ROI [10] was used as the pulsatile tissue measure in this experiment and noted at the upper right of each image. The pulsatile tissue image clearly revealed different distribution of pulsatile tissues slice by slice as well as case by case. The pulsatile tissue measure revealed a significant difference between HIE positive and negative groups. It should be emphasized that the symptom of HIE was diagnosed a few weeks after birth, while the US movie was taken just after birth. Therefore, this result strongly suggests the applicability of pulsatile tissue measures to predict the prognosis of HIE in the early postnatal stage.
5 Conclusion
This research presented a layer-based analysis system to deal with a large-scale archive of neonatal cranial US movies while successfully eliminating the intensive and redundant computations and inconsistent data. The system was designed to include a primary layer for fully-automated preprocessing such as frequency spectrum analysis and several consequent layers for semi-automated retrospective analysis such as detection of probe-stabilized scene and assessment of pulsatile tissues. Four test cases were examined by the conventional and novel systems to evaluate the performance. The results demonstrated that the novel system performed essential tasks three to five times faster, while also providing new capabilities that the conventional system lacks. Besides, the solution also showcases the potential applications for analysis of periodical phenomena in other mass movie repositories than our archive. In summary, the proposed system successfully handled the large-scale archive of US movies, accelerating its retrospective analysis and revealing the clinical applicability of the system.
References
Kurinczuk, J.J., White-Koning, M., Badawi, N.: Epidemiology of neonatal encephalopathy and hypoxic-ischaemic encephalopathy. Early Hum. Dev. 86(6), 329–338 (2010). https://doi.org/10.1016/j.earlhumdev.2010.05.010
Christian, E.A., et al.: Trends in hospitalization of preterm infants with intraventricular hemorrhage and hydrocephalus in the United States, 2000–2010. J. Neurosurg. Pediatr. 17(3), 260–269 (2016). https://doi.org/10.3171/2015.7.PEDS15140
de Figueiredo Vinagre, L.E., et al.: Temporal trends in intraventricular hemorrhage in preterm infants: a Brazilian multicenter cohort. Eur. J. Paediatr. Neurol. 39, 65–73 (2022). https://doi.org/10.1016/j.ejpn.2022.05.003
Leijser, L.M., Vries, L.S., Cowan, F.M.: Using cerebral ultrasound effectively in the newborn infant. Early Hum. Dev. 82(12), 827–835 (2006). https://doi.org/10.1016/j.earlhumdev.2006.09.018
Yamada, M., et al.: Pulsation detection from noisy ultrasound-echo moving images of newborn baby head using Fourier transform. Jpn. J. Appl. Phys. 34, 2854–2856 (1995). https://doi.org/10.1143/jjap.34.2854
Fukuzawa, M., Kitsunezuka, Y., Yamada, M.: A real-time processing system for pulsation detection in neonatal cranial ultrasonogram. Jpn. J. Appl. Phys. 37, 3106–3109 (1998). https://doi.org/10.1143/jjap.37.3106
Fukuzawa, M., Kubo, H., Kitsunezuka, Y., Yamada, M.: Motion analysis of artery pulsation in neonatal cranial ultrasonogram. In: Proceedings of the SPIE 3661, Medical Imaging 1999: Image Processing, California, United States (1999). https://doi.org/10.1117/12.348519
Fukuzawa, M., Yamada, M., Nakamori N., Kitsunezuka, Y.: A new imaging technique on strength and phase of pulsatile tissue-motion in brightness-mode ultrasonogram. In: Proceedings of the SPIE 6513, Medical Imaging 2007: Ultrasonic Imaging and Signal Processing, 65130B, California, United States (2007). https://doi.org/10.1117/12.709354
Tabata, Y., Fukuzawa, M., Izuwaki, Y., Nakamori, N., Kitsunezuka, Y.: Time-frequency analysis of neonatal cranial ultrasonic movies for selective detection of pulsatile tissues by avoiding probe-motion artifact. In: Proceedings of the SPIE 9419, Medical Imaging 2015: Ultrasonic Imaging and Tomography, 941913, Florida, United States (2015). https://doi.org/10.1117/12.2081383
Fukuzawa, M., Takahashi, K., Tabata, Y., Kitsunezuka, Y.: Effect of echo artifacts on characterization of pulsatile tissues in neonatal cranial ultrasonic movies. In: Proceedings of the SPIE 9790, Medical Imaging 2016: Ultrasonic Imaging and Tomography, 979014, California, United States (2016). https://doi.org/10.1117/12.2216751
Acknowledgement
The authors would like to thank Mr. Daiki Terai for his significant contribution to the initial stages of this study.
This work was supported by JSPS Core-to-Core Program (grant number: JPJSCCB20230005).
Author information
Authors and Affiliations
Corresponding authors
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2024 The Author(s), under exclusive license to Springer Nature Singapore Pte Ltd.
About this paper
Cite this paper
Nguyen, M.N., Kawamidori, R., Kitsunezuka, Y., Fukuzawa, M. (2024). Retrospective Analysis of a Large-Scale Archive of Ultrasonic Movies for Ischemic Diseases of Neonatal Brain. In: Thai-Nghe, N., Do, TN., Haddawy, P. (eds) Intelligent Systems and Data Science. ISDS 2023. Communications in Computer and Information Science, vol 1949. Springer, Singapore. https://doi.org/10.1007/978-981-99-7649-2_5
Download citation
DOI: https://doi.org/10.1007/978-981-99-7649-2_5
Published:
Publisher Name: Springer, Singapore
Print ISBN: 978-981-99-7648-5
Online ISBN: 978-981-99-7649-2
eBook Packages: Computer ScienceComputer Science (R0)