Summary
Failure of intraventricular pressure (IVP) measurement in case of catheter blockage is believed to be eliminated by using intraventricular microtransducers. We report about an avoidable methodological error, which may affect the reliability of IVP measurement with these devices. Intraventricular fiberoptic or solid state devices were implanted in 43 patients considered to be at risk for catheter occlusion. Two different types were used: devices where the transducer is placed inside the ventriculostomy catheter (Type A), and devices where the transducer is integrated in the external surface of the catheter (Type B).
Of the 15 patients treated with Type A devices, no reliable pressure recording could be obtained in three patients where ventricular puncture was not successful. In four cases of the remaining 12 patients, periods of erroneous pressure readings were revealed. After opening of CSF drainage, all Type A devices failed to reflect real IVP. In patients treated with Type B devices, no erroneous pressure recording could be identified, irrespective if CSF drainage was performed or not.
Transducers, which are simply placed inside the ventriculostomy catheter require fluid coupling. They may fail, either during CSF drainage or when the catheter is blocked or placed within the parenchyma.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Similar content being viewed by others
References
Feldman Z, Narayan RK (1993) Intracranial pressure monitoring: techniques and pitfalls. In: Cooper PR (ed) Head injury. Williams and Wilkins, Baltimore, pp 247–274
Luerssen TG, Shields PF, Vos HR, Marshall LF (1989) Clinical experience with fiberoptic brain parenchymal pressure monitor. In: Hoff JT, Betz AL (eds) Intracranial pressure VII. Springer, Berlin Heidelberg New York Tokyo, pp 35–37
Ostrup RC, Luerssen TG, Marshall LF, Zornow MH (1987) Continuous monitoring of intracranial pressure with a miniaturized fiberoptic device. J Neurosurg 67: 206–209
Piek J, Raes P (1996) Pressure-controlled drainage of cerebrospinal-fluid — clinical-experience with a new-type of ventricular catheter (ventcontrol mtc) and an integrated piezoresistive sensor at its tip-technical note. Neurosurgery 38: 216–218
Sundbarg G, Nordstrom CH, Messeter K, Soderstrom S (1987) A comparison of intraparenchymatous and intraventricular pressure recording in clinical practice. J Neurosurg 67: 841–845
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 1998 Springer-Verlag Wien
About this paper
Cite this paper
Raabe, A., Stöckel, R., Hohrein, D., Schöche, J. (1998). An Avoidable Methodological Failure in Intracranial Pressure Monitoring Using Fiberoptic or Solid State Devices. In: Marmarou, A., et al. Intracranial Pressure and Neuromonitoring in Brain Injury. Acta Neurochirurgica Supplements, vol 71. Springer, Vienna. https://doi.org/10.1007/978-3-7091-6475-4_19
Download citation
DOI: https://doi.org/10.1007/978-3-7091-6475-4_19
Publisher Name: Springer, Vienna
Print ISBN: 978-3-7091-7331-2
Online ISBN: 978-3-7091-6475-4
eBook Packages: Springer Book Archive