4.2 Tip and Pearls

  • Recommended laser settings in phonosurgery without the scanner are SuperPulse or UltraPulse, 0.1 second exposure duration of 2-3 W, with a focused beam with a 250 µm spot size for a working distance of 400 mm.

  • If microvasculature coagulation is required (e.g., chorditis vocalis or Reinke’s edema) before incision of the epithelium, laser is set at 0.05-second pulsed exposure of 1 W with a slightly defocused beam.

  • With the scanning system, the proposed parameters for phonosurgery are a depth of 0.2 mm, 10 W, single pulse, 0.10 second for SuperPulse; and two passes, 10 W, single pulse, 0.10 second for UltraPulse. The length of the incision line or the diameter of the surface to ablate was 1-2 mm depending on the lesion.

  • Continuous mode should never be used because of increased irradiance of tissues in this mode.

  • Beam parameters are for guidance only. A number of factors affect the ultimate effect of the beam on the pathological (and normal) tissue. The type of laser used, the lifespan of the device, the make, and the experience of the surgeon have a bearing on the outcome.