Increasing numbers of medical specialists are performing fluoroscopy-guided interventional procedures (FGIP) (Ann ICRP 30(2):7–67, 2000; Vano, Curr Radiol Rep 1:11–22, 2013). The use of medical ionizing radiation in the USA was reported sevenfold higher in 2006 compared to 1980, when the amount due to FGIP increased 33 times (American College of Radiology, ACR technical standard for management of the use of radiation in fluoroscopic procedures. Practice guidelines and technical standards 2003, American College of Radiology, Reston, 2003; National Council on Radiation Protection & Measurements (Bethesda, MD). Ionizing radiation exposure of the population of the United States. Report No.:160. Available from: http://www.ncrponline.org/PDFs/2012/DAS_DDM2_Athens_4-2012.pdf). The new international recommendations on radiation safety have led to national and international efforts to promote patient and staff radiation safety. Basic understanding of radiation safety principles in daily practice is a key to safe interventional radiology practice. This chapter will review basic principles of radiation physics, new technology aimed at dose reduction to staff and patients, as well as the basis of biological effects of ionizing radiation.
Access provided by CONRICYT-eBooks. Download chapter PDF
Similar content being viewed by others
Keywords
- Radiation protection
- Optimization
- Patient dosimetry
- Occupational dosimetry
- X-ray system
- Radiation risks
- Pregnancy
- Pediatrics
Introduction
Increasing numbers of medical specialists are performing fluoroscopy-guided interventional procedures (FGIP) [1, 2]. The use of medical ionizing radiation in the USA was reported sevenfold higher in 2006 compared to 1980, when the amount due to FGIP increased 33 times [3, 4]. The new international recommendations on radiation safety have led to national and international efforts to promote patient and staff radiation safety.
Key Points
-
Specialties that utilize image guidance:
-
Interventional radiology
-
Diagnostic radiology
-
Urology
-
Gastroenterology
-
Orthopedic surgery
-
Vascular surgery
-
Trauma and general surgery
-
Anesthesiology
-
Cardiology
Inherent in the growing use of medical radiation is a better understanding of the potential stochastic risks for cancer and the methods to monitor and reduce the risk of deterministic effects for skin injury. Modern angiography systems allow virtually unlimited exposure. CT or MR angiography are routinely used for most endovascular procedures. It is generally believed that the exposure to the staff is not significant and does not represent a real hazard. In fact, there is real risk to operators and staff of both tumor formation and damage to the eyes. Planning of each and every intervention should comprise radiation protection measures as part of the procedure [5]. Lack of radiation protection training of those working with fluoroscopy can increase the radiation risk to workers and patients alike. Patient dose monitoring is essential whenever fluoroscopy is used. The International Commission on Radiological Protection (ICRP) recommended that manufacturers should develop systems to indicate patient dose indices with the possibility of producing patient dose reports and shielding screens that can be effectively used for the protection of workers using fluoroscopy without hindering the clinical task [6].
Obesity is recognized worldwide as an epidemic causing devastating or fatal health disorders, such as diabetes and heart disease [7, 8]. The scatter radiation exposure to the operator’s waist increases dramatically with obese patients. It doubles with each additional 5 cm (1.97 in) of patient thickness; patient entrance air kerma increases by a factor of 8.4 when thickness increased from 24 to 34 cm [9,10,11,12,13,14,15]. Complex FGIP are associated with high radiation doses. These procedures can result in patient skin doses that are high enough to cause radiation injury and an increased risk of cancer [16].
Pediatric patients have a higher average risk of developing cancer compared with adults receiving the same radiation dose. The longer life expectancy in children allows more time for any harmful effects of radiation to manifest, and developing organs and tissues are more sensitive to the effects of radiation. Special attention is required to optimize appropriate protocols for pediatric patients. Major pediatric interventional procedures should be performed by experienced pediatric interventional operators, preferably with additional training in radiological protection [17, 18].
The Society of Interventional Radiology (SIR) and the Cardiovascular and Interventional Radiology Society of Europe (CIRSE) have jointly produced several guidelines that should be part of the education material for trainees aiming to be interventionists:
-
1.
Patient Radiation Dose Management [19]
-
2.
Occupational Radiation Protection in Interventional Radiology [20]
-
3.
Radiation Management for Interventions using Fluoroscopic or Computed Tomographic Guidance during Pregnancy [21]
-
4.
Occupational Radiation Protection of Pregnant or Potentially Pregnant Workers in Interventional Radiology [22]
X-ray Systems for Interventional Radiology
X-ray and imaging systems for interventional radiology are complex and have several modes to acquire images using different levels of radiation dose depending on the required image quality and diagnostic information for the clinical task. New technology in interventional imaging systems allows for substantial reduction in patient doses while maintaining enough image quality and diagnostic information, thanks to advanced image processing and refined selection of technical parameters during the imaging acquisition. During the commissioning of x-ray systems, some basic information about the modes of operation should be obtained [23].
Basic Radiation Physics Units
-
Absorbed dose is the energy absorbed per unit mass. The unit of absorbed dose is the gray (Gy); 1 gray is 1 Joule per kilogram.
-
Air kerma is the kinetic energy released in a mass of air. For the x-ray energies utilized in interventional procedures, the air kerma is numerically equal to the absorbed dose in air. The units for air kerma are the gray (Gy) or milligray (mGy) (Fig. 3.1).
-
The dose-area product (DAP) also called the kerma-area product (KAP) is the sum of the products of the incident doses and the areas of the x-ray fields for all segments of an interventional procedure. It can be determined at any convenient location between the x-ray source and the patient. The practical used unit for DAP is Gy·cm2. This quantity is presented by most of the interventional x-ray systems during the procedures, and the cumulative value is reported at the end of the procedure (Fig. 3.1).
-
Air kerma at the patient entrance reference point. This “patient entrance reference point” is located 15 cm from the isocenter in the direction of the focal spot for C-arm interventional x-ray equipment (Fig. 3.2). These two quantities (DAP and air kerma) are the most used by the x-ray systems to show interventionists the radiation dose received by the patients [24].
-
Equivalent dose is derived from the absorbed doses in specific tissues, weighted by the relative effect of the type and energy of the radiation encountered. For x-rays used in interventional procedures, the weighting factor is 1. Dose limits for occupational exposures are expressed in equivalent doses for deterministic effects in specific tissues. It is measured in Sievert (Sv).
-
Effective dose measures the global risk of the person exposed to ionizing radiation and takes into account the equivalent doses in the different tissues and the radiosensitivity of that tissue (Table 3.1). This quantity is used to determine radiation exposure risk to cancer development. Dose limits for occupational exposures are expressed as effective dose for stochastic effects throughout the body [26]. It is also measured in Sievert (Sv).
-
Personal dose equivalent is the operational quantity for individual monitoring and represented by Hp(d) (Fig. 3.3). It is the dose equivalent in soft tissue at an appropriate depth, d, below a specific point on the human body. The specified point is normally taken at 10 mm, termed Hp(10) for monitoring the effective dose. For the assessment of the dose to the skin and to the hands and feet, Hp(0.07) is used. A depth of 3 mm is adequate for monitoring the dose to the lens of the eye. In practice, Hp(0.07) and Hp(10) can be used for monitoring occupational doses during interventions guided by radiological imaging. A typical personal dosimeter provides two values, Hp(0.07) and Hp(10). Hp(0.07) from the collar dosimeter worn over protective garments (apron, thyroid shield) which provides a reasonable estimate of the dose delivered to the surface of the unshielded skin and to the lens of the eye. A single under-lead dosimeter does not provide any information about eye dose [27].
Summary of Biological Effects of Ionizing Radiation
The biological effects of radiation can be grouped into two types: deterministic effects (tissue reactions) and stochastic effects (cancer and heritable effects).
Key Points
Deterministic effects: Side effect occurs above a threshold radiation dose and severity increases with increasing dose.
Stochastic effects: Risk of developing side effect increases above a certain dose but the severity does not.
Deterministic Effects
Deterministic effects describe a relationship between radiation and side effects which occur above a certain threshold. With increasing doses above the threshold, the probability of occurrence will rise steeply to l00% (i.e., every exposed person will show the effect), and the severity of the effect will increase with dose. Such effects can occur in some complex interventional procedures [26, 28]. In FGIP, the tissues of concern for deterministic effects are the skin and the lens of the eye.
Key Points
Deterministic effects that occur above a threshold absorbed dose:
-
Fetal abnormality: 0.1–0.5 Gy
-
Sterility: 2–3 Gy
-
Skin erythema: 2–5 Gy
-
Hair loss: 2–5 Gy
-
Lethality (whole body): 3–5 Gy
-
Cataracts: 5 Gy
-
Irreversible skin damage: 20–40 Gy
Stochastic Effects
There is good evidence from cellular and molecular biology that radiation damage to the DNA in a single cell can lead to a transformed cell that is still capable of reproduction. Despite the cellular repair mechanisms, there is a small probability that this type of damage can lead to a malignant condition termed the somatic effect . For stochastic effects, a simple linear non-threshold dose-response relationship is assumed for radiological protection purposes. At higher doses and dose rates, the probability of developing cancer increases. At even higher doses, close to the thresholds of deterministic effects (tissue reactions), the probability increases more slowly and may begin to decrease, because of the competing effect of cell killing. These effects, both somatic and heritable, are called “stochastic.” The probability of such effects is increased when ionizing radiation is used in medical procedures [28].
Effects of In Utero Irradiation
There are radiation-related risks to the embryo/fetus during pregnancy that are related to the stage of pregnancy and the absorbed dose to the embryo/fetus. At doses below 100 mGy, lethal effects are extremely infrequent, and there is no reason to believe that exposure will result in any fetal abnormalities. During the period of major organogenesis, conventionally taken to be from the third to the eighth week after conception, malformations can occur, particularly in the organs under development at the time of exposure. These effects have a threshold of approximately 100 mGy [28, 29].
Radiation Protection System in Medicine
Several features of radiation exposure in medicine require an approach to radiation protection that is somewhat different from that for other types of radiation exposure. Medical uses of radiation for patients is voluntary in nature, with an expectation of direct individual health benefit to the patient.
In medicine, the goal is to use the appropriate radiation dose to obtain the desired image or desired therapy without excess exposure. In this regard, the ICRP introduced the use of diagnostic reference levels for imaging procedures. Radiation protection should be part of the quality assurance (QA) programs in interventional radiology (Fig. 3.4).
Radiation protection in medicine serves to identify the minimal dose for patients while allowing appropriate diagnosis or therapy and optimizing protection. The term ALARA (as low as reasonably achievable) is used to identify the optimization principle. ALARA is only part of the concept of optimization. The entire concept implies, more precisely, keeping patient exposure to the minimum necessary to achieve the required medical objective, both diagnostic and therapeutic. That said, dose to a patient should not be limited if effective diagnosis and treatment are imperiled. The physicians and other health professionals involved in the procedures that irradiate patients should always be trained in the principles of radiological protection, including the basic principles of physics and biology [17]. Physicians, radiographers, and medical physicists all play an essential role in the safe use of fluoroscopy in medical practice [30].
Key Points
“As Low As Reasonably Achievable” (ALARA) is based on the safety principle of minimizing radiation dose and limiting radioactive materials into the environment by employing all reasonable methods. The three major principles for a good protection are:
-
1.
Time
-
2.
Distance
-
3.
Shielding
Radiation Protection of Patients (And Diagnostic Reference Levels)
Diagnostic reference levels (DRLs ) are used in medical imaging to indicate whether, in routine conditions, the levels of patient dose from a specified imaging procedure are unusually high or low for that procedure. If so, a local review should be initiated to determine whether protection has been adequately optimized or whether corrective action is required [26].
DRLs should be reviewed at intervals that represent a compromise between the necessary stability and the long-term changes in the observed patient dose distributions [31, 32]. National DRLs should be set as the seventy-fifth percentile of median values obtained in a sample of representative centers. Median values of the DRL quantity for medical imaging procedures should be compared with DRLs to identify whether the data are substantially higher or lower than might be anticipated [32].
To protect a patient from excess radiation, the patient should be placed as far as possible away from the x-ray tube (portion underneath the table) and as close as possible to the image receptor (part above the table). Tight collimation also decreases patient dose and improves image quality by reducing scatter.
Radiation Protection of Staff (Including Pregnant Women)
There are different theories regarding possible dangers of exposure to personnel. It is extremely important to adapt the behavior and a safe working culture to the new powerful x-ray machines. Over time, longer procedures can lead to cumulative damage to our eyes if the proper protection is not regularly used. Reports on the radiosensitivity of the eye that can lead to visual impairment are available [33, 34].
In 2010, joint guidelines on protection of personnel were published by SIR (North American Society of Interventional Radiology) and CIRSE (Cardiovascular Interventional Radiology Society of Europe) in the Journals of both Societies (JVIR and CVIR) [20]. These guidelines provide a comprehensive overview that includes detailed instructions on why and how to protect IR from occupational exposure. These guidelines should become an integral part of any IR training program as well as routine practice in IR Labs.
Effective use of occupational radiation protection methods requires both appropriate education and training in radiation protection for all interventional radiology personnel and the availability of appropriate protective tools and equipment. Regular review and investigation of personnel monitoring results, accompanied by changes in how procedures are performed and equipment used, will ensure continual improvement in the practice of radiation protection in the interventional suite [35].
Passive and Active Personnel Radiation Protection
Personnel radiation protection process includes passive and active tools (Table 3.2). Passive radiation protection is based on the equipment in the IR lab. Active radiation protection is based on the passive protection tools and is about adapting our behavior to the “unfriendly” environment in the fluoroscopy room.
Active protection tools include protective drapes suspended from the table and from the ceiling. Table-suspended drapes hang from the side of the patient table, between the under-Table X-ray tube and the operator. They should always be employed, as they have been shown to substantially reduce operator dose.
Key Point
0.5 mm lead blocks approximately 95–99.5% of 70- to 100-kVp X-rays. Leaded glasses reduce exposure by a factor of 8–10.
It is not enough to have the protective tools available, but they must be used appropriately in order to safely protect all staff and patients within an interventional suite. The use of these tools must also be judged against their impedance to performing the procedure. Protective resources such as radiation protection gloves could lengthen the procedure in some cases and thus compromise the security and protection of the patient, as the tactile sensation of the catheter is reduced. In addition, the use of a leaded screen suspended from the ceiling could inhibit the movement of the C-arm x-ray system in some cases. Staff exposure drops dramatically with distance from the x-ray source. The inverse square law describes the proportional reduction in radiation density by the square of the distance.
Key Points
Inverse square law:The intensity of radiation exposure is inversely proportional to the distance from the source.
Hybrid rooms present additional radiation protection challenges [30]. Multidisciplinary teams of diverse staff members using different surgical and endovascular tools, imaging with fluoroscopy or DSA, cone beam CT, C-arm angulations, and isocentric positioning of the central beam create greater need for behavioral adaptation and increased awareness of radiation exposure. A small symphonietta should be orchestrated as these teams work shoulder to shoulder.
Personnel Dose Limits
The limit on effective dose for exposed workers should be 100 mSv in a consecutive 5-year period, subject to a maximum effective dose of 50 mSv in any single year. The limit on equivalent dose for the lens of the eye should be 150 mSv in a year. The limit on equivalent dose for the skin should be 500 mSv in a year. The limit on equivalent dose for the hands, forearms, feet, and ankles should be 500 mSv in a year. The current limit for the annual dose to the lens of the eye is 150 mSv, but recently based on the reports on the potential eye damages, the ICRP recommended about sevenfold less limit of 20 mSv/year for the eyes (or 100 mSv in 5 years with a maximum value of 50 mSv in a single year) [36].
Key Points
Limit of effective dose for exposed workers:
-
50 mSv max in a single year
-
100 mSv in a consecutive 5-year period
-
20 mSv equivalent dose to the lens of the eye, recently lowered from 150 mSv
-
500 mSv equivalent dose to the skin
-
1 mSv to fetus during pregnancy
Dosimetry badges are assessed periodically thus operators learn of their exposures in retrospect, sometimes weeks later. This delayed feedback may limit changes in staff habits. To better implement changes in work practices, it can be helpful for the individual to receive frequent feedback on dose levels through a real-time dosimeter. This may have a real impact radiation practice and influence behavioral change [37].
Pregnant Personnel
For pregnant workers, fetal dose is usually estimated using a dosimeter placed on the mother’s abdomen, under her radiation protective garments. For women who may be pregnant, the ICRP recommends that the additional dose to the embryo/fetus does not exceed about 1 mSv during the pregnancy [26].The restriction of a dose of 1 mSv to the embryo/fetus of a pregnant worker after declaration of pregnancy does not mean that it is necessary for a pregnant woman to avoid work with radiation completely or that she must be prevented from entering or working in designated radiation areas. It does, however, imply that the employer should review the exposure conditions of pregnant women carefully [6, 22].
Key Point
A pregnant female does not need to stop working in radiation areas; instead extra care should be taken to protect the fetus including extra shielding.
Particular Consideration for Pediatrics and Pregnancy
There are several important considerations for the pediatric population. UNSCEAR has recently published a new report of radiation risks for pediatrics [38] concluding that children may be at increased risk, the same risk or less risk than adults for development of malignancy depending upon the tumor type. The attributable lifetime risk of death (total cancers) in young children is higher than in adults, perhaps by a factor of 2 or 3. Appropriate weight bands are recommended by ICRP for establishing pediatric DRLs [38]. The settings and imaging protocols for interventional procedures in pediatrics require specific evaluation and regular updates for optimization [39].
Radiation exposure to both the patient and staff within the interventional suite when pregnant is an important and justified concern. In any circumstances involving the potential or actual use of fluoroscopically or CT-guided interventional procedures, a pregnant patient may be extremely concerned about the outcome of the pregnancy, and a counseling session with the mother (and father if possible) is often useful based on dose and risk to the fetus. If possible, pre-procedure and post-procedure counseling should take place [21].
Medical radiation procedures on pregnant patients should be justified and tailored to reduce fetal dose. Termination of pregnancy at fetal doses of <100 mGy is not justified based upon radiation risk [29].
Radiation Protection in CT Fluoroscopy (CTF)-Guided Interventions
CT-guided procedures and particularly growing use of CT fluoroscopy (CTF) guidance are an important contributor to the patient, as well as operator, exposure. Careful management of CT scanner parameters is required. Combination of fluoroscopy and CT with real-time image control over the entire body has high geometric accuracy, no significant interfering artifacts, increased target accuracy, reduced intervention times, and improved needle visualization.
Practical Recommendations for a Good Practice Minimizing Radiation Risks
Radiation dose management requires a comprehensive approach including preprocedural planning, intraprocedural management, and postprocedural care. It also includes periodic quality assessment [40]. The informed consent process supplies patients with sufficient information to make an appropriate decision regarding the proposed procedure. Participation by the radiologist in the follow-up of patients at risk is an integral part of radiation dose management. Close follow-up, with monitoring and management of radiation-induced injury or referral to another specialist, is appropriate for the interventional radiologist [19].
Key Points
Steps for safe radiation practice [20]:
-
Minimize fluoro time and number of spot images.
-
Use available shielding, both personal and equipment.
-
Use collimation.
-
Plan the procedure ahead of time as much as possible.
-
Position yourself in a low-scatter area.
-
Obtain appropriate training.
-
Wear your dosimeter and know your dose!
References
Avoidance of radiation injuries from medical interventional procedures. ICRP Publication 85. Ann ICRP. 2000;30(2):7–67.
Vano E. Reduction of exposure of patients and staff to radiation during fluoroscopically guided interventional procedures. Curr Radiol Rep. 2013;1:11–22. Available at: https://springerlink.bibliotecabuap.elogim.com/content/pdf/10.1007%2Fs40134-012-0001-9.pdf.
American College of Radiology. ACR technical standard for management of the use of radiation in fluoroscopic procedures. Practice guidelines and technical standards 2003. Reston: American College of Radiology; 2003.
National Council on Radiation Protection & Measurements (Bethesda, MD). Ionizing radiation exposure of the population of the United States. Report No.:160. Available from: http://www.ncrponline.org/PDFs/2012/DAS_DDM2_Athens_4-2012.pdf.
Verstandig A, Shamieh B, Shraibman V, Raveh D. Radiation dose reduction in fluoroscopic procedures: left varicocele embolization as a model. Eur Radiol. 2015;25(6):1639–45.
ICRP Publication 117. Radiological protection in fluoroscopically guided procedures performed outside the imaging department. Ann ICRP. 2010;40(6):1–102.
Smith-Bindman R, Miglioretti DL, Johnson E, Lee C, Feigelson HS, Flynn M, et al. Use of diagnostic imaging studies and associated radiation exposure for patients enrolled in large integrated health care systems, 1996–2010. JAMA. 2012. Available from: http://jama.jamanetwork.com/article.aspx?articleid=1182858&resultClick=3.
Centers for Disease Control and Prevention [Internet]. Atlanta (GA) [updated 2014 Sept 9; cited 2014 Sep 24]. Available from: http://www.cdc.gov/obesity/data/adult.html.
World Health Organization [Internet]. c2014 [updated 2014 Aug; cited 2014 Sept 24]. Obesity and overweight. Available from: http://www.who.int/mediacentre/factsheets/fs311/en/.
Schueler BA, Vrieze TJ, Bjarnason H, Stanson AW. An investigation of operator exposure in interventional radiology. Radio Graphics. 2006;26:1533–41. https://doi.org/10.1148/rg.265055127.
Miller DL, Balter S, Cole PE, Lu HT, Schueler BA, Geisinger M, et al. Radiation doses in interventional radiology procedures: the Rad-IR study part I: overall measures of dose. J Vasc Interv Radiol. 2003;14:711–27.
Miller DL, Balter S, Cole PE, Lu HT, Berenstein A, Albert R, et al. Radiation doses in interventional radiology procedures: the Rad-IR study part II: skin dose. J Vasc Interv Radiol. 2003;14:977–90.
O’Dea TJ, Geise RA, Ritenour ER. The potential for radiation induced skin damage in interventional neuroradiological procedures: a review of 522 cases using automated dosimetry. Med Phys. 1999;26:2027–33.
Koenig TR, Wolff D, Mettler FA, Wagner LK. Skin injuries from fluoroscopically guided procedures, part 1, characteristics of radiation injury. AJR. 2001;177(1):3–11.
Koenig TR, Mettler FA, Wagner LK. Skin injuries from fluoroscopically guided procedures, part 2, review of 73 cases and recommendations for minimizing dose delivered to patient. AJR. 2001;177(1):13–20.
Jaschke W, Schmuth M, Trianni A, Bartal G. Radiation-induced skin injuries to patients: what the interventional radiologist needs to know. Cardiovasc Interv Radiol. 2017;40(8):1131–40.
Education and training in radiological protection for diagnostic and interventional procedures. ICRP Publication 113. Ann ICRP. 2009;39(5):1–68.
ICRP Publication 121. Radiological protection in paediatric diagnostic and interventional radiology. Ann ICRP. 2013;42(2):1–63.
Stecker MS, Balter S, Towbin RB, Miller DL, Vañó E, Bartal G, et al. Guidelines for patient radiation dose management. SIR Safety and Health Committee; CIRSE Standards of Practice Committee. J Vasc Interv Radiol. 2009;20(7 Suppl):S263–73.
Miller DL, Vano E, Bartal G, Balter S, Dixon R, Padovani R, et al. Occupational radiation protection in interventional radiology: a joint guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology. J Vasc Interv Radiol. 2010;21(5):607–15.
Dauer LT, Thornton RH, Miller DL, Damilakis J, Dixon RG, Marx MV, et al. Radiation management for interventions using fluoroscopic or computed tomographic guidance during pregnancy: a joint guideline of the Society of Interventional Radiology and the Cardiovascular and Interventional Radiological Society of Europe with Endorsement by the Canadian Interventional Radiology Association. J Vasc Interv Radiol. 2012;23(1):19–32.
Dauer LT, Miller DL, Schueler B, Silberzweig J, Balter S, Bartal G, et al. Occupational radiation protection of pregnant or potentially pregnant workers in IR: a joint guideline of the Society of Interventional Radiology and the Cardiovascular and Interventional Radiological Society of Europe. Society of Interventional Radiology Safety and Health Committee; Cardiovascular and Interventional Radiological Society of Europe Standards of Practice Committee. J Vasc Interv Radiol. 2015;26(2):171–81.
Vano E, Geiger B, Schreiner A, Back C, Beissel J. Dynamic flat panel detector versus image intensifier in cardiac imaging: dose and image quality. Phys Med Biol. 2005;50(23):5731–42.
International Electrotechnical Commission. Medical electrical equipment. Part 2–43. Particular requirements for the safety of X-ray equipment for interventional procedures. Report IEC 60601–2-43 (2010). Geneva, Switzerland.
Bartal G, Paulo G, Roguin A. Minimizing radiation risk to patients and staff. Endovasc Today. 2016;15(8):56–62.
The 2007 recommendations of the international commission on radiological protection. ICRP Publication 103. Ann ICRP. 2007;37(2–4):1–332.
Miller DL, Balter S, Schueler BA, Wagner LK, Strauss KJ, Vañó E. Clinical radiation management for fluoroscopically guided interventional procedures. Radiology. 2010;257(2):321–32.
Radiological protection in medicine. ICRP Publication 105. Ann ICRP. 2007;37.
Pregnancy and medical radiation. ICRP Publication 84. Ann ICRP. 30(1).(6): 1–63.
Bartal G, Vano E, Paulo G, Miller DL. Management of patient and staff radiation dose in interventional radiology: current concepts. Cardiovasc Intervent Radiol. 2014;37(2):289–98.
Publication 129. Radiological protection in cone beam computed tomography (CBCT). Ann ICRP. 2015;44(1):7–127.
Publication 135. Diagnostic reference levels in medical imaging. Ann ICRP. (in press).
Vano E, Gonzalez L, Fernandez J, Haskal Z. Eye lens exposure to radiation in interventional suites: caution is warranted. Radiology. 2008;248(3):945–53.
Vano E, Kleiman NJ, Duran A, Romano-Miller M, Rehani MM. Radiation-associated lens opacities in catheterization personnel: results of a survey and direct assessments. J Vasc Interv Radiol. 2013;24(2):197–204.
Duran A, Hian SK, Miller DL, Le Heron J, Padovani R, Vano E. Recommendations for occupational radiation protection in interventional cardiology. Catheter Cardiovasc Interv. 2013;82(1):29–42.
ICRP Publication 118: ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs – threshold doses for tissue reactions in a radiation protection context. Ann ICRP. 2012;41(1–2):1–322.
Chiriotti S, Ginjaume M, Vano E, et al. Performance of several active personal dosemeters in interventional radiology and cardiology. Radiat Meas. 2011;46:1266–70.
United Nations Scientific Committee on Effects of Atomic Radiations Source and Effects of Ionizing Radiation (UNSCEAR). Report to the General Assembly with Scientific Annexes. 2013. Annex B: Effects of radiation exposure of children. Available at: http://www.unscear.org/docs/reports/2013/UNSCEAR2013Report_AnnexB_Children_13-87320_Ebook_web.pdf.
Ubeda C, Vano E, Miranda P, Aguirre D, Riquelme N, Dalmazzo D, et al. Patient and staff doses in paediatric interventional cardiology derived from experimental measurements with phantoms. Phys Med. 2016;32(1):176–81.
Vano E. Radiation exposure to cardiologists: how it could be reduced. Heart. 2003;89(10):1123–4.
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2018 Springer International Publishing AG, part of Springer Nature
About this chapter
Cite this chapter
Bartal, G., Vano, E. (2018). Radiation Safety. In: Keefe, N., Haskal, Z., Park, A., Angle, J. (eds) IR Playbook. Springer, Cham. https://doi.org/10.1007/978-3-319-71300-7_3
Download citation
DOI: https://doi.org/10.1007/978-3-319-71300-7_3
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-71299-4
Online ISBN: 978-3-319-71300-7
eBook Packages: MedicineMedicine (R0)