Measurement of extremity volumes with plethysmography or indirect via circumference measurement is essential for assessing the size of the lymphedema and for treatment monitoring. It is important that both the normal and diseased limb is measured in exactly the same way at all time points to eliminate natural volume variations. The normal arm volume variations are between 2 and 13 % in healthy individuals. The dominant arm is usually about 1.5 % larger. Lymphedema is usually defined as a relative extremity volume difference > 10 % (the edematous extremity is 10 % larger than the normal). Both plethysmography and circumference measurements are useful and show satisfactory validity and reliability; plethysmography is recommended if only one method is used.
Access provided by Autonomous University of Puebla. Download chapter PDF
Similar content being viewed by others
Keywords
Key Points
-
Outcomes of surgical treatment of lymphedema very seldom present results as excess volume or excess volume reduction. Instead circumference measurements, taken at random sites along the extremity, are used making it very difficult to estimate the true outcome as well as to compare various studies.
-
Volume measurement, either using plethysmography or based on circumference measurements, is an easy and quick method to objectively assess treatment outcome and to increase the scientific impact when presenting results.
Introduction
Measurement of extremity volumes with plethysmography or indirect via circumference measurement is essential for assessing the size of the lymphedema and for treatment monitoring. It is important that both the normal and diseased limb is measured in exactly the same way at all time points to eliminate natural volume variations. The normal arm volume variations are between 2 and 13 % in healthy individuals. The dominant arm is usually about 1.5 % larger [1]. Lymphedema is usually defined as a relative extremity volume difference > 10 percent (the edematous extremity is 10 % larger than the normal) [2, 3]. Both plethysmography and circumference measurements are useful and show satisfactory validity and reliability [4–9], plethysmography is recommended if only one method is used [10].
Volume Measurements Using Plethysmography
The most reliable way to measure the volume of an extremity is plethysmography according to Archimedes’ principle, i.e., the limb is immersed in a water bath and the drained water is weighed [1–11]. The weight in grams equals the volume in milliliter and also includes the hand and foot volume. For description of the volume meter see Figs. 11.1 and 11.2.
For both the arm and leg volume meter it is important that the draining pipe is be placed high enough so that the entire limb can be measured. The diameter of the draining hose should be large so that the water can flow quickly. It is important that the limb is immersed into the volume meter at exactly the same depth each time.
Arm Volume Meter (Fig. 11.1)
The hose is connected to a standard faucet with a threaded bayonet mount and the volume meter is filled with water through a hose at the bottom of tank. The arm is brought down into the volume meter with the palm against the wall to avoid surge. The fingers are held outstretched and brought down until the tip of the middle finger reaches the bottom. Wait a few seconds to allow the water to flow out. There are gradation lines with 1-cm intervals for patients who do not reach down to the bottom of the volume meter. Measurements can be reproduced from time to time by noting the distance between the volume meter’s bottom and the tip of the middle finger. The drained water is collected in a plastic tray (Orth Plast, for example, model 101-8: 40 cm × 34 cm × 16 cm), which holds about 17 L. To empty the volume meter the hose is led down a drain in the floor or to a sink so the water can drain. An electric pump accelerates the runoff (Eheim 1046, Eheim, http://www.eheimparts.com). A drawing of a volume meter is shown in Fig. 11.1. The scale should have an accuracy of 5 g (digital max 30 kg/5 g). The volume meter has an average measurement error (CV%) of 0.24.
Leg Volume Meter (Fig. 11.2)
To compensate for different leg lengths, different spacers are put into the volume meter so that the entire leg volume can be measured up to the crotch. The measuring cylinder’s height minus the leg length = the height of the spacer. An electric pump accelerates the runoff when emptying (Eheim 1060, Eheim, http://www.eheimparts.com). To hoist the patient a commercially available “lift” (Santo Lift SC 200, PMH International AB, http://www.pmh.se) is used. The water is collected in a large plastic laundry basket (Idealplast; model 0519; 33 cm × 55 cm), which holds about 40 L (Fig. 11.2).
Volume Measurements Based on Circumference Measurements
Another method is circumference measurements on well-defined distances along the limb. Circumference measurements also provide information of the localization of the swelling. Hand and foot volumes cannot be calculated. To only make single circumference measurements (for example the middle of the upper arm, elbow, middle of the forearm) is not adequate either for clinical use or scientific study.
Volume measurements based on plethysmography or circumference measurements are useful and show satisfactory validity and reliability [4–9], plethysmography is recommended if only one method is used [10].
How Volumes Are Calculated Using Circumference Measurements
The circumferences of the limb measured every 4 cm along the limb. The volume of each segment is calculated according to the formula of the truncated and added together to get the limb volume (Fig. 11.3):
You need:
-
A long ruler to mark every the 4 cm
-
Marker pen (eyeliner pen)
-
Narrow tape measure for circumference measurements
-
Protocol.
How to measure:
-
Arm lymphedema: The patient sits on a chair with the arm straight and abducted 90 ° on a table
-
Highlight the 0 point in the wrist level
-
The tape is placed around the extremity distal to each mark
-
Read the circumferences measured and record
-
Leg lymphedema: The patient sits with the hip flexed at 90° with extended knee and foot in 0-position.
The measurement programs for arms and legs can be downloaded from: https://lu.box.com/s/9abvigfbx2rw7afgk9kz or by contacting the first author.
Other Methods
The following methods are not used routinely, but mainly for research purposes.
Perometry
This is an optoelectronic measuring method using a square frame with multiple perpendicular light beams. The frame is moved along the limb and the cross-sectional area is calculated continuously and thereby the volume. The equipment is expensive, but the method is accurate and fast [12]. Foot and hand volume are not measured [13, 14].
Bioelectric Impedance (BIS)
Extracellular fluid can be measured by bioelectrical impedance, which is recorded with a low current with different frequencies, transmitted through the tissue [15]. The method is particularly useful for diagnosing early lymphedema of the arm after breast cancer treatment [16].
Local Tissue Fluid, Tissue Dielectric Constant (TDC)
Tissue fluid may be measured by means of electromagnetic waves transmitted via a measuring head, which is set against the skin. The instrument measures a few millimeters into the skin and a constant (tissue dielectric constant), directly proportional to the tissue fluid content, is calculated. The method is useful for measuring tissue water anywhere in the body where the skin is not too close to the skeleton. The method has been used to measure edema in the extremities [17], legs [18, 19].
Other methods include computed tomography (CT) [20] and magnetic resonance imaging (MRI) [21].
Volume Calculations
A large number of computational methods have been presented for reporting treatment results [5, 7, 8]. The excess volume of a unilateral lymphedema is calculated as the difference between the edematous and the contralateral healthy extremity.
The absolute excess volume (volume of the affected limb − volume of the unaffected limb) of a lymphedema [22, 23] together with the relative value in percent (affected limb/unaffected limb × 100) gives optimal information before and after treatment. The relative value provides information about the extent of the edema [1, 10]. The same absolute excess volume provides greater relative value (swollen extremity/healthy extremity) in a skinny person as compared to a corpulent [1, 10].
Absolute Excess Volume and Percentage Reduction
Edematous arm (EA), Normal arm (NA), Before treatment (bt), After treatment (at).
EA | NA | Excess volume | Excess volume reduction (%) | |
---|---|---|---|---|
bt | 5,000 ml | 2,000 ml | ∆3,000 ml | |
67 % | ||||
at | 3,000 ml | 2,000 ml | ∆1,000 ml |
By measuring the normal arm at the same time as the edematous arm one takes into account the daily variations in arm volumes as well as weight gain or loss. It is very important that this is done at every time when volumes are measured.
Follow-up controls
EA | NA | Excess volume | % Excess volume reduction | |
---|---|---|---|---|
bt | 5,000 ml | 2,000 ml | ∆3,000 ml | |
at1 | 4,000 ml | 2,200 ml | ∆1,800 ml | \( \begin{array}{l}\frac{3,000\;\mathrm{ml}-1,800\;\mathrm{ml}}{3,000\;\mathrm{ml}}\times 100\\ {}=\mathbf{40}\;\%\end{array} \) |
at2 | 3,000 ml | 1,800 ml | ∆1,200 ml | \( \begin{array}{l}\frac{3,000\;\mathrm{ml}-1,200\;\mathrm{ml}}{3,000\;\mathrm{ml}}\times 100\\ {}=\mathbf{60}\;\%\end{array} \) |
at3 | 2,000 ml | 2,000 ml | ∆0 ml | \( \begin{array}{l}\frac{3,000\;\mathrm{ml}-0\;\mathrm{ml}}{3,000\;\mathrm{ml}}\times 100\\ {}=\mathbf{100}\;\%\end{array} \) |
The same calculation applies also for a leg.
In bilateral lymphedema the excess volume cannot be determined. Instead the change in volume, also in percent, for each limb is calculated, i.e.,
When scientific material is presented, the mean excess volume for all patients at baseline is calculated, as well as at follow-up. The percentage reduction for each patient is calculated and the mean percent edema reduction for all patients is calculated at follow-up.
The Relative Value of the Excess Volume
The excess volume can also be calculated in relation to the normal arm as a ratio, i.e., edematous arm/normal arm.
EA | NA | Ratio | |
---|---|---|---|
bt | 5,000 ml | 2,000 ml | 2.5 |
\( \frac{{\mathrm{EA}}_{\mathrm{bt}}}{{\mathrm{NA}}_{\mathrm{bt}}}=\frac{5,000\;\mathrm{ml}}{2,000\;\mathrm{ml}}\times 100=2.5 \) | |||
at | 3,000 ml | 2,000 ml | 1.5 |
\( \frac{{\mathrm{EA}}_{\mathrm{at}}}{{\mathrm{NA}}_{\mathrm{at}}}=\frac{3,000\;\mathrm{ml}}{2,000\;\mathrm{ml}}\times 100=1.5 \) |
The absolute excess volume is not always sufficient to describe a lymphedema. The same absolute excess volume of 200 ml (difference between the edematous arm and the normal arm) in a lean and in a corpulent person results in 10 % relative excess volume in a lean subject as opposed to 5 % in a corpulent subject.
Lean subject | Corpulent subject | ||||
---|---|---|---|---|---|
EA | NA | Excess volume | EA | NA | Excess volume |
\( \begin{array}{l}2,200\;\mathrm{ml}--2,000\;\mathrm{ml}\\ {}=200\;\mathrm{ml}\end{array} \) | \( \begin{array}{l}4,200\;\mathrm{ml}--4,000\;\mathrm{ml}\\ {}=200\;\mathrm{ml}\end{array} \) |
Excess volume in percent:
Potential Sources of Errors
Potential Sources of Error
A common source of error is when the normal limb volume is measured only at baseline or not at all. This leads to incorrectly calculated values when the normal limb increases or decreases in volume [11].
Diurnal variation and weight loss or gain are sources of error, especially in patients with a small lymphedema.
Examples
Example of calculations of percent excess volume reduction when even the normal limb volume decreases (see Table 11.1).
The percentage reduction is calculated correctly when the normal limb volume is measured both before treatment and at follow-up . The left side of the table shows data from one arm while the right shows data from a leg. Note that if the initial normal limb volume is used at all subsequent measurements incorrect values are achieved. In the example of the arm an incorrect reduction of 53.9 % is calculated instead of 32.5 %, which is the correct excess percentage reduction. In the example of the leg an erroneous reduction of 91.2 % is calculated instead of the correct value of 7.3 % (see Table 11.1).
Example of calculations of percent excess volume reduction when even the normal limb volume increases (see Table 11.2):
The percentage reduction will be calculated correctly by the normal limb volume measured both at initial measurement and the final measurement . The left side of the table shows data from one arm while the right shows data from a bone. Note that if the normal limb initial measured value is used all the time to get the wrong values. In the example on an arm improperly obtained an edema reduction of 44.5 % instead of 52.3 %, which is the correct edema reduction. In the example on one leg got an error reduction of 50.7 % instead of the correct 64.6 % (see Table 11.2).
Conclusion
Volume measurement should be used when presenting outcomes of lymphedema treatment.
References
Godal R, Swedborg I. A correction for the natural asymmetry of the arms in the determination of the volume of oedema. Scand J Rehabil Med. 1982;14(4):193–5.
Swedborg I. Effects of treatment with an elastic sleeve and intermittent pneumatic compression in post-mastectomy patients with lymphoedema of the arm. Scand J Rehabil Med. 1984;16(1):35–41.
Borup Christensen S, Lundgren E. Sequelae of axillary dissection vs. axillary sampling with or without irradiation for breast cancer. A randomized trial. Acta Chir Scand. 1989;155(10):515–9.
Kuhnke E. Volumenbestimmung aus Umfangsmessungen. Folia Angiologica. 1976;24:228–32.
Kuhnke E. Die Volumenbestimmung entrundeter Extremitäten aus Umfangsmessungen. Lymphologie. 1978;2:35–44.
Stranden E. A comparison between surface measurements and water displacement volumetry for the quantification of leg edema. J Oslo City Hosp. 1981;31(12):153–5.
Casley-Smith JR. Measuring and representing peripheral oedema and its alterations. Lymphology. 1994;27(2):56–70.
Sitzia J. Volume measurement in lymphoedema treatment: examination of formulae. Eur J Cancer Care (Engl). 1995;4(1):11–6.
Brorson H, Höijer P. Standardised measurements used to order compression garments can be used to calculate arm volumes to evaluate lymphoedema treatment. J Plast Surg Hand Surg. 2012;46(6):410–5.
Swedborg I. Voluminometric estimation of the degree of lymphedema and its therapy by pneumatic compression. Scand J Rehabil Med. 1977;9(3):131–5.
Bernas MWM, Witte C, Belch D, Summers P. Limb volume measurements in lymphedema: issues and standards. Lymphology. 1996;29(Suppl):199–202.
Stanton AW, Northfield JW, Holroyd B, Mortimer PS, Levick JR. Validation of an optoelectronic limb volumeter (Perometer). Lymphology. 1997;30(2):77–97.
Goltner E, Fischbach JU, Monter B, Kraus A, Vorherr H. Objective measurement of lymphedema after mastectomy. Dtsch Med Wochenschr. 1985;110(24):949–52.
Bednarczyk JH, Hershler C, Cooper DG. Development and clinical evaluation of a computerized limb volume measurement system (CLEMS). Arch Phys Med Rehabil. 1992;73(1):60–3.
Cornish BH, Thomas BJ, Ward LC, Hirst C, Bunce IH. A new technique for the quantification of peripheral edema with application in both unilateral and bilateral cases. Angiology. 2002;53(1):41–7.
Cornish BH, Chapman M, Thomas BJ, Ward LC, Bunce IH, Hirst C. Early diagnosis of lymphedema in postsurgery breast cancer patients. Ann N Y Acad Sci. 2000;904:571–5.
Mayrovitz HN, Weingrad DN, Davey S. Local tissue water in at-risk and contralateral forearms of women with and without breast cancer treatment-related lymphedema. Lymphat Res Biol. 2009;7(3):153–8.
Jensen MR, Birkballe S, Norregaard S, Karlsmark T. Validity and interobserver agreement of lower extremity local tissue water measurements in healthy women using tissue dielectric constant. Clin Physiol Funct Imaging. 2012;32(4):317–22.
Mayrovitz HN, Davey S, Shapiro E. Localized tissue water changes accompanying one manual lymphatic drainage (MLD) therapy session assessed by changes in tissue dielectric constant inpatients with lower extremity lymphedema. Lymphology. 2008;41(2):87–92.
Uhl JF. 3D multislice CT to demonstrate the effects of compression therapy. Int Angiol. 2010;29(5):411–5.
Sagen A, Karesen R, Skaane P, Risberg MA. Validity for the simplified water displacement instrument to measure arm lymphedema as a result of breast cancer surgery. Arch Phys Med Rehabil. 2009;90(5):803–9.
Tracy GD, Reeve TS, Fitzsimons E, Rundle FF. Observations on the swollen arm after radical mastectomy. Aust N Z J Surg. 1961;30:204–8.
Treves N. An evaluation of the etiological factors of lymphedema following radical mastectomy; an analysis of 1,007 cases. Cancer. 1957;10(3):444–59.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2015 Springer International Publishing Switzerland
About this chapter