Abstract
Acoustic radiation force elasticity imaging (ARFI) methods have become a viable clinical option to noninvasively evaluate the stiffness of the liver to evaluate fibrosis stage. Unlike external vibration-based methods such as transient elastography, acoustic radiation force excitations can be focused directly in the tissue of interest, while also providing real-time, ultrasonic B-mode imaging guidance and clinical evaluation. Acoustic radiation force methods are also available to screen for and characterize liver masses. Efforts to standardize and provide more consistent measurements between different manufacturer systems is being addressed by consensus documents and guideline documents, and the RSNA QIBA Ultrasonic Shear Wave Speed group is taking important steps toward characterizing the performance of these systems in calibrated elastic and viscoelastic media.
Access provided by Autonomous University of Puebla. Download chapter PDF
Similar content being viewed by others
Keywords
- Acoustic radiation force
- Acoustic radiation force impulse imaging
- ARFI
- Acoustic attenuation
- Liver stiffness
- Shear wave propagation
- Young’s modulus
- Shear wave elastography imaging
- SWEI
- Point shear wave elastography
- pSWEI
- Elastography
- Liver disease imaging
- Elasticity
- Viscosity
- Shear wave elasticity imaging
- SWS
- Shear wave speed
- Strain imaging
Introduction
Methods to characterize the elastic properties of the liver typically have two means of introducing mechanical perturbations into liver to then monitor the liver’s response and estimate its stiffness: (1) external vibration—such as transient elastography and MR elastography—and (2) an internally applied acoustic radiation force. For more details see also other chapters of the book section II “Techniques to measure liver stiffness.” While external sources of vibration can be well-controlled at fixed frequencies with external vibrators and couple relatively strong waves into the body, these waves must couple from the skin surface, through superficial tissues (i.e., skin, subcutaneous fat, muscle), into the liver. These propagating external waves can be distorted through interactions with these tissues surrounding the liver, and in some circumstances, such as abdominal ascites, cannot couple through fluids into the liver. Given some of the challenges that external vibration sources can face when characterizing the liver, there was parallel development in the 1990s and early 2000s to develop acoustic radiation force methods that would generate sources of mechanical perturbation in the focal zone of an ultrasonic excitation inside the target tissue of interest and not relying on coupling external mechanical energy into the patient. These acoustic radiation force methods have been developed and deployed on standard diagnostic ultrasound scanners as software features, allowing clinicians to evaluate the liver without any additional hardware.
What Is Acoustic Radiation Force?
Acoustic radiation force is a phenomenon that was first described in the acoustics literature by Nyborg in 1965 in the context of acoustic streaming [1] and further by Apfel and Chu in 1985 [2]. Acoustic radiation force is generated in the direction of propagation of an acoustic wave in a lossy (attenuating) medium, where the loss of momentum of the propagating wave results in an impulse transfer to the tissue. The direction and magnitude of acoustic radiation force (\( \overset{\rightharpoonup }{F} \)) can be represented as:
where a is the acoustic attenuation of the tissue, \( \overset{\rightharpoonup }{I} \)is the acoustic intensity vector, and c is the sound speed of the tissue [1, 3, 4].
The application of acoustic radiation force to tissue results in a transient displacement of the tissue, where the magnitude of that displacement is related to the magnitude of the acoustic radiation force applied to the tissue, and the stiffness of the tissue. Stiffer tissues resist deformation than more compliant tissues and, therefore, experience less induced displacement. The transient, impulsive application of acoustic radiation force also leads to the generation of shear waves that emanate out from the region of acoustic radiation force application, and the speed of these shear waves can also be related to the stiffness of the tissue they are propagating in [5,6,7]. While acoustic radiation force is associated with all ultrasonic insonification of tissue (e.g., B-mode and Doppler imaging), the ultrasonic pulses for acoustic radiation force-based elasticity imaging must be intense and/or long enough to generate tissue displacements that can be estimated ultrasonically (on the order of microns) [8]. Commercial implementations of acoustic radiation force imaging methods that are described in this chapter have traditionally adhered to U.S. Food and Drug Administration diagnostic ultrasound output limits [9, 10], but new research efforts are exploring the benefits of using elevated acoustic output for more robust imaging performance, especially in the difficult to image demographic, such as those with high Body Mass Indices [11].
Acoustic Radiation Force Impulse (ARFI) Imaging
ARFI imaging refers to images of tissue displacement (or metrics related to tissue displacement, such a time-to-peak displacement or maximum displacement) that result from the application of impulsive acoustic radiation force excitations [3, 12]. The “impulsive” nature of the excitation refers to an insonification time that is less than the mechanical response time of tissue, which is related to the tissue’s stiffness, with stiffer tissues reacting faster [12]. For liver tissue, this typical impulsive excitation lasts for less than 1 ms [12].
A single A-line of an ARFI image typically involves the following sequence:
-
1.
A single, B-mode-like tracking pulse to determine the RF data associated with the tissue pre-ARFI-induced deformation;
-
2.
A impulsive ARFI excitation either at a single focal depth, or over a range of focal depths to extend the effective depth-of-field of the excitation energy [13];
-
3.
A series of B-mode-like tracking pulses at a relatively high pulse repetition frequency (typically 5–10 kHz) to track the tissue displacement and recovery from the ARFI excitation using correlation, phase-shift, or more complex displacement estimation methods [8, 14,15,16,17,18,19].
To form a 2D ARFI image, the A-line sequence is repeated at laterally offset positions from one another to form an image. Displacement estimation can be affected by motion (and other) artifacts, include those related to respiratory and cardiac motion. While clinical imaging protocols may recommend techniques like suspended breathing to minimize these effects [20, 21], most scanner post-processing includes motion filtering, through means of temporal profile shape-fitting [22], or frequency-domain filtering. While ARFI images can utilize elasticity as a mechanism of contrast not present in B-mode images, ARFI images are not typically used to estimate absolute stiffness of tissues (i.e., relating displacement amplitude to elastic modulus) since the magnitude of the acoustic radiation force is a function of acoustic attenuation [23], which is not easily estimated for different imaging targets/tissues. Additionally, the acoustic radiation force magnitude is not constant as a function of depth, and instead varies as a function of focal depth, requiring some depth-dependent normalization scheme to be applied to compensate for these force gradients [3].
Shear Wave Elasticity Imaging (SWEI)
SWEI was first described in the literature by Saravazyan et al. [24] as novel approach to generating shear wave with acoustic radiation force and measuring the resultant shear wave propagation and shear wave speed to ultimately reconstruct an elastic modulus. Unlike ARFI images that can generate images of relative tissue stiffness differences, SWEI allows for absolute metrics of stiffness to be estimated in tissue. The typical assumptions surrounding the reconstruction of an elastic modulus (E, Young’s modulus, or μ, shear modulus) from shear wave speed (cT) include the tissue being linear, isotropic, incompressible, and having a density of water (ρ = 1.0 g/cm3), such that the following relationships can hold [25]:
The incompressibility assumption allows the Young’s modulus to be simply related to the shear modulus by a factor of 3. For more details see also book Part II “Techniques to measure liver stiffness.” It should be noted that the relationship between elastic moduli and shear wave speed is quadratic, and therefore, when statistical analyses are performed on data acquired on systems that report different metrics, retrospective conversion of reported thresholds for diagnostic purposes should not be attempted with simple linear scaling. Instead, thresholds for significance and confidence intervals should be recalculated as they may change with this nonlinear relationship. Like ARFI imaging, the acoustic radiation force excitations used for SWEI can involve single or multiple focal zones, depending the depth-of-field being characterized by the shear wave propagation. Rapid firing of multiple focal zone excitations such that a virtual extended shear wave front is launched is referred to as a “supersonic” shear excitation, which was popularized by SuperSonic Shear Wave™ Elastography [26].
One challenge with SWEI can be the distance over which shear wave propagate with displacements that can be reliably tracked. Greater distances can be achieved with stronger radiation force excitations [27, 28], more advanced displacement estimation approaches [29, 30], or creative implementations of interspersed acoustic radiation force excitation, tracking, and directional filtering to tease apart the complexities of intersecting propagating wave fields [31, 32].
Unlike MR elastography, which can measure displacement components in three dimensions and reconstruct a resultant (complex) shear modulus from these data using the Helmholtz equation [33], ultrasonic shear wave methods have a much higher resolution for displacement estimation in the single direction orthogonal to the transducer face, and have instead utilized time-of-flight methods to estimate shear wave speed [26, 34,35,36,37,38,39]. These shear wave speed estimation methods applied over 2D regions of interest (ROI) can be used to generate two types of quantitative images:
-
1.
Point Shear Wave Elastography (pSWE) utilizes all the propagation data in the 2D ROI to estimate a consensus shear wave speed in that region [3, 21]. Typically, a singleton shear wave speed metric is reported per measurement, and clinical studies have supported using a median value across 12 repeated measurements to report as a representative measurement for diagnostic purposes [20, 21]. Quality metrics can be reported along with the shear wave speed to increase diagnostic confidence.
-
2.
2D-shear wave Elastography (2D-SWE) breaks the ROI into smaller shear wave reconstruction kernels to generate 2D images of shear wave speed to characterize local variabilities in elasticity [21, 40].
As the confounding factors—viscosity, nonlinearity, anisotropy, structural boundaries—surrounding accurate reconstruction of an elastic modulus from shear wave speed became more well understood, many derivatives of SWEI have settled on directly reporting shear wave speed instead of making additional assumptions to report an elastic modulus. Thus, additional details about the conditions under which an elastic modulus has been estimated must be provided (e.g., the frequency of excitation used in transient elastography or MR elastography) [41, 42]. Given the complexities and nuances surrounding the different commercial implementations of SWEI for liver characterization, guidelines have been established for its recommended clinical usage by the World Federation of Ultrasound in Medicine and Biology [20] and the Society of Radiologists in Ultrasound [21].
Viscosity
Tissue viscosity is a material property that makes the shear wave speed dependent on the frequency content of the shear wave, and is a confounding factor when assuming that liver tissue is purely elastic (shear wave speed is independent of the shear wave frequency content) [43]. It is known that different elasticity imaging methods can generate shear waves of differing frequency content [21] (see also Table 4.1). The frequency differences can lead to different reconstructed group shear wave speeds in viscoelastic media [44], which can be a source of discrepancy when comparing measurements with different elasticity imaging modalities [41]. Additionally, the processing methods used by each manufacturer—specifically the use of displacement versus velocity data—can also influence the estimated speeds in the presence of viscosity. Velocity data are effectively high pass filtered displacement profiles, and this higher frequency bias can lead to an increase in the estimated shear wave speed.
Some investigators have sought to characterize the viscosity of liver tissue as a measurement of hepatic steatosis or inflammation [45,46,47], but no conclusive conclusions have been drawn to date.
In addition to the elastography guidelines and consensus documents that have been published, the Radiological Society of North America (RSNA) Quantitative Imaging Biomarker Alliance (QIBA) establishes an Ultrasonic Shear Wave Speed (US SWS) working group in 2012 to study the factors that influence the consistent reconstruction of the shear wave speed metric across different manufacturer systems [41, 48, 49]. This working group is composed of international researchers, manufacturers, and regulatory members, and a profile guiding best practices for manufacturers to achieve consistent measurements across different systems is available for consultation.
Another resource generated through the RSNA QIBA US SWS effort has been the generation and public release of digital phantom data generated through finite element method models of shear wave propagation in elastic and viscoelastic materials for shear wave reconstruction development validation [50]. Additionally, standardized sequences for generating and processing shear waves using the Verasonics ultrasound research platform have also been released for public use [51].
One confounding factor discovered by the QIBA group in calibrated elastic and viscoelastic phantom studies has been the presence of a negative bias in shear wave speed as a function of increasing focal depth when using curvilinear arrays. The relative magnitude of these biases has been <4% for focal depths ranging from 3 to 7 cm (https://github.com/RSNA-QIBA-US-SWS/).
Another resource generated through the RSNA QIBA US SWS effort has been the generation and public release of digital phantom data generated through finite element method models of shear wave propagation in elastic and viscoelastic materials for shear wave reconstruction development validation [50]. Additionally, standardized sequences for generating and processing shear waves using the Verasonics ultrasound research platform have also been released for public use [51]. The weblinks to this information is provided in Table 4.2.
Conclusions
Acoustic radiation force elasticity imaging methods have become a viable clinical option to noninvasively evaluate liver stiffness. Unlike external vibration-base methods, acoustic radiation force excitations can be focused directly in the tissue of interest, while also providing real-time, ultrasonic B-mode imaging guidance and clinical evaluation. Acoustic radiation force methods are also available to screen for and characterize liver masses. Efforts to standardize and provide more consistent measurements between different manufacturer systems is being addressed by consensus documents and guideline documents, and the RSNA QIBA Ultrasonic Shear Wave Speed group is taking important steps toward characterizing the performance of these systems in calibrated elastic and viscoelastic media.
References
Nyborg WLM, Litovitz T, Davis C. Acoustic streaming, vol. IIA. New York: Academic Press; 1965. p. 265–331.
Apfel RE, Chu BT. Acoustic radiation pressure produced by a beam of sound. J Acoust Soc Am. 1982;72(6):1673–87.
Doherty JR, Trahey GE, Nightingale KR, Palmeri ML. Acoustic radiation force elasticity imaging in diagnostic ultrasound. IEEE Trans Ultrason Ferroelectr Freq Control. 2013;60(4):685–701.
Shiina T, Nightingale KR, Palmeri ML, Hall TJ, Bamber JC, Barr RG, et al. WFUMB guidelines and recommendations for clinical use of ultrasound elastography: part 1: basic principles and terminology. Ultrasound Med Biol. 2015;41(5):1126–47.
Sandrin L, Tanter M, Gennisson JL, Catheline S, Fink M. Shear elasticity probe for soft tissues with 1-D transient elastography. IEEE Trans Ultrason Ferroelectr Freq Control. 2002;49(4):436–46.
Sandrin L, Tanter M, Catheline S, Fink M. Shear modulus imaging with 2-D transient elastography. IEEE Trans Ultrason Ferroelectr Freq Control. 2002;49(4):426–35.
Sandrin L, Fourquet B, Hasquenoph J-M, Yon S, Fournier C, Mal F, et al. Transient elastography: a new non-invasive method for assessment of hepatic fibrosis. Ultrasound Med Biol. 2003;29(12):1705–13.
Pinton GF, Dahl JJ, Trahey GE. Rapid tracking of small displacements with ultrasound. IEEE Trans Ultrason Ferroelectr Freq Control. 2006;53(6):1103–17.
Nightingale KR, Church CC, Harris G, Wear KA, Bailey MR, Carson PL, et al. Conditionally increased acoustic pressures in nonfetal diagnostic ultrasound examinations without contrast agents: a preliminary assessment. J Ultrasound Med. 2015;34(7):1–41.
Ncrp. report No. 113: exposure criteria for medical diagnostic ultrasound: I. Criteria based on thermal mechanisms: National Council on Radiation Protection and Measurements; 2002.
Deng Y, Palmeri ML, Rouze NC, Rosenzweig SJ, Abdelmalek MF, Nightingale KR. Analyzing the impact of increasing mechanical index and energy deposition on shear wave speed reconstruction in human liver. Ultrasound Med Biol. 2015;41(7):1948–57.
Palmeri ML, McAleavey SA, Fong KL, Trahey GE, Nightingale KR. Dynamic mechanical response of elastic spherical inclusions to impulsive acoustic radiation force excitation. IEEE Trans Ultrason Ferroelectr Freq Control. 2006;53(11):2065–79.
Rosenzweig SJ, Palmeri ML, Nightingale KR. Analysis of rapid multi-focal zone ARFI imaging. IEEE Trans Ultrason Ferroelectr Freq Control. 2015;62(2):280–9.
Kasai C, Koroku N, Koyano A, Omoto R. Real-time two-dimensional blood flow imaging using an autocorrelation technique. IEEE Trans Ultrason Control. 1985;32(3):458–63.
Loupas T, Powers JT, Gill RW. An axial velocity estimator for ultrasound blood flow imaging, based on a full evaluation of the Doppler equation by means of a two-dimensional autocorrelation approach. IEEE Trans Ultrason Ferroelectr Freq Control. 1995;42(4):672–88.
Pesavento A, Perrey C, Krueger M, Ermert H. A time-efficient and accurate strain estimation concept for ultrasonic elastography using iterative phase zero estimation. IEEE Trans Ultrason Ferroelectr Freq Control. 1999;46(5):1057–67.
Byram B, Trahey G, Palmeri M. Bayesian speckle tracking. Part I: an implementable perturbation to the likelihood function for ultrasound displacement estimation. IEEE Trans Ultrason Ferroelectr Freq Control. 2013;60(1):132–43.
Byram B, Trahey G, Palmeri M. Bayesian speckle tracking. Part II: biased ultrasound displacement estimation. IEEE Trans Ultrason Ferroelectr Freq Control. 2013;60(1):144–57.
Palmeri M, McAleavey S, Trahey G, Nightingale K. Ultrasonic tracking of acoustic radiation force-induced displacements in homogeneous media. IEEE Trans Ultrason Ferroelectr Freq Control. 2006;53(7):1300–13.
Ferraioli G, Filice C, Castera L, Choi BI, Sporea I, Wilson SR, et al. WFUMB guidelines and recommendations for clinical use of ultrasound elastography: part 3: liver. Ultrasound Med Biol. 2015;41(5):1161–79.
Barr RG, Ferraioli G, Palmeri ML, Goodman ZD, Garcia-Tsao G, Rubin J, et al. Elastography assessment of liver fibrosis: society of radiologists in ultrasound consensus conference statement. Ultrasound Q. 2016;32(2):94–107.
Giannantonio DM, Dumont DM, Trahey GE, Byram BC. Comparison of physiological motion filters for in vivo cardiac ARFI. Ultrason Imaging. 2011;33(2):89–108.
Palmeri ML, Frinkley KD, Oldenburg KG, Nightingale KR. Characterizing acoustic attenuation of homogeneous media using focused impulsive acoustic radiation force. Ultrason Imaging. 2006;28(2):114–28.
Sarvazyan AP, Rudenko OV, Swanson SD, Fowlkes JB, Emelianov SY. Shear wave elasticity imaging: a new ultrasonic technology of medical diagnostics. Ultrasound Med Biol. 1998;24(9):1419–35.
Lai WM, Rubin DEK. Introduction to continuum mechanics. Woburn: Butterworth-Heinmann; 1999.
Bercoff J, Tanter M, Fink M. Supersonic shear imaging: a new technique for soft tissue elasticity mapping. IEEE Trans Ultrason Ferroelectr Freq Control. 2004;51(4):396–409.
Deng Y, Palmeri ML, Rouze NC, Trahey GE, Haystead CM, Nightingale KR. Quantifying image quality improvement using elevated acoustic output in B-mode harmonic imaging. Ultrasound Med Biol. 2017;43(10):2416–25.
Deng Y, Palmeri ML, Rouze NC, Haystead CM, Nightingale KR. Evaluating the benefit of elevated acoustic output in harmonic motion estimation in ultrasonic shear wave elasticity imaging. Ultrasound Med Biol. 2018;44(2):303–10.
Dumont DM, Byram BC. Robust tracking of small displacements with a Bayesian estimator. IEEE Trans Ultrason Ferroelectr Freq Control. 2016;63(1):20–34.
Dumont DM, Walsh KM, Byram BC. Improving displacement signal-to-noise ratio for low-signal radiation force elasticity imaging using Bayesian techniques. Ultrasound Med Biol. 2016;42(8):1986–97.
Song P, Zhao H, Manduca A, Urban M, Greenleaf J, Chen S. Comb-push ultrasound shear elastography (CUSE): a novel method for two-dimensional shear elasticity imaging of soft tissues. IEEE Trans Med Imaging. 2012;31:1821–32.
Lipman SL, Rouze NC, Palmeri ML, Nightingale KR. Evaluating the improvement in shear wave speed image quality using multidimensional directional filters in the presence of reflection artifacts. IEEE Trans Ultrason Ferroelectr Freq Control. 2016;63(8):1049–63.
Huwart L, Sempoux C, Vicaut E, Salameh N, Annet L, Danse E, et al. Magnetic resonance elastography for the noninvasive staging of liver fibrosis. Gastroenterology. 2008;135(1):32–40.
Palmeri ML, Wang MH, Dahl JJ, Frinkley KD, Nightingale KR, Zhai L. Quantifying hepatic shear modulus in vivo using acoustic radiation force. Ultrasound Med Biol. 2008;34(4):546–58.
Wang MH, Palmeri ML, Rotemberg VM, Rouze NC, Nightingale KR. Improving the robustness of time-of-flight based shear wave speed reconstruction methods using RANSAC in human liver in vivo. Ultrasound Med Biol. 2010;36(5):802–13.
Rouze NC, Wang MH, Palmeri ML, Nightingale KR. Robust estimation of time-of-flight shear wave speed using a Radon sum transformation. IEEE Trans Ultrason Ferroelectr Freq Control. 2010;57(12):2662–70.
Song P, Manduca A, Zhao H, Urban MW, Greenleaf JF, Chen S. Fast shear compounding using robust 2-d shear wave speed calculation and multi-directional filtering. Ultrasound Med Biol. 2014;40(6):1343–55.
Song P, Urban MW, Manduca A, Greenleaf JF, Chen S. Coded excitation plane wave imaging for shear wave motion detection. IEEE Trans Ultrason Ferroelectr Freq Control. 2015;62(7):1356–72.
Song P, Macdonald M, Behler R, Lanning J, Wang M, Urban M, et al. Two-dimensional shear-wave elastography on conventional ultrasound scanners with time-aligned sequential tracking (TAST) and comb-push ultrasound shear elastography (CUSE). IEEE Trans Ultrason Ferroelectr Freq Control. 2015;62(2):290–302.
Rouze NC, Wang MH, Palmeri ML, Nightingale KR. Parameters affecting the resolution and accuracy of 2-D quantitative shear wave images. IEEE Trans Ultrason Ferroelectr Freq Control. 2012;59(8):1729–40.
Palmeri M, Nightingale K, Fielding S, Rouze N, Deng YF, Lynch T, et al. RSNA QIBA ultrasound shear wave speed phase II phantom study in viscoelastic media. IEEE Int Ultra Sym. 2015.
Rouze NC, Palmeri ML, Nightingale KR, editors. Estimation of model parameters characterizing dispersion in ARFI induced shear waves in in vivo human liver. In: 2014 IEEE International Ultrasonics Symposium; 2014.
Nightingale K, Rouze N, Rosenzweig S, Wang M, Abdelmalek M, Guy C, et al. Derivation and analysis of viscoelastic properties in human liver: impact of frequency on fibrosis and steatosis staging. IEEE Trans Ultrason Ferroelectr Freq Control. 2015;62(1):165–75.
Rouze NC, Deng Y, Trutna CA, Palmeri ML, Nightingale KR. Characterization of viscoelastic materials using group shear wave speeds. IEEE Trans Ultrason Ferroelectr Freq Control. 2018;65(5):780–94.
Asbach P, Klatt D, Hamhaber U, Braun J, Somasundaram R, Hamm B, et al. Assessment of liver viscoelasticity using multifrequency MR elastography. Magn Reson Med. 2008;60:373–9.
Chen S, Sanchez W, Callstrom MR, Gorman B, Lewis JT, Sanderson SO, et al. Assessment of liver viscoelasticity by using shear waves induced by ultrasound radiation force. Radiology. 2012;266(3):964–70.
Chen X, Shen Y, Zheng Y, Lin H, Guo Y, Zhu Y, et al. Quantification of liver viscoelasticity with acoustic radiation force: a study of hepatic fibrosis in a rat model. Ultrasound Med Biol. 2013;39(11):2091–102.
Hall TJ, Milkowski A, Garra B, Carson P, Palmeri M, Nightingale K, et al. RSNA/QIBA: Shear wave speed as a biomarker for liver fibrosis staging. In: 2013 IEEE International Ultrasonics Symposium (IUS); 2013, pp. 397–400.
Deng Y, Rouze NC, Palmeri ML, Nightingale KR. On system-dependent sources of uncertainty and bias in ultrasonic quantitative shear-wave imaging. IEEE Trans Ultrason Ferroelectr Freq Control. 2016;63(3):381–93.
Palmeri ML, Qiang B, Chen S, Urban MW. Guidelines for finite-element modeling of acoustic radiation force-induced shear wave propagation in tissue-mimicking media. IEEE Trans Ultrason Ferroelectr Freq Control. 2017;64(1):78–92.
Deng Y, Rouze NC, Palmeri ML, Nightingale KR. Ultrasonic shear wave elasticity imaging sequencing and data processing using a verasonics research scanner. IEEE Trans Ultrason Ferroelectr Freq Control. 2017;64(1):164–76.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2020 Springer Nature Switzerland AG
About this chapter
Cite this chapter
Palmeri, M.L. (2020). Characterizing Liver Stiffness with Acoustic Radiation Force. In: Mueller, S. (eds) Liver Elastography. Springer, Cham. https://doi.org/10.1007/978-3-030-40542-7_4
Download citation
DOI: https://doi.org/10.1007/978-3-030-40542-7_4
Published:
Publisher Name: Springer, Cham