Abstract
Spinal cord stimulation is commonly performed for chronic low back, neck, leg, and arm pain. This chapter provides a step-by-step guide on how place a thoracic lead safely and shows all the relevant C-arm and needle positions that need to be demonstrated during the Fellow of Interventional Pain Practice (FIPP) exam administered by the World Institute of Pain (WIP). Both native and edited high-quality images are included, so the fluoroscopy anatomy is more easily understood.
Possible complications of the procedures and common reasons for failure at the FIPP exam are also outlined. Evidence for the procedure is provided based on the available literature reviewed by the Benelux section of WIP and also by the American Society of Interventional Pain Physicians (ASIPP).
Access provided by Autonomous University of Puebla. Download chapter PDF
Similar content being viewed by others
Keywords
- Spinal cord stimulation
- low back pain
- lumbar radiculopathy
- cervicalgia
- neck pain
- cervical radiculopathy
- FIPP exam
- fluoroscopic needle placement
Equipment and Monitoring
-
Standard ASA monitoring
-
Fluoroscopy
-
Sterile prep, and drape
-
Skin local anesthesia prior to any needle larger than 25G (unless sedation is used)
-
Coaxial view is always used to advance needle, unless otherwise specified
-
CPR equipment and medications available
-
IV antibiotics single-dose preoperative is recommended
-
13–15G 3.5 inch (90 mm) to 6 inch (150 mm) straight or Coude epidural needle
-
Saline 5–15 ml
-
Spinal cord stimulator lead, anchors, tunneling tools, extension cables, and IPG/receiver
-
Loss of resistance (LOR) syringe
-
2.0 nylon/vicryl, 2.0 silk, 3.0 vicryl sutures
-
Occlusive bandage–if possible silver impregnated
-
Bacitracin in saline for wound irrigation
Anatomy
Target is the posterior epidural space, to stimulate the dorsal columns with traditional tonic stimulation
-
Back pain
-
Epidural entry level: L2–3 or L3–4
-
Level of top of the lead: T8–11
-
-
Leg pain
-
Epidural entry level: L2–3, L3–4
-
Level of tip of the lead: generally, T9–10
-
-
Foot pain–epidural entry level: L2–3, L3–4
-
Level of tip of lead: T11-L1
-
-
Arm pain
-
Epidural entry level: T2–3 to T5–6
-
Level of tip of the lead: C3–4
-
-
Neck pain
-
Epidural entry level: T4–5, T5–6
-
Level of tip of the lead: C2–3
-
-
Abdominal pain
-
Epidural entry level: L2–3, L1–2
-
Level of tip of the lead: T4–5 to 5–6
-
The above levels reflect general concepts. The levels of epidural entry and target level for the lead may vary based on patient size, anatomy, previous spinal surgery, pain location, or other circumstances. For traditional tonic stimulation, it is important to overlap the areas of pain with a sensation of paresthesia. More recently, a number of novel waveforms and frequencies that do not require a paresthesia have become clinically available. Placement for these devices tends to have a set anatomic location. Burst and high frequency are placed between T8 and T9 for low back pain and at C2–3 for neck pain with or without radiculopathy.
Structures to Keep in Mind and Possible Complications
-
Dura puncture → spinal headache
-
Dural puncture and intrathecal medication administration → sudden onset motor block, cardiac/respiratory arrest, death
-
Subdural puncture → spinal cord compression, or medication related sudden onset motor block, cardiac/respiratory arrest, death
-
Epidural blood vessels → bleeding, spinal cord compression and epidural hematoma formation, possible onset even after procedure is completed. This can also occur with insertion or removal of the electrodes
-
Direct spinal cord damage/injection → loss of bladder/bowel function, various paresthesias, motor loss, depending on injected amount, syrinx formation, quadriplegia, death
-
Epidural infection → abscess
-
Exiting nerve roots → nerve injury
-
Procedure should not be performed while patient is taking certain anticoagulants
-
Heavy sedation should be preferably avoided for routine procedures
-
Infection
-
Bleeding
-
Postprocedure pain
-
Vasovagal reaction
-
Allergic reaction
Fluoroscopy Technique, Target Localization
-
Patient in prone position , with pillows under abdomen to decrease lumbar lordosis
-
True anteroposterior (AP) image, spinous processes in midline (Fig. 14.1)
-
Slight caudad tilt in difficult cases may facilitate visualizing the interlaminar opening (Fig. 14.2a)
-
Non-coaxial view technique :
-
Entry 2 vertebral bodies below target interlaminar space, in order to reach epidural space utilizing a shallow angle, entering in the midline. This will facilitate driving the lead midline in the posterior epidural space
-
-
Coaxial view technique :
-
30° caudad tilt of the C-arm
-
Skin entry just paramedian of the spinous process, immediately below the target interlaminar space
-
Procedure Steps
-
Target the lamina caudad from the target interlaminar epidural place, in a paramedian approach (Fig. 14.2b, c)
-
Walk needle off the cephalad edge of lamina to engage in ligamentum flavum
-
Advance needle with LOR technique
-
After LOR, check lateral view to see the depth of the needle tip (Fig. 14.3)
-
Thread lead to confirm epidural position (Fig. 14.4a, b)
-
Advance electrode cephalad in the posterior epidural space (Fig. 14.5)
-
Use a forefinger/thumb “pill rolling” technique as the lead is advanced to maintain lead position as close as possible to the anatomical midline
-
After lead placement, for tonic stimulation test, stimulation is performed with the patient awake. For paresthesia stimulation, the leads are left in an anatomic location
-
Analgesia/paresthesia is reported by the patient and should cover the painful area for tonic stimulation
-
Secure leads to fascia or skin once best pain area coverage is obtained
-
When leads are in final position, check and save lateral and AP images (Figs. 14.6, 14.7a, b, and 14.8)
-
For cervical levels: Consider using contralateral oblique view (CLO) where the tip of the needle can be advanced to the ventrolaminar line (VILL) without LOR. At interlaminar line, LOR should be employed. Detailed description provided in cervical epidural chapter
Clinical Pearls
-
If the needle angle is too acute at epidural entry, it will be very difficult to keep the lead/s in the posterior epidural space
-
Always confirm lead position in lateral view especially if patient experiences motor stimulation. This is done to rule out the possibility of anterior migration of the leads
-
Repositioning of leads can be performed with a combination of AP and lateral views at regular intervals
-
More than one lead may need to be placed for adequate pain area coverage
Unacceptable, Potentially Harmful Needle Placement on Exam
-
Not checking lateral view
-
Compromising spinal cord
-
Rough needle/catheter manipulation, which may compromise spinal cord
-
Obvious signs of lack of understanding thoracic spinal anatomy, for example, placing the lead in the subcutaneous area, posterior to the spinal canal, and believing it is well positioned
Unacceptable, But Not Harmful Needle Placement on Exam
-
Not reaching the epidural space, but lateral view and LOR technique was carefully utilized to prevent spinal cord damage
-
Lead in anterior epidural space
-
Abandoned procedure
Evidence
Suggested Reading
Jang H-D, Kim M-S, Chang C-H, Kim S-W, Kim O-L, Kim S-H. Analysis of failed spinal cord stimulation trials in the treatment of intractable chronic pain. J Korean Neurosurg Soc. 2008;43(2):85–9.
Levy RM. Anatomic considerations for spinal cord stimulation. Neuromodulation: Technology at the Neural Interface. 2014;17:2–11.
Lee RA, van Zundert AAJ, Botha CP, Lataster LMA, van Zundert TCRV, van der Ham WGJM, Wieringa PA. The anatomy of the thoracic spinal canal in different postures. Reg Anesth Pain Med. 35(4):364–9.
Sitzman BT, Provenzano DA. Best practices in spinal cord stimulation. Spine. 42:S67–71.
Yampolsky C, Hem S, Bendersky D. Dorsal column stimulator applications. Surg Neurol Int. 2012;3:275.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Additional information
This chapter was reviewed by Alan Berkman; Fabricio D. Assis; Andrea M. Trescot; Milan Stojanovic; Peter S. Staats; Agnes R. Stogicza; Andre M. Mansano.
Rights and permissions
Copyright information
© 2020 Springer Nature Switzerland AG
About this chapter
Cite this chapter
Akbas, M., Stogicza, A.R. (2020). Spinal Cord Stimulator. In: Stogicza, A.R., Mansano, A.M., Trescot, A.M., Staats, P.S. (eds) Interventional Pain . Springer, Cham. https://doi.org/10.1007/978-3-030-31741-6_14
Download citation
DOI: https://doi.org/10.1007/978-3-030-31741-6_14
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-030-31740-9
Online ISBN: 978-3-030-31741-6
eBook Packages: MedicineMedicine (R0)