9.1 Introduction

Human blood is a complex fluid composed mainly of suspended deformable red blood cells (RBCs) within the plasma flow, which are responsible for the supply of oxygen and nutrients to, and removal of carbon dioxide and metabolic wastes from tissues. Phenomena of blood flow in microcirculation depend on several combined effects such as cell deformability, flow shear rates and vessel wall, together with micro-scale biochemical and physiological factors. Since the availability of microscopic techniques scientists have started using optical methods mainly because they are less invasive for measuring the flow field. However, the fact that measurements of blood flow in the microcirculation need to be performed at very small scales has limited the number of flow measurement techniques suitable for this purpose. For instance, MRI and ultrasonography are not suitable for obtaining quantitative flow information in micro-vessels mainly due to their poor spatial resolution. The majority of the works performed in the past to measure blood flow in micro-vessels and micro-channels are double-slit photometry [4,12], video microscopy and image analysis [2,6,1315,41,43] and laser-Doppler anemometry [5,7,9,16,49]. Recently, the considerable progress in computing capability, optics, high-speed imaging and image processing techniques have made it possible to successfully combine a conventional PIV system with an inverted epifluorescent microscope [19,39,45,51]. This combination, known as micro-PIV, has greatly increased the resolution of conventional PIV; as a result, this technique has gained widespread use to investigate blood flow behaviour in both micro-channels and micro-vessels. Sugii and his co-workers, by using a conventional micro-PIV system, have used red blood cells as tracer markers to measure their velocities in both straight [46] and bifurcated arterioles [40] and they found that the velocity profiles were markedly blunt in the central region. Later they measured both tracer particles and RBCs through a 100 μm glass capillary and they reported that by using in vitro blood with 20% hematocrit (Hct) the velocity profiles were parabolic [47]. More recently, by using liposome tracer particles the blood-plasma velocity has been measured in the beating heart of a chicken embryo [50]. Kim and Lee [18] have analysed the flow behaviour of blood through a circular opaque micro-channel by using an X-ray PIV technique. Their measurements have shown typical non-Newtonian flow characteristics of blood such as yield stress and shear-thinning effects. Generally, most of the findings relating to blood flow dynamics using conventional micro-PIV have yielded conflicting results. In addition, such measurements do not give detailed information on the micro-scale flow behaviour of RBCs in micro-vessels. Hence, there is a need to apply other kinds of optical methods better able to obtain further insight into the complex flow behaviour of blood in the microcirculation.

Recently, the ability to combine both conventional particle image velocimetry (PVI) and particle tracking velocimetry (PTV) systems with a spinning disc confocal microscope (SDCM) has proven an extremely powerful technique to obtain detailed qualitative and quantitative measurements of in vitro blood flowing through micro-channels. The current chapter will present the most relevant theoretical and technical issues related to both methods and also a comparison between them. Additionally, our most recent confocal micro-PIV/PTV results of blood flow behaviour in both glass and polydimethylsiloxane (PDMS) micro-channels are reviewed.

9.2 PIV/PTV Principles

9.2.1 Calculation Methodology

The principle of calculating several physical parameters (displacement, velocity, shear stress, etc.) related to fluid mechanics is common to both PIV and micro-PIV. The traditional way consists of measuring the displacement of fluorescent tracer particles flowing within the working fluid. However, to measure physiological fluids, such as blood, it is also common to use labelled blood cells as natural tracer particles. By applying a powerful light source, the tracer particles or cells of interest are illuminated and as a result objects with a known time interval (Δt) can be recorded by a high speed camera. By using a short Δt, it can be assumed that the magnitude and direction of the object velocities are constant. As a result, the location of a particle on two consecutive images can be used to estimate its instantaneous velocity as:

$$ \vec{u} = \frac{{\vec{x}_{{t + \Delta t}}^i - \vec{x}_t^i}}{{\Delta t}} $$
(9.1)
$$ \vec{v} = \frac{{\vec{y}_{{t + \Delta t}}^i - \vec{y}_t^i}}{{\Delta t}} $$
(9.2)

where u and v are respectively the velocity components of the particle in the x and y direction. Figure 9.1 shows an overview of the PIV/PTV principle.

Fig. 9.1
figure 00091

Principle for both PIV and PTV (adapted from [24])

The approach used to calculate the velocity of one particle can be extended to the entire flow field of interest. In microfluidics it is common practice to average the instantaneous velocities over a large number of recorded image pairs (N usually greater than 20). Hence, the time-average mean velocity vector of the flow can be defined as

$$ \vec{U} = \frac{1}{N}\sum\limits_{{t = 1}}^N {{{\vec{u}}_t}} $$
(9.3)
$$ \vec{V} = \frac{1}{N}\sum\limits_{{t = 1}}^N {{{\vec{v}}_t}} $$
(9.4)

where N is the number of instantaneous velocity measurements, u and v are the time-average mean velocity components of the particle in the x and y directions, respectively.

9.2.2 PIV vs. PTV Methods

The density of trace particles in the images determines the most suitable PIV methodology to calculate the velocity fields. For high concentration of particles, i.e. the interrogation window contains at least three particles, the method is known as high-image-density PIV mode (see Fig. 9.2) [1]. Several researchers consider this method as the most reliable way of determining the velocity field which is estimated by using a statistical technique known as cross-correlation. It is described with much more detail elsewhere [24,27], the overall method being known as PIV cross-correlation, or simply PIV.

Fig. 9.2
figure 00092

Confocal high density image with pure water (left image) and confocal low density image with blood 30 % Hct (right image)

In some specific situations such as physiological fluids with high concentrations of cells, the amount of tracer particles captured within the fluid is often low (see Fig. 9.2). One way to overcome the problem of this particular condition is to measure the displacements by tracking individual particles or cells (see Fig. 9.3). This low-image-density PIV methodology is often referred to as particle tracking velocimetry (PTV) or as single particle tracking (SPT) [1,44]. By using this method detailed quantitative information can be obtained on the motion of particles and cells flowing within the working fluid. PTV is, however, limited in its ability to compute automatically particle trajectories, especially when there are multiple collisions of particles or cells. One way to overcome this limitation is by using manual tracking methods.

Fig. 9.3
figure 00093

Trajectory of RBC flowing in a micro-channel by using a manual particle tracking technique from ImageJ

9.3 Confocal Micro-PIV/PTV System

9.3.1 Conventional Micro-PIV System

A conventional micro-PIV system consists essentially of a microscope, an objective lens, optical filters, a light source for flow illumination and a high speed camera. Figure 9.4 gives a schematic illustration of a conventional micro-PIV system. Briefly, the light enters the microscope and is reflected 90° upwards by a dichromatic mirror to be transmitted through the objective lens which illuminates the entire flow volume. The objective lens collects the light emitted from the particles which goes back to the dichromatic mirror and to a high speed camera to record the signals from the trace particles. Finally, the recorded images are transferred to a computer to use a PIV post-processing method such as the cross-correlation technique. It is worth mentioning that the resolution of a micro-PIV system is influenced by several factors such as: out-of-focus particle images due to the volume illumination, and density and size of the tracer particles. By using a confocal system, most of the out-of-focus particles can be removed, consequently reducing the errors generated in the velocity field measurements.

Fig. 9.4
figure 00094

Experimental setup of a conventional micro-PIV system.

9.3.2 Confocal Micro-PIV System

A confocal micro-PIV system consists usually of an inverted microscope combined with a confocal scanning unit (CSU), a high resolution objective lens, a high power light source (usually laser) and a high speed camera, as shown in Fig. 9.5. In a confocal system the light enters the CSU and then is conducted to the microscope to illuminate the sample from below the microscope stage. The light emitted from the fluorescent trace particles goes back into the CSU and then to a high-speed camera to capture the confocal images.

Fig. 9.5
figure 00095

Experimental setup of a confocal micro-PIV system

9.3.3 Comparison Between Conventional and Confocal Micro-PIV

Figures 9.4 and 9.5 show the main components of both conventional and confocal micro-PIV systems. It is clear that the main difference between the two systems is the existence of a CSU in the confocal system. A study to compare the systems was performed under the same flow conditions. The flow measurements were performed in a 100 μm glass square micro-channel where pure water seeded with 0.15% of tracer particles was pressure driven by means of a syringe pump.

Figure 9.6, clearly demonstrates that by using a CSU it is possible to obtain much clearer image definition of the individual trace particles. Hence, a confocal system reduces the out-of-focus background noise and therefore increases the contrast and definition of an image. By contrast, images recorded by the conventional system were largely blurred, since the out-focus-light reduces contrast and decreases resolution making it difficult to identify the tracer particles within the flow. Note that the conventional microscopic images were obtained by using an epifluorescent microscope equipped with a mercury lamp and colour filters. These images may be improved by using a more powerful light source, such as a laser or metal halide.

Fig. 9.6
figure 00096

Comparison of particle images from (a) conventional and (b) confocal micro-PIV system at 20 μm depth for pure water

9.4 Confocal Micro-PIV/PTV Measurements and Results

This section shows the most recent studies on blood flow behaviour in micro-channels performed by our confocal micro-PIV/PTV system. The main emphasis is on the confocal micro-PIV/PTV results obtained from both glass and polydimethylsiloxane (PDMS) micro-channels.

9.4.1 Working Fluids, Micro-channels, Experimental Set-up and Image Analysis

Confocal micro-PIV measuements were performed with different working fluids such as physiological saline (PS) with fluorescent trace particles, and in vitro blood containing trace particles and human RBCs with a hematocrit (Hct) up to 20%. Confocal micro-PTV experiments were carried out also with different working fluids such as dextran 40 (Dx-40) containing about 3% Hct up to 35% Hct of human RBCs. For this latter system, rather than employing fluorescent particles labelled RBCs were used to study the flow in micro-channels. A full description of the procedure for labeling human RBCs can be found in Lima et al. [28].

For both confocal systems experiments were performed with different kinds of micro-channels, such as 100-μm and 50-μm circular and square borosilicate glass micro-channels as fabricated by Vitrocom (Mountain Lakes, NJ, USA) [23,25,31]. By using a soft lithographic technique it was also possible to fabricate rectangular polydimensiloxane (PDMS) micro-channels with complex geometries such as stenoses, bifurcations, confluences and hyperbolic contractions [11,2022,27,3436,52,53]. Additionally, by using a wire casting technique [29] it was possible to fabricate 75-μm circular PDMS micro-channels.

Generally both confocal systems used in our flow studies consisted of an inverted microscope (IX71; Olympus, Tokyo, Japan) combined with a confocal scanning unit (CSU22; Yokogawa, Tokyo, Japan), a diode-pumped solid state (DPSS) laser (Laser Quantum Ltd., Stockport, UK) and a high-speed camera (Phantom v7.1;Vision Research, NJ, USA) (as in Fig. 9.5). All the micro-channels were placed on the stage of the microscope where the flow rate of the working fluid was kept constant using a syringe pump (KD Scientific Inc., Holliston, MA, USA). A thermo-plate controller (Tokai Hit, Shizuoka, Japan) was also used: this achieved a surrounding temperature around the micro-channel of about 37 °C. More detailed information about this system can be found elsewhere [23,24,27,28].

All the measurements were digitised and transferred to a computer for evaluation using Phantom camera control software (PH607). The PIV images of the flowing particles were processed and the flow velocity was determined using the cross-correlation PIV method (PivView). A full description and evaluation of the confocal micro-PIV system used in our studies can be found in Lima et al. [23]. For the case of the confocal micro-PTV measurements, series of confocal images were evaluated in Image J [3] using the manual tracking MtrackJ [38] plug-in. Generally, the motions of the labelled RBCs were manually tracked through successive images using the bright centroid criteria available at the MtrackJ. Using this method, it was possible to track labelled RBCs even when two cells were in close proximity. After obtaining a series of x and y positions, data were exported for the determination of several physical quantities such as velocity, radial displacement, and dispersion coefficient.

9.4.2 Confocal Micro-PIV Results

The ability of a confocal micro-PIV system to measure both pure water and diluted suspensions of RBCs was demonstrated by Lima et al. [23]. This study was performed with a square glass micro-channel and it was found that there was good agreement between the measured velocity profiles of pure water and an established analytical solution. Further work was performed by Lima et al. this time to measure both physiological saline (PS) and in vitro blood (20% Hct) in a rectangular PDMS micro-channel [27]. As expected for a long rectangular micro-channel, the velocity profiles were markedly smooth and flat in the centre of the micro-channel (as in Fig. 9.7). However, for the in vitro blood with 20% Hct, small fluctuations in the shape of the ensemble velocity profiles were observed (see Fig. 9.7). Possible reasons for these fluctuations are the interactions between neighbouring RBCs, the formation of a cell-free layer, abrupt increase in the shear rate in the vicinity of the wall, Hct temporal variation and the light scattered and absorbed from the RBCs.

Fig. 9.7
figure 00097

Ensemble velocity profiles with PS (left image) and in vitro blood 20 % Hct (right image) in a rectangular PDMS micro-channel (from [27])

Lima and his colleagues have performed another study to measure both ensemble and instantaneous velocity profiles for in vitro blood (Hct up to 17%) flowing through a 100-μm square micro-channel [25]. All the measurements were made in the middle plane at a constant flow rate and low Reynolds number (Re = 0.025). Although the ensemble velocity profiles were markedly parabolic, some fluctuations in the instantaneous velocity profiles were found to be closely related to the Hct increase. This study has also shown strong evidence that the Root Mean Square (RMS) increases with the Hct increase suggesting that the concentration of RBCs within the plasma flow strongly influences the measurements of the instantaneous velocity fields. Possible reasons for the RMS increase are the motion and interaction of RBCs and the light scattered and absorbed by the RBCs. This latter behaviour seems to be predominant at an Hct value of about 17%. More detailed information about these results can be found in Lima et al. [25].

9.4.3 Confocal Micro-PTV Results

Confocal micro-PIV experiments have shown the ability to measure with good accuracy in vitro blood with Hct up to 9%, in a 100-μm square micro-channel. However, for Hct bigger than 9%, the light absorbed by the RBCs contributes to diminishing the concentration of tracer particles in the recorded confocal images. The low density images become more evident for Hct bigger than 20%, which generates spurious errors in the velocity fields [25]. Therefore, Lima and his colleagues [24,26,28] have applied a new approach, known as confocal micro-PTV, to track the trajectories of individual labelled RBCs at high Hcts. Figures 9.8 and 9.9 show the ability of this confocal method to measure the motion of blood cells at both diluted (3% Hct) and high (20% Hct) suspensions of RBCs, respectively. Additionally, successful measurements were performed in a 75-μm circular PDMS micro-channel as shown in Fig. 9.10.

Fig. 9.8
figure 00098

Labelled RBC trajectory at diluted suspensions of cells (3 % Hct) obtained by the proposed confocal micro-PTV system (adapted from [26])

Fig. 9.9
figure 00099

Trajectory of a labelled RBC at high suspensions of cells (20 % Hct) obtained by the proposed confocal micro-PTV system. Experiments were carried out in glass micro-channels (adapted from [26])

Fig. 9.10
figure 000910

Trajectory of labelled RBCs at low (3 % Hct) and high suspensions of RBCs (23 % Hct) obtained by the proposed confocal micro-PTV system. Experiments were carried out in circular PDMS micro-channel (adapted from [29])

The ability of the confocal system to generate thin in-focus planes has allowed both qualitative and quantitative measurements in flowing blood at concentrated suspensions (up to 35% Hct) of: cell–cell hydrodynamic interaction, RBC orientation and RBC radial dispersion at different depths. Hence, to evaluate the capability of the confocal micro-PTV system to track RBCs, motions of labelled RBCs were followed at several Hcts (3–37%). The measurements of complex micro-rheological events in flowing blood (such as interaction and orientation of blood cells) were performed near the wall of the micro-channel (z = 20 μm) with Hct ~ 20% and Re ~0.007.

9.4.3.1 RBC–RBC Interactions in Flowing Blood

The effect of hemodynamic interactions on the motion of RBCs depend on multi-physics factors, such as shear rate, deformability, plasma layer and wall constriction. Figure 9.11 shows the trajectories of two-RBC interactions close to the cell-free layer. This figure shows clearly the radial disturbance effect enhanced by the collision of a neighbouring RBC. The radial displacement (ΔR) of the RBCs increases from two to six times of its initial radial displacement. For the case of RBC1, ΔR continues to increase, due not only to the interaction above but also to others which may occur with neighbouring cells.

Fig. 9.11
figure 000911

Two-RBC interactions at different time intervals (adapted from [28])

9.4.3.2 RBC–WBC Interactions in Flowing Blood

The hemodynamic interaction effect of WBC on the motion of RBCs was also investigated by using confocal micro-PTV. Figure 9.12 shows the interaction of a RBC with the centre upper part of a WBC. Note that the RBC is located in the in-focus plane (high intensity) whereas part of the WBC is under the in-focus plane (lower intensity). It is possible to observe that transversal RBC displacement tends to increase when a collision occurs with a neighbouring WBC.

Fig. 9.12
figure 000912

Lateral RBC–WBC interaction at different time intervals (adapted from [32])

9.4.3.3 Translational and Rotational Motion of RBCs

The orientation of RBCs was extensively studied at low Hct (values <1%), but this behaviour differed from that for moderate and high Hcts. Figures 9.13 and 9.14 show the orientation of a RBC without and with an interaction, for flow close to the wall of a glass micro-channel. By adjusting the image contrast, it was possible to quantify both translational and rotational motion. The translational motion was measured at the centre of the RBC whereas the rotational was measured along the membrane: these are as shown in Figs. 9.13 and 9.14. The RBC motion without interaction (Fig. 9.13) was measured along the wall of the micro-channel. This RBC did not suffer any interaction with a neighbouring RBC so that its translational radial displacement was almost constant: as a result the RBC rotates as a biconcave disc in a clockwise direction with periodically varying angular velocity. By contrast, for RBCs that interact with neighbouring cells (Fig. 9.14) both translational and rotational motions change dramatically. In this case the rotation of the RBC is no longer regular and periodic but rather in an erratic manner.

Fig. 9.13
figure 000913

Translational and rotational motion of a RBC rolling on the wall without interaction (adapted from [28])

Fig. 9.14
figure 000914

Translational and rotational motion of RBC with interaction (adapted from [28])

9.4.3.4 Effect of Hct on RBC Radial Dispersion

The randomlike transverse motions of RBCs can be analysed by using a radial dispersion coefficient (D yy ), given by:

$$ {D_{{yy}}} = \sum\limits_{{t = 0}}^n {\frac{{\left\langle {{{\left( {{R_y}(t) - {R_y}(0)} \right)}^2}} \right\rangle }}{{2t}}} $$
(9.5)

where R y and t are the radial displacement and time interval respectively. Hence, in vitro blood with several Hcts was used, the confocal micro-PTV system enabling the paths of hundreds of labelled RBCs to be recorded in the centre plane of 100 and 50 μm glass capillaries and 75 μm PDMS circular micro-channels. The results demonstrated that for RBCs the D yy tends to increase with increasing Hct (as shown in Figs. 9.15 and 9.16). For instance Fig. 9.15 shows clearly that RBCs D yy at Hct of 23% is almost one order magnitude bigger than D yy with 3% Hct.

Fig. 9.15
figure 000915

RBCs radial dispersion coefficient (D yy ) of 75 μm PDMS circular micro-channel for 3 % Hct and 23 % Hct (adapted from [29])

Fig. 9.16
figure 000916

Effect of the microtube geometry on the RBC \( D_{{yy}}^c \) (adapted from [29])

These research findings are evidence that the RBCs flowing in a crowded environment tend to undergo multi-body collisions which increase the amplitude of the RBC’s lateral motion and consequently RBC D yy . Hence, RBCs at high concentrations tend to exhibit higher erratic radial displacement compared to dilute suspensions of RBCs. Additionally, results given in Fig. 9.16 demonstrate that RBC radial dispersion, at both moderate and high Hcts, tends to decrease with the diameter. The main reasons for this are as follows: Hct reduction with diameter (Faharaeus effect) and the constrictive effect of the geometry, the latter limiting the amplitude of the radial displacements of the RBCs. Additional detailed results relating to these data can be found elsewhere [26,28,29].

Very recently, in a study performed by Saadatmand et al. [37] fluid particle diffusion was examined in concentrated suspensions of human RBCs (Hct up to 20%) through a capillary tube. The results demonstrated significant enhancement of the particle diffusion, due to a micron-scale flow-field generated by the RBC motions, whereas the dispersion of tracer particles was about 0.7 times that of RBCs. The results also showed that by increasing the flow rate particle dispersion tends to increase almost linearly under constant Hct levels. Moreover, the particle dispersion showed near linear dependency on Hct up to 20%.

9.4.3.5 Effect of Hct on the Cell-free Layer (CFL)

Recent studies have investigated the Hct effect on the cell-free layer (CFL) thickness of both straight circular glass and PDMS micro-channels [8,10,33]. Labelled RBCs flowing near the CFL were tracked manually by using the MtrackJ plugin from Image J. Figure 9.17 gives an example of two trajectories of labelled RBCs flowing at the edge of the CFL. By using MtrackJ, the positions of the tracked RBCs are measured and the average thickness of the CFL is calculated [8,10,30,33].

Fig. 9.17
figure 000917

Trajectory of labelled RBCs flowing nearby the cell-free layer (adapted by [29])

Examination of Fig. 9.18 reveals an overall enhancement of the CFL thickness as Hct decreases. For the case of the PDMS micro-channel, the CFL thickness decreases almost four fold as Hct is increased from 3 to 37% [30,33]. The results are clear evidence that the CFL thickness tends to reduce as the diameter of the micro-channel decreases.

Fig. 9.18
figure 000918

Average thickness of the cell-free layer at different Hcts and diameters [30,33]

9.4.3.6 In Vitro Blood Flow Through Complex Geometries

Past studies on blood flow through glass micro-channels and in micro-vessels have yielded conflicting results with respect to flow resistance and deformability of RBCs [42,48]. However, observed in vivo/in vitro discrepancies have not yet been convincingly explained by the research, which is restricted to straight rigid glass micro-channels. One possible way of understanding observed in vivo/in vitro discrepancies is by using PDMS micro-channels fabricated by a soft lithographic technique [27,34]. By using this kind of technique it is possible to fabricate biomimetic PDMS micro-channels with complex geometries similar to human blood arterioles and capillary networks. In a very recent study performed by Leble et al. [2022] the effects were investigated of bifurcations (diverging and converging) on the flow behaviour of both fluorescent particles suspended in pure water (PW) and RBCs in concentrated suspensions (14% Hct). A very dominant phenomenon was observed in blood flow behaviour at the region of the converging bifurcation apex. At this region, the formation of a clear triangular CFL was observed, the layer tending to propagate along the centre of the micro-channel. Ishikawa et al. [17] and Leble et al. ([2022], 2012) give more detailed information about this phenomenon.

Recently, Fujiwara et al., by using a confocal micro-PTV system have investigated the flow behaviour of RBCs in a micro-channel with a stenosis [11]. The influence of a stenosis (70%) on the trajectories of individual RBCs was investigated in a concentrated suspension of up to 20% Hct. For 10% Hct the results indicated that healthy RBC trajectories have a tendency to become asymmetric before and after the stenosis whereas trajectories of particles suspended in pure water tend to be almost symmetric. Furthermore, the influence of the deformability of RBCs on the CFL thickness was also investigated. Fujiwara et al. have demonstrated that deformability plays a crucial role on the asymmetry of the CFL thickness. Generally, they have concluded that the motions of RBCs are strongly influenced not only by RBC deformability but also by the Hct and channel geometry [11].

9.5 Conclusion and Future Directions

The recent developments in computing, digital image processing techniques and microscopy have made it possible to combine both particle image velocimetry (PIV) and particle tracking velocimetry (PTV) system with confocal microscopes, in so-called confocal micro-PIV/PTV. In this chapter we have presented the most relevant aspects of research using this approach: the theoretical and technical issues for both conventional and confocal micro-PIV/PTV methods and hemodynamic studies in micro-channels are facilitated by using systems based on these methods.

By using a 100-μm square micro-channel, the confocal micro-PIV system has shown good accuracy in measuring blood plasma flow with Hct up to 9%. Nevertheless, for Hct higher than 9%, the light absorbed and scattered by the RBCs contributes to a diminution of the concentration trace particles and as a consequence generates spurious errors in the velocity profiles. This effect becomes much more evident for Hct above 20%. Owing to its optical sectioning ability and consequent improvement of the image contrast and definition, a confocal micro-PTV system was then proposed to track individual blood cells at both low and high Hcts. By using such a system it was indeed found possible to measure cell–cell hydrodynamic interaction, RBC orientation and RBC radial dispersion at different depths and Hcts. Generally, the results suggest that the RBC paths are strongly dependent on the Hct and therefore the RBC dispersion coefficient tends to increase with increase in Hct. In addition, the confocal micro-PTV system has proved to be a powerful tool to obtain further insight into the flow behaviour of blood through complex geometries such as bifurcations, confluences and stenoses. Moreover, by culturing endothelial cells within PDMS microfluidic devices, we expect to develop in the near future a flow system device that closely mimics the in vivo micro-vessel environment.