Abstract
An S-band cavity BPM is designed for a new injector in National Synchrotron Radiation Laboratory. A re-entrant position cavity is tuned to the TM110 mode as position cavity. Cut-through waveguides are used as pickups to suppress the monopole signal. Theoretical resolution of this design is 31 nm. A prototype cavity BPM system is manufactured for off-line cold tests. The wire scanning method is used to calibrate the BPM and estimate the performance of the on-line BPM system. A cross-talk problem has been detected during the cold test. Racetrack cavity BPM design can be used to suppress the cross-talk. With the nonlinear effect being ignored, transform matrix can be used to correct cross-talk. Analysis of cold test results shows that the position resolution of prototype BPM is better than 3 μm.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Winick H. The linac coherent light source (LCLS): A fourth-generation light source using the SLAC linac. J Electron Spectroscopy Related Phenomena, 1995, 75: 1–8
Sargsyan V. Tesla Reports 2004-03, 2004
Li J H, Sun B G, He D H, et al. Research and measurement of BPM system at HLS LINAC. HEP & NP (in Chinese), 2007, 31(01): 96–100
Lorenz R. Cavity beam position monitors. In: Proceedings of 1998 Beam Instrumentation Workshop. SLAC, Stanford, 1998. 53–73
Maesaka H, Otake Y, Togowa K, et al. Beam position monitor at the SCSS prototype accelerator. In: Proceedings of APAC 2007, Indore, 2007. 387–389
Walston S, Boogert S, Chung C, et al. Performance of a high resolution cavity beam position monitor system. Nucl Instrum Methods Phys Res A, 2007, 578: 1–22
Lunin A, Romanov G, Solyak N, et al. Design of a submicron resolution cavity BPM for the ILC main linac. In: Proceedings of the DIPAC’07, Venice, 2007. 192–194
Sargsyan V. Cavity Beam Position Monitor for the TESLA-cryomodule. Cross-Talk minimization. Dissertation for the Doctoral Degree. Berlin: Technische Universität Berlin, 2003
Luo Q, Sun B G, He D H, et al. Beam position monitor with racetrack cavity. High Power Laser Particle Beams, 2010, 22(07): 1635–1639
Li Z. S-band cavity BPM for ILC. In: Proceedings of 2005 ILC Physics and Detector Workshop and 2nd ILC Accelerator Work-shop (Snowmass 2005), 2005. http://www.hep.ucl.ac.uk/lc/T474/Documentation/ILCAW0901_TALK.PDF
Luo Q, Sun B G, He D H. Design of s-band re-entrant cavity BPM. Nucl Sci Tech, 2009, 20(03): 133–139
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Luo, Q., Sun, B., Jia, Q. et al. Design and cold test of S-BAND cavity BPM for HLS. Sci. China Phys. Mech. Astron. 54 (Suppl 2), 292–295 (2011). https://doi.org/10.1007/s11433-011-4555-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11433-011-4555-y