Abstract
The noninvasive transient clearance method provides absolute quantitative measurement of skin blood flow. It is based on thermally insulating the skin under investigation and measuring the time constant of the resultant exponential skin temperature increase. Conventional assessment of the time constant, by measuring the temperature increase until final equilibrium skin temperature is achieved, has the disadvantage of long time of measurement. A procedure shortening the measurement time is presented. The time constant of the exponential temperature increase is calculated without obtaining the final equilibrium temperature. The method can be used for calculating the time constant of an exponential change of any variable, even though the asymptotic value of that variable is not known.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Abbreviations
- F :
-
regional blood flow (ml blood/ml tissue minute)
- t :
-
time (min)
- T :
-
temperature (°C)
- T o :
-
initial temperature
- T b :
-
final temperature
- τ:
-
time constant
- I 12 :
-
integral ofT fromt 1 tot 2
References
Fronek, A. Noninvasive diagnostics in vascular disease. New York: McGraw-Hill Book Co., 1989.
Muller-Schauenburg, W.; Benzing, H.; Betz, E.; Blum, B. (eds.), Measurement of blood flow and local tissue energy production by thermal methods. Proc. Int. Symp. Jerusalem, 1982. XII European Conference on Microcirculation (Stuttgart: George Thieme Verlag).
Nitzan, M.; Anteby, S.O.; Mahler, Y. Transient heat clearance method for regional blood flow measurements. Phys. Med. Biol. 30:557–561; 1985.
Nitzan, M.; Mahler, Y. Theoretical analysis of the transient thermal clearance method for regional blood flow measurement. Med. Biol. Eng. Comp. 24:597–601; 1986.
Nitzan, M.; Mahler, Y.; Roberts, V.C. The transient thermal clearance method for regional blood flow measurement—The influence of tissue heat conduction. Clin. Phys. Physiol. Meas. 9:339–346; 1988.
Nitzan, M.; Fairs, S.L.E.; Roberts, V.C. Simultaneous measurement of skin blood flow by the transient thermal clearance method and laser Doppler flowmetry. Med. Biol. Eng. Comp. 26:407–410; 1988.
Nitzan, M.; Brama, I.; Mahler, Y. Measurement of nasal mucosal blood flow by a thermal clearance method. IEEE Tr. BME. 32:1063–1066; 1985.
Nitzan, M.; Mahler, Y.; Gross, C.; Bar-On, H. The dependence of foot skin blood flow and specific microvascular resistance on systolic blood pressure. Proc. World Congress on Medical Physics and Biomedical Engineering. Kyoto, Japan; 1991: p. 735.
Nitzan, M.; Weshler, Z.; Mahler, Y.; Gross, C.; Gimmon, Z. Quantitative measurement of blood flow in breast undergoing radiotherapy treatment, for the assessment of tissue sensitivity to irradiation. Proc. World Congress on Medical Physics and Biomedical Engineering. Kyoto, Japan; 1991: p. 735.
Weinbaum, S.; Jiji, L. M.; Lemons, P.E.. Theory and experiment for the effect of vascular microstructure of surface heat transfer. I. Anatomical foundation and model conceptualization. J. Biomech. Eng. 106:321–330; 1984.
Mahler, Y.; Lifshitz, N.; Nitzan, M. Microprocessor automatic multiple temperature measurement device. Med. Biol. Eng. Comp. 26:554–556; 1988.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Nitzan, M., Mahler, Y. & Lifshitz, N. Faster procedure for deriving regional blood flow by the noninvasive transient thermal clearance method. Ann Biomed Eng 21, 259–262 (1993). https://doi.org/10.1007/BF02368181
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF02368181