Abstract
A remote-controlled mitral valve was designed and constructed to occlude the mitral or tricuspid orifice at any time in the cardiac cycle. It is used to study ventricular properties in the anesthetized dog by controlling ventricular filling, atrial properties by controlling atrial emptying and interaction of the two chambers by uncoupling them. The device can produce transient or steady-state perturbations in filling volume which make possible studies of intrinsic control of cardiac output. When filling volume is transiently reduced, stroke volume and end-systolic volume are reduced. A compensatory increase in stroke volume of the next cycles occurs due to increased ventricular preload and reduced afterload. The compensation continues until the lost stroke volume is regained, at which time the system returns to its previous steady state.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Gabbay, S., D.M. McQueen, E.L. Yellin, R.M. Becker and R.W.M. Frater. In vitro hydrodynamic comparison of mitral valve protheses at high flow rates.J. Thorac. Cardiovasc. Surg. 76:771–787, 1978.
Hori, M., E.L. Yellin and E.H. Sonnenblick. Left ventricular diastolic suction as a mechanism of ventricular filling.Jpn. Circ. J. 46:124–129, 1982.
Laniado, S., E.L. Yellin, H. Miller and R.W.M. Frater. Temporal relation of the first heart sound to closure of the mitral valve.Circulation 47:1006–1014, 1973.
Meisner, J.S., S. Nikolic, K.K. Nelson, C. Picker, R.W.M. Frater and E.L. Yellin. Remote-controlled mitral valve (RCMV): A new research tool (Abstract)Proc. 38th ACEMB, 27:333, 1985.
Starling, E.H.Linacre Lecture on the Law of the Heart. London: Longmans, Green and Co., 1918.
Suga, H. and K. Sagawa. End-diastolic and end-systolic ventricular volume clamper for isolated canine heart.Am. J. Physiol. 233:H718-H722, 1977.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Meisner, J.S., Nikolić, S., Tamura, T. et al. Development and use of a remote-controlled mitral valve. Ann Biomed Eng 14, 339–349 (1986). https://doi.org/10.1007/BF02367407
Issue Date: