Abstract
A new dye laser spectrometer utilizing a non-tunable reference resonator is described. The resonator consists of two Zerodur mirrors optically contacted to a Zerodur spacer. Frequency scanning of the laser is provided by acoustooptic modulation. Residual drifts of the resonator frequency — measured on line — are compensated automatically by corresponding corrections of the modulation frequency. The stability during several hours and the resettability of the dye laser frequency are±2.5 kHz and±10 kHz, respectively.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
J. Hough, D. Hils, M.D. Rayman, Ma L.-S., L. Hollberg, J.L. Hall: Appl. Phys. B33, 179 (1984)
U. Grzesik: Dissertation, University of München (1985)
R.V. Pound: Rev. Sci. Instrum.17, 490 (1946)
R.W.P. Drever, J.L. Hall, F.V. Kowalski, J. Hough, G.M. Ford, A.J. Munley, H. Ward: Appl. Phys. B31, 97 (1983)
G.C. Bjorklund: Opt. Lett.5, 15 (1980)
J. Helmcke, S.A. Lee, J.L. Hall: Appl. Opt.21, 1686 (1982)
Documents Concerning the New Definition of the Metre: Metrologia19, 163 (1984)
J.-M. Chartier, J. Helmcke, A.J. Wallard: IEEE Trans. IM-25, 450 (1976)
B. Burghardt, W. Jitschin, G. Meisel: Appl. Phys20, 141–146 (1979)
Mention of this product is for technical orientation only and does not constitute and endorsement. Furhtermore, it is not implied that other products are necessarily less well-suited to our applications
F. Bayer-Helms, H. Darnedde, G. Exner: Metrologia21, 49 (1985)
J.A. Barnes, et al.: IEEE Trans. IM-20, 105–120 (1971)
J.C. Bergquist, R.L. Barger, D.J. Glaze: InLaser Spectroscopy IV, ed. by H. Walther, K.W. Rothe. Springer Ser. Opt. Sci.21 (Springer, Berlin, Heidelberg 1979) p. 120
J. Helmcke, D. Zevgolis, B.Ü. Yen: Appl. Phys. B28, 83 (1982)
J.L. Hall, S.A. Lee: Appl. Phys. Lett.29, 367 (1976)
A. Morinaga, J. Helmcke: To be published
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Helmcke, J., Snyder, J.J., Morinaga, A. et al. New ultra-high resolution dye laser spectrometer utilizing a non-tunable reference resonator. Appl. Phys. B 43, 85–91 (1987). https://doi.org/10.1007/BF00692821