The irradiation dose to the lens of the eye during CT scanning has been measured in 16 patients. The dose is greatly influenced by changes in gantry angulation, the position of the patient, the thickness and number of slices, KV, and milliamperage. The total lens dose has been measured to be higher than previously reported. Repeated CT scans imply the risk of irradiation cataract, especially in infants and children.
Article PDF
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
References
Bergström K (1975) Computer tomography of the orbits. Acta Radiol [Suppl] 346:155–160
Bergström K, Dahlin H, Gustafsson M, Nylén O (1972) Eye lens doses in carotid angiography. Acta Radiol (Diagn) 12: 134–140
Merriam GR, Focht EF (1957) A clinical study of radiation cataracts and the relationship to dose. AJR 77:759–785
Charles MW, Lindop PJ (1980) Skin and eye irradiations. Examples of some limitations of international recommendations in radiological protection. IAEA-SR-36/6:547–561
Recommendations of the International Commission on Radiological Protection (1977) ICRP 26:12–13; 21
Schmidt T, Stieve FE (1980) Radiation exposure of infants and children in computed tomography. Ann Radiol 23:143–149
Bhave G, Kelsey CA, Burstein J, Brogdon BG (1977) Scattered radiation doses to infants and children during EMI head scans. Radiology 124:379–380
Brasch RC, Boyd DP, Gooding CA (1978) Computed tomographic scanning in children; comparison of radiation dose and resolving power of commercial CT scanners. AJR 131: 95–101
Brasch RC, Korobkin M, Gooding CA (1978) Computed body tomography in children: evaluation of 45 patients. AJR 131: 21–25
Erasmie U, Bergström M (1979) Computed tomography of the head in children. Acta Radiol (Diagn) 20:282–288
Horsley RJ, Peters VG (1976) Radiation exposure from EMI scanners and multiple scans. Br J Radiol 49:810–811
Isherwood I, Young IM, Bowker KW, Bramall GK (1975) Radiation dose to the eyes of the patient during neuroradiological investigations. Neuroradiology 10:137–141
McCullogh EC, Payne JT (1978) Patient dosage in computed tomography. Radiology 129:457–463
Wall BF, Green DAC, Veerappan R (1979) Radiation dose to patients from EMI brain and body scanners. Br J Radiol 52: 189–196
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Lund, E., Halaburt, H. Irradiation dose to the lens of the eye during CT of the head. Neuroradiology 22, 181–184 (1982). https://doi.org/10.1007/BF00341246
Received:
Revised:
Issue Date:
DOI: https://doi.org/10.1007/BF00341246