Abstract
Background
We have shown in acute lung injury patients that lung elastance can be determined by a positive end-expiratory pressure (PEEP) step procedure and proposed that this is explained by the spring-out force of the rib cage off-loading the chest wall from the lung at end-expiration. The aim of this study was to investigate the effect of the expanding chest wall on pleural pressure during PEEP inflation by building a model with an elastic recoiling lung and an expanding chest wall complex.
Methods
Test lungs with a compliance of 19, 38, or 57 ml/cmH2O were placed in a box connected to a plastic container, 3/4 filled with water, connected to a water sack of 10 l, representing the abdomen. The space above the water surface and in the lung box constituted the pleural space. The contra-directional forces of the recoiling lung and the expanding chest wall were obtained by evacuating the pleural space to a negative pressure of 5 cmH2O. Chest wall elastance was increased by strapping the plastic container. Pressure was measured in the airway and pleura. Changes in end-expiratory lung volume (ΔEELV), during PEEP steps of 4, 8, and 12 cmH2O, were determined in the isolated lung, where airway equals transpulmonary pressure and in the complete model as the cumulative inspiratory-expiratory tidal volume difference. Transpulmonary pressure was calculated as airway minus pleural pressure.
Results
Lung pressure/volume curves of an isolated lung coincided with lung P/V curves in the complete model irrespective of chest wall stiffness. ΔEELV was equal to the size of the PEEP step divided by lung elastance (EL), ΔEELV = ΔPEEP/EL. The end-expiratory “pleural” pressure did not increase after PEEP inflation, and consequently, transpulmonary pressure increased as much as PEEP was increased.
Conclusions
The rib cage spring-out force causes off-loading of the chest wall from the lung and maintains a negative end-expiratory “pleural” pressure after PEEP inflation. The behavior of the respiratory system model confirms that lung elastance can be determined by a simple PEEP step without using esophageal pressure measurements.
Similar content being viewed by others
Background
Transpulmonary driving pressure is a key factor in ventilator-induced lung injury and also for rational setting of positive end-expiratory pressure (PEEP) [1–3]. The transpulmonary pressure is traditionally calculated as the product of end-inspiratory airway pressure and the ratio of lung to respiratory system elastance [4–6]. The ratio of lung and respiratory system elastance is in turn estimated from the change in airway and esophageal pressure during tidal volume inflation. Unfortunately, due to difficulties in positioning of the catheter as well interpretation of data, esophageal pressure measurements at the bedside is still rarely performed [7].
We have recently shown that it is possible to determine lung elastance without using esophageal pressure measurements using a simple PEEP step procedure [8, 9]. We have proposed that this phenomenon is explained by the spring-out force of the rib cage off-loading the chest wall from the lung at end-expiration. The consequence is that the end-expiratory pleural pressure does not change when end-expiratory lung volume is increased by raising PEEP. Consequently, end-expiratory transpulmonary pressure increases as much as PEEP is increased. This has previously been described by Rahn and coworkers 1946, where the chest wall pressure/volume curve was determined during end-expiratory relaxation at different lung volumes [10] without using esophageal or pleural pressure measurements. The fundamental effects of the expansive chest wall on lung mechanics have been discussed [11], but skepticism concerning the clinical effects seems to prevail.
The aim of the present study was to examine the hypothesis that the expansive chest wall is the cause of the unchanged pleural pressure after PEEP inflation in contrast to the changes in pleural pressure during tidal inflation. To study this, a respiratory system model based on Newtonian laws of mechanics [12], with an elastic recoiling lung, an expanding chest wall, and a “pleural” space in between, was designed and studied during tidal and PEEP inflation.
Methods
The respiratory system model
A model including the contra-directional forces of the recoiling lung and the spring-out force of the rib cage and the abdomen, which can be regarded as a mainly fluid filled container being attached to the caudal rim of the rib cage [10, 13, 14] was designed (see Fig. 1). In this model, the spring-out force of the rib cage created a negative pressure in the pleural space. The abdominal content is displaced caudally during inflation, rather than being inflated like the lung [15]. Test lungs (Maquet CC, Solna, Sweden) with a nominal compliance of 19 ml/cmH2O were used as the elastic recoiling lung in the model. The elastic recoil of the test lungs used is not achieved by using an elastic balloon but rather by having a fairly non-elastic balloon placed between two elastic Teflon plates.
One, two, or three test lungs were connected in parallel obtaining a nominal lung compliance of 19, 38, or 57 ml/cmH2O. The test lungs were placed in an airtight box, which was connected with a 22 mm diameter tubing to the upper part of a semi-stiff plastic container, which was three quarter filled with water.
In a patient less than half of a volume increase, tidal or PEEP inflation is directed towards the diaphragmatic area and thereby affecting the abdomen. The rest of the volume increase occurs towards the chest wall. The displacement of the diaphragm in caudal direction causes a slight rise of the upper surface of the abdomen. This rise is dependent on the size of the upper surface area of the abdomen. In the model, all volume increase is directed towards the “diaphragm” and the abdominal container as it was not possible to build a separate chest wall and diaphragm. To compensate for this, the diaphragmatic surface area was made 1500 cm2, twice the size in an adult person. A water-tight sack with good plasticity and a volume of 10 l, equivalent to the volume of the abdominal cavity and an upper surface area of 1800 cm2, double the upper surface of the abdomen of an adult person [16, 17], was connected to the bottom of the plastic container by a 22-mm diameter tubing with a valve with adjustable resistance. The space above the water surface in the plastic container, the tubing, and the box in which the test lungs were placed constituted the pleural space.
Chest wall elastance could be increased by strapping hard board on the outside of the pleural space, i.e., the plastic container and the lung box.
The contra-directional forces of the recoiling lung and the expanding chest wall were obtained by evacuating the pleural space through suctioning to a pressure of −5 cmH2O, which caused a higher water level in the plastic container than in the abdominal sack. As a result of the negative pleural pressure, the lung was “inflated” to functional residual capacity (FRC), where the end-expiratory transpulmonary pressure is equal to the end-expiratory airway pressure minus the pleural pressure, 0 − −5 = 5 cmH2O (Fig. 1).
Measurements
Airway pressure was measured by the Flow-I anesthesia machine (Maquet Critical Care, Solna, Sweden) and presented by the Maquet software. End-expiratory pressure was used for calculations of lung elastance and transpulmonary pressures. Pleural pressure was measured in the pleural space by a standard pressure receptor (PVB Medizintechnik, Kirchseeon, Germany) and monitor (S/5, GE Healthcare, Helsinki, Finland). For volume measurements, the pneumotachograph of the ventilator of the Flow-I anesthesia machine (Maquet Critical Care, Solna, Sweden) was used. The change in end-expiratory lung volume (ΔEELV) following an increase/decrease in PEEP was determined as the cumulative difference in inspiratory and expiratory volume during the first 15 breaths following a PEEP change using a dedicated software provided by Maquet CC [18]. The software collects data from the Servo-I or the Flow-I about pressures and flow at a frequency of 100 Hz. The software analyzes the pressure signal and provides the actual measured level of PEEP and senses for changes. If there is a change of PEEP any contribution to the ΔEELV are summed up during 15 breaths. The inspiratory (VTi) and expiratory (VTe) tidal volumes are calculated by the summation of flow samples from the inspiratory and expiratory phase, respectively. The contribution to the ΔEELV (VTi-VTe) after a change of PEEP are compensated for any offset, and the eventual offset are decided during the five breaths following the 15 included in the summation at the new PEEP level.
Calculations
Tidal total respiratory system, chest wall, and lung elastance were calculated as tidal airway pressure variation, tidal pleural pressure variation, and the difference between tidal airway and pleural pressure variation divided by tidal volume (ΔPAW/VT, ΔPPL/VT, (ΔPAW − ΔPPL)/VT), respectively.
Lung elastance was also calculated as the change in end-expiratory airway pressure divided by the change in end-expiratory lung volume (ΔPEEP/ΔEELV) induced by an increase or a decrease in PEEP.
Experimental procedure
The respiratory system model was ventilated by a Flow-I anesthesia machine (Maquet CC, Solna, Sweden) in volume control ventilation with a tidal volume of 300 ml, respiratory rate of 15/min, I:E ratio of 1:2, and an end-inspiratory pause of 10 % of the breathing cycle.
Isolated lung
FRC was determined in isolated lung by measuring change in lung volume (ΔEELV) induced by a PEEP step from zero to 5 cmH2O. Lung compliance was determined by ventilation of 1, 2, and 3 test lungs disconnected from the plastic container and “abdominal” sack, i.e., with isolated lung, where airway pressure equals transpulmonary pressure. As the transpulmonary pressure at FRC was ≈5 cmH2O, PEEP steps 5-9-5, 5-13-5, and 5-17-5 cmH2O were performed in the isolated lung, corresponding to 0-4-0, 0-8-0, and 0-12-0 cmH2O with connected pleural space and abdomen. The PEEP steps were performed with one, two, and three test lungs connected, except when one test lung was used, where the 5-17-5 cmH2O steps were omitted as 17 cmH2O in PEEP resulted in end-inspiratory volume above the capacity of a single test lung.
Lung(s) together with a normal or stiff chest wall
PEEP steps 0-4-0, 0-8-0, and 0-12-0 cmH2O were performed with one, two, or three test lungs connected. When one test lung was used, the 0-12-0 cmH2O step was omitted as 12 cmH2O in PEEP resulted in end-inspiratory volume above the capacity of a single test lung. All PEEP step sequences were performed with “normal” chest wall and with stiff chest wall.
Data analysis and statistics
Each PEEP step procedures were run three times, and the average of these runs was used as the result for a PEEP step. As the “pleural” pressure might not be exactly the same after changing between one, two, and three lungs and stiff and normal chest wall, SD is reported as the variation in results when mean results are reported from the totally six procedures: one lung stiff chest wall (CW), one lung normal CW, two lungs stiff CW, two lungs normal chest wall, three lungs stiff CW, and three lungs normal chest wall. Correlation between measured ΔEELV using spirometry and ΔEELV calculated as change in PEEP divided by lung elastance (determined as the difference in airway and pleural driving pressure divided by tidal volume) was analyzed using linear regression analysis.
Results
In figures, all curves are starting from FRC but for clarity depicted as starting from zero volume. Pressure/volume curves from ventilation of an isolated “lung,” a lung with intact chest wall ventilated at different PEEP levels as well as lungs with a stiff chest wall are shown in Fig. 2. Note that the isolated lung P/V curve without chest wall, where airway pressure is equal to transpulmonary pressure, coincides with the end-expiratory airway P/V points irrespective of the stiffness of chest wall. The airway pressure/volume curves at increasing PEEP levels were seemingly successively left-shifted irrespective of normal or stiff chest wall in the same way as seen in patients [19–25] (Fig. 2, mid panels). The end-inspiratory airway P/V points are right shifted from the lung (isolated) P/V curve as a consequence of the influence of the chest wall [26] (Fig. 2).
Figure 3 shows the changes in airway pressure, transpulmonary pressure, and pleural pressure and lung volume in the lung model. After increasing PEEP, there was an initial increase in end-expiratory pleural pressure. From this, still negative end-expiratory pleural pressure level, the pressure subsided back towards the baseline negative value of −5 cmH2O. Simultaneously, the end-expiratory lung volume measured with spirometry continued to increase with a concomitant increase in end-expiratory transpulmonary pressure (Fig. 3).
The increase in end-expiratory transpulmonary pressure when increasing PEEP from zero end-expiratory pressure (ZEEP) to 4.4 ± 0.1, 8.4 ± 0.2, and 12.2 ± 0.1 cmH2O was 4.2 ± 0.1, 8.0 ± 0.2, and 11.4 ± 0.2 cmH2O, respectively with normal and stiff chest wall when the model was used with one, two, or three test lungs in parallel with a nominal lung compliance of 19, 38, and 57 ml/cmH2O. A step change of PEEP of 0-4-8-12 cmH2O, where each step was 4.1 ± 0.3 cmH2O, resulted in a step change in transpulmonary pressure of 3.9 ± 0.3 cmH2O. The corresponding increase in end-expiratory pleural pressure was 0.2 ± 0.1 cmH2O.
The change in end-expiratory lung volume (ΔEELV) of all PEEP steps with one, two, and three lungs and normal and stiff chest wall measured with the ventilator pneumotachograph and ΔEELV calculated as the change in PEEP divided by the lung elastance, ΔPEEP/EL, where lung elastance was determined as the difference in airway and pleural driving pressure divided by the tidal volume (ΔPAW – ΔPPL)/VT, were closely correlated, Y = 1.05X, r 2 = 0.99.
In the isolated lung without a chest wall, a baseline PEEP of 5 cmH2O was applied to achieve a volume corresponding to lung volume (FRC) at a pleural pressure of −5 cmH2O in the complete model with lung pleura and abdominal compartment. A PEEP of 5 cmH2O in the isolated lung resulted in a FRC of 92, 167, and 242 ml with one, two, or three test lungs with a nominal lung compliance of 19, 38, and 57 ml/cmH2O respectively. Lung P/V curves obtained during ventilation of an isolated lung with one, two, or three test lungs were similar to the corresponding lung P/V curves when the chest wall complex was connected, irrespective of if the chest wall was normal or stiff (Fig. 4).
Tidal airway P/V curves at increasing PEEP levels are seemingly successively left-shifted, but in reality, it is the end-inspiratory airway P/V points that are right shifted from the lung P/V curve (Figs. 5 and 6).
Tidal chest wall elastance of the model was 18 cmH2O/L with normal chest wall stiffness and 24 cmH2O/L with increased chest wall stiffness, irrespective of PEEP level (Fig. 7).
End-expiratory chest wall elastance, between 0 and 12 cmH2O in PEEP, calculated as the change in end-expiratory pleural pressure divided by the change in end-expiratory lung volume, ΔPPLEE/ΔEELV, was 1.5 cmH2O/L, corresponding to an end-expiratory chest wall compliance of 637 ± 72 ml/cmH2O, irrespective of tidal chest wall elastance.
Discussion
We have constructed a model of the respiratory system with a recoiling lung and an expanding chest wall, where tidal airway (total respiratory system) pressure/volume curves are seemingly left-shifted when increasing PEEP in accordance with what is seen in patients [21, 22, 24, 25, 27]. A change in end-expiratory airway pressure, PEEP with an ensuing inflation of the lung, causes a negligible increase in end-expiratory “pleural” pressure, indicating that end-expiratory elastance of the chest wall is close to zero, i.e., the end-expiratory chest wall compliance is extremely high. As a consequence, the end-expiratory transpulmonary pressure increases as much as PEEP is increased. The increase in end-expiratory lung volume following a PEEP step is determined by the size of the change in PEEP and lung elastance. If the change in volume is measured, lung elastance can be determined as the change in PEEP divided by the change in end-expiratory lung volume, ΔPEEP/ΔEELV. Thus, we do not need to know pleural or esophageal pressure to estimate lung elastance and changes in transpulmonary pressure.
Tidal and end-expiratory chest wall elastance
The expanding chest wall has a fundamental physiological role in counteracting the recoil of the lung at end-expiration, i.e., functional residual capacity (FRC). This has previously been described in a number of classical studies and in textbooks of physiology [10, 13, 28–31] but have only recently been introduced in studies of mechanical ventilation [8, 9, 11].
The outward directed force of the chest wall has a pulling effect on the lung, which in itself strives to recoil to a lower volume, the minimal volume, slightly below the residual volume. The net effect of these contra-directional forces is that a negative pleural pressure and an equally large positive transpulmonary pressure are created, 5–10 cmH2O [29–32]. Thus, instead of compressing, squeezing, or leaning at the lung at end-expiration, the chest wall tends to expand the lung at end-expiration as long as the end-expiratory lung volume, irrespective of the reason for the increased volume, emphysema or PEEP inflation, is below the resting volume of the chest wall at 70–90 % of total lung capacity (TLC) [10, 30, 31, 33–36]. The resilience of the chest wall originates in the ribs, ligaments, and cartilages of the rib cage, which are unaffected in severe acute lung injury (ARDS) [11]. This means that the resting volume of the chest wall, i.e., the volume where the chest wall does not strive outwards any more, is around 3 l above FRC, also in respiratory failure patients. As the mean pleural pressure at FRC is around −5 and 0 cmH2O at the chest wall resting volume, the end-expiratory chest wall compliance is 3000/5 ≈ 600 ml/cmH2O estimated from the chest wall relaxation P/V curve as described by Rahn and coworkers in 1946 [10]. In emphysema patients, who were subjected to lung reduction surgery, end-expiratory chest wall compliance is estimated to be 700 ml/cmH2O [37]. In the present lung/chest wall/abdomen model, the end-expiratory chest wall compliance was also very high, i.e., around 600 ml/cmH2O, confirming that the geometry of the model reflects chest wall mechanics in humans. Between FRC and the resting volume of the chest wall pleural pressure is negative, and the chest wall strives outwards at end-expiration, irrespective of the pressure inside the lung, of which there are no receptors in the chest wall [10, 31, 37, 38]. This indicates that the chest wall has a negligible influence on the lung during end-expiration as a result of the rib cage spring-out force.
Tidal pleural pressure variations, the chest wall-driving pressure, reflect the force needed to displace a weight, the chest wall, and the abdominal content during inspiration, rather than inflating a recoiling structure. As the weight of the chest wall and the abdomen does not change with changing PEEP, tidal chest wall elastance is almost constant during increasing PEEP levels in the respiratory system model, as also reported for patients in several studies [8, 9, 25, 39, 40] (Fig. 7). During an inspiration from ZEEP/FRC pressure/volume equilibrium, pleural pressure increases from a negative pressure level to a less negative pressure level, or if the chest wall is very stiff, to a positive end-inspiratory pleural pressure (Figs. 3 and 7). But, during the expiration, the lung will recoil back to the baseline, and from a volume point of view, the interior volume of the chest wall is always essentially equal to the lung exterior volume; the lung will pull the chest wall back to baseline volume. In doing so, pleural pressure will return to a negative level because of the rib cage spring-out force opposing the recoil of the lung.
When PEEP is stepwise increased, the ventilator expiratory valve is closing at the selected, new PEEP level during the first expiration after increasing PEEP, and a volume corresponding to total respiratory system compliance times the change in PEEP is retained in the lung (Fig. 3). As long as this first expiration lung volume increase does not cause the end-expiratory lung volume to exceed the chest wall resting volume at 70–80 % of TLC, the end-expiratory pleural pressure will be negative when the expiratory valve closes (Figs. 3 and 7), indicating that the rib cage spring-out force is active. During the ensuing breaths, the end-expiratory lung volume increases breath by breath and the end-expiratory pleural pressure subsides back to or close to the baseline negative pleural pressure level and a new P/V equilibrium is reached. PEEP inflation of the lung is a pressure control inflation of the lung with constant airway pressure, and end-expiratory airway pressure is the sum of end-expiratory transpulmonary and end-expiratory pleural pressure. As end-expiratory transpulmonary pressure increases breath by breath, when the end-expiratory lung volume increases breath by breath, end-expiratory pleural pressure must decrease after the initial first expiration increase, as much as the transpulmonary pressure increases (Fig. 3). If the chest wall behaved as an elastic entity recoiling to a lower volume like the lung, end-expiratory chest wall elastance, ΔPPLEE/ΔEELV, would be equal to tidal chest wall elastance, ΔPPL/VT, like end-expiratory lung elastance, ΔPTPEE/ΔEELV, is equal to tidal lung elastance ΔPTP/VT. Consequently, total end-expiratory elastance would be the sum of tidal chest wall and tidal lung elastance, total respiratory system elastance (ETOT) = ΔPPL/VT + ΔPTP/VT and the change in end-expiratory lung volume change following a PEEP increase could not exceed the change in PEEP (=ΔPAWEE) divided by the total respiratory system elastance, ΔPEEP/ETOT. However, such a low change in end-expiratory lung volume is only present in isolated lung, where total respiratory system elastance is the same as lung elastance, as seen when ventilating isolated test lungs in the model (Fig. 2). In patients, the change in end-expiratory lung volume is always larger than ΔPEEP/ETOT, except in isolated lung, as reported in multiple studies [20, 21, 22, 27] and determined by the size of the PEEP step and lung elastance, ΔEELV = P ΔPEEP/EL, as seen in the model and in accordance with findings in previous studies in pigs and acute lung injury (ALI) patients [8, 9].
Transpulmonary pressure
The consequence of the mean end-expiratory pleural pressure remaining at a baseline negative level when PEEP is increased is that the end-expiratory transpulmonary pressure increases as much as PEEP is increased. This is in accordance with findings in a porcine and in an ALI patient study, where the transpulmonary pressure variations of a tidal volume equal to the change in end-expiratory lung volume was closely related to the change in PEEP [8, 9]. As the transpulmonary pressure increases as much as the end-expiratory airway pressure changes, the static end-expiratory and end-inspiratory transpulmonary pressure at a certain lung volume level is independent of whether this volume is reached by tidal or PEEP inflation (Fig. 8). Consequently, end-expiratory and end-inspiratory transpulmonary pressure/volume points are aligned on a common, single transpulmonary (lung) P/V curve, as seen in the model, irrespective of whether one, two, or three test lungs are used, or the chest wall is normal or stiff (Figs. 2, 4, 5, 6, and 7).
An indication that the model is physiologically adequate would be if it could be shown, also in patients, that the increase in end-expiratory transpulmonary pressure following a PEEP increase is equal to the ΔPEEP. To confirm this, we calculated the increase in end-expiratory transpulmonary pressure of an end-expiratory lung volume increase following a PEEP increase as ΔEELV × EL from the reported mean values in 8 patients representative of patients with healthy lungs, 8 patients representative of moderate acute lung injury (ALI), and 8 patients representative of severe acute lung injury (ARDS) of a study by Pesenti and coworkers [39], and in 9 patients with pulmonary and 12 patients with extrapulmonary ARDS from a study by Gattinoni and coworkers [25]. In these representative “patients,” PEEP was increased from ZEEP to 5, 10, and 15 cmH2O in random order and the increase in lung volume was determined during a prolonged expiration to ZEEP from the different PEEP levels. In addition, the same calculation was performed in a group of 26 patients, representative of patients with mixed ALI, during incremental PEEP steps of 5 cmH2O, from 5 to 40 and back to 5 cmH2O [41]. In all three studies, lung elastance was determined as the tidal difference in airway pressure minus the tidal difference in esophageal pressure divided by the tidal volume, EL = (ΔPAWLP − ΔPPLLP)/VT, avoiding uncertainties of the validity of absolute end-expiratory esophageal pressure in relation to absolute pleural pressure [6, 15, 42, 43, 44]. Twenty-eight PEEP steps in six mean “patients” representative of the most extreme lung conditions, from healthy lungs to the most severe pulmonary and extrapulmonary ARDS, were included and showed that the transpulmonary pressure increased 5.1 ± 0.7 cmH2O, when the PEEP was increased by 5 cmH2O and that end-expiratory pleural pressure, as a consequence did not increase, −0.1 ± 0.7 cmH2O. This is a strong indication that the model performs in accordance with human physiology as the increase in end-expiratory “pleural” pressure in the model also was negligible, 0.2 ± 0.1 cmH2O.
Clinical implications
At end-expiration at any PEEP/EELV level below the resting volume of the chest wall, which is more than 3 l above FRC [30, 33], the spring-out force of the resilient rib cage makes the chest wall strive to a higher volume, while the end-expiratory airway pressure keeps the lung distended. The end-expiratory transpulmonary pressure and the end-expiratory pleural pressure at supine at FRC is zero at the most dorsal region of the open lung due to the gravitational pleural pressure gradient [11]. Thus, at end-expiratory pressure/volume equilibrium at increased PEEP, the end-expiratory transpulmonary pressure of most dorsal region of the open lung is equal to the end-expiratory airway pressure, as the chest wall is “lifted” from the lung, i.e., the chest wall does not squeeze or lean on the lung at end-expiration. It has been shown in a porcine study that the static transpulmonary pressure, PTPEE, is less damaging than the dynamic, transpulmonary driving pressure during tidal inspiration, ΔPTP [2, 45]. In humans, the key factor in ventilator-induced lung injury is the airway driving pressure [1], but in an accompanying editorial, it was pointed out that the culprit was not the total respiratory system (airway) but rather the transpulmonary driving pressure [3]. As the transpulmonary (lung) driving pressure can constitute anything between 50 % in mainly extrapulmonary ARDS and 90 % of the total respiratory system (airway) driving pressure in mainly direct, pulmonary ARDS, it is of great importance to be able to determine the lung driving pressure, both to know when there is a possibility to use higher airway driving pressure without harming the lung and when the tidal volume should be reduced to prevent harmful lung driving pressures. The present model shows that the ΔPTP can be calculated by a simple PEEP step procedure, where the change in end-expiratory lung volume is determined following a PEEP change. Then, lung elastance for the lung volume range between the two PEEP levels can be calculated as the change in PEEP divided by the change in end-expiratory lung volume, EL = ΔPEEP/ΔEELV. Tidal chest wall elastance can be determined as the difference between total respiratory system elastance (ΔPAW/VT) and lung elastance, chest wall elastance (ECW) = ETOT − ΔPEEP/ΔEELV. As tidal chest wall elastance is essentially constant when increasing PEEP, transpulmonary driving pressure at any PEEP level can be calculated as total respiratory system driving pressure minus tidal chest wall elastance times the tidal volume, ΔPAW − ECW × VT.
Conclusions
Tidal chest wall elastance is related to the force needed to displace the abdominal content (weight) during inspirations. In contrast, end-expiratory chest wall elastance is related to the spring-out force of the rib cage, which maintains end-expiratory pleural pressure negative when increasing PEEP and is unaffected by conditions of the lung and the abdomen and remains mainly unchanged and close to zero in acute respiratory failure. Inspiratory and expiratory chest wall elastance are thus totally different entities, where inspiratory chest wall elastance is a measure of how much the chest wall and abdomen impedes lung inflation, while expiratory chest wall elastance indicates that the chest wall is off-loading the lung and consequently has no constraining effect on the lung at end-expiration.
As a consequence, we do not need to measure pleural or esophageal pressures to estimate lung elastance and changes in transpulmonary pressure.
Abbreviations
- ALI:
-
Acute lung injury
- ARDS:
-
Severe acute lung injury
- CW:
-
Chest wall
- ECW:
-
Chest wall elastance
- EE:
-
End-expiratory
- EELV:
-
End-expiratory lung volume
- EL:
-
Lung elastance
- ETOT:
-
Total respiratory system elastance
- FRC:
-
Functional residual capacity
- LP:
-
Low PEEP
- LV:
-
Lung volume
- P/V:
-
Pressure/volume
- PAW:
-
Airway pressure
- PEEP:
-
Positive end-expiratory pressure
- PPL:
-
Pleural pressure
- PPLEE:
-
End-expiratory pleural pressure
- PTP:
-
Transpulmonary pressure
- PTPEE:
-
End-expiratory transpulmonary pressure = static transpulmonary pressure
- TLC:
-
Total lung capacity
- VT:
-
Tidal volume
- ZEEP:
-
Zero end-expiratory pressure
- Δ…:
-
Delta… = change in…
References
Amato MB, Meade MO, Slutsky AS, Brochard L, Costa EL et al (2015) Driving pressure and survival in the acute respiratory distress syndrome. N Engl J Med 372:747–755
Protti A, Andreis DT, Monti M, Santini A, Sparacino CC et al (2013) Lung stress and strain during mechanical ventilation: any difference between statics and dynamics? Crit Care Med 41:1046–1055
Loring SH, Malhotra A (2015) Driving pressure and respiratory mechanics in ARDS. N Engl J Med 372:776–777
Grasso S, Terragni P, Birocco A, Urbino R, Del Sorbo L et al (2012) ECMO criteria for influenza A (H1N1)-associated ARDS: role of transpulmonary pressure. Intensive Care Med 38:395–403
Staffieri F, Stripoli T, De Monte V, Crovace A, Sacchi M et al (2012) Physiological effects of an open lung ventilatory strategy titrated on elastance-derived end-inspiratory transpulmonary pressure: study in a pig model*. Crit Care Med 40:2124–2131
Gattinoni L, Chiumello D, Carlesso E, Valenza F (2004) Bench-to-bedside review: chest wall elastance in acute lung injury/acute respiratory distress syndrome patients. Crit Care 8:350–355
Akoumianaki E, Maggiore SM, Valenza F, Bellani G, Jubran A, et al. (2014) The application of esophageal pressure measurement in patients with respiratory failure. Am J Respir Crit Care Med 189:520–531
Lundin S, Grivans C, Stenqvist O (2015) Transpulmonary pressure and lung elastance can be estimated by a PEEP-step manoeuvre. Acta Anaesthesiol Scand 59:185–196
Stenqvist O, Grivans C, Andersson B, Lundin S (2012) Lung elastance and transpulmonary pressure can be determined without using oesophageal pressure measurements. Acta Anaesthesiol Scand 56:738–747
Rahn H, Otis AB et al (1946) The pressure-volume diagram of the thorax and lung. Am J Physiol 146:161–178
Stenqvist O, Gattinoni L, Hedenstierna G (2015) What’s new in respiratory physiology? The expanding chest wall revisited! Intensive Care Med 41:1110–1113
Newton I (1687, 1729) The mathematical principles of natural philosophy.
Agostini E, Hyatt R (1986) The respiratory system. Mechanics of breathing. Handbook of physiology. American Physiological Society, Bethesda, pp 113–130
Froese AB (1989) Role of lung volume in lung injury: HFO in the atelectasis-prone lung. Acta Anaesthesiol Scand Suppl 90:126–130
Hedenstierna G (2012) Esophageal pressure: benefit and limitations. Minerva Anestesiol 78:959–966
Hedenstierna G, Strandberg A, Brismar B, Lundquist H, Svensson L et al (1985) Functional residual capacity, thoracoabdominal dimensions, and central blood volume during general anesthesia with muscle paralysis and mechanical ventilation. Anesthesiology 62:247–254
Grimby G, Hedenstierna G, Lofstrom B (1975) Chest wall mechanics during artificial ventilation. J Appl Physiol 38:576–580
Grivans C, Lundin S, Stenqvist O, Lindgren S (2011) Positive end-expiratory pressure-induced changes in end-expiratory lung volume measured by spirometry and electric impedance tomography. Acta Anaesthesiol Scand 55:1068–1077
Nunes S, Uusaro A, Takala J (2004) Pressure-volume relationships in acute lung injury: methodological and clinical implications. Acta Anaesthesiol Scand 48:278–286
Valta P, Takala J, Eissa NT, Milic-Emili J (1992) Effects of PEEP on respiratory mechanics after open heart surgery. Chest 102:227–233
Lu Q, Constantin JM, Nieszkowska A, Elman M, Vieira S et al (2006) Measurement of alveolar derecruitment in patients with acute lung injury: computerized tomography versus pressure-volume curve. Crit Care 10:R95
Maggiore SM, Richard JC, Brochard L (2003) What has been learnt from P/V curves in patients with acute lung injury/acute respiratory distress syndrome. Eur Respir J Suppl 42:22s–26s
Richard JC, Maggiore SM, Mercat A (2004) Clinical review: bedside assessment of alveolar recruitment. Crit Care 8:163–169
Ranieri VM, Giuliani R, Fiore T, Dambrosio M, Milic-Emili J (1994) Volume-pressure curve of the respiratory system predicts effects of PEEP in ARDS: “occlusion” versus “constant flow” technique. Am J Respir Crit Care Med 149:19–27
Gattinoni L, Pelosi P, Suter PM, Pedoto A, Vercesi P et al (1998) Acute respiratory distress syndrome caused by pulmonary and extrapulmonary disease. Different syndromes? Am J Respir Crit Care Med 158:3–11
Stenqvist O, Lundin S (2014) Lung elastance and transpulmonary pressure may be determined without using esophageal pressure measurements. Am J Respir Crit Care Med 190:120
Dellamonica J, Lerolle N, Sargentini C, Beduneau G, Di Marco F et al. (2011) PEEP-induced changes in lung volume in acute respiratory distress syndrome. Two methods to estimate alveolar recruitment. Intensive Care Med 37:1595–1604
Agostoni E, Mead J (1973) Statics of the respiratory system. In: Fenn WO, Rahn H (eds) Handbook of physiology. American Physiological Society, Washington, pp 387–409
Donders FC (1853) Beiträgezum Mechanismus der Respiration und Circulation in gesunden und kranken Zustände. Z Ration Med 3. Zeit Ration Med N F 3:287–319
Nunn J (1995) Elastic forces and lung volumes. In: N JF (ed) Nunn’s applied respiratory physiology, 4th edn. Butterworth-Heinemann, Oxford, pp 36–52
West JB (1985) Mechanics of breathing. In: West JB (ed) Respiratory physiology, 3rd edn. Williams & Wilkins, Baltimore, Hong Kong, London, Sydney, pp 85–111
Mead J, Gaensler EA (1959) Esophageal and pleural pressures in man, upright and supine. J Appl Physiol 14:81–83
Agostoni E (1961) A graphical analysis of thoracoabdominal mechanics during the breathing cycle. J Appl Physiol 16:1055–1059
Christie RV (1934) The elastic properties of the emphysematous lung and their clinical significance. J Clin Invest 13:295–321
Finucane KE, Colebatch HJ (1969) Elastic behavior of the lung in patients with airway obstruction. J Appl Physiol 26:330–338
Hurtado A, Kaltreider NL, Fray WW, Brooks WD, McCann WS (1934) Studies of total pulmonary capacity and its Sub-divisions. Vi. Observations on cases of obstructive pulmonary emphysema. J Clin Invest 13:1027–1051
Sciurba FC, Rogers RM, Keenan RJ, Slivka WA, Gorcsan J 3rd et al (1996) Improvement in pulmonary function and elastic recoil after lung-reduction surgery for diffuse emphysema. N Engl J Med 334:1095–1099
Agostoni EMJ (1986) Handbook of physiology. American Physiological Society, Bethesda
Pelosi P, Cereda M, Foti G, Giacomini M, Pesenti A (1995) Alterations of lung and chest wall mechanics in patients with acute lung injury: effects of positive end-expiratory pressure. Am J Respir Crit Care Med 152:531–537
Talmor D, Sarge T, Malhotra A, O’Donnell CR, Ritz R et al (2008) Mechanical ventilation guided by esophageal pressure in acute lung injury. N Engl J Med 359:2095–2104
Garnero A, Tuxen D, Ducros L, Demory D, Donati SY et al (2015) Non-invasive assessment of lung elastance in patients with acute respiratory distress syndrome. Minerva Anestesiol 81:1096–1104
Gulati G, Novero A, Loring SH, Talmor D (2013) Pleural pressure and optimal positive end-expiratory pressure based on esophageal pressure versus chest wall elastance: incompatible results*. Crit Care Med 41:1951–1957
Pelosi P, Goldner M, McKibben A, Adams A, Eccher G et al (2001) Recruitment and derecruitment during acute respiratory failure: an experimental study. Am J Respir Crit Care Med 164:122–130
Talmor DS, Fessler HE (2010) Are esophageal pressure measurements important in clinical decision-making in mechanically ventilated patients? Respir Care 55:162–172, discussion 172-164
Protti A, Andreis DT, Milesi M, Iapichino GE, Monti M et al (2015) Lung anatomy, energy load, and ventilator-induced lung injury. Intensive Care Med Exp 3:34
Acknowledgements
We are grateful for the dedicated software for the estimation of lung volume changes provided by Magnus Hallbäck Ph.D, M.Sc, Maquet Critical Care, Solna, Sweden
Funding
This study was supported by grants from the Sahlgrenska Academy at the University of Gothenburg, the Medical Society of Gothenburg, Sweden, and the Swedish Heart and Lung Foundation.
Authors’ contributions
OS and SL designed the study. PP and OS performed the experimental research and were responsible for data acquisition. PP, SL, and OS analyzed the data. OS, PP, and SL drafted and revised the manuscript. All authors read and approved the final manuscript.
Competing interests
Stefan Lundin and Ola Stenqvist hold stocks in The Lung Barometry Company Sweden AB. Per Persson has no competing interests.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Persson, P., Lundin, S. & Stenqvist, O. Transpulmonary and pleural pressure in a respiratory system model with an elastic recoiling lung and an expanding chest wall. ICMx 4, 26 (2016). https://doi.org/10.1186/s40635-016-0103-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s40635-016-0103-4