INTRODUCTION

The hardware–software complex of a Gamma-spectrometer was developed in 2014 at the Dolgoprudnenskaya Scientific Station (DNS) of the Lebedev Physical Institute within the framework of international cooperation with scientists from Brazil and Argentina, which is designed to detect and determine the energy spectra of secondary γ-quants [1]. This equipment is installed at the CASLEO observatory (Argentina, S31.47°, W69.17°, at an altitude of 2550 m above sea level and a geomagnetic cutoff rigidity of Rc = 9.8 GV), where the complex of cosmic ray detectors that was previously developed at the FIAN DNS is currently operating. CARPET, the first facility that records the charged component of cosmic rays, was launched in 2006 [27]. In 2015, the Gamma-spectrometer and the neutron detector device for recording the neutron component of cosmic rays were launched [8, 9].

This paper presents the circuitry solutions used in the development of the Gamma-spectrometer and its characteristics. The background radiation calibration technique is also presented.

THE HARDWARE AND SOFTWARE COMPLEX

The Gamma-spectrometer contains four detecting modules (Fig. 1), which are based on the 12S12/3.VD.HVG.PA scintillation assembly manufactured by ScintiTech (http://www.scintitech.com/). The scintillation assembly includes a Hamamatsu R1307 photomultiplier tube (PMT) (https://www.hamamatsu. com/jp/en/product/type/R1307/index.html) 76 mm in diameter, an NaI (Tl) scintillator with a height of 76.2 mm and an electronic unit, which consists of a high-voltage converter, a voltage divider and a preamplifier. A voltage of ±12 V is required to power the electronic unit.

Fig. 1.
figure 1

A block diagram of the Gamma-spectrometer. ADC, analog-to-digital converter; PC, personal computer.

The interface unit we developed (Fig. 2) is designed to connect four detecting modules via power connectors (XR1–XR4) and signal connectors (XR6–XR9). Primary supply voltages ±12 V are generated on TML 05212 converters (DA1 and DA2). Each transducer supplies two detection modules. The output voltages of the high-voltage converters of the scintillation assemblies are adjustable in the range from –200 V to –1500 V, which makes it possible to set a high voltage for each PMT individually using potentiometers R1R4.

Fig. 2.
figure 2

A schematic diagram of the gamma-spectrometer interface module. DA1, DA2—TML05212; DA3, DA4— LM1117-5.0.

Using the XR5 connector, signals from the detection modules are fed to an analog-to-digital converter (ADC) USB3000 (http://www.r-technology.ru/products/adc/usb3000.php), where they are converted to a sequence of two-byte samples. Each count is a number in the range from –8000 to +8000 (maximum amplitude resolution), which corresponds to the signal range from –5 to +5 V. Since the signals from the detecting modules are positive, the output ADC a sequence of samples is obtained in the range from 0 to 8000. Therefore, the signal quantization step is –625 μV.

Data transfer from ADC to PC is carried out via the USB bus. In the software settings one can set the ADC sampling rate (time resolution), the step of signal quantization by amplitude (number of channels), and the duration of the data accumulation time for each file.

The software of the PC sequentially generates files containing data received during a specified time interval (duration). Under the current conditions of the experiment, measurements are carried out with a specified duration of data files of 600 s and a sampling rate of 500 kHz, with an energy resolution of 128 channels.

Each data file in the header contains information about the start time (UTC), the sampling rate ADC and the duration of the data collection interval. Next, a table is written, which indicates the number of the energy channel, the number of pulses (γ-quants) that entered the channel with the number N (differential spectrum), the total number γ-quanta that entered the channels starting from the number N + 1 and higher (integral spectrum), and the γ-quant energy corresponding to the channel number, as calculated by preliminary calibration with 60Co and 137Cs sources.

SPECTROMETER CALIBRATION

The detecting modules of the Gamma-spectrometer were pre-calibrated at the LPI using sources 60Co γ-radiation (1.17 and 1.33 MeV) and 137Cs (661.7 keV). Initial calibration was necessary in order to correctly select the output voltages of the high-voltage converters: 775 V for module 0, 789 V for module 1, and 728 V for modules 2 and 3, which roughly set the ranges of the detected γ-quanta energies of 50 keV–3.5 MeV for modules 0 and 1 and 50 keV–5 MeV for modules 2 and 3. However, this calibration cannot be considered final, since it was carried out in laboratory conditions and does not take the instrumental effect into account, that is, the effect of temperature on the output voltage of high-voltage converters.

To illustrate the temperature effect, Fig. 3 shows the graphs of two differential spectra for the detecting module 0: spectrum 1, obtained by superimposing epochs of all 10-min spectra for January 2020, spectrum 2, as of June 2020

Fig. 3.
figure 3

Graphs of the differential spectrum of the module 0 Gamma-spectrometer: 1, according to data for January 2020 (summer period), 2, for June 2020 (the winter period).

It was necessary to develop a method for continuous calibration of the spectrometer according to the available data, which takes the temperature effect into account and does not require additional devices, since the installation operates in an autonomous mode. For calibration, one can use the spectral lines of radioactive substances in soil. The isotopes 40K, 214Bi, 208Tl were used as calibration sources for γ-radiation [10]. Figure 3 shows the approximate neighborhoods of the spectral lines of these isotopes. Compton scattering of energetic γ-quant, therefore it is not used during calibration in the energy range of ~0–0.4 MeV. The penultimate energy peak, which lies in the range of approximately the (65–75)th channels is a consequence of the formation of electron–positron pairs and is also not taken into account [11].

The graphs in Fig. 3 show the discrepancy between the lines of isotopes in the spectrum, which increases with increasing energy (channel number).

The initial condition is added to the existing six spectral lines: E(U) = 0 at U = 0. The result is seven known points En: E0(U0 = 0), E1(U1), ..., E6(U6), dividing the entire spectral range of the detecting module into six intervals, the energy values within which we will fill with a piecewise linear function:

$$E(U) = {{E}_{n}} + {{k}_{n}}(U - {{U}_{n}}),$$
(1)

where E(U) is the energy value corresponding to the channel number U; En is the initial energy value in this interval, corresponding to the channel number Un; and kn is the slope coefficient of the given interpolation line:

$${{k}_{n}} = \frac{{{{E}_{{n + 1}}} - {{E}_{n}}}}{{{{U}_{{n + 1}}} - {{U}_{n}}}},$$
(2)

where En + 1 is the final energy value in this interval, corresponding to the channel number Un + 1.

To demonstrate the technique, we consider an arbitrary time interval, for example, January 1, 2018 from 01:00 to 02:00 UTC (Fig. 4). According to the most prominent peaks corresponding to 40K, it is clearly seen that the sensitivity of detecting modules 0 and 1 is higher than the sensitivity of detecting modules 2 and 3, due to the higher voltages set across the PMT.

Fig. 4.
figure 4

The initial differential spectra for four detecting modules (0–3) according to the data for January 1, 2018 from 01:00 to 02:00 UTC. The numbers at the spectra correspond to the modules.

Using this technique, the spectra obtained by the detecting modules were divided into six intervals, in each of which the energy values were obtained using formulas (1) and (2) γ-quanta depending on the channel number (Fig. 5). In the graphs in Fig. 5 some difference in the absolute values of the counting rates of the detecting modules is noticeable, which, if necessary, can be eliminated by mutual normalization according to the data of simultaneous measurements.

Fig. 5.
figure 5

The differential spectra for four detecting modules (0–3), after energy recalculation of γ-quants, according to data for January 1, 2018 from 01:00 to 02:00 UTC.

CONCLUSIONS

This paper presented a description and the schematic implementation of an installation for detecting flows. The γ-radiation spectrometer has been continuously functioning in the CASLEO astronomical complex from 2015 to the present. For a comprehensive analysis of the experimental data, a technique has been developed for the calibration and recalculation of the measured values of the channel numbers of the analog-to-digital converter into the energy of γ-quanta based on natural background sources of γ-radiation.

This setup is of particular interest for the study of generation processes of γ-radiation in thunderclouds and from lightning discharges (so-called TGF- and TLE-events) [12].