Abstract
To develop and investigate the feasibility of sub-second temporal resolution volumetric T1-weighted four-dimensional (4D-) MRI in comparison with 4D-CT for respiratory-correlated motion assessment using an MRI/CT-compatible phantom. Sub-second high temporal resolution (0.5 s) gradient-echo T1-weighted 4D-MRI was developed using a volumetric acquisition scheme with compressed sensing. An MRI/CT-compatible motion phantom (simulated liver tumor) with three sinusoidal movements of amplitudes and two respiratory patterns was introduced and imaged with 4D-MRI and 4D-CT to investigate the geometric accuracy of the target movement. The geometric accuracy, including centroid position, volume, similarity index of dice similarity coefficient (DSC), and Hausdorff distance (HD), was systematically evaluated. Proposed 4D-MRI achieved a similar geometric accuracy compared with 4D-CT regarding the centroid position, volume, and similarity index. The observed position differences of the absolute average centroid were within 0.08 cm in 4D-MRI and 0.03 cm in 4D-CT, less than the 1-pixel resolution for each modality. The observed volume difference in 4D-MRI/4D-CT was within 0.73 cm3 (4.5%)/0.29 cm3 (2.1%) for a large target and 0.06 cm3 (11.3%)/0.04 cm3 (11.6%) for a small target. The observed DSC values for 4D-MRI/4D-CT were at least 0.93/0.95 for the large target and 0.83/0.84 for the small target. The maximum HD values were 0.25 cm/0.31 cm for the large target and 0.21 cm/0.15 cm for the small target. Although 4D-CT potentially exhibit superior numerical accuracy in phantom studies, the proposed high temporal resolution 4D-MRI demonstrates sub-millimetre geometric accuracy comparable to that of 4D-CT. These findings suggest that the 4D-MRI technique is a viable option for characterizing motion and generating phase-dependent internal target volumes within the realm of radiotherapy.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
Four-dimensional CT (4D-CT) is essential for determining respiration-related target motion while avoiding tumour under-dosage and explaining in detail organ at risk toxicity when adopted for radiotherapy planning1. Generally, an internal target volume derived using 4D-CT is used to assess respiratory motion and determine treatment margins for a moving target2. However, for a liver tumour, the poor soft-tissue contrast in conventional 4D-CT does not provide sufficient information to determine the target or assess its motion. Contrast-enhanced CT is generally used in clinical practice to obtain additional information3. However, a 4D-CT scan is usually inappropriate because acquiring the target region requires several minutes (mostly the entire liver). In addition, the axial direction of CT scan introduces undesirable motion artefacts, particularly in the superior-inferior direction, during high respiratory rate or irregular respiration4.
Respiratory-correlated 4D-MRI is a promising modality for evaluating the target motion of liver cancer in radiotherapy because of its high soft-tissue contrast and longer visualization capturing target motion without radiation exposure. This approach also provides a non-axial scanning direction, more desirable with fewer motion artefacts in the superior-inferior direction. Compared with 2D multi-slice acquisition-based MRIs5,6, 3D or volumetric acquisition-based 4D-MRIs have exhibited better geometric corrections, motion averaging, and image acceleration options7,8,9. However, previously reported volumetric 4D-MRI presented limitations regarding gaining sufficient temporal-spatial resolution while maintaining the image quality for tumour-to-tissue contrast7,10,11. Moreover, previous studies primarily applied single-shot T2-weighted or balanced steady-state free-precession sequences, where boundaries between the tumour and background liver tissue cannot be clearly defined12.
In order to address the limitations of poor soft tissue contrast in the liver site when determining moving targets using 4D-CT, as well as the challenges in achieving sufficient temporal-spatial resolution in previously reported 4D-MRI techniques, we have developed a 4D-MRI methodology based on gradient-echo 3D T1-weighted images (hepatobiliary phase) that attains high temporal-spatial resolution while maintaining optimal tumor-to-liver contrast.The gradient-echo 3D T1-weighted fast imaging is one of the basic sequences in liver MRI13. This sequence is used for hepatobiliary phase images that are acquired approximately 20 min after administration of gadoxetic acid contrast agent. The hepatobiliary phase image shows the highest tumour-to-liver contrast, lesion delineation, and continuously contrast enhancement in several hours14. In proposed 4D-MRI sequence, recently developed compressed sensing, enabling accelerated MRI acquisitions using the reconstruction of sparse images from highly under-sampled data15,16, was implemented to achieve sub-second temporal resolution. To the best of our knowledge, there are only a limited number of studies17,18 that have evaluated the T1-weighted 4D-MRI sequence in the existing literature. The proposed 4D-MRI method adopts a distinct approach from those previously reported. The aim of the present study is twofold: firstly, to employ the proposed 4D-MRI methodology to achieve sub-second temporal resolution in conjunction with high spatial resolution and image contrast; and secondly, to assess its accuracy by evaluating respiratory-correlated motion and comparing it with 4D-CT.
Materials and methods
This study is Institutional Review Board exempt from the Kobe University Hospital as no actual patient data was used. All methods were carried out in accordance with relevant guidelines and regulations in our institution.
Phantom preparation
Phantom geometry
A commercial MRI/CT-compatible motion phantom (QUASA MRI4D Motion Phantom, ModusQA Medical Device), which measures 20 × 30 cm in diameter and 20 cm in length, was employed to conduct the evaluation systematically. The default phantom insert included a 3-cm-diameter spherical object as a simulated tumour. Additionally, a 1-cm-diameter insert to simulate a small-size tumour was also fabricated in this study (Fig. 1). External surrogate signals were used in both modalities for consistent comparison; a bellow (an airbag with a pressure sensor) was used for 4D-MRI, and an infrared reflective cube sensor for 4D-CT. The performances of several respiratory patterns in terms of different respiratory periods and motion amplitudes were quantified using two shapes of the simulated target.
Contrast agent solution
The contrast agent gadolinium-ethoxybenzyl-diethylenetriamine-pentaacetic-acid (Gd-EOB-DTPA, SH L 569 B, Primovist®) was used to prepare the optimal solution inside the phantom to gain target-to-tissue contrast similar to the target and normal tissue in gradient-echo 3D-T1-TFE (hepatobiliary phase image for liver cancer). Target-to-tissue in 4D-CT was necessary for comparison purposes using the same phantom settings. Hence, a contrast medium solution was prepared using 1000 mL of distilled water, 2 mL of Gd-EOB-DTPA (EOB Primovist® Inj Syringe, BAYER), and 12 mL of Iohexol (OMNIPAQUE® 300 inject syringe, GE healthcare) to simulate the hepatobiliary phase image of the liver in MRI and equilibrium phase image in CT (the solution was referred to in a previous study19 and modified). For a 3-cm-diameter target in MRI, the simulated tumour was filled with distilled water and its exterior with a contrast agent solution. In clinical practice, the simulated tumour is present at a lower signal without EOB, unlike a normal liver, which is present at a higher signal. For a 1-cm-diameter target in the MRI, the simulated tumour was injected with the EOB/Iohexol contrast agent solution and its exterior with water because creating a spherical object with a 1-cm-diameter was challenging. A possible workaround could be creating a 1-cm-diameter cavity inside a cylindrical stick to simulate the small-sized simulated tumour. In this case, only the spherical cavity would be present at a higher signal.
Motion pattern
The sinusoidal movements of amplitudes 1, 2, and 3 cm along the longitudinal axis were used in two respiratory settings: a regular respiratory pattern with 5 s/cycle and a fast respiratory pattern with 3 s/cycle. The sinusoidal motion patterns were modelled as the fourth power of a sinusoidal wave function, as reported in20.
where A and T represent the amplitude of the motion and respiratory cycle, respectively.
4D-MRI image acquisition
To achieve sub-second high temporal resolution, our developed sequence features a single-shot 3D acquisition technique, which incorporates both half-scan and compressed sensing techniques. The 4D-MRI acquisition process is presented in Fig. 2. However, a potential limitation of this approach is the reduction in image contrast resulting from an increased TFE (Turbo Field Echo) factor. Consequently, the use of contrast agent in 4D-MRI was motivated by the desire to attain high temporal resolution imaging without compromising image contrast. Volumetric (3D) MRI data were acquired using a clinical MRI 3 T scanner (Ingenia ver. 5.4, Philips Medical Systems) with the following parameters: 3D-T1-TFE; TFE prepulse: saturate; TFE shot duration: 832 ms (849 ms) for a large (small) target; FOV 256 × 256 mm2; temporal resolution: 0.5 s; spatial resolution: 1.45 × 1.45 × 3 mm3 with 15 slices (1.45 × 1.14 × 2 mm3 with 12 slices) for a large (small) target; flip angle: 5°; TR/TE = 2.5/1.26 ms (2.6/1.25 ms) for a large (small) target; readout bandwidth: 1393 Hz/pixel; no fat suppression; compressed SENSE (Philips Medical Systems); factor: 5; and de-noising: strong. The half Fourier factor values utilized in this study were Y = 0.7 and Z = 0.8. The TFE (Turbo Field Echo) factor was set to 204, while the gradient mode was adjusted to its maximum setting. In this study, we employed a commercial application of the compressed sensing technique (Compressed SENSE). This approach integrates Compressed Sensing15,16 and SENSE technologies, and allows for user-defined reduction rates and denoising levels. When applied to 3D data acquisition, Compressed SENSE minimizes motion artifacts by setting a high reduction rate, which maintains image quality while significantly reducing acquisition time. To acquire 10-phase volumetric data based on each respiratory cycle, we used an external bellow sensor (an airbag with a pressure sensor) to monitor the respiratory cycles and acquire data using the triggered signal set at the inhale phase. Figure 3a illustrates the setup for the bellow sensor, configured to stably detect movement using the phantom insert. The delay interval for acquiring the next phase was specifically set for each respiratory pattern (e.g., 500 ms for 5 s/cycle) to acquire the 10-phase volumetric data (thrice). The 10-phase image data were sequentially binned based on the respiratory profile to generate the 4D image series. A static image series without motion was used as a reference.
4D-CT image acquisition
The 4D-CT images were acquired using the clinical multi-slice (16 detector rows) CT scanner (Aquilion LB, Canon Medical Systems). For CT scan protocol, the following clinical exposure settings were used: 120 kVp; 100 mA (200 mA) for a large (small) target; FOV: 500 × 500 mm2; gantry rotation speed: 0.5 s; and spatial resolution: 0.98 × 0.98 × 2 mm3. The respiratory phase was detected using the Varian real-time position management (RPM) system (Varian Medical Systems, Inc.) with an infrared reflective cube sensor positioned as in Fig. 3b. Similarly, a static image series without motion was used as a reference.
Image analysis and statistical analyses
All the images were transferred to the MIM Maestro version 6.9.6 (MIM Software Inc, Cleveland, OH, USA) for target segmentation. To minimize the potential bias arising from the differing resolutions of the two modalities, we resampled images for both modalities to a resolution of 1 × 1 × 1 mm3 prior to performing ROI generation. This ensured that the same resolution was employed for delineating ROIs in both modalities. The reference target was contoured based on the static image series. A region growth method was used for objective and repeatable segmentation of each image series. The upper and lower thresholds of region growth were determined by generating the reference target region of interest (ROI). Minimal manual modification was used when the method did not perform properly.
Position accuracy evaluation
The target position was determined from the centroid of the target ROI for each image set. The exhale phase was normalised for reference, and the displacement distances of other phases were calculated with respect to it. The target and ground truth (GT) positions derived from the sinusoidal model were plotted and compared to evaluate their accuracy.
Target volume and similarity index evaluation
We utilized static images as a reference for each modality to assess volume and shape variations of the imaged object due to motion artifacts. The rationale behind using static images as a reference for each modality was the difficulty in discerning the plastic wall of the target phantom insert (as illustrated in Fig. 1) and the stem using MRI. Consequently, we defined the reference ground truth volume in the static image for each modality; specifically, for 4D-MRI/4D-CT, the large and small target volumes were 16.25 cc/14.13 cc and 0.57 cc/0.33 cc, respectively. The volume difference between the measurements and the reference GT was calculated as:
where Vmeasured and Vref represent the measured and actual volumes of the spherical object, respectively.
The volume difference could not reveal the extent of the shape difference between ROIs. Hence, the similarity index of each respiratory phase was evaluated using the Dice similarity coefficient (DSC)21 and Hausdorff distance (HD)22. The ROI of each phase was shifted with a constant displacement (based on the centroid position) in the superior-inferior direction with respect to the reference position. Moreover, the similarity index between the shifted and reference ROIs was calculated. The Mann–Whitney U-test was used to compare the performances of the two modalities (n = 60). Statistical calculations were performed in MATLAB 2022a (MathWorks, Natick, MA, USA). For statistical analyses, the reported p values were two-tailed; a p value < 0.05 was considered significant. A flowchart was included in Supplementary Material 2 to illustrate the methodology employed in this study.
Results
Figure 4 shows the acquired 4D-MRI and 4D-CT images with ten respiratory phases in a regular (5 s/2 cm) and fast (3 s/3 cm) motion pattern. The static image for each modality is presented on the left of the figure. In a regular motion pattern for large targets, 4D-MRI and 4D-CT represent visually comparable target-to-tissue contrasts with a sharp boundary in each respiratory phase. In a fast-motion pattern for large targets, a fast-moving phase (e.g., ExIn3, InEx1) revealed a slightly deformed target shape and blurred boundary in both modalities. For small targets in 4D-MRI, the fast-moving phase showed a large deformed shape along the movement direction while maintaining the target-to-tissue contrast with a sharp boundary. Contrarily, the target-to-tissue contrast for 4D-CT decreased in the fast-moving phase and the blurred target boundary (e.g., ExIn2-3, InEx1-2). Supplementary Material 1 presents a zoomed-out view of the imaged phantom with as much detail as possible.
Position accuracy evaluation
The results of position accuracy analysis are plotted in Fig. 5. The averaged position difference and standard deviation from GT are summarised in Table 1. For a regular motion pattern, the absolute averaged position difference was observed within 0.06 and 0.02 cm for 4D-MRI and 4D-CT, respectively. Moreover, for the fast-motion pattern, the absolute averaged position difference was observed within 0.08 and 0.03 cm for 4D-MRI and 4D-CT, respectively. For both image modalities, the position accuracy of the small target was comparable to that of the large target. Although the average difference (0.05 cm) was less than 1-pixel resolution in both modalities, the results of the Mann–Whitney test suggested that 4D-CT outperformed the position accuracy of 4D-MRI (p < 0.01 for both large and small targets) in each case.
Target volume and similarity index evaluation
The results of the target volume difference are plotted in Fig. 6. The absolute and percentage volume differences are summarised in Table 2. The volume difference in 4D-MRI and 4D-CT for the regular motion pattern was within 0.48 (3.0%) ± 0.25 cm3 and 0.19 (1.3%) ± 0.06 cm3 for the large target and 0.06 (11.3%) ± 0.03 cm3 and 0.03 (7.6%) ± 0.02 cm3 for the small target, respectively. For the fast-motion pattern, the volume difference in 4D-MRI and 4D-CT was observed within 0.72 (4.5%) ± 0.26 cm3 and 0.29 (2.1%) ± 0.19 cm3 for the large target, and 0.06 (11.2%) ± 0.04 cm3 and 0.03 (7.6%) ± 0.02 cm3 for the small target, respectively. The Mann–Whitney test showed that 4D-MRI was statistically equivalent to 4D-CT (p = 0.13) for large target, but 4D-CT showed improved performance (p < 0.01) for small target.
The results of the target similarity index (DSC and HD) analysis for each respiratory phase are plotted in Fig. 7. Small targets with fast-motion patterns were less robust than large targets with regular motion patterns. Phases in ExIn2-3 and InEx1-2, which involved larger movements, exhibited poor similarity index values (shown as low DSC or high HD). For the two similarity indices summarised in Tables 3 and 4, 4D-MRI and 4D-CT showed comparable results. In particular, the minimum average DSC in 4D-MRI and 4D-CT was 0.93 ± 0.01 and 0.95 ± 0.02 for the large target and 0.83 ± 0.07 and 0.84 ± 0.06 for the small target, respectively. Contrastingly, the maximum average HD values in 4D-MRI and 4D-CT were 0.25 ± 0.06 cm and 0.31 ± 0.07 cm for the large target and 0.21 ± 0.06 cm and 0.15 ± 0.06 for the small target, respectively. Although the two modalities have equivalent similarity indices, the results of the Mann–Whitney test showed that 4D-CT has numerically improved performance compared to 4D-MRI; specifically, p < 0.01 and p = 0.09 for large and small targets in terms of DSC, respectively, and p < 0.01 for both targets in terms of HD. Included in Supplementary Material 3 is a detailed statistical summary of our study.
Discussion
In this study, we developed a sub-second high temporal-spatial resolution T1-weighted 4D-MRI using a volumetric acquisition scheme with compressed sensing. Additionally, we systematically investigated the 4D-MRI feasibility for target motion assessment using a motion phantom. The 4D-MRI exhibited similar geometrical accuracy compared with 4D-CT in terms of the centroid position, volume, and similarity index. The results suggest that the proposed volumetric T1-weighted 4D-MRI sequence is feasible for accurate assessment of target movements.
The position accuracy results between in 4D-MRI and 4D-CT were comparable to those reported in previous phantom studies. For a 3-cm target, Yue et al.19 calculated the error as 0.02 and 0.04 cm for fast and relatively slow motions. In the present study, we calculated the position error as 0.02 and 0.03 cm for fast and regular motions. These results confirmed that a volumetric-scan-based 4D-MRI with compressed sensing could achieve excellent position accuracy comparable to a 2D multi-slice-basis 4D-MRI. For a 1-cm small target that was rarely investigated, the average errors were observed within 0.08 cm in volumetric T1-weighted 4D-MRI, which was larger than those of 4D-CT (within 0.03 cm). This is because of the insufficient temporal resolution and relatively slow 3D-readout process in volumetric 4D-MRI, affecting the accuracy of the centroid position derived from the target ROI. To mitigate the potential influence of resolution discrepancies between two modalities, we resampled images for both modalities to a resolution of 1 × 1 × 1 mm3 prior to performing ROI generation. Furthermore, our analysis of position accuracy revealed that the displacement of the target ROI for both modalities was within 1 mm, which is deemed acceptable since it falls below the 1-pixel resolution used for ROI generation. The results of the volume difference and similarity index indicate that the volumetric T1-weighted 4D-MRI and 4D-CT can achieve similar accuracies. In particular, the similarity index was applied in our study because the target volume might underestimate the motion-induced shape change (e.g., same volume but different shape). The results show that the volume difference is usually correlated with DSC and HD, compared to the DSC, providing more direct insights into the motion-induced artefacts. In DSC for small targets in the proposed 4D-MRI, phases ExIn3 and InEx1 were observed to be most vulnerable to the motion artefacts along the motion direction. The insufficient temporal resolution and relatively slow process for 3D-readout in the proposed sequence might cause the artefacts to appear as a shape deformation along the movement direction. In contrast, the movement trajectory was also captured in the volumetric-based MR scan, suggesting that 4D-MRI has no limitations in evaluating the target movement range and defining internal target volume.
The literature contains several studies discussing the use of 4D-MRI utilizing 3D sequences. Feng et al.23 developed a framework called iGRASP for image reconstruction, which enhances image quality and reduces motion-blurring artifacts in liver dynamic contrast-enhanced MRI. Oar et al.17 employed the stack-of-stars radial trajectory24, which acquires data through the center of k-space, to generate respiratory phase-resolved 4D-MRI with reduced motion artifacts. Although, these studies did not provide details on the temporal resolution, their temporal resolution was limited to more than 10 s23. To achieve sub-second high temporal resolution, our developed sequence features a single-shot 3D acquisition technique, which incorporates both half-scan and compressed sensing techniques. Despite its merits, the approach presents potential impediments, including a decrease in contrast due to an elevated TFE factor and a limited slice availability (between 2.4 and 4.5 cm in the current study). Therefore, we utilized an enhanced hepatobiliary phase T1-weighted MRI, ensuring high-temporal-resolution imaging without compromising image contrast. The 3D acquisition was performed in the sagittal plane, covering a FOV size of 256 × 256 mm2 to capture the mostly respiratory-induced motion direction (superior-inferior(S-I) and anterior–posterior(A-P) direction). Though respiratory motion typically exerts less influence on the left–right(L-R) direction, the available slice range could impact assessments of movements exceeding 7 mm in the L-R direction under the current setting. Nevertheless, their findings suggest a strong potential for combining the motion-robustness of radial imaging with compressed sensing techniques, which could be integrated into our sequence. Biederer et al.25 utilized an artificial chest phantom to investigate the diaphragm dome displacements between 4D-MRI and 4D-CT. Despite the significant spatial resolution difference between the two modalities in their study, the authors emphasized the importance of temporal resolution in evaluating 4D-MRI. We concur with this notion, and it served as the driving force for our current work. Besides, our findings showed that 4DCT exhibits better numerical accuracy compared to 4DMRI, such as with a target ROI displacement of 0.03 cm in 4D-CT and 0.08 cm in 4D-MRI for small targets. However, it is important to note that both modalities show sub-millimeter ROI displacement, which falls below the resolution used for ROI generation and is considered comparable. Similar results were observed when comparing the volume differences and similarity indices between the two modalities. Therefore, the conclusion of our study is not to assert the superiority of 4DMRI over 4DCT but to state that 4DMRI, in its current stage, can achieve a level of accuracy that is comparable to 4DCT. Further, Yuan et al.18 demonstrated a rapid 4D MR acquisition using a 3D scheme, achieving a sub-second to 0.6-s temporal resolution. The reported spatial resolution (2.7 × 2.7 × 4 mm3) might limit the accuracy of evaluation for small targets (e.g., 1 cm in diameter), as only 2–3 slices may be available for target delineation in a 4 mm out-of-plane resolution. It's generally recognized that an enhancement in spatial resolution might correspondingly deteriorate image quality. However, considering that the principal objective of employing 4D-MRI in radiotherapy is to assess the motion of the target, rather than mapping the entire abdominal area, our approach strategically confines the FOV and slice range to a suitable degree, without hindering the evaluation of motion in the most significant directions. Through this optimized methodology, we've managed to achieve a 0.5-s temporal resolution, which exceeds the performance of previous studies while simultaneously maintaining high spatial resolution.
In terms of clinical implications, the findings of our study have several potential applications. Increasingly, the evidence-based advantages of radiotherapy, notably SBRT, in liver cancer treatment are acknowledged, providing exceptional local control, improved survival rates, and minimal adverse effects26,27.Notably, tumors smaller than 5 cm have been shown to have particularly favorable outcomes26,28. Our proposed methodology enhances this approach by offering a more intricate and comprehensive evaluation of tumor motion. Moreover, it holds significant potential for managing smaller tumors (e.g., 1 cm in size), a task that presents substantial challenges to current techniques. The proposed method could assist in the precision of target margin setting and treatment planning, resulting in a potential enhancement of the therapy's effectiveness and a reduction in its side effects. Nevertheless, the current method has a relatively restricted scan scope in out-of-plane could limit the assessment of larger tumors under high spatial resolution conditions. Concurrently, the potential usability of MRI-linac using 4D MRI has also been highlighted, with improved target coverage and reduced dose to organs at risk compared to conventional radiotherapy29. Despite uncertainties related to dosimetric impact of the contrast agent, our approach offers a promising imaging guide sequence for future 4D-MRI guided online adaptation by MRI-linac. Moreover, many hepatocellular carcinoma patients have chronic liver diseases, posing potential risks of liver decompensation post-SBRT30. Recent suggestions31 propose individualized adaptive SBRT using the indocyanine green retention test as a liver function marker. Contrastingly, our method could provide a non-invasive technique for evaluating both overall and local liver functionality, extending its utility beyond merely a motion management strategy13,32.
The present study has a few limitations that need to be addressed in future investigations. First, given the experiment's magnitude, we had to restrict our study to a single-axis (in S-I direction) sigmoidal motion, which typically represent the most prominent direction of movements. It's noteworthy that in-plane resolution (S-I and A-P direction) surpasses that of out-of-plane (L-R direction), suggesting that motion occurring within the in-plane directions is likely to be measured with a higher degree of accuracy. Multi-axis movement, including a higher proportion of out-of-plane motion, might detrimentally affect the precision of positioning and volume estimation. Further investigations are required to evaluate the impacts of multi-axis motion, which is frequently encountered in real patient scenarios. In addition, surrogate signals using different approaches were not investigated. Prior studies have shown that internal navigator or self-navigation techniques might have fewer binning motion artefacts and more robustness against irregular breaths33,34. However, Stemkens et al.7 argued that the self-navigation method has limitations, including phase instability and partial saturation bands, particularly in the case of 2D image-based self-navigators. It is vital to understand that in a phantom study, surrogate indicators, whether external or internal, have a minimal impact on the outcome due to the rigid and consistent nature of phantom motion. Although exploring various surrogate signals is beyond the scope of current phantom study, we wish to underline that, in actual patients, the external signal might differ from the external signal. Moreover, the implementation of a respiratory phase sorting/reconstruction algorithm is expected to deal with irregular breathing patterns and better simulate real patient cases17,35. Unfortunately, we are currently unable to incorporate a sorting/reconstruction algorithm into our sequence since the techniques are still in development. Therefore, in this study, we utilized a respiratory gating sensor to acquire 10-phase MR images, which were binned in a sequential manner based on the respiratory profile. A comparable approach was also reported in a previous study18. While our study focused on achieving high temporal resolution in 4DMRI acquisition, we acknowledge the need for a respiratory phase sorting/ reconstruction algorithm to enhance its clinical utility. High temporal-spatial resolution MRI with satisfactory target-to-background contrast was verified using the introduced scanning parameters. However, scanning a patient with large body sizes or more dynamic tissue densities might be challenging. Further investigation is required to ensure whether the proposed 4D-MRI could provide clinically acceptable image quality for real patients. Moreover, Further improvements in temporal-spatial resolution and image quality in 4D-MRI might not necessarily be limited by hardware development or image acquisition techniques. The trade-off between escalating temporal-spatial resolution and preserving image quality could be significantly optimized by employing artificial intelligence-enhanced image processing techniques36,37,38. A recent study36 highlights a deep learning-driven super-resolution method for MRI, achieving 2-3x (potentially up to 4x) enhanced resolution while maintaining image quality. This indicates the feasibility of achieving sub-millimeter spatial resolution in both in-plane and out-of-plane directions with our proposed method. Moreover, it suggests an alternative approach to expanding the scan scope in the out-of-plane direction of our proposed sequence by adjusting the scan resolution (e.g., 5 mm) for larger targets, while maintaining a final output at 2-mm spatial resolution. In addition, a variety of deep learning-oriented techniques for noise and artifact reduction in MRI have been proposed, tested on diverse datasets, and demonstrated substantial enhancements in image quality37. This suggests the potential to further increase the temporal resolution without compromising image contrast. Further, study from Liu et al.38 extends the advantages of using 4D MRI over 4DCT, not only for motion assessment but also for dose calculation. The creation of synthetic 4DCT based on 4DMRI enables direct dose calculation for treatment planning and also shows potential for online adaptation with MRI-linac.
In conclusion, while 4D-CT may demonstrate superior numerical accuracy in phantom studies, the proposed high temporal resolution 4D-MRI demonstrates sub-millimetre geometric accuracy comparable to that of 4D-CT. These findings suggest that the 4D-MRI technique is a viable option for characterizing motion and generating phase-dependent internal target volumes within the realm of radiotherapy.
Data availability
The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.
Abbreviations
- 4D-CT:
-
Four-dimensional computed tomography
- 4D-MRI:
-
Four-dimensional magnetic resonance imaging
- DSC:
-
Dice similarity coefficient
- HD:
-
Hausdorff distance
- GT:
-
Ground truth
- Vmeasured :
-
Measured volume
- Vref :
-
Reference volume
- ExIn:
-
Respiratory phase from exhaling to inhaling
- InEx:
-
Respiratory phase from inhaling to exhaling
- SBRT:
-
Stereotactic body radiation therapy
References
Low, D. A. et al. A method for the reconstruction of four-dimensional synchronized CT scans acquired during free breathing. Med. Phys. 30, 1254–1263 (2003).
Xi, M. et al. Defining internal target volume (ITV) for hepatocellular carcinoma using four-dimensional CT. Radiother. Oncol. 84, 272–278 (2007).
Francis, I. R. et al. Multidetector CT of the liver and hepatic neoplasms: Effect of multiphasic imaging on tumor conspicuity and vascular enhancement. AJR Am. J. Roentgenol. 180, 1217–1224 (2003).
Keall, P. J. et al. The management of respiratory motion in radiation oncology report of AAPM Task Group 76. Med. Phys. 33, 3874–3900 (2006).
Han, F. et al. Respiratory motion-resolved, self-gated 4D-MRI using rotating Cartesian k-space (ROCK). Med. Phys. 44, 1359–1368 (2017).
Paganelli, C., Summers, P., Bellomi, M., Baroni, G. & Riboldi, M. Liver 4DMRI: A retrospective image-based sorting method. Med. Phys. 42, 4814–4821 (2015).
Stemkens, B., Paulson, E. S. & Tijssen, R. H. N. Nuts and bolts of 4D-MRI for radiotherapy. Phys. Med. Biol. 63, 21TR01 (2018).
Mickevicius, N. J. & Paulson, E. S. Investigation of undersampling and reconstruction algorithm dependence on respiratory correlated 4D-MRI for online MR-guided radiation therapy. Phys. Med. Biol. 62, 2910–2921 (2017).
Freedman, J. N. et al. T2-Weighted 4D magnetic resonance imaging for application in magnetic resonance–guided radiotherapy treatment planning. Investig. Radiol. 52, 563–573 (2017).
Blackall, J. M. et al. MRI-based measurements of respiratory motion variability and assessment of imaging strategies for radiotherapy planning. Phys. Med. Biol. 51, 4147–4169 (2006).
Tokuda, J. et al. Adaptive 4D MR imaging using navigator-based respiratory signal for MRI-guided therapy. Magn. Reson. Med. 59, 1051–1061 (2008).
Van den Bos, I. C. et al. MR imaging of hepatocellular carcinoma: Relationship between lesion size and imaging findings, including signal intensity and dynamic enhancement patterns. J. Magn. Reson. Imaging 26, 1548–1555 (2007).
Murakami, T., Sofue, K. & Hori, M. Diagnosis of hepatocellular carcinoma using Gd-EOB-DTPA MR imaging. Magn. Reson. Med. Sci. 21, 168–181 (2022).
Vogl, T. J. et al. Liver tumors: Comparison of MR imaging with Gd-EOB-DTPA and Gd-DTPA. Radiology 200, 59–67 (1996).
Jaspan, O. N., Fleysher, R. & Lipton, M. L. Compressed sensing MRI: A review of the clinical literature. Br. J. Radiol. 88, 20150487 (2015).
Usman, M. et al. Motion corrected compressed sensing for free-breathing dynamic cardiac MRI. Magn. Reson. Med. 70, 504–516 (2013).
Oar, A. et al. Comparison of four dimensional computed tomography and magnetic resonance imaging in abdominal radiotherapy planning. Phys. Imaging Radiat. Oncol. 7, 70–75 (2018).
Yuan, J. et al. A fast volumetric 4D-MRI with sub-second frame rate for abdominal motion monitoring and characterization in MRI-guided radiotherapy. Quant. Imaging Med. Surg. 9, 1303 (2019).
Yue, Y. et al. Geometric validation of self-gating k-space-sorted 4D-MRI vs 4D-CT using a respiratory motion phantom. Med. Phys. 42, 5787–5797 (2015).
Akino, Y. et al. Evaluation of the accuracy of the CyberKnife synchrony™ respiratory tracking system using a plastic scintillator. Med. Phys. 45, 3506–3515 (2018).
Al-Mayah, A. et al. Biomechanical-based image registration for head and neck radiation treatment. Phys. Med. Biol. 55, 6491–6500 (2010).
Huttenlocher, D. P., Klanderman, G. A. & Rucklidge, W. J. Comparing images using the Hausdorff distance. IEEE Trans. Pattern Anal. Mach. Intell. 15, 850–863 (1993).
Feng, L. et al. XD-GRASP: Golden-angle radial MRI with reconstruction of extra motion-state dimensions using compressed sensing. Magn. Reson. Med. 75, 775–788 (2016).
Winkelmann, S., Schaeffter, T., Koehler, T., Eggers, H. & Doessel, O. An optimal radial profile order based on the Golden Ratio for time-resolved MRI. IEEE Trans. Med. Imaging 26, 68–76 (2007).
Biederer, J. et al. 4D-Imaging of the lung: Reproducibility of lesion size and displacement on helical CT, MRI, and cone beam CT in a ventilated ex vivo system. Int. J. Radiat. Oncol. Biol. Phys. 73, 919–926 (2009).
Mahadevan, A. et al. Stereotactic Body Radiotherapy (SBRT) for liver metastasis–clinical outcomes from the international multi-institutional RSSearch® Patient Regstry. Radiat. Oncol. 13(1), 1–11 (2018).
Apisarnthanarax, S. et al. External beam radiation therapy for primary liver cancers: An ASTRO clinical practice guideline. Pract. Radiat. Oncol. 12(1), 28–51 (2022).
Kuo, H.-T. et al. Impact of tumor size on outcome after stereotactic body radiation therapy for inoperable hepatocellular carcinoma. Medicine 96(50), e9249 (2017).
Hall, W. A. et al. Initial clinical experience of stereotactic body radiation therapy (SBRT) for liver metastases, primary liver malignancy, and pancreatic abdominal nodal recurrence with 4D-MRI based online adaptation and real-time MRI monitoring using a 1.5 Tesla MR-Linac. Int. J. Radiat. Oncol. Biol. Phys. 105(1), S90 (2019).
Shiani, A., Narayanan, S., Pena, L. & Friedman, M. The role of diagnosis and treatment of underlying liver disease for the prognosis of primary liver cancer. Cancer Control 24, 1073274817729240 (2017).
Feng, M. et al. Individualized adaptive stereotactic body radiotherapy for liver tumors in patients at high risk for liver damage: A phase 2 clinical trial. JAMA Oncol. 4, 40–47 (2018).
Tsujita, Y. et al. Prediction of post-hepatectomy liver failure using gadoxetic acid-enhanced magnetic resonance imaging for hepatocellular carcinoma with portal vein invasion. Eur. J. Radiol. 130, 109189 (2020).
Li, G. et al. Direct comparison of respiration-correlated four-dimensional magnetic resonance imaging reconstructed using concurrent internal navigator and external bellows. Int. J. Radiat. Oncol. Biol. Phys. 97, 596–605 (2017).
van de Lindt, T. et al. A self-sorting coronal 4D-MRI method for daily image guidance of liver lesions on an MR-LINAC. Int. J. Radiat. Oncol. Biol. Phys. 102, 875–884 (2018).
Liu, Y. et al. Accuracy of respiratory motion measurement of 4D-MRI: A comparison between cine and sequential acquisition. Med. Phys. 43, 179–187 (2016).
Park, S. et al. Autoencoder-inspired convolutional network-based super-resolution method in MRI. IEEE J. Transl. Eng. Health Med. 9, 1–13 (2021).
Tamada, D. Review: Noise and Artifact Reduction for MRI Using Deep Learning. arXiv preprint arXiv:2002.12889 (2020).
Liu, Y. et al. MRI-based treatment planning for liver stereotactic body radiotherapy: Validation of a deep learning-based synthetic CT generation method. Br. J. Radiol. 92, 20190067 (2019).
Funding
This work was partly supported by the Grant-in-Aid for Scientific Research of the Japan Society for the Promotion of Science (grant numbers 21K07620, 22H03026 and 19K17235).
Author information
Authors and Affiliations
Contributions
T.W. wrote the main manuscript text and prepared figures and tables. K.S. is the corresponding author, gave advices through the whole study and revised the main manuscript, figures and tables. R.Shimada. participated in the imaging acquisition with MRI scanner, prepared Fig. 2a and revised the manuscript. T.I. gave the advices for the imaging evaluation. R.Y. participated in the preparation of motion phantom. M.M. participated in the imaging acquisition with CT scanner. R.Sasaki., T.M. reviewed the manuscript and gave advices.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Wang, T., Sofue, K., Shimada, R. et al. Comparative study of sub-second temporal resolution 4D-MRI and 4D-CT for target motion assessment in a phantom model. Sci Rep 13, 15685 (2023). https://doi.org/10.1038/s41598-023-42773-z