Abstract
Pulmonary hypertension is a hemodynamic disorder defined by an abnormal elevation of pulmonary artery pressure (PAP). Current options for measuring PAP are limited in clinical practice. The aim of this study was to evaluate if electrical impedance tomography (EIT), a radiation-free and non-invasive monitoring technique, can be used for the continuous, unsupervised and safe monitoring of PAP. In 30 healthy volunteers we induced gradual increases in systolic PAP (SPAP) by exposure to normobaric hypoxemia. At various stages of the protocol, the SPAP of the subjects was estimated by transthoracic echocardiography. In parallel, in the pulmonary vasculature, pulse wave velocity was estimated by EIT and calibrated to pressure units. Within-cohort agreement between both methods on SPAP estimation was assessed through Bland–Altman analysis and at subject level, with Pearson’s correlation coefficient. There was good agreement between the two methods (inter-method difference not significant (P > 0.05), bias ± standard deviation of − 0.1 ± 4.5 mmHg) independently of the degree of PAP, from baseline oxygen saturation levels to profound hypoxemia. At subject level, the median per-subject agreement was 0.7 ± 3.8 mmHg and Pearson’s correlation coefficient 0.87 (P < 0.05). Our results demonstrate the feasibility of accurately assessing changes in SPAP by EIT in healthy volunteers. If confirmed in a patient population, the non-invasive and unsupervised day-to-day monitoring of SPAP could facilitate the clinical management of patients with pulmonary hypertension.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Introduction
Pulmonary hypertension (PH) is a pathophysiological condition characterized by a persistent and abnormal elevation of mean pulmonary artery pressure (PAP) above 25 mmHg, as assessed by right heart catheterization1. Chronic PH, with left heart disease as its most common cause, affects at least 0.3% of the general population and more than 60% of patients with moderate or severe heart failure2,3. The long-term follow-up of patients with PH relies in part on spot PAP measurements performed during clinic visits at intervals of several months1. However, assessing changes in PAP on a daily basis has shown to improve outcomes in patients with left heart failure equipped with an implantable monitor4, as increases in intracardiac pressures occur days or weeks before the onset of symptoms5. The availability of day-to-day trends in PAP for the clinician allows anticipating worsening conditions, thereby significantly reducing hospitalization rates4. However, while implantable monitors provide valuable insights in the patient’s response to therapy and allow earlier adaptation of treatment, they remain a highly invasive solution. Transthoracic echocardiography (TTE)—although non-invasive—is incompatible with daily unsupervised monitoring because of its dependency on a trained operator to perform the measurement and its inapplicability in 20–50% of patients due to unmeasurable tricuspid regurgitation6,7. A non-invasive solution able to provide PAP estimates on a day-to-day basis without supervision does currently not exist.
In this context, we have recently proposed a non-invasive and unsupervised method for the continuous monitoring of PAP8 and demonstrated its feasibility in a model-based simulation9. Our technique is based on electrical impedance tomography (EIT), a non-invasive and radiation-free medical imaging method allowing continuous monitoring at the bedside10. The main strength of EIT resides in its ability to monitor changes in the intra-thoracic distribution of electrical impedance with a high temporal resolution. Thus, impedance changes induced by respiratory and cardiovascular activity (e.g. changes in air volume in the lungs or in blood volume in the heart and arteries) can be monitored in real-time. In particular, changes in the pulsatility of the pulmonary arteries have shown to produce variations in EIT images in the pulmonary region, allowing assessment of PAP-related information from EIT11. We present hereafter the first experimental evidence of our method in a preclinical controlled hypoxemia study in 30 healthy participants.
Methods
The study was conducted in accordance with the amended Declaration of Helsinki. The protocol was approved by the Ethical Committee of the Canton of Bern (KEK), Switzerland (KEK Application 200/12), and registered at ClinicalTrials.gov (NCT02969486). All participants provided written informed consent.
Measurement principle
The velocity at which a pressure wave travels along an artery—the pulse wave velocity (PWV)—is linked to the elastic properties of the arterial wall12. As the arterial wall stiffens, the PWV increases and a larger blood pressure pulse is required to expand the wall. Thus, PWV and blood pressure are intrinsically linked. The PWV between two locations separated by a distance D along an arterial pathway can be calculated as PWV = D/PTT, where PTT is the pulse transit time. For blood pressure-related applications, estimating D—which is considered constant—is unnecessary as a calibration function f(·) is eventually required for converting the estimated PWV values to pressure units (mmHg). Therefore, monitoring changes in PAP using the PWV directly relies on monitoring changes in PTT: PAP = f(PTT). In our approach, the first arterial location is the pulmonary valve, and the second one a distal location in the lungs. As further explained hereafter, we use EIT to detect the arrival of the pressure pulse in the lungs. However, detecting in a non-invasive way the time of opening of the pulmonary valve is particularly complex. This is why in practice the PTT is often approximated by the pulse arrival time, which uses the R-wave peak of the electrocardiogram (ECG) as reference time instead of the opening of the valve13. In the present study, an ECG signal was thus measured along with the EIT data, and the pulse arrival time was used as a surrogate measure of the PTT.
Participants and experimental protocol
We enrolled thirty healthy participants in our study. Exclusion criteria were age below eighteen, known heart or pulmonary disease, pacemaker or implantable cardioverter defibrillator, pregnancy or lactation.
In order to induce PAP variations, the subjects breathed nitrogen-enriched air through a mask connected to an altitude simulator (AltiTrainer, SMTEC, Switzerland). Normobaric hypoxemia causes vasoconstriction in the pulmonary arterial tree and therefore induces elevations of PAP. The peripheral capillary oxygen saturation (SpO2) of the subjects was continuously monitored using the Radical-7 device (Masimo, USA), whereas their systolic pulmonary artery pressure (SPAP) was measured periodically by TTE. Each subject was lying in supine position and underwent seven 2-min EIT and ECG measurements at various oxygen saturation levels: two measurements in normoxemia (~ 540 m, SpO2 ≥ 95%), two measurements in mild hypoxemia (~ 4200 m, 90% > SpO2 ≥ 80%), two measurements in profound hypoxemia (~ 5400 m, 80% > SpO2 ≥ 65%), and at the end again one baseline measurement in normoxemia. Immediately after each EIT measurement, the subjects were asked to turn on their left side to measure their SPAP by TTE.
The echocardiographic measurements were performed with a real-time, phased-array sector scanner (Acuson Sequoia 512, Siemens, Germany) with an integrated color Doppler system and transducers containing crystal sets for 2-dimensional imaging (5.0 MHz with second harmonic technology) and for continuous-wave Doppler recording (2.5 MHz). After localization of tricuspid regurgitation via Doppler color flow imaging, the maximal tricuspid regurgitation velocity vmax was measured by means of continuous-wave Doppler ultrasound. The right atrioventricular pressure gradient ΔP was estimated using the modified Bernoulli equation (ΔP = 4 vmax2), as previously validated by our group in hypoxemia against invasive measurements14. The SPAP was estimated as ΔP + RAP, where RAP (right atrial pressure) was assumed to be 3 mmHg15. To reduce the impact of TTE’s imprecision16, each reported SPAP value is the average of at least three consecutive measurements. The maximal deviation among averaged values was verified to be systematically ≤ 4 mmHg. The EIT data were acquired using the 16-electrode Goe-MF II device (CareFusion, Germany) in combination with 16 gel electrodes (Ambu BR-50-K, Ambu, Denmark) placed around the participants’ thorax in a transversal plane at the level of the fifth intercostal space. An ECG was synchronously recorded using the ECG100C module (Biopac Systems, Inc., USA) and fed into the auxiliary input of the EIT device. The EIT data were acquired using an injection frequency of 100 kHz and a frame rate of 25 per second. The ECG signal was sampled at 325 Hz. The EIT data were reconstructed into images of 64 × 64 pixels with Matlab (MathWorks, Natick, USA) using the EIDORS toolbox17. The reconstruction was performed using the 2.5D human thorax model of EIDORS and the GREIT algorithm with the recommended settings18.
Estimation of the pulmonary PTT
The estimation of the pulmonary PTT for each 2-min EIT recording was fully automated and is illustrated in Fig. 1. As a first step, the EIT time signal at each pixel was high-pass filtered (0.65 Hz) using zero-phase filtering. Each 2-min EIT signal at each pixel was then averaged to one representative pulse through ECG-gated ensemble averaging19. Then, a pulmonary region of interest (ROI), defined as those pixels depicting a lung-like behavior and a significant pulsatility, was automatically segmented from the EIT images. To do so, a similar approach to the one proposed previously was implemented9,20. The pulmonary ROI was defined as those pixels showing coherent phase with the average cardiac-related EIT signal (average signal over all pixels). The ROI was then limited to pixels showing a pulsatile amplitude exceeding a threshold equal to the first quartile of the distribution of amplitude values. The pulmonary PTT was estimated at each pixel in the ROI from the ensemble average pulse using the maximum of its first time derivative. Possible outliers were automatically rejected using the median absolute deviation method21. The final pulmonary PTT was obtained by averaging all non-rejected PTT estimates.
EIT-based SPAP estimation
In order to translate our PTT estimates into actual SPAP values, we made use of a calibration function f(·) linking both quantities as SPAP = f(PTT). Although the function f(·) is typically an inverse exponential22,23, monitoring relatively small PAP changes around a certain baseline value allows to locally approximate f(·) by a simple affine function: SPAP = α·PTT + β24. In this study we considered the practical use case where only the β coefficient is adjusted for each subject, while the α coefficient remains the same for all subjects. The value of the α coefficient was obtained through linear regression of the PTT values onto their corresponding SPAP values at a cohort level, leading to an α value of − 710 mmHg/s. Being a simple offset, the β coefficient does not require inducing PAP variations to be calibrated and was therefore adjusted for each subject using the baseline measurements in normoxemia only, as β = average (SPAP − α·PTT). We also investigated the calibration scenario where the α coefficient is adjusted at a subject level; although impractical as it requires inducing PAP variations in the patient, this analysis allows quantifying the best achievable agreement using our proposed approach.
Statistical analysis
The agreement between the estimated EIT-derived SPAP values (SPAPEIT) and the TTE-derived SPAP values (SPAPTTE) at a cohort-wise level was evaluated via Bland–Altman analysis25. Using SPAPTTE as reference, the accuracy of SPAPEIT was evaluated using the statistical bias (average value of the differences between both methods), and its precision was evaluated using the standard deviation (SD) of the differences. These accuracy and precision assessments were also performed after grouping all measurements into three groups (normoxemia, mild hypoxemia, and profound hypoxemia). Paired-sample t-tests were used to determine if the difference between both methods was significant (at the 5% significance level). The agreement of SPAPEIT with SPAPTTE was also evaluated at a subject-wise level. Finally, the ability of EIT in tracking SPAP changes in the individual participant was evaluated using Pearson’s correlation coefficient (5% significance level). The distribution of the subject-wise performance values was reported using the median, the first (Q1) and third (Q3) quartiles, and the range (minimal and maximal values).
Results
Three volunteers had no measurable tricuspid regurgitation—needed for the estimation of SPAPTTE—and in three subjects the EIT signals were non-interpretable for technical reasons, leaving us with a total of 24 subjects. In two participants, the recording of the ECG signal during the first measurement in normoxemia failed, leaving us with 6 measurements instead of 7 for these two subjects. Thus, a total of 166 paired TTE-EIT data points were obtained (70 in normoxemia, 48 in mild hypoxemia, and 48 in profound hypoxemia). SPAPTTE was significantly increased (P < 0.001) during mild (25.1 ± 4.3 mmHg) and profound hypoxemia (32.0 ± 5.2 mmHg) as compared to baseline normoxemic values (19.5 ± 2.9 mmHg). The characteristics of the study participants are presented in Table 1. None of the participants suffered from PH, nor any complication during the tests. Systemic blood pressure and cardiac function were within the normal range in all participants.
Figure 2 (upper left) depicts the Bland–Altman plot for the entire cohort. Each SpO2 level (normoxemia, mild hypoxemia, and profound hypoxemia) is represented by a different symbol. The solid line shows the cohort-wise accuracy (bias) of − 0.1 mmHg, and the two dashed lines encompass the 95% limits of agreement (bias ± 1.96 SD), thus depicting a cohort-wise precision (SD) of 4.5 mmHg. Bland–Altman plots for each SpO2 level are shown in the lower panel of Fig. 2. Finally, the upper right panel depicts the correlation plot between SPAPEIT and SPAPTTE, with a cohort-wise Pearson’s correlation coefficient of 0.76 (P \(\ll\) 0.001).
Table 2 reports the cohort-wise agreement between SPAPEIT and SPAPTTE at various levels of oxygen saturation. In all cases, the null hypothesis that the paired differences between both methods come from a normal distribution with mean equal to zero could not be rejected at the 5% significance level.
Table 3 shows the agreement and correlation between SPAPEIT and SPAPTTE for all participants at the subject level. The distributions of all values are depicted using the three quartiles (Q1, median, and Q3), and the range (minimal and maximal values). Thus, a median per-subject difference of 0.7 ± 3.8 mmHg (bias ± SD) was found between both methods, and a median Pearson’s correlation coefficient of 0.87. The correlation was found to be significant at the 5% significance level for all participants but four, three of which with P < 0.10 and one, with the lowest per-subject correlation of 0.63, with P = 0.12.
In order to evaluate the best achievable agreement between SPAPEIT and SPAPTTE, the calibration scenario where the α coefficient is adjusted at a subject level was also investigated. The cohort-wise accuracy (bias) and precision (SD) decreased to 0.0 and 2.8 mmHg, respectively, and the cohort-wise Pearson’s correlation coefficient increased to 0.90 (P \(\ll\) 0.001). At a subject level, a median per-subject difference of 0.0 ± 2.6 mmHg (bias ± SD) was found between SPAPEIT and SPAPTTE. The subject-wise Pearson’s correlation coefficients being unaffected by the value of the α coefficient, remained unchanged.
Discussion
There is currently no non-invasive and unsupervised solution for the day-to-day monitoring of pulmonary artery pressure in patients with chronic pulmonary hypertension. This study presents experimental evidence of the feasibility to monitor SPAP non-invasively using EIT in 24 healthy subjects through the measurement of the pulmonary PTT. Changes in SPAP were induced by hypoxemic vasoconstriction and compared to SPAP values obtained from echocardiographic measurements. Results demonstrate no significant difference between both methods, irrespective of the level of SPAP, from normoxemia to profound hypoxemia. This agreement was also verified at a subject level, along with strong and significant subject-wise Pearson’s correlation coefficients, thereby confirming the ability of our proposed approach in tracking intra-subject changes in PAP via EIT. When adjusting the α calibration coefficient at a subject level, a general ~ 35% precision improvement was observed; however, given the impractical applicability of such a calibration procedure, these results only serve to illustrate the best achievable agreement obtainable with our method.
While our results demonstrate excellent accuracy (bias of − 0.1 mmHg), their moderate precision (standard deviation of 4.5 mmHg) may lead to occasional large differences in individual measurements. A part of this imprecision may be attributed to our use of TTE as reference16, highlighting the need for further comparison studies against an invasive reference. However, because EIT measures continuously and without supervision, its precision can easily be improved by repeating and averaging measurements, as averaging \(N\) measurements decreases the standard deviation of the error by a factor \(\sqrt{N}\)26. For instance, averaging 3 measurements already brings the standard deviation down from 4.5 to 2.6 mmHg. Considering 2-min measurements as is the case in our study, this suggests that wearing an EIT system for only a few minutes is sufficient to obtain averages with higher precision.
Cardiovascular applications of EIT have received increasing attention over the last decade, from investigations on cardiac output to pulmonary perfusion27,28,29,30. The use of EIT for monitoring patients with PH has been proposed previously by Smit et al.11. However, although they, with high accuracy, could discriminate between patients and healthy subjects, no correlation was found between EIT-based features and the severity of the disease. Furthermore, as all their patients suffered from advanced disease, they could not extrapolate their results to early stages. In contrast, our PWV-based approach for PAP monitoring is thought to be particularly well-suited for detecting early changes in pulmonary hemodynamics. This is due to the direct link between PWV and arterial distensibility12, which is an even stronger predictor of survival than PAP itself31,32. In addition, the inverse exponential nature of the PTT-PAP relationship suggests higher measurement sensitivity at lower pressures, from mild to moderate PH. Conversely, our approach may suffer from lower sensitivity at severe hypertensive stages. From a model-based study9, this decrease in the sensitivity of our approach may begin at mean PAP values of ~ 40–50 mmHg (SPAP ~ 60–75 mmHg), which we did not investigate in the present study. This is concordant with findings from an intravascular ultrasound study23, where arterial distensibility compared to PAP progressively reached a plateau starting from a mean PAP of ~ 40 mmHg, as the arteries were approaching their elastic limit. A direct consequence of the non-linear nature of the PTT-PAP relationship is an expectable decrease in the magnitude of the α calibration coefficient as PAP increases, suggesting that applying our proposed approach to a large patient population might require prior training of the α calibration coefficient over a large range of PAP values covering the physiological range.
When compared to the only currently available non-invasive PAP measuring solution that is TTE, EIT has several advantages, such as its lower cost, and its ability to provide unsupervised monitoring. While TTE requires an experienced echocardiographer for accurate measurement of SPAP, no particular training other than attaching an electrode belt to a patient is required for EIT. Furthermore, the applicability of EIT exceeds that of TTE, for which 20–50% of patients have no measurable tricuspid regurgitation6,7. Lastly, because EIT provides continuous measurements, our method has the potential of detecting minute variations in pulmonary hemodynamics, otherwise missed by intermittent methods such as TTE. This could prove particularly useful in contexts of care such as intensive care units, where the use of right heart catheterization has been sharply reduced over the last decades after several randomized controlled trials failed to demonstrate any decrease in morbidity and mortality following its use in critically ill and hemodynamically unstable patients33,34,35,36. In this context, EIT could be used to carry on the monitoring of PAP after earlier removal of the catheter, thereby reducing the risks associated with its maintenance. In patients lacking invasive monitoring, EIT could provide valuable information by measuring continuous PAP trends.
A potentially limiting aspect of our approach is that it does not directly estimate absolute values of PAP, but absolute changes of PAP around an arbitrary baseline value. Setting this baseline value requires a calibration, such as a TTE-derived measurement of the patient’s SPAP during a visit in the clinics. In other contexts such as the aforementioned intensive care settings, the EIT device could simply be calibrated using the invasive PAP before catheter removal.
Another potential limitation of our approach is the use of the pulse arrival time as a surrogate of the PTT, which may make it sensitive to factors affecting the pre-ejection period, such as ventricular contractility changes13. Also, EIT belt position changes (e.g. rotations or vertical shifts) may affect the accuracy of our approach, mainly at severe hypertensive levels37. The use of 3D EIT38, shirt-embedded EIT belts39, or torso shape detection methods40 are various possible strategies that should help minimize the influence of such belt displacements. The well-known limited signal-to-noise ratio of cardiovascular-related EIT signals10 is another aspect that may occasionally challenge our approach, as observed in the three subjects rejected for this reason prior to analysis in our study (10% failure rate). Also, considering one single global PTT value in our approach for the entire lung vasculature may lead to inaccuracies in cases of heterogeneous PTT distributions, e.g. in case of ventilation/perfusion mismatch. In such cases, an analysis based on subregions of interest should be performed, but is out of the scope of the present study, where healthy subjects were investigated. Lastly, the performance of our approach needs to be assessed versus the invasive gold standard, and on a patient population, for a formal validation. Preliminary results compared to the invasive method proved promising in neonatal lambs41.
In conclusion, the results of our study suggest that changes in PAP may be reliably monitored by a novel, non-invasive, continuous and unsupervised approach based on EIT. If confirmed in a patient population, these findings may open new perspectives for the day-to-day follow-up of patients with pulmonary hypertension, offering a better control of early disease progression and symptoms. Our method may also be particularly useful in intensive care settings, where the benefit-to-risk ratio associated with an invasive monitoring approach often dissuade clinicians.
References
Galiè, N. et al. 2015 ESC/ERS guidelines for the diagnosis and treatment of pulmonary hypertension. Eur. Heart J. 37, 67–119 (2016).
Strange, G. et al. Pulmonary hypertension: prevalence and mortality in the Armadale echocardiography cohort. Heart 98, 1805–1811 (2012).
Ghio, S. et al. Independent and additive prognostic value of right ventricular systolic function and pulmonary artery pressure in patients with chronic heart failure. J. Am. Coll. Cardiol. 37, 183–188 (2001).
Abraham, W. T. et al. Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trial. Lancet 377, 658–666 (2011).
Zile, M. R. et al. Transition from chronic compensated to acute decompensated heart failure: pathophysiological insights obtained from continuous monitoring of intracardiac pressures. Circulation 118, 1433–1441 (2008).
Currie, P. J. et al. Continuous wave Doppler determination of right ventricular pressure: a simultaneous Doppler-catheterization study in 127 patients. J. Am. Coll. Cardiol. 6, 750–756 (1985).
De Meester, P., Van De Bruaene, A., Herijgers, P., Voigt, J.-U. & Budts, W. Tricuspid valve regurgitation: prevalence and relationship with different types of heart disease. Acta Cardiol. 67, 549–556 (2012).
Adler, A., Brunner, J. X., Ferrario, D., Solà, J. & Proença, M. Method and apparatus for the non-invasive measurement of pulse transit times (PTT). US9999357B2 (2018).
Proença, M., Braun, F., Solà, J., Thiran, J.-P. & Lemay, M. Noninvasive pulmonary artery pressure monitoring by EIT: a model-based feasibility study. Med. Biol. Eng. Comput. 55, 949–963 (2016).
Frerichs, I. et al. Chest electrical impedance tomography examination, data analysis, terminology, clinical use and recommendations: consensus statement of the TRanslational EIT developmeNt stuDy group. Thorax 72, 83–93 (2017).
Smit, H. J., Vonk-Noordegraaf, A., Boonstra, A., de Vries, P. M. & Postmus, P. E. Assessment of the pulmonary volume pulse in idiopathic pulmonary arterial hypertension by means of electrical impedance tomography. Respiration 73, 597–602 (2006).
Nichols, W. W., O’Rourke, M. F. & Vlachopoulos, C. McDonald’s Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles (Hodder Arnold, London, 2011).
Mukkamala, R. et al. Towards ubiquitous blood pressure monitoring via pulse transit time: theory and practice. IEEE Trans. Bio.-Med. Eng. 62, 1879–1901 (2015).
Allemann, Y. et al. Echocardiographic and invasive measurements of pulmonary artery pressure correlate closely at high altitude. Am. J. Physiol. Heart C 279, H2013–H2016 (2000).
Rudski, L. G. et al. Guidelines for the echocardiographic assessment of the right heart in adults: a report from the American Society of Echocardiography: endorsed by the European Association of Echocardiography, a registered branch of the European Society of Cardiology, and the Canadian Society of Echocardiography. J. Am. Soc. Echocardiogr. 23, 685–713 (2010).
D’Alto, M. et al. Accuracy and precision of echocardiography versus right heart catheterization for the assessment of pulmonary hypertension. Int. J. Cardiol. 168, 4058–4062 (2013).
Adler, A. & Lionheart, W. R. B. Uses and abuses of EIDORS: an extensible software base for EIT. Physiol. Meas. 27, S25–S42 (2006).
Adler, A. et al. GREIT: a unified approach to 2D linear EIT reconstruction of lung images. Physiol. Meas. 30, S35–S55 (2009).
Rangayyan, R. M. Biomedical Signal Analysis: A Case-Study Approach (Wiley, Hoboken, 2002).
Proença, M. et al. Noninvasive monitoring of pulmonary artery pressure from timing information by EIT: experimental evaluation during induced hypoxia. Physiol. Meas. 37, 713–726 (2016).
Leys, C., Ley, C., Klein, O., Bernard, P. & Licata, L. Detecting outliers: do not use standard deviation around the mean, use absolute deviation around the median. J. Exp. Soc. Psychol. 49, 764–766 (2013).
Sanz, J. et al. Evaluation of pulmonary artery stiffness in pulmonary hypertension with cardiac magnetic resonance. J. Am. Coll. Cardiol. Imaging 2, 286–295 (2009).
Lau, E. M. et al. Abnormal pulmonary artery stiffness in pulmonary arterial hypertension: in vivo study with intravascular ultrasound. PLoS ONE 7, e33331 (2012).
Proença, M. Non-invasive Hemodynamic Monitoring by Electrical Impedance Tomography (EPFL Swiss Federal Institute of Technology, Lausanne, 2017).
Bland, J. M. & Altman, D. G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 327, 307–310 (1986).
Riffenburgh, R. Statistics in Medicine (Elsevier, Amsterdam, 2005).
Braun, F. et al. Noninvasive measurement of stroke volume changes in critically ill patients by means of electrical impedance tomography. J. Clin. Monit. Comput. 34, 903–911 (2020).
Pikkemaat, R., Lundin, S., Stenqvist, O., Hilgers, R. D. & Leonhardt, S. Recent advances in and limitations of cardiac output monitoring by means of electrical impedance tomography. Anesth. Analg. 119, 76–83 (2014).
Borges, J. B. et al. Regional lung perfusion estimated by electrical impedance tomography in a piglet model of lung collapse. J. Appl. Physiol. 112, 225–236 (2012).
Nguyen, D. T., Jin, C., Thiagalingam, A. & McEwan, A. L. A review on electrical impedance tomography for pulmonary perfusion imaging. Physiol. Meas. 33, 695–706 (2012).
Lankhaar, J. W. et al. Pulmonary vascular resistance and compliance stay inversely related during treatment of pulmonary hypertension. Eur. Heart J. 29, 1688–1695 (2008).
Lau, E. M., Manes, A., Celermajer, D. S. & Galiè, N. Early detection of pulmonary vascular disease in pulmonary arterial hypertension: time to move forward. Eur. Heart J. 32, 2489–2498 (2011).
Connors, A. F. et al. The effectiveness of right heart catheterization in the initial care of critically ill patients. J. Am. Med. Assoc. 276, 889–897 (1996).
Harvey, S. et al. Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet 366, 472–477 (2005).
Richard, C. et al. Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial. J. Am. Med. Assoc. 290, 2713–2720 (2003).
Sandham, J. D. et al. A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N. Engl. J. Med. 348, 5–14 (2003).
Braun, F. et al. Limitations and challenges of EIT-based monitoring of stroke volume and pulmonary artery pressure. Physiol. Meas. 39, 014003 (2018).
Grychtol, B. et al. Thoracic EIT in 3D: experiences and recommendations. Physiol. Meas. 40, 074006 (2019).
Frerichs, I. et al. Multimodal remote chest monitoring system with wearable sensors: a validation study in healthy subjects. Physiol. Meas. 41, 015006 (2020).
de Gelidi, S. et al. Torso shape detection to improve lung monitoring. Physiol. Meas. 39, 074001 (2018).
Braun, F. et al. EIT measurement of pulmonary artery pressure in neonatal lambs. in 20th International Conference on Biomedical Applications of Electrical Impedance Tomography, London, UK, July 1-3, 2019. Proceedings (eds. Boyle, A., Aristovich, K., Witkowska-Wrobel, A. & Holder, D.) 33 (Medical Physics and Biomedical Engineering, 2019)
Funding
This work was supported in part by grants from the Swiss National Science Foundation (SNSF, no. 205321-153364) and the SNSF/Nano-Tera project OBESENSE (no. 20NA21-1430801).
Author information
Authors and Affiliations
Contributions
Study design: M.P., F.B., J.S., M.L., J.P.T., S.R. and E.R. Data acquisition: M.P., F.B., J.S., A.A., T.R., S.R. and E.R. Data processing: M.P. and F.B. Data analysis: M.P. and F.B. Data interpretation: M.P., F.B. and E.R. Figure preparation: M.P. Funding acquisition: M.P., J.S., M.L. and J.P.T. Writing, original draft: M.P. and E.R. Writing, critical review and editing: M.P., F.B., M.L., J.S., A.A., T.R., J.P.T., S.R., F.M. and E.R. All authors have read and approved the final version of the manuscript.
Corresponding author
Ethics declarations
Competing interests
Martin Proença, Fabian Braun, Mathieu Lemay and Josep Solà are with CSEM, the assignee of US patent 9,999,357 B2, which protects the approach presented here, and of which Martin Proença, Andy Adler and Josep Solà are inventors. Although CSEM has financial relationship with companies that might be interested by said approach, none of these companies were involved in any way in the present study. There are no financial or other relations that could lead to a conflict of interest. All other authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Proença, M., Braun, F., Lemay, M. et al. Non-invasive pulmonary artery pressure estimation by electrical impedance tomography in a controlled hypoxemia study in healthy subjects. Sci Rep 10, 21462 (2020). https://doi.org/10.1038/s41598-020-78535-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-020-78535-4
- Springer Nature Limited
This article is cited by
-
Absolute values of regional ventilation-perfusion mismatch in patients with ARDS monitored by electrical impedance tomography and the role of dead space and shunt compensation
Critical Care (2024)
-
Assessing Pulmonary Function Parameters Non-invasively by Electrical Bioimpedance Tomography
Journal of Medical and Biological Engineering (2024)