Abstract
Manipulating mechanism of particle’s motion has been extensively studied for the sample preparation in microfluidic applications including diagnostics, food industries, biological analyses and environmental monitoring. However, most of conventional methods need additional external forces such as electric field or pressure and complicated channel designs, which demand highly complex fabrication processes and operation strategies. In addition, these methods have inherent limitations of dilution or mixing during separation or preconcentration step, respectively, so that a number of studies have reported an efficient selective preconcentration process, i.e. conducting the separation and preconcentration simultaneously. In this work, a power-free spontaneous selective preconcentration method was suggested based on leveraging convective flow over diffusiophoresis near the water-absorbing nanoporous ion exchange medium, which was verified both by simulation and experiment. Especially, the velocity of the convective flow by an imbibition deviated from the original tendency of t−1/2 due to non-uniformly patterned nanoporous medium that has multiple cross-sectional areas. As a result, the direction of particle’s motion was controlled at one’s discretion, which led to the spontaneous selective preconcentration of particles having different diffusiophoretic constant. Also, design rule for maximizing the efficiency was recommended. Thus, this selective preconcentration method would play as a key mechanism for power-free lab on a chip applications.
Similar content being viewed by others
Introduction
Separation and preconcentration are the key sample preparation step in a wide range of chemical, environmental, and biological processes. Especially, those sample preparation technologies using distinct physical properties of particles or cells are particularly important in microfluidic applications including diagnostics1,2,3,4, food industries5,6,7, biological analyses8,9 and environmental monitoring10,11. However, these processes have mutual contradiction since aggressive separations inevitably dilute a sample and sample preconcentration mixes all components in the sample so that a number of studies aim to develop an efficient selective preconcentration process, i.e. conducting the separation and preconcentration simultaneously12,13,14. Previously developed selective preconcentration methods usually require external stimuli such as electric field14,15,16, magnetic field4,17,18, acoustic waves19,20,21, etc. However, these active selective preconcentration methods highly demand complex/expensive fabrication processes and chip design for the external forces, preventing the practical applications where energy input is of critical concern. This has resulted in the development of several passive method without any external forces, for example, capillary force22,23, pinched flow fractionation24,25,26, inertial focusing4,27,28, and inertia-elastic focusing29,30,31, etc. However, complicated microchannel design or specific chemicals are often required for their processes, which induces practical difficulty of real applications. In addition, external pump is still utilized in order for careful control of the input flow rate. It means that these methods are unsuitable to portable and cost-effective devices.
Recently, diffusiophoresis have been suggested as the alternative method of manipulation of particles without external forces32,33,34,35,36,37,38. Diffusiophoresis refers to the manipulation of particle’s motion with a concentration gradient induced either by two miscible liquids of different concentrations39,40,41,42,43 or by ion exchange44,45,46,47,48. Similar with our mechanism, two studies have reported about utilizing the balance between diffusiophoresis and convective flow for separating/preconcentrating colloid particles. Using the concentration gradient generated by injecting two different concentrations of electrolyte, Friedrich et al. and Ault et al. reported that colloidal particles could be preconcentrated and multiple particles could be separated, respectively36,49. However, these works should still need additional apparatus for the concentration gradient generation and external pressure-driven pump to induce convective flow.
Meanwhile, the concentration gradient induced by ion exchange is spontaneously generated if the ion exchange medium instantly meets the solution containing non-protonic cation. When the nanoporous medium is wetted with water, surface group on the medium is dissociated. The proton inside the medium would be exchanged with non-protonic cation dissolved in water through the Brownian motion50. It was reported that the ion concentration near the medium was unchanged because ion exchange process is 1:1 process51. However, due to the difference of diffusivity between proton and cation, natural ion depletion layer is generated near the nanoporous medium45,52,53. In the presence of concentration gradient near the medium, the electrical double layer of a charged particle is deformed and internal electric field is induced around the particle. Thus, the particle is effectively transported by the diffusiophoresis.
However, water permeance through the nanoporous medium in such diffusion-dominant environment should be considered since the medium should absorb the water to have the dissolved protons which will be exchanged with other cations53. Considering the water permeance through the nanoporous medium, the concentration profile near the medium has three types (ion depletion, ion accumulation and intermediate) depending on the permselectivity and the water-permeability of the medium53. Under the medium that has sufficiently low water permeance, a charged particle’s moving velocity (Up) is independently affected by two velocities as shown in Fig. 1(a): diffusiophoretic velocity (UDP) and convective velocity (Uμ). Thus Up is expressed by the sum of two flow velocities as
where
and
Here, diffusiophoretic constant, DDP is already a function of the zeta potential of a particle, viscosity of fluid, temperature, diffusivity of ions45,53,54 and the radius of a particle43, and ctot is the total ion concentration. It is reported that UDP is proportional to t−1/2 48. An is the cross-sectional area of a wetted nanoporous medium, Aμ is the cross-sectional area of a micro-channel, φp is the porosity of a nanoporous medium, S is the absorbing parameter53, Un is the velocity of wetting through nanoporous medium obtained by Darcy’s law, and t is the time. Uμ is derived from the flow continuity condition at the interface between the nanoporous medium and the micro-channel. Both of the velocities in Eq (2–3) are inversely proportional to the square root of t, so that the direction of the particle’s motion should be unidirectional. However, Uμ is able to be manipulated when nanoporous medium is non-uniformly patterned as shown in Fig. 1(b). Owing to the non-uniformity, An is no longer constant and becomes a function of t since the water is absorbed through the expanding pathway as shown in Fig. 1(b). Consequently, Uμ becomes a saturating function other than t−1/2 55,56,57,58,59, so that the direction of the particle’s motion is able to be leveraged. In this work, by adjusting the Uμ, manipulating motion of multiple particles was presented and the spontaneous selective preconcentration mechanisms were verified both by simulation and experiment. Moreover, our method used simple straight and dead-end microchannel design and require no externally connected devices such as pump or electric power source so that it would be an effective mean to be developed as a portable analytical device in resource-limited settings.
Results and Discussions
Mechanism of leveraging convective flow over diffusiophoresis by non-uniformly patterned nanoporous medium
With the non-uniformly patterned nanoporous ion exchange medium as depicted in Fig. 1(b), the flow velocity induced by the imbibition through the medium (Uμ) is numerically estimated in Fig. 2(a). The simulated flow velocity is calculated based on the Richard’s equation59 and more information on the simulation is available in Supplementary Note 1. Until the critical time (tc), the flow velocity follows the 1-dimensional Darcy’s law since the water is absorbed through the fixed An. After tc, however, the flow velocity deviates from the straight line and leads to the saturated velocity due to the increasing An with an expanding water-pathway57. In Fig. 2(a), diffusiophoretic constant of particle 1 was higher than that of particle 2, and the comparison among the velocities was UDP1 > UDP2 > Uμ, before the time of direction switching (ts), which meant that both particles would move from the nanoporous medium to the bulk. After ts, the comparison among the velocities was UDP1 > Uμ > UDP2, which meant that particle 1 would still move to the bulk but the particle 2 would reversely move to the nanoporous medium. Thus, the particles would be separated due to their different moving direction without external power.
Langevin dynamics (LD) simulation was conducted for confirming this separation. The force balance for each particle included the Brownian motion of a particle itself and the drag force from Uμ and UDP. Figure 2(b) showed the LD results with the black particles representing the particle 1 and the gray particles representing the particle 2. The critical time (tc) when saturated flow velocity started were set to be 10,000 seconds (2.8 hours). Before tc, both particles were pushed away from the ion exchange medium due to the diffusiophoretic force. During this period, particles near the ion exchange medium were depleted since the diffusiophoretic force were stronger near the medium. At the same time, particles near the reservoir were drawn towards the medium by the fluid flow. Due to these two forces acting simultaneously, particles were not only depleted from the medium but also preconcentrated at the depletion boundary marked as arrows in Fig. 2(b). These depletion boundaries extended proportional to t1/2 until tc and the particle 1 having the higher diffusiophoretic constant were depleted further than particle 2. Between tc and ts, the depletion boundaries of both types of particle gradually stopped developing because of the saturated Uμ. After ts (which was estimated to be around 18,000 seconds (5 hours) in the simulation), only particle 2 (gray) which had the lower diffusiophoretic constant switched their direction of motion towards the medium, while particle 1 (black) kept its direction toward reservoir, leading to a simultaneous separation and preconcentration, i.e. selective preconcentration. Details on the simulation are available in Supplementary Note 2 and see Supplementary Video 1.
Experimental demonstration of particles’ selective preconcentration
Aforementioned LD analysis was experimentally demonstrated as shown in Fig. 3. Nafion as a nanoporous medium was used because Nafion has high perm-selectivity with the low water permeance for the generation of ion depletion near the medium53. The microfluidic device made of polydimethyl-siloxane (PDMS) had a microchannel that was connected with non-uniformly patterned Nafion at the end of the channel and a reservoir at the other end (Fig. 3(a)). The microchannel was filled with 1 mM KCl electrolyte solution containing the carboxylate functionalized fluorescent particles (negatively charged and the diameters of 2 μm and 0.04 μm). Larger particle has higher diffusiophoretic constants60. In Fig. 3(b), the yellow particles (2 μm) with relatively high diffusiophoretic constant gradually moved to the bulk. However, the green particles (0.04 μm) with relatively low diffusiophoretic constant moved to the bulk until around 6 hr but after that time, they moved back towards the ion exchange medium as arrows indicated. Moreover, in the region of constant concentration (i.e. near reservoir), particles were forced only by convective flow so that particles were continuously provided from the reservoir, leading an efficient preconcentration. In Fig. 3(c), the intensity of each particle was plotted by analyzing the fluorescent signal. As time passed, the distance between the peaks of yellow particles and green particles increased (separation), and the intensity of each particle simultaneously increased (preconcentration). See Supplementary Video 2. The separation resolution (Rs) calculated by (peak to peak distance)/(average width of bands) led to 1.2161. This number indicates almost perfect separation (i.e. no overlap) under assumption that the distribution follows Gaussian distribution. In addition, the average pixel intensity of yellow particle reached up to 171.24 at 18 hr from 0.50 at 6 hr, which denotes that yellow particles were preconcentrated over 300-fold. Also, the average pixel intensity of green particle reached up to 171.28 at 18 hr from 1.83 at 6 hr, which denotes that yellow particles were preconcentrated over 90-fold. In this sense, our method is comparable to previous researches utilizing an external electric field to selective-preconcentrate multiple particles14,15,62 if only preconcentration factor was considered. This comparison was irrelevant if a short operation time is demanded.
Selective preconcentration of three types of particles was also demonstrated in Fig. 3(d) using the red particle of 0.2 μm which had intermediate diffusiophoretic constant between yellow particle and green particle (i.e. DDP_yellow > DDP_red > DDP_green). As expected, three types of particles were selectively preconcentrated in the order of diffusiophoretic constant value. See Supplementary Video 3.
Design rule for maximizing efficiency of selective preconcentration
As mentioned earlier, when the nanoporous medium was being wetted through the straight portion of nanoporous medium, the Uμ followed 1-dimensional Darcy’s law due to the constant An fixed as An,0. Then, An increased due to the expanding water-pathway would cause Uμ to be saturated. That is, Uμ was described as
and
Due to the saturated Uμ, it intersected with the UDP at the time of direction switching (ts) as
Here, tc is a function of L1 (the length of the straight region prior to the flow expanding pathway as shown in the inset of Fig. 4) according to the Darcy’s law as
Thus, the relationship between ts and L1 was obtained as
and indicated by a dashed line as shown in Fig. 4, which was experimentally verified by adjusting L1 (see gray circles). Using the microchip, ts was obtained by measuring the time when the particles (carboxylate 0.2 μm) switched their moving direction. The experimental values of ts were proportional to L12 until L1 = 700 μm, but were saturated at a certain constant for L1 longer than the critical length (Lc) due to the diffusion of ions from the reservoir. The diffusion from the reservoir mitigated the concentration gradient formed by the ion exchange, which was usually neglected since the length of microchannel (L2) was commonly assumed as infinitely long. Considering the diffusion from the bulk, the restoration of total concentration was simulated in Supplementary Note 3. The experimentally estimated time of limitation (tlim) when the particles changed their moving direction without the help of increasing An was around 14,000 seconds. That is, ts would be limited to tlim when L1 was longer than Lc, which meant that the effect of L2 became dominant over L1. Conclusively, L1 is the value for determining the time of starting separation and, thus, it should be chosen smaller than Lc, for maximizing the efficiency of selective preconcentration of particles.
Conclusions
Selective preconcentration plays an important role for sample preparation step in a wide range of biochemical microfluidic applications. However, conventional selective preconcentration methods usually require additional devices (or apparatus) for inducing external stimuli such as electric field or pressure. This has resulted in the necessity and development of power-free selective preconcentration mechanism. In this work, spontaneous selective preconcentration method was presented based on leveraging convective flow induced by imbibition through nanoporous medium over diffusiophoresis. While traditional imbibition vs. diffusiophoresis mechanism conveyed a unidirectional force field, we successfully demonstrated a bidirectional field utilizing non-uniformly patterned nanoporous medium. These mechanisms were verified both by simulation and experiment. Consequently, the selective preconcentration of two or three particles having different diffusiophoretic constant was demonstrated and a design rule was also suggested through a simple analysis for maximizing the efficiency of power-free selective preconcentration. Although these mechanisms have the limitation of slow processing in comparison to other methods using external fields, this method can be useful for a time-insensitive lab on a chip application such as environmental monitoring and food monitoring, etc. In order to overcome this drawback, one could employ a paper device that has faster imbibition. In addition to this, our method is unsuitable for recovering the selective preconcentrated sample due to the dead-end channel. However, we are expecting the applications of this work for the portable diagnosis devices where one can directly observe the selectively preconcentrated sample inside the channel.
Materials and Methods
As building blocks of each microchannel for filling samples and Nafion (Sigma Aldrich, USA), polydimethyl-siloxane (PDMS, Sylgard 184 silicone elastomer kit, Dow Corning, USA) was used as shown in Fig. 3(a). PDMS base and curing agent (Sylgard 184 silicone elastomer kit, Dow Corning, USA) were mixed in a ratio of 10:1 and degassed in a vacuum chamber for one hour. The mixed solution was poured onto a silicon wafer which had patterned microchannels and was cured in an oven for four hours at 75 °C.
The Nafion as a nanoporous medium was patterned on the glass slide following a previously reported surface patterning method with the prepared PDMS block for filling Nafion63,64. The Nafion is divided into two regions of a narrow Nafion (dimension: width 200 μm × depth 0.5 μm × length 500 μm) and a wide Nafion (dimension: width 600 μm × depth 0.5 μm × length 1,100 μm). On the Nafion-patterned glass slide, the prepared PDMS block for filling samples was irreversibly bonded to a designated position using plasma bonder (CuteMP, Femto Science, Korea) as shown in Fig. 3(a). The microchannel has the dimension of width 200 μm × depth 15 μm × length 5 mm and filled with a mixture of 1 mM KCl solution (Sigma Aldrich, USA) with the negatively charged fluorescent carboxylate particles (diameter = 2 μm, 0.2 μm, and 0.04 μm, Invitrogen, USA). The motions of fluorescent particles were captured by an inverted fluorescence microscope (IX53, Olympus, Japan) and the CellSens program (Olympus, Japan). The fluorescent intensity was analyzed by ImageJ.
References
Suresh, S. et al. Connections between single-cell biomechanics and human disease states: gastrointestinal cancer and malaria. Acta biomaterialia 1, 15–30 (2005).
Fan, R. et al. Integrated barcode chips for rapid, multiplexed analysis of proteins in microliter quantities of blood. Nature biotechnology 26, 1373 (2008).
Alshareef, M. et al. Separation of tumor cells with dielectrophoresis-based microfluidic chip. Biomicrofluidics 7, 011803 (2013).
Ozkumur, E. et al. Inertial focusing for tumor antigen–dependent and–independent sorting of rare circulating tumor cells. Science translational. medicine 5, 179ra147–179ra147 (2013).
Law, W. S., Kubáň, P., Zhao, J. H., Li, S. F. Y. & Hauser, P. C. Determination of vitamin C and preservatives in beverages by conventional capillary electrophoresis and microchip electrophoresis with capacitively coupled contactless conductivity detection. Electrophoresis 26, 4648–4655 (2005).
Kubáň, P. & Hauser, P. C. Application of an external contactless conductivity detector for the analysis of beverages by microchip capillary electrophoresis. Electrophoresis 26, 3169–3178 (2005).
Skurtys, O. & Aguilera, J. Applications of microfluidic devices in food engineering. Food Biophysics 3, 1–15 (2008).
Toner, M. & Irimia, D. Blood-on-a-chip. Annu Rev Biomed Eng 7, 77–103, https://doi.org/10.1146/annurev.bioeng.7.011205.135108 (2005).
Gijs, M. A., Lacharme, F. & Lehmann, U. Microfluidic applications of magnetic particles for biological analysis and catalysis. Chemical reviews 110, 1518–1563 (2009).
Gardeniers, J. G. & van den Berg, A. Lab-on-a-chip systems for biomedical and environmental monitoring. Analytical and bioanalytical chemistry 378, 1700–1703 (2004).
Marle, L. & Greenway, G. M. Microfluidic devices for environmental monitoring. TrAC Trends in Analytical Chemistry 24, 795–802 (2005).
Tong, A., Akama, Y. & Tanaka, S. Selective preconcentration of Au (III), Pt (IV) and Pd (II) on silica gel modified with γ-aminopropyltriethoxysilane. Analytica chimica acta 230, 179–181 (1990).
Cole, L. J. & Kennedy, R. T. Selective preconcentration for capillary zone electrophoresis using protein G immunoaffinity capillary chromatography. Electrophoresis 16, 549–556 (1995).
Choi, J. et al. Selective preconcentration and online collection of charged molecules using ion concentration polarization. RSC Advances 5, 66178–66184, https://doi.org/10.1039/c5ra12639h (2015).
Cheow, L. F. & Han, J. Continuous signal enhancement for sensitive aptamer affinity probe electrophoresis assay using electrokinetic concentration. Analytical chemistry 83, 7086–7093 (2011).
Lee, H. et al.dCas9-mediated Nanoelectrokinetic Direct Detection of Target Gene for Liquid Biopsy. Nano Letters 18, 7642–7650 (2018).
Pamme, N. & Manz, A. On-chip free-flow magnetophoresis: continuous flow separation of magnetic particles and agglomerates. Analytical chemistry 76, 7250–7256 (2004).
Xia, N. et al. Combined microfluidic-micromagnetic separation of living cells in continuous flow. Biomedical microdevices 8, 299 (2006).
Laurell, T., Petersson, F. & Nilsson, A. Chip integrated strategies for acoustic separation and manipulation of cells and particles. Chemical Society Reviews 36, 492–506 (2007).
Shi, J., Huang, H., Stratton, Z., Huang, Y. & Huang, T. J. Continuous particle separation in a microfluidic channel via standing surface acoustic waves (SSAW). Lab on a Chip 9, 3354–3359 (2009).
Destgeer, G., Lee, K. H., Jung, J. H., Alazzam, A. & Sung, H. J. Continuous separation of particles in a PDMS microfluidic channel via travelling surface acoustic waves (TSAW). Lab on a Chip 13, 4210–4216 (2013).
Oh, Y., Lee, H., Son, S. Y., Kim, S. J. & Kim, P. Capillarity ion concentration polarization for spontaneous biomolecular preconcentration mechanism. Biomicrofluidics 10, 014102, https://doi.org/10.1063/1.4939434 (2016).
Park, S. et al. Capillarity ion concentration polarization as spontaneous desalting mechanism. Nat Commun 7, 11223, https://doi.org/10.1038/ncomms11223 (2016).
Yamada, M., Nakashima, M. & Seki, M. Pinched flow fractionation: continuous size separation of particles utilizing a laminar flow profile in a pinched microchannel. Analytical chemistry 76, 5465–5471 (2004).
Sai, Y., Yamada, M., Yasuda, M. & Seki, M. Continuous separation of particles using a microfluidic device equipped with flow rate control valves. Journal of chromatography A 1127, 214–220 (2006).
Morijiri, T., Sunahiro, S., Senaha, M., Yamada, M. & Seki, M. Sedimentation pinched-flow fractionation for size-and density-based particle sorting in microchannels. Microfluidics and nanofluidics 11, 105–110 (2011).
Di Carlo, D., Irimia, D., Tompkins, R. G. & Toner, M. Continuous inertial focusing, ordering, and separation of particles in microchannels. Proceedings of the National Academy of Sciences 104, 18892–18897 (2007).
Kuntaegowdanahalli, S. S., Bhagat, A. A., Kumar, G. & Papautsky, I. Inertial microfluidics for continuous particle separation in spiral microchannels. Lab Chip 9, 2973–2980, https://doi.org/10.1039/b908271a (2009).
Yang, S., Kim, J. Y., Lee, S. J., Lee, S. S. & Kim, J. M. Sheathless elasto-inertial particle focusing and continuous separation in a straight rectangular microchannel. Lab Chip 11, 266–273, https://doi.org/10.1039/c0lc00102c (2011).
Lim, E. J. et al. Inertio-elastic focusing of bioparticles in microchannels at high throughput. Nat Commun 5, 4120, https://doi.org/10.1038/ncomms5120 (2014).
Lu, X. & Xuan, X. Continuous microfluidic particle separation via elasto-inertial pinched flow fractionation. Analytical chemistry 87, 6389–6396 (2015).
Ebel, J., Anderson, J. L. & Prieve, D. Diffusiophoresis of latex particles in electrolyte gradients. Langmuir 4, 396–406 (1988).
Anderson, J. L. & Prieve, D. C. Diffusiophoresis caused by gradients of strongly adsorbing solutes. Langmuir 7, 403–406 (1991).
Kar, A., Chiang, T.-Y., Ortiz Rivera, I., Sen, A. & Velegol, D. Enhanced transport into and out of dead-end pores. ACS nano 9, 746–753 (2015).
Palacci, J., Abécassis, B., Cottin-Bizonne, C., Ybert, C. & Bocquet, L. Colloidal motility and pattern formation under rectified diffusiophoresis. Physical review letters 104, 138302 (2010).
Friedrich, S. M., Burke, J. M., Liu, K. J., Ivory, C. F. & Wang, T.-H. Molecular rheotaxis directs DNA migration and concentration against a pressure-driven flow. Nature Communications 8, 1213 (2017).
Shin, S., Shardt, O., Warren, P. B. & Stone, H. A. Membraneless water filtration using CO 2. Nature. Communications 8, 15181 (2017).
Guha, R. et al. Modulation of Spatiotemporal Particle Patterning in Evaporating Droplets: Applications to Diagnostics and Materials Science. ACS applied materials & interfaces 9, 43352–43362 (2017).
Banerjee, A., Williams, I., Azevedo, R. N., Helgeson, M. E. & Squires, T. M. Soluto-inertial phenomena: Designing long-range, long-lasting, surface-specific interactions in suspensions. Proceedings of the National Academy of Sciences of the United States of America 113, 8612–8617, https://doi.org/10.1073/pnas.1604743113 (2016).
Boulogne, F., Shin, S., Dervaux, J., Limat, L. & Stone, H. A. Diffusiophoretic manipulation of particles in a drop deposited on a hydrogel. Soft Matter 13, 5122–5129, https://doi.org/10.1039/c7sm00915a (2017).
Shin, S., Ault, J. T., Feng, J., Warren, P. B. & Stone, H. A. Low-Cost Zeta Potentiometry Using Solute Gradients. Advanced Materials 29 (2017).
Shin, S., Ault, J. T., Warren, P. B. & Stone, H. A. Accumulation of Colloidal Particles in Flow Junctions Induced by Fluid Flow and Diffusiophoresis. Physical Review X 7 (2017).
Shin, S. et al. Size-dependent control of colloid transport via solute gradients in dead-end channels. Proc. Natl. Acad. Sci. 113, 257–261 (2016).
Chai, B., Mahtani, A. G. & Pollack, G. H. Unexpected Presence of Solute-Free Zones at Metal-Water Interfaces. Contemp Mater 3, 1–12, https://doi.org/10.7251/COM1201001C (2012).
Lee, H., Kim, J., Yang, J., Seo, S. W. & Kim, S. J. Diffusiophoretic exclusion of colloidal particles for continuous water purification. Lab Chip 18, 1713–1724, https://doi.org/10.1039/c8lc00132d (2018).
Chai, B. H., Yoo, H. & Pollack, G. H. Effect of Radiant Energy on Near-Surface Water. Journal of Physical Chemistry B 113, 13953–13958, https://doi.org/10.1021/jp908163w (2009).
Klyuzhin, I., Symonds, A., Magula, J. & Pollack, G. H. New method of water purification based on the particle-exclusion phenomenon. Environmental Science & Technology 42, 6160–6166, https://doi.org/10.1021/es703159q (2008).
Florea, D., Musa, S., Huyghe, J. M. & Wyss, H. M. Long-range repulsion of colloids driven by ion exchange and diffusiophoresis. Proceedings of the National Academy of Sciences, 201322857 (2014).
Ault, J. T., Shin, S. & Stone, H. A. Diffusiophoresis in narrow channel flows. Journal of Fluid Mechanics 854, 420–448 (2018).
Helfferich, F. G. Ion exchange. (McGraw-Hill, 1962).
Helfferich, F. G. Ion exchange chromatography. (McGraw-Hill, 1962).
Florea, D., Musa, S., Huyghe, J. M. R. & Wyss, H. M. Long-range repulsion of colloids driven by ion exchange and diffusiophoresis. Proceedings of the National Academy of Sciences of the United States of America 111, 6554–6559, https://doi.org/10.1073/pnas.1322857111 (2014).
Lee, J. A., Lee, D., Park, S., Lee, H. & Kim, S. J. Non-negligible Water-permeance through Nanoporous Ion Exchange Medium. Sci Rep 8, 12842 (2018).
Keh, H. J. Diffusiophoresis of charged particles and diffusioosmosis of electrolyte solutions. Current Opinion in Colloid & Interface Science 24, 13–22, https://doi.org/10.1016/j.cocis.2016.05.008 (2016).
Benner, E. M. & Petsev, D. N. Potential flow in the presence of a sudden expansion: Application to capillary driven transport in porous media. Physical Review E 87, https://doi.org/10.1103/PhysRevE.87.033008 (2013).
Elizalde, E., Urteaga, R. & Berli, C. L. Rational design of capillary-driven flows for paper-based microfluidics. Lab Chip 15, 2173–2180, https://doi.org/10.1039/c4lc01487a (2015).
Mendez, S. et al. Imbibition in porous membranes of complex shape: quasi-stationary flow in thin rectangular segments. Langmuir 26, 1380–1385, https://doi.org/10.1021/la902470b (2010).
Hong, S., Kwak, R. & Kim, W. Paper-Based Flow Fractionation System Applicable to Preconcentration and Field-Flow Separation. Anal Chem 88, 1682–1687, https://doi.org/10.1021/acs.analchem.5b03682 (2016).
Perez-Cruz, A., Stiharu, I. & Dominguez-Gonzalez, A. Two-dimensional model of imbibition into paper-based networks using Richards’ equation. Microfluidics and Nanofluidics 21, https://doi.org/10.1007/s10404-017-1937-0 (2017).
Kirby, B. J. Micro-and nanoscale fluid mechanics: transport in microfluidic devices. (Cambridge university press, 2010).
Giddings, J. C. Unified Separation Science. (Wiley, 1991).
Cheow, L. F., Ko, S. H., Kim, S. J., Kang, K. H. & Han, J. Increasing the sensitivity of enzyme-linked immunosorbent assay using multiplexed electrokinetic concentrator. Analytical chemistry 82, 3383–3388 (2010).
Lee, J. H., Song, Y. A. & Han, J. Y. Multiplexed proteomic sample preconcentration device using surface-patterned ion-selective membrane. Lab Chip 8, 596–601, https://doi.org/10.1039/b717900f (2008).
Son, S. Y., Lee, S., Lee, H. & Kim, S. J. Engineered nanofluidic preconcentration devices by ion concentration polarization. BioChip Journal 10, 251–261, https://doi.org/10.1007/s13206-016-0401-7 (2016).
Acknowledgements
This work is supported by Basic Research Laboratory Project (NRF-2018R1A4A1022513), Basic Science Research Program (2016R1A1A1A05005032 and 2016R1A6A3A11930759) and the Center for Integrated Smart Sensor funded as Global Frontier Project (CISS- 2011-0031870) by the Ministry of Science and ICT and Korean Health Technology RND project from the Ministry of Health and Welfare Republic of Korea (HI13C1468, HI14C0559). Also all authors acknowledged the financial supports from BK21 Plus program of the Creative Research Engineer Development IT, Seoul National University. S. J. Kim acknowledged the financial support from LG Yonam Foundation, Korea.
Author information
Authors and Affiliations
Contributions
D. Lee fabricated the micro/nanofluidic device and mainly conducted experiments, J.A. Lee numerically analyzed the diffusiophoretic migration, H. Lee and S.J. Kim supervised the project. All authors wrote the manuscript.
Corresponding authors
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Lee, D., Lee, J.A., Lee, H. et al. Spontaneous Selective Preconcentration Leveraged by Ion Exchange and Imbibition through Nanoporous Medium. Sci Rep 9, 2336 (2019). https://doi.org/10.1038/s41598-018-38162-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-018-38162-6
- Springer Nature Limited
This article is cited by
-
Effect of evaporation through nanoporous medium on diffusiophoresis
Micro and Nano Systems Letters (2020)
-
Spontaneous diffusiophoretic separation in paper-based microfluidic device
Micro and Nano Systems Letters (2020)
-
Nanoelectrokinetic Selective Preconcentration Based on Ion Concentration Polarization
BioChip Journal (2020)