Introduction

Myocardial perfusion single photon emission computed tomography (SPECT) is widely used for the diagnosis of ischemic heart disease, myocardial viability, risk stratification, adaptation of revascularization and evaluation of therapy. However, myocardial perfusion SPECT is subject to image quality due to the use of different acquisition protocols, image reconstruction parameters and image display settings among each institution [1]. To standardize these parameters, scientific conferences of various countries have published guidelines [25]. For example, the guidelines of both the Society of Nuclear Medicine and Molecular Imaging (SNMMI) and the European Association of Nuclear Medicine (EANM) have recommended that the necessary average counts to maintain images quality must only be obtained with a minimum of 500,000 counts per each view (a minimum of 400 counts/cm2 for normal myocardium) [2, 3]. However, SPECT projections commonly contain uptake other than myocardium (e.g. lung, liver, gallbladder and intestine), which makes it difficult to judge whether these satisfy these guidelines. A guideline of the Japan Society of Nuclear Medicine Technology (JSNMT) has recommended that the average myocardial counts of left anterior oblique (LAO) 45° projection data must have more than 100 counts/pixel [5]. However, these criteria cannot be used for measurement estimation after the examination has taken place. Therefore, actual myocardial counts may contain significant variability due to patient’s condition and examination protocols (e.g. injection dose, the intervals from injection until acquisition start, uptake ratio of myocardium and washout rate).

In general, image reconstruction parameters of myocardial perfusion SPECT are subjectively or objectively determined by operators. If image reconstruction parameters are determined subjectively, intra- and inter-operator repeatability decreases. Thus, image reconstruction parameters need to be set objectively to improve operator repeatability. SNMMI and EANM guidelines have recommended that image reconstruction parameters change as a function of injection dose, which should not be varied routinely [2, 3]. However, even if radioactivity dose was identical in all subjects, myocardial counts is inversely proportional to the patient’s weight [6].

One of the principal image reconstruction parameters is the Butterworth filter, a parameter strongly affecting the quality of myocardial images. Onishi et al. have reported a procedure to decide optimal Butterworth filter parameters by each collimator from spatial domain and frequency domain assessments using 99mTc and 201Tl agents [7]. Furthermore, they have reported optimal Butterworth filter parameters must change by a difference of radioactivity dose [8, 9]. However, the correlation between the difference of myocardial counts and optimal image reconstruction parameters variability have never been revealed. The objective of this study was to formulate a flowchart to decide optimal Butterworth filter parameters by filtered back projection (FBP), ordered subset expectation maximization (OSEM) and collimator-detector response compensation OSEM (CDR-OSEM) methods [10] using the evaluation system of the myocardial image based on technical grounds (EMIT) phantom [11].

Materials and methods

EMIT phantom design

EMIT phantom was developed to reduce a difference of image quality among institutions [11]. The appearance and myocardial region of the phantom are shown in Fig. 1. The thorax phantom was constructed with an elliptical shape 300 mm in the major axis, 220 mm in the minor axis and a height of 223 mm, which included the lungs containing beads of polystyrene foam and myocardium. Myocardium was constructed from cylinder shapes of 223 mm in height and 80 mm in diameter, and the myocardial thickness and diameter of the left ventricular (LV) chamber were 10 and 40 mm, respectively. The myocardial and LV chamber volumes were approximately 375 and 265 mL, which can fill with liquid such as water and radiopharmaceuticals. Furthermore, the myocardium has normal part and eight defect part with four sizes of defect (5, 10, 15 and 20 mm; thickness and length were constant at 10 and 30 mm) and four thicknesses of defect (10, 7.5, 5 and 2.5 mm; size and length were constant at 20 and 30 mm) simultaneously. The four defects of different sizes were set on same walls with 15 mm interval, and the four defects of different thickness were set on the opposite with contralateral myocardial walls having 15 mm interval. The major characteristic of the phantom was to shift the defect location, so that it can simulate the four patterns: anterior, septal, inferior and lateral walls. Furthermore, the phantom can be used to simulate and acquire eight different defects simultaneously.

Fig. 1
figure 1

The structure of the evaluation system of myocardial image based on technical grounds (EMIT) phantom. The myocardium and left ventricle (LV) volume are approximately 375 and 265 mL, respectively. a The alignment of lungs and myocardium for z axis view. The length of z axis view is 233 mm, the lungs contain polystyrene foam beads. b X–Y view alignment. The length of x and y axis views are 300 and 220 mm, respectively. c The four sizes of defect (5, 10, 15 and 20 mm; thickness is constant at 10 mm). d The four thicknesses of defect (2.5, 5, 7.5 and 10 mm; size is constant at 20 mm)

Acquisition protocols

SPECT images were obtained using a dual-headed SPECT scanner (Infinia3, GE Healthcare Co., Ltd., Tokyo, Japan). Technetium-99m agent was inserted into the myocardial part of EMIT phantom. SPECT study was acquired using low energy high resolution (LEHR) collimator, 90° apart head configuration, 64 × 64 matrix, 1.6 zoom, 30 projections, circular orbit of 180° rotation and pixel size of 5.52 mm. Normal myocardial counts were determined by average counts obtained from regional of interests (ROI) on the normal myocardial parts of LAO 45° projection data. The average normal myocardial counts were approximately 10, 20, 40, 60, 80, 100 and 120 counts/pixel, which was adjusted by acquisition time. The defect locations were set on four patterns (anterior, lateral, inferior and septal walls of myocardium). Finally, SPECT images were obtained from 28 projection datasets (seven acquisitions of the difference of normal myocardial counts using four different patterns in each acquisition).

Image reconstruction parameters

The SPECT images were reconstructed by FBP, OSEM and CDR-OSEM methods. The FBP reconstruction was performed with a Ramp filter and the Butterworth filter. Butterworth filter was derived by Eq. 1.

$${\text{B}}({\text{f}}) = \sqrt {\frac{1}{{1 + ({\text{fc}}/{\text{F}})^{\text{p}} }}}$$
(1)

where F indicates the cut-off frequency, and n is the power (constant value: 10). Cut-off frequencies were used for seven patterns of 0.30, 0.35, 0.40, 0.45, 0.50, 0.55 and 0.60 cycles/cm to validate optimal parameters among FBP, OSEM and CDR-OSEM methods. The subset and iteration of OSEM and CDR-OSEM methods used six and five for OSEM method, six and twenty for CDR-OSEM method, respectively. Scatter and attenuation corrections were not employed in all methods.

Image assessments

Short axis images were distinctly examined for defect and normal parts, which were evaluated by visual and quantitative assessments of the defect part and visual assessment of the normal part. The defect sizes using visual and quantitative assessments were 15 mm for FBP and OSEM methods and 10 mm for CDR-OSEM method, and thickness was constant at 10 mm in all methods. The visual assessment was scored on a scale of 1–5 (5: excellent, 4: good, 3: even, 2: bad, 1: poor) by five observers, and the average score of acceptance criterion defined by a total of more than four points. The quantitative assessment was evaluated using EMIT programs [11]. EMIT programs can work on the windows OS. As the first step, image registration was performed between reconstructed SPECT image and template magnetic resonance image (MRI), and a polar map calculated from SPECT image after registration was subsequently generated. Template ROI was created with eight defect parts using MRI, which was set on SPECT polar map after image registration. A profile curve in template ROI was drawn using each defect part on polar map (Fig. 2). The %count of myocardial defects for each size and thickness was calculated in Eq. 2.

$$\% {\text{count}} = \frac{\text{Volume of area under count profile curve of each defect in template ROI}}{\text{Volume of each defect using template ROI decided from MRI}} \times 100$$
(2)
Fig. 2
figure 2

An algorithm of the evaluation system of the myocardial image based on technical grounds (EMIT) programs. The left image indicates the polar map for MRI, and the right image is for SPECT. The blue boxes indicate the template ROI for each defect size. The profile curve of each defect size is obtained from the four red lines in MRI and SPECT images. The four count profile curves were combined by normalization of max counts. An aqua box indicates each defect volume using template ROI determined from MRI, and yellow areas indicate the volume of area under the count profile curve of each defect in the template ROI

The %count defined a defect as less than 2 standard deviations (SD) of normal database of Mid-A, Mid-I, Mid-IL, and Mid-IS. We used a normal database created by Nakajima et al. [12]. The optimal cut-off frequency was determined by comprehensive assessment including visual and quantitative assessments, and created regression equations described the correlation between normal myocardial counts and optimal cut-off frequency. Finally, we formulated a flowchart indicating the image reconstruction procedure to optimize SPECT images based on comprehensive assessment. In addition, the utility of the flowchart was validated in retrospectively using clinical studies compared with guidelines. Permission for clinical data used in this study was obtained from the hospital ethics committee.

Results

The visual score of defect assessment for all methods had the optimal cut-off frequency ranges (0.35–0.60 cycles/cm) in all normal myocardial counts (Fig. 3). The %count of FBP and OSEM methods optimized cut-off frequencies of 0.30–0.60 cycles/cm and CDR-OSEM method optimized frequencies of 0.40–0.60 cycles/cm (Fig. 4). Normal assessment had optimal cut-off frequency in normal myocardium of 40–120 counts/pixel for FBP and CDR-OSEM methods, and 60–120 counts/pixel for OSEM method. When the normal myocardial count was low, the normal part of the myocardial image was poor, with 10 and 20 counts/pixel for FBP method, 10, 20 and 40 counts/pixel for OSEM method and 10 counts/pixel for CDR-OSEM method (Fig. 5). The optimal cut-off frequency ranges among three methods were determined by a comprehensive assessment including normal and defect parts (FBP: 40, 60, 80, 100 and 120 counts/pixel were 0.35, 0.35–0.45, 0.35–0.50, 0.35–0.50 and 0.35–0.50 cycles/cm, OSEM: 60, 80, 100 and 120 counts/pixel were 0.30–0.35, 0.30–0.35, 0.30–0.40 and 0.35–0.40 cycles/cm, CDR-OSEM: 40, 60, 80, 100 and 120 counts/pixel were 0.35–0.40, 0.40–0.45, 0.35–0.45, 0.40–0.45 and 0.35–0.45 cycles/cm). However, the 10 and 20 counts/pixel for the FBP and CDR-OSEM methods and the 10, 20 and 40 counts/pixel of the OSEM method had no optimal cut-off frequency ranges.

Fig. 3
figure 3

Visual score of defect assessment among three methods (a FBP, b OSEM, c CDR-OSEM). Dotted line indicate acceptance criteria, and error bar describes standard error of the mean

Fig. 4
figure 4

%count of defect assessment among three methods (a FBP, b OSEM, c CDR-OSEM). Dotted line indicates acceptance criteria

Fig. 5
figure 5

Visual score of normal assessment among three methods (a FBP, b OSEM, c CDR-OSEM). Dotted line indicates acceptance criteria, and error bar describes standard error of the mean

The correlation between optimal cut-off frequencies and normal myocardial counts are shown in Fig. 6. Correlation between normal myocardial counts and optimal cut-off frequency could be represented as regression equations in all methods, which had high or medium coefficient of determination as follows: FBP: y = −0.0000446x2 + 0.00889x + 0.07, R2 = 0.98, OSEM: y = 0.001x + 0.285, R2 = 0.80, CDR-OSEM: y = −00.000018x2 + 0.0031x + 0.55, R2 = 0.55, where x and y indicate optimal cut-off frequency and normal myocardial counts, respectively. We formulated a flowchart to optimize image reconstruction parameters based on comprehensive assessment to improve inter- and intra-operator reproducibility (Fig. 7). Myocardial short axis images were created using optimal cut-off frequency guided from the flowchart procedure, so that those images correctly described defect and normal parts (Fig. 8).

Fig. 6
figure 6

Correlation between cut-off frequency and normal myocardial counts among three methods. All methods were represented as regression equations, which had high or medium coefficient of determination

Fig. 7
figure 7

A flowchart to indicate image reconstruction procedure to optimize the cut-off frequency based on comprehensive assessment. The first section confirms normal myocardial counts. If normal myocardial counts have less than 40 counts/pixel for FBP and CDR-OSEM methods or 60 counts/pixel for OSEM method, we recommend re-examination by changing the acquisition parameters. The next section recommends processing the projection data using regression equations, and thereby determining the final images. If it has a problem with image quality, an explanation of the regression equations is warranted

Fig. 8
figure 8

Myocardial images processed using the optimal cut-off frequency using three methods. These images indicate myocardial image slice used for normal and defect assessments regarding three methods. All of the images have described correctly the defect and normal locations

Case report

An 80-year-old woman was referred to our department for exercise myocardial perfusion SPECT because of suspected angina pectoris. She had been treated for diabetes mellitus and hypertension. The radiopharmaceutical agent of 99mTc-tetrofosmin 296 MBq was injected at the half time during the six minutes adenosine infusion, and myocardial perfusion SPECT acquisition was undertaken with the patient in the supine position after a 30 min injection. The imaging was performed with a dual-head gamma-camera equipped with a LEHR collimator, and energy windows were 140 keV ± 10 %. The projection data was acquired in 30 steps around a 180° arc for 40 s per projection starting with the right anterior oblique view and ending with the left posterior oblique view. The total acquisition time was approximately 12 min. The acquisition was in step-and-shoot mode with a zoom factor of 1.6, and the images were recorded in a 64 × 64 matrix (pixel size: 5.52 mm). In addition, a coronary angiography (CAG) was performed after 8 days, and results of CAG showed significant stenosis on the proximal and middle of the right coronary artery, in addition to the distal and first diagonal branch of the left coronary artery. Myocardial images were processed by means of the guideline and the flowchart as shown in Fig. 9 (a: Guideline, b: FBP, c: OSEM, d: CDR-OSEM). Normal myocardial counts in this case had the 79 counts/pixel. In FBP, OSEM and CDR-OSEM methods, myocardial images had a mild hypo-perfusion area in the mid and apical anterior walls, basal and mid inferior walls. However, the myocardial images based on EANM guidelines showed no hypo-perfusion areas.

Fig. 9
figure 9

Short axis images of myocardial perfusion SPECT (a guideline, b FBP, c OSEM, d CDR-OSEM). Normal myocardial counts in this case had 79 counts/pixel. Myocardial images processed using the flowchart had a mild hypo-perfusion area in the mid and apical anterior walls, basal and mid inferior walls. However, myocardial images using the EANM guidelines showed no hypo-perfusion areas

Discussion

Myocardial perfusion SPECT is widely used for diagnosis of ischemic and viability, risk stratification, adaptation of revascularization and therapy evaluation [13]. In particular, it plays an important role in diagnosis of angina pectoris and myocardial infarction, which is used to judge whether medical therapy or revascularization is required [14]. However, myocardial perfusion SPECT study produces different images depending on the different acquisition protocols, image reconstruction parameters and image display settings used by each institution, which has been reported in a study [1]. Therefore, it is possible to generate the artifact, so that the myocardial image generates a false-positive or false-negative. Butterworth filter parameter setting is possibly one of the causes of the artifacts, which has to set for each patient. However, optimal parameters for each patient have not been investigated in detail in previous studies. Therefore, we have attempted to determine optimal image reconstruction parameters using normal myocardial counts in this study.

Normal myocardial count ranges are determined in reference to clinical studies. In addition, to validate optimal Butterworth filter parameters, cut-off frequency ranges were set to 0.20–0.60 cycles/cm corresponding to 20–70 % of the Nyquist frequency, which was determined by the ranges in reference to past studies [79]. The subset and iteration of reconstruction parameters for OSEM and CDR-OSEM methods were determined from past studies [1518]. Furthermore, Maru and Yanagisawa have reported that it was not necessary to change subset and iteration by the difference of acquisition counts [19]. Thus, we did not change subset and iteration in this study. The estimated defect size among three methods were determined by full width at half maximum (FWHM) from a phantom study based on National Electrical Manufactures Association (NEMA) [20]. Estimated defect sizes in this study were determined as 15 mm for FBP and OSEM methods, 10 mm for CDR-OSEM method corresponding to the smallest size in defect size more than the FWHM which is 11.8, 11.4 and 5.3 mm for FBP, OSEM and CDR-OSEM methods, respectively. The %count defined a defect as less than 2 SD of the normal database in our study. The %count theoretical value is normally 0 % in all of myocardial walls in order to simulate a complete defect. However, these criteria have taken into account the influence of both partial volume effect and limitation of spatial resolution. However, to approximate the %count theoretical value, the optimal cut-off frequency by normal myocardial counts determined the highest of cut-off frequency ranges as indicated by comprehensive assessment. Normal myocardial counts needed for optimal imaging was a minimum of 40 counts/pixel for FBP and CDR-OSEM method or 60 counts/pixel for OSEM method. The OSEM algorithm has a characteristic of enhancing statistical noise when normal myocardial counts are less, so that it was required to obtain higher normal myocardial counts as compared to other methods. However, the CDR-OSEM method could provide optimal image until lower counts in spite of OSEM algorithm. The CDR-OSEM method has reconstructed using maximum a posteriori-expectation maximization (MAP-EM) algorithm incorporates Bayes’ theorem, which reportedly decreases the statistical noise [21]. Therefore, the CDR-OSEM method had optimal image until lower counts, a higher cut-off frequency than OSEM method. The correlation between normal myocardial counts and optimal cut-off frequency was expressed as regression equations by three methods. Furthermore, we formulated a flowchart to produce optimal image quality using regression equations. Consequently, the flowchart allowed us to objectively process against different normal myocardial counts.

In the clinical case, we performed image reconstruction using EANM guidelines and the flowchart. Myocardial images reconstructed using EANM guidelines tended to reduce the detection of anterior and inferior walls, whereas myocardial images reconstructed using the flowchart incorporating the three methods detected mild hypo-perfusion of the anterior and inferior walls. As a result of this case, we could demonstrate the usefulness of the image reconstruction using the flowchart.

Our study has some limitations. Firstly, the EMIT phantom does not simulate scatter and attenuation, however human bodies are susceptible to the effects of scatter and attenuation. Should a phantom simulate these effects, the flowchart may need to be changed. Furthermore, different acquisition parameters were not validated in this study, however, Onishi et al. have reported that on varied optimal image reconstruction parameters by different acquisition parameters [79]. These limitations will be validated in the next study.

Conclusion

We formulated a flowchart to optimize cut-off frequency among three methods using the EMIT phantom, and were able to determine the optimal cut-off frequency for normal myocardial counts. Furthermore, the usefulness of image reconstruction using the flowchart was demonstrated by a clinical case. Finally, it has been possible to objectively process image reconstruction using the flowchart.