1 Introduction

Microdosimetry focuses on studying how energy is randomly deposited in tiny, sub-nuclear cellular structures like chromosomes. It is important to note that microdosimetry is not merely small-scale dosimetric-investigation; it is about characterizing the random fluctuations of energy deposited event by event.

Figure 1 illustrates the difference in energy deposition patterns between low and high Linear Energy Transfer (LET) radiation. The larger circles in the figure symbolize microscopic target volumes, while red dots represent fixed-sized energy depositions. On the left, each target has a single dot, demonstrating a homogeneous distribution. The background is uniform light-yellow, emphasizing the uniformity in energy distribution. On the right, some areas have multiple red dots (representing high energy deposition), while others have only a few or none, indicating a heterogeneous distribution. Areas with higher energy deposition are shaded in red, indicating clustered DNA damage due to high Linear Energy Transfer (LET) radiation. Despite both sides having an average of, let’s say, 1 Gy of absorbed dose, the distribution pattern plays a significant role. Small energy deposits generally result in simple DNA damages that are easily repairable. Conversely, larger energy depositions over shorter distances can cause clustered DNA damage that cells struggle to repair effectively [1]. The severity of this damage is represented in Fig. 1 by the color intensity of the targets.

Fig. 1
figure 1

Representation of two irradiations at 1 Gy. Left: Low Linear Energy Transfer (LET) radiation shows a uniform microscopic distribution. Right: High LET radiation displays a heterogeneous distribution of energy deposition, where some targets experience substantial energy deposition events while others have few or no energy depositions. The color intensity represents the severity of biologically induced damage

In microdosimetry, a commonly used quantity is the lineal energy, \(y\). It is defined as the ratio of the energy imparted by a single event, \({\varepsilon }_{s}\), to the mean chord length of the target volume, \(\overline{l }\). The formula is:

$$y=\frac{{\varepsilon }_{s}}{\overline{l} }$$
(1)

Lineal energy is expressed in units of keV µm−1. Being a stochastic quantity, the lineal energy is characterized by a probability density distribution, \(f(y)\). The dose probability density of \(y\), \(d(y)\), is even more important and is obtained as the product of the frequency of a given event size by the size of the event, normalized to unit area:

$$d(y)=\frac{yf(y)}{\int_{0}^{\infty}yf\left(y\right){\text{d}}y}$$
(2)

More details on microscopic quantities definitions and their relation to macroscopic quantities like the absorbed dose and the LET can be found in the ICRU Reports on Microdosimetry [2, 3]. Notably, the new ICRU Report 98 recommends considering the stochastic nature of radiation interactions, when the mean number of charged particles, including delta rays originating outside the site, interacting with the volume of interest is small. Furthermore, it is explained that in mixed radiation fields, the fluence of the higher LET components should be considered. This specific case applies to BNCT, because the high LET components related to alpha particles and lithium ions emerging from the neutron capture reaction have a small fluence, albeit contributing significantly to the dose. The situation is illustrated in Fig. 2, which is divided into two parts labeled “High fluence” and “Low fluence”. In both parts, small circles represent volumes of interest where radiation interactions occur. Under “High fluence”, there are many arrows indicating many low LET charged particles interacting with each volume. In this case microdosimetry is not necessary because the resulting distribution of the imparted energy is rather homogenous. Under “Low fluence”, fewer arrows indicate that a small number of high LET charged particles interact with each volume, resulting in a heterogeneous distribution of energy imparted. Here, microdosimetry is necessary to study the stochastic nature of these interactions. The color intensity within circles under “Low fluence” varies, visually representing the heterogeneous distribution of energy.

Fig. 2
figure 2

Schematic illustration of when a microdosimetric study of the stochastic nature of radiation interactions is recommended. The volume of interest is represented by each small circle. Left: the mean number of charged particles interacting with the volume of interest is large, albeit the corresponding absorbed dose is rather low due to low LET of particles. Right: the mean number of high LET charged particles interacting with the volume of interest is small and results in a heterogeneous distribution of the energy imparted. In this case the stochastic nature of radiation interactions should be considered

In the literature, the majority of studies focus on computational microdosimetry, utilizing Monte Carlo tools for investigating various irradiation conditions. For example, Nichols et al. [4] developed an enhanced tissue model to elucidate observed differences in clinical outcomes based on tissue cellular geometry and the influence of neighboring cells. More recently, Sato et al. [5] proposed a novel microdosimetric model for biological effectiveness in BNCT, considering intra- and intercellular heterogeneity in 10B distribution. Hu et al. [6] successfully validated the PHITS Monte Carlo simulation code against experimental data obtained using a TEPC. They concluded that PHITS can accurately calculate thermal neutron fluxes, gamma-ray dose rates, and RBE when physical measurements pose challenges. Moreover, building on the pioneering work by Wuu et al. [7], several experimental studies were conducted in reactor-based BNCT fields using both standard-sized TEPCs [8,9,10,11,12] and miniaturized TEPCs suitable for intense radiation fields [13,14,15]. Silicon microdosimeters have also demonstrated success [16, 17], and more recently, experimental techniques have been implemented for characterizing accelerator-based neutron sources for BNCT [18,19,20]. This work provides a comprehensive survey of the methodology and sheds light on the potential advantages that microdosimetry offers in terms of determining dose components and assessing RBE.

2 Methods

Experimental microdosimetry in BNCT relies on the dual-counter technique initially introduced by Wuu et al. [7]. This method entails conducting paired measurements with two TEPCs that are identical except for their boron content. One detector lacks boron, while the other has cathode walls doped with boron. The experimental setup is illustrated in Fig. 3, which displays two identical TEPCs, one without boron and the other with boron-doped cathode walls, as depicted in the right-side drawings.

Fig. 3
figure 3

The microdosimetric dual-counter technique uses two TEPCs that are identical, differing only in boron content: one with boron-doped cathode walls and the other without. The two microdosimeters shown on the left side of the figure were constructed at Legnaro National Laboratories (LNL) of Istituto Nazionale di Fisica Nucleare (INFN) and feature a cylindrical sensitive volume measuring 1 mm in diameter and height

Microdosimetric lineal energy spectra measured in a thermal neutron field without and with boron at 1 μm simulated site sizes are shown in Fig. 4. More details on the experimental setup and radiation field can be found in [19]. The dose probability distributions were normalized, as per Eq. (2); therefore, in this graphical representation, equal areas under the curve correspond to equal fractions of the total dose [2, 3]. In both spectra the low LET component, mainly due to unavoidable gamma rays, is distinguishable below approximately 15 keV mm−1, while the events above this threshold are due to neutron interactions.

Fig. 4
figure 4

Microdosimetric spectra measured in a mixed thermal neutron and gamma radiation field at a simulated site size of 1 mm. The left spectrum was measured with a TEPC without boron, while the right spectrum was measured with a TEPC with cathode walls doped at 100 ppm of 10B. Data adapted from [19]

The lineal energy distributions presented in Fig. 4 are analyzed, considering that the radiation spectra (below approximately 15 keV μm−1) should be nearly identical, irrespective of boron content. Figure 5 illustrates this. The left panel shows different gamma components, with the gamma portion being lower in the spectrum with boron. This is a consequence of the normalization procedure on \(d(y)\). The areas subtended by the two curves (0 and 100 ppm of 10B) are equal and correspond to an absorbed dose of, let’s say, 1 Gy. However, the absolute value of the dose contribution due to gamma radiation should be the same in both cases. If we want to calculate the relative dose enhancement due to the boron neutron capture (BNC) reactions, the spectrum with boron must be multiplied by a constant factor to align its gamma region to that without boron, as shown in the right side of Fig. 5. In this case the area subtended by the curve without 10B is still 1 Gy, while the additional area subtended by the curve with 100 ppm of 10B represents the dose enhancement due to the BNC reactions.

Fig. 5
figure 5

Microdosimetric spectra measured with and without boron. In the left panel both distributions \(d\left(y\right)\) are normalized to the unit dose. In the right panel the spectrum with boron has been multiplied by a constant factor to align its gamma component to that without boron. Data adapted from [19]

By subtracting the spectrum without 10B from the one with 10B, the resulting distribution represents the contribution of BNC products, specifically alpha particles and lithium ions. Figure 6 shows this distribution, labelled as the “pure BNC component”.

Fig. 6
figure 6

The “pure BNC component” is obtained by subtracting the spectrum without boron from the one with boron. Data adapted from [19]

A recent work by Conte et al. [21] confirmed experimentally that the dose enhancement due to BNC interactions is proportional to the 10B concentration, while the individual microdosimetric distribution of the lineal energy is invariant with boron content. The proportionality factor has not a unique value but depends on the total distribution, and specifically on the gamma and fast neutron components of the investigated radiation field. For instance, for the distribution shown in Fig. 6, the BNC dose enhancement amounts to approximately 84% of the dose without boron. Referring to the 10B concentration of 100 ppm, this number leads to a proportionality factor of about 7.7 × 10−3 per ppm of boron. Based on these findings, the impact of BNCT in the tumor treatment region as well as in the healthy tissue can be easily evaluated. In this work, a desirable [23] tumor:normal tissue ratio of 3.5:1 was considered, specifically with concentrations of 20 and 70 ppm of 10B in normal and tumor tissues, respectively.

In addition to the pure BNC component, the gamma and total neutron components (both with and without BNC reactions) can also be evaluated. The gamma component is obtained by merging the shoulder of the spectrum corresponding to the e-edge with a pure gamma spectrum, for instance, that produced from a 137Cs gamma source. The neutron components are obtained by subtracting the gamma portion from the entire spectra. An example of the gamma and neutron components are highlighted in Fig. 7.

Fig. 7
figure 7

The gamma and neutron components of the microdosimetric spectra measured without (left) and with 100 ppm of 10B (right). Data adapted from [19]

Finally, the relative biological effectiveness (RBE) of the radiation field can be estimated based on the measured microdosimetric spectra. The procedure consists of weighting the total and the individual lineal energy distributions \(d\left(y\right)\) with a lineal energy dependent weighting function \(r\left(y\right)\) to determine the effective RBE, neutron_RBE, photon_RBE and BNC_RBE:

$$RBE=\underset{0}{\overset{\infty}{\int}}r\left(y\right)d\left(y\right){\text{d}}y$$
(3)

Weighting functions \(r\left(y\right)\) are determined by combining data from biological experiments and microdosimetric distributions, and in general depend on the biological end point under consideration. In this work we used the weighting function derived from Tilikidis [22] using RBE values at 2 Gy for jejunum crypt cell survival obtained in therapeutic fast neutron beams and in low-energy neutron beams. This function has been already used in other studies focused on BNCT, leading to microdosimetric assessment of the RBE that is in reasonable agreement with biological data [6, 10, 11]. Minor differences have been evidenced when using other weighting functions [20].

3 Results and discussion

Following the procedure outlined in the "Methods" section and assuming a tumor-to-normal (T/N) tissue ratio of 3.5:1 for the 10B concentration, with a concentration of 20 ppm in normal tissue, the concentration in tumor cells was calculated to be 70 ppm. This ratio has been assumed because it is the typical value obtained in clinical trials using boronpèhenylalanine, which has demonstrated to ensure a safe and effective BNCT treatment [23, 24]. Maintaining a T/N ratio of at least 3.5:1 is considered a vital threshold to achieve the desired therapeutic outcomes.

Furthermore, achieving a sufficiently high concentration in the tumor is equally essential. Approximately 109 atoms of 10B per cell, equivalent to a concentration range of 20–50 ppm of 10B, are required to facilitate a substantial number of boron neutron capture reactions within a single cell. This high concentration is pivotal in guaranteeing the therapeutic effect and maximizing the potential for tumor cell destruction [24].

Figure 8 illustrates the impact of varying 10B concentrations (0, 20, and 70 ppm) on three distributions. The latter two distributions were derived from the 100 ppm distribution, accounting for individual components (photons, neutrons without boron, and BNC). The pure BNC component was scaled based on the 10B concentration. Note that only the no-boron distribution \(d(y)\) is normalized to the unit dose, while the others are scaled to the no-boron spectrum for alignment with the gamma regions. The primary distinction is observed in the high lineal energy values (approximately 200 to 500 keV mm−1), attributed to alpha particles and Li ions generated through the BNC reaction in the detector walls. These events overlap with contributions from residual fast neutrons and thermal neutron capture reactions in nitrogen, visible in the blue spectrum (without 10B) above around 20 keV mm−1.

Fig. 8
figure 8

The microdosimetric distributions evaluated at 0, 20 and 70 ppm of 10B

As illustrated in Fig. 6, the individual components contributing to the total dose can also be evaluated. This analysis is presented in Fig. 9 for 20 ppm of 10B (left) and 70 ppm of 10B (right). The total and individual dose components, relative to the absorbed dose in the absence of boron content, \({D}_{0}\), specifically the total dose \({D}_{tot}/{D}_{0}\), photon dose \({D}_{ph}/{D}_{0}\), neutron dose \({D}_{n}/{D}_{0}\) and pure BNC dose \({D}_{\text{BNC}}/{D}_{0}\), are reported in Table 1.

Fig. 9
figure 9

The individual photon, neutron and BNC components at 20 ppm (left) and 70 ppm (right) of 10B

Table 1 The total and the individual dose components relative to the absorbed dose in case of no boron content, for boron concentrations of 0, 20 and 70 ppm

The dose enhancement due to the addition of 10B amounts to 17% at 20 ppm and 59% at 70 ppm. To assess the biological impact, it is crucial to consider the increased biological effectiveness of high-LET particles. The distributions at 0, 20, and 70 ppm can be weighted using the Tilikidis weighting function to account for this. The resulting weighted distributions are depicted in Fig. 10, where the integral under each curve corresponds to the RBE-weighted dose. The resulting values, relative to the physical dose without boron, \({D}_{0}\), are 1.7, 2.1 and 2.7 for 0, 20 and 70 ppm, respectively.

Fig. 10
figure 10

The microdosimetric distributions at 0, 20 and 70 ppm of 10B, weighted with the Tilikidis weighting function

As a result of the weighting, the influence of the photon component decreases, while that of the alpha and lithium ions increases. If each individual distribution (photon, neutron, and BNC) is normalized to unity, Eq. (3) enables the calculation of the microdosimetric RBE for each component of the radiation field. The individual RBE values for the photon, neutron, and pure BNC components are 1.2, 4.2, and 4.4, respectively.

It is crucial to emphasize that the presented results are specific to the particular radiation field studied, and variations can be anticipated for different gamma/thermal and fast neutron/thermal neutron dose ratios. Furthermore, our investigation assumes a homogeneous distribution of 10B atoms within the walls of the sensitive volume, representing the cytoplasm surrounding critical nuclear structures. It is important to acknowledge that the actual scenario may differ significantly, considering the potential internal uptake of 10B atoms within the critical target. Notably, the measured BNC component with a TEPC uniquely corresponds to the cellular effect when 10B atoms are located in the tissue surrounding the sensitive target rather than within it. This uniqueness stems from the release of the BNC dose inside the cell, involving the simultaneous emission of two ions at an angle of 180°, presenting methodological challenges. The TEPC, designed to capture ionization events resulting from only one of the two ions produced in the A-150 walls, inherently yields event sizes generally smaller than those produced by BNC reactions within the biological target. Additionally, the potential existence of a heterogeneous distribution of the boron compound within cells and tissues introduces a complex dimension, leading to diverse biological outcomes [5].

4 Conclusion

Despite the acknowledged challenges related to the distribution of boron within and around biological targets, the microdosimetric approach stands as a powerful tool, offering critical insights into the intricate dynamics of Boron Neutron Capture Therapy (BNCT) within both cancerous and healthy tissues.

The potential applications of microdosimetry extend into the domain of treatment planning, providing opportunities for innovative strategies. Experimental microdosimetric spectra, measured at various depths within phantoms, serve as invaluable input data for advanced treatment planning methodologies [25].

To further explore the nuanced impact of different boron atom placements within biological tissue at the micrometer scale, especially considering heterogeneous distributions, we recommend the utilization of Monte Carlo simulations. These simulations offer a robust tool for comprehensive exploration and analysis. Additionally, the insights derived from Tissue-Equivalent Proportional Counter (TEPC) measurements play a crucial role in validating and refining Monte Carlo models, ensuring their accuracy and reliability in simulating the complex dynamics of BNCT.

Moreover, the inherent heterogeneity found in tumors, increasingly recognized as a major challenge in the treatment of solid tumors [26, 27], adds another layer of complexity. While replicating the real situation with experimental setups poses considerable challenges, microdosimetry, as a simplified model of reality, remains an essential tool for validating more versatile Monte Carlo models.

In conclusion, the application of microdosimetry in BNCT not only deepens our understanding of the underlying biological effects but also opens pathways for advancements in treatment planning and modeling. The integration of experimental data and sophisticated simulations is key to unlocking the full potential of BNCT as a targeted and effective cancer therapy.