Background

Dementia with Lewy bodies (DLB) is the second most common form of neurodegenerative dementia after Alzheimer’s disease, accounting for 5-10% of clinically diagnosed cases.1,2 Accurate diagnosis is important for clinical management, prognosis, and carer well-being,3,4,5 but initial misdiagnosis outside the specialist setting is common.2,6,7 Schneider et al found that about 15% of dementia cases at autopsy have Lewy bodies in the brain;8 further evidence that many people with DLB are being misdiagnosed during life.

Cardiac 123I-metaiodobenzylguanidine (MIBG) sympathetic innervation imaging is now included as an indicative biomarker in the fourth DLB consensus criteria, alongside 123I-FP-CIT SPECT.6 Cardiac MIBG studies are quantified on planar images using the heart-to-mediastinum ratio (HMR) as a diagnostic indicator, where HMR is the ratio between the count density in the cardiac left ventricle region (mean pixel value) and the count density in a mediastinum region of non-specific uptake. Abnormal scans have reduced cardiac uptake with a low HMR. Although the imaging technique is straightforward technically to implement in any hospital with a nuclear medicine service, the interpretation depends critically on establishment of a normal range, or cut-off for HMR values. We previously found that, provided that an appropriate normal HMR cut-off is used, HMR analysis is more accurate than visual rating.9 The HMR normal range will depend on the following:

  • Acquisition factors, including the camera and collimator

  • Processing factors, such as region placement

  • Population-specific factors, for example obesity and heart disease

We addressed processing factors in our previous publication,9 with a standardized 6-cm circular cardiac ROI recommended. We discussed the effects of camera calibration and population-specific factors9,10 and now address these here using images from older adults from a typical UK older population to generate a normal range.

It has long been known that we need to standardize cardiac MIBG image acquisition and processing parameters in order for cardiac uptake thresholds to be applicable between centers.11,12,13,14,15,16 Medium-energy collimators are recommended by the European Association for Nuclear Medicine for cardiac 123I-MIBG imaging as they give superior image quality for 123I, due to reduced septal penetration of high energy emissions.17 However, LEHR collimators are usually used in the USA, probably as they were used in the large ADMIRE-HF study in heart failure.18

Unfortunately, due to differences between vendors and collimator specification, it is not always possible to generalize results from any medium-energy collimator, as shown by the phantom studies of Nakajima et al19 and Verschure et al.14 Phantom calibration method can adjust for technical factors, for example collimator differences, but not for population differences.

In this study, we aim to

  • establish, for the first time, a normal HMR cut-off based on older adults without cognitive impairment in a typical UK population;

  • compare the diagnostic accuracy of this UK cut-off for detecting Lewy body disease in a cohort of dementia patients to that of calibrated cut-off values from the literature.

Methods

Phantom Calibration

Before starting cardiac 123I-MIBG imaging at our center, a calibration check on our gamma cameras was performed with help from colleagues in Japan and the Netherlands, using a planar phantom. The method developed by Nakajima et al11,19,20 was applied, which would allow us to see whether HMR cut-off values from Japanese studies needed to be modified for use locally. This calibration procedure revealed that the performance of our Siemens cameras fitted with medium-energy low-penetration (MELP) collimators was very similar to that of typical medium-energy general purpose (MEGP) collimators in Japan (personal communication, Prof. K. Nakajima, Kanazawa University, Japan). Calibrated HMR cut-offs published by Komatsu et al21,22 of 2.51 for early images and 2.20 for delayed images are therefore appropriate for use with our Siemens cameras with MELP collimators.

Subjects

Older Controls Without Cognitive Impairment

Thirty-one individuals aged over 60 years with normal cognition were recruited as part of a Newcastle University study into the use of cardiac MIBG in a representative UK population of older adults. They underwent a detailed neurological examination by a research physician, including rating for parkinsonism with the Unified Parkinson’s Disease Rating Scale (UPDRS), a thorough neurocognitive examination which confirmed they had normal cognition and an MRI brain scan reported by a radiologist as normal. Volunteers were excluded if they presented with cognitive impairment, major cerebrovascular disease on MRI brain imaging, Class II or worse heart failure according to the New York Heart Association classification, had had a myocardial infarction in the previous year, were taking labetalol, or were unlikely to be able to tolerate imaging. Subjects were not excluded if they had risk factors for cardiac disease because we aimed to recruit typical older adults with normal cognition. Diabetes has been shown to be associated with reduced cardiac MIBG uptake in a small proportion of patients with clinical autonomic neuropathy23 but similar to healthy controls in another study of patients with impaired glucose tolerance.24 In our clinical studies in Lewy body disease, we aim to assess cardiac MIBG in a “real-world” setting, where diabetes and risk factors for cardiac disease are extremely common. However, a detailed medical history including assessment of cardiac risk factors was taken.

Older People with Dementia

Seventeen patients meeting the clinical criteria for probable dementia with Lewy bodies (DLB) and 15 meeting criteria for dementia due to Alzheimer’s disease (AD) were recruited from memory clinics, as described in our previous publications.9,10 Briefly, patients had their dementia diagnoses confirmed by an expert panel of old-age psychiatrists. They were categorized as having probable DLB if two or more of the core consensus criteria for DLB6 (fluctuations, visual hallucinations, REM (rapid eye movement) sleep behavior disorder and parkinsonism) were present, and AD if none of these were present. The same broad general inclusion criteria as discussed above for the controls, including being aged over 60 years, were applied to ensure the patients were typical of the local population.

Image Acquisition and Processing

Subjects were administered 111 MBq I-123-MIBG via slow intravenous injection. Potassium iodate tablets (170mg) were given before and after injection to minimize uptake of free iodine by the thyroid. Ten minute anterior planar images were acquired at 20 minutes (early) and at 4 hours (± 30 minutes) after injection (delayed). All images were acquired on a dual-headed Siemens Symbia Intevo or Siemens Symbia gamma camera (Siemens Healthcare, Munich, Germany) with medium-energy low-penetration (MELP) collimators with energy window of 159 keV ± 10%, matrix size 128 × 128, and no zoom applied.

All images were processed on a Hermes workstation (Hermes Medical Ltd, Stockholm). Planar images were analyzed to obtain the heart-to-mediastinum ratio (HMR) using a 6 cm circular ROI placed over the left ventricle and 4 × 3 cm rectangular ROI between the lungs in the mediastinum, as described in our previous publication.9 Images were processed by a Medical Physics Expert (GR). Normal ranges based on control images were taken to be two standard deviations below the mean value and were calculated separately for early and delayed images.

Results

Control and Patient Demographics

Of the 31 controls recruited, 2 were excluded due to their image appearance (see below). The remaining 29 people included were aged between 62 and 94 years (mean 75.2). Most were male (22/29) and white (28/29). Body mass indices (BMI) ranged from 22 to 38 kg/m2, with the mean value of 27.2 in the overweight category. The demographics for controls and patients are summarized in Table 1. The controls and dementia patients are taken from concurrent clinical studies, so not all the same information was collected for dementia patients.

Table 1 Demographics summary for the 29 controls, 17 DLB, and 15 AD patients included in the study

Healthy Control HMR Results

Of the 31 healthy older adults recruited, one had a markedly abnormal appearance on early and delayed cardiac MIBG imaging and one had an abnormal appearance on delayed imaging. The HMRs were also clearly reduced for these patients, shown as empty symbols on the plot in Figure 1. It is well recognized that occult Lewy body disease (usually termed incidental Lewy body disease) is present in many older people,25 which may explain these findings. These two control scans with abnormal appearance were therefore excluded from analysis to determine local normal HMR cut-offs. Figure 1 also shows the distribution of early and delayed control HMRs and the change in HMR between early and delayed.

Figure 1
figure 1

Left: early HMR values (blue triangles) and delayed (orange circles) with means of 2.72 (SD 0.32) and 2.78 (SD 0.46), respectively, for the 29 included individuals. Excluded individuals are shown as empty symbols. Right: individual change in HMR between early (left) and delayed (right) images for each of the 29 controls

Eight of the 29 controls had HMR values below the Japanese cut-off of 2.51 on early images and 5/29 below 2.20 on delayed images.

Means, standard deviation and normal cut-offs (mean-2SD) derived from the control HMR data are shown in Table 2. The cut-off on the early images is higher due to the lower variation between individuals on the early images—the mean HMR is actually lower on the early images than delayed. With these cut-offs applied none of the 29 control subjects included were classified as abnormal.

Table 2 HMR data from healthy control subjects with normal scan appearances

Dementia Patient Results

The mean HMR values for the patients with Alzheimer’s disease and those with dementia with Lewy bodies from our previous work10 are given in Table 3 with the control results included for comparison. Figure 2 shows plots of the HMRs for controls, AD patients and DLB patients for early and delayed images, with the cut-offs determined from controls shown as red lines and the Japanese cut-offs as purple lines. Three of the AD patients have uptake below the local cut-off (i.e., unexpectedly low) on early images and two on delayed. Six of the DLB patients have uptake above the local cut-off on early images, five of which are well above and 1 borderline (HMR = 2.15). This latter patient’s HMR was 1.61, clearly abnormal, on the delayed image. The other five patients retained a normal HMR on the delayed images, and the images appeared normal visually. The corresponding sensitivity, specificity, and overall accuracy for DLB vs AD diagnosis using the local cut-offs are given in Table 4, alongside the results with the higher cut-off offs taken from the Komatsu et al multicenter follow-up study.22 The third row shows the results with the optimal cut-offs for this cohort that we determined via ROC analysis in our previous work,10 which was 1.61 to 1.70 for delayed images. Similar analysis for early images gave the highest accuracy for cut-offs between 1.77 and 1.80 for early images.

Table 3 HMR results (mean and standard deviation) for DLB and AD patients
Table 4 Sensitivity, specificity, and overall accuracy for DLB vs AD differential diagnosis for early and delayed images with HMR cut-offs taken from local controls, Japanese multicenter study, and ROC analysis of the data
Figure 2
figure 2

Heart-to-mediastinum ratios for controls, Alzheimer’s disease, and dementia with Lewy bodies patients for early images (left) and delayed images (right). The red lines show the local normal cut-off values, the green lines the optimal DLB vs AD cut-offs for data from our previous study,10 and the purple lines the calibrated cut-off taken from the Japanese multicenter 3-year follow-up study22

Discussion

Early and Delayed Cut-Offs

Eight of the 29 controls included in our study had HMR values below the early image cut-off of 2.51 from the Japanese follow-up DLB study22 on early images and five of 29 below 2.20 on delayed images, providing strong evidence for lower thresholds for this population. We note that the value from the Japanese normal database (patient-based) gives a normal range of 2.2 to 4.0 for early images (mean 3.1) and 2.2 to 4.4 for delayed images (mean 3.3).26 These HMR ranges seen in Japanese patients without Lewy body disease do not seem to fit our UK population. If typical true HMR results vary between populations phantom calibration will not be sufficient, and measurements of local controls will be required. In our previous work, we saw that when we applied a normal HMR cut-off derived from Japanese studies to our dementia patients it seemed much too high, with several Alzheimer’s disease patients falling into the abnormal category.10 Our cut-offs of 2.07 (early) and 1.86 (delayed) reflect the higher variation in normal delayed HMR results compared with early results, rather than a difference in the mean, which is in line with previous studies.27,28,29 These results were obtained using Siemens cameras fitted with MELP collimators, and as such cannot be directly applied to other centers using other collimators.

For early images, the use of the local control cut-off rather than the cut-off of Komatsu et al (2.07 vs 2.51) resulted in two fewer DLB patients being correctly assigned as abnormal, one of whom had a less equivocal HMR on delayed imaging. The specificity was greatly increased with the local cut-off as only seven AD patients had early HMRs above the Japanese cut-off of 2.51. Use of the Japanese normal database result of 2.2 instead of 2.51 would not have affected either the sensitivity or specificity of early imaging. With delayed imaging, the sensitivity and specificity was the same with both the local cut-off and that of Komatsu et al (1.86 vs 2.20) as there were five DLB patients with clearly normal looking scans (lowest HMR = 2.39) and there were no AD patients falling between the two cut-offs. It is worth noting however that five control subjects have HMR values between the two thresholds, suggesting our local values are more appropriate.

Comparison with Other Studies

At our center, we focus on the use of cardiac MIBG imaging for the diagnosis of Lewy body disease. However, as well as the differential diagnosis of Lewy body disease from other conditions,30,31,32 cardiac MIBG scintigraphy is also used for the estimation of risk of sudden cardiac death in heart failure patients and predicting those who could benefit from an implantable cardiac defibrillator (ICD)18,33 and this is currently the principal application in Europe and the US.34 The largest study of healthy controls recruited to a cardiac MIBG research study (n = 94) was that of Jacobsen et al35 carried out as part of the ADMIRE-HF heart failure study.18 The controls were aged between 29 to 82 years (mean 58.5 ± 10.6 years, with BMI between 20.4 and 44.3 (mean 28.7 ± 5.1). These individuals were carefully screened to exclude coronary heart disease or Type 1 or 2 diabetes, making them healthier on average than the typical population. Although a substantial proportion (39/94) were aged over 60 years, the inclusion requirement of having an annual cardiac risk of less than 10% would further increase the difference in health between these individuals and a typical elderly population. We regard the inclusion of adults with diabetes and coronary risk factors as a strength of our study in relation to the diagnosis of Lewy body disease, rather than a weakness. In recruiting older adults that aside from having no cognitive impairment are similar in general health to that of the general population, we are determining a true normal range. However, despite this, the controls from the US ADMIRE study have lower HMR values on average than those from our study, even accounting for the different collimator. Using LEHR collimators, their result was 1.78 ± 0.22, which would correspond to normal cut-off of 1.34 with LEHR (two standard deviations below the mean of 1.78). This cut-off converts to around 1.6 with our camera with MELP (determined using calibration data for LEHR collimators from Ref. 36). One possible reason for the lower HMRs in the United States may be differences in the method used to delineate the cardiac ROI, with a freehand method used in the ADMIRE-HF study. We found previously that this tends to give lower HMR values than a 6-cm circular region.9 Interestingly, the a priori cut-off used in the ADMIRE-HF study was much higher than subsequently seen in the controls: 1.6 with LEHR collimators, equivalent to around 2.1 for our camera with MELP (again, determined using calibration data for LEHR collimators from Ref. 36). The a priori cut-off for the ADMIRE-HF study was determined by pooling the literature on cardiac MIBG values in healthy controls (n = 202).37 This was done without population or camera differences being taken into account, which may account for the discrepancy.

To our knowledge, the only other recent UK study in controls is by Asghar et al in Manchester, again using LEHR collimators and excluding people any history of diabetes or cardiac disease.27 In 14 healthy adults (mean age 54.6 ± 5.4 years, mean BMI 27.0 ± 3.1), they found a mean early HMR of 1.67 ± 0.13 and delayed 1.73 ± 0.16, which would give cut-offs (mean − two standard deviations) of 1.41 for both early and delayed images. This cut-off is equivalent to around 1.7 for our camera with MELP collimation36 and therefore similar to our result for delayed imaging of 1.86. Since there is thought to be little dependence of HMR on age38,39 we would expect any differences between our controls and this study to be due to the inclusion criteria, as it seems unlikely that there would be large differences between the population of Manchester and Newcastle, two post-industrial cities in northern England. The fact that our mean and cut-off values are slightly higher rather than lower than the Manchester study, after accounting for camera differences, provides further indirect support of our previous findings that cardiac disease is unlikely to have a big impact on HMR,10 and further supports the small body of evidence suggesting HMR is not strongly linked to age.

Effect of Cardiovascular Disease on HMR

It is known that regional cardiac sympathetic denervation assessed with SPECT or PET sympathetic imaging is associated with both areas of myocardial infarction (MI) and also with ischaemia in patients without previous MI.40,41,42 The impact of MI and coronary artery disease (CAD) on global uptake assessed via planar cardiac MIBG scintigraphy is not well established; however, most studies in healthy controls exclude those with CAD. To our knowledge there are no large studies that evaluate the impact of regional sympathetic denervation associated with MI and CAD on planar HMR. In this study and our previous work,10 one control and three AD patients had experienced an MI, all more than 1 year prior to scanning. The control showed clear regional defects on SPECT imaging but borderline-normal delayed planar HMR (2.14). The three AD cases showed defects of varying magnitude and extent on SPECT and planar HMRs of 2.99, 2.16, and 2.59. The numbers are too low to establish a correlation between the extent of regional denervation and effect on global HMR but it seems likely that regional denervation would have to be widespread for planar HMR to be abnormally low. Given the variation between healthy individuals, it is unlikely that MI and CAD alone would give an abnormal HMR result in individual cases. However, it is likely that both the mean and standard deviation of a group of subjects would be affected by heart disease, suggesting that differences in the prevalence of cardiovascular disease between populations could contribute to differences in HMR cut-offs. In the UK, the prevalence of cardiovascular diseases increases with age, with 8% of people between 60 and 70 years of age and 24% of those over 80 reporting at least one symptom.43 In England, the prevalence of coronary heart disease over all ages in 2018 was reported by the National Health Service to be 3.1%.44 Equivalent data for Japan is not published but we note that a large cohort study (Akita-Osaka study) showed 0.2% incidence of new coronary heart disease cases in urban Japanese men aged between 60 and 69 between 1996 and 200345 suggesting a much lower prevalence in this age group than in the UK.

Effect of Obesity on HMR

Obesity has been linked to decreased apparent cardiac MIBG uptake by two previous studies,38,46 and is thought to be due to increased attenuation and scatter in larger individuals, rather than to true differences in cardiac uptake. The ADMIRE control study discussed above38 had a wide BMI range of 20.4 to 44.3 kg/m2 (mean 28.7) and reported a “modest” decrease in HMR with BMI. This was a study into aging and the incidental finding of a relationship between BMI and HMR was not discussed in detail. Pellegrino et al reported lower HMRs in patients with BMI > 30 (n = 10) compared to those with BMI < 30 (n = 35).46 This was in heart failure patients and the results have not yet been replicated in healthy controls but do suggest patient size may affect apparent cardiac MIBG uptake. In our study, the mean BMI of the controls was 27.8 kg/m2 which is representative of the mean value of 27.3 for 119,230 men aged 40-69 recruited into the UK biobank study between 2006 and 2010.47 The Manchester study had a similar mean BMI of 27.0. In contrast, the mean BMI in a Japanese longitudinal study of 104,928 individuals was 22.6 in men and 22.9 in women.48 A relatively higher proportion of overweight individuals may therefore contribute to the lower mean HMRs in the UK and US compared to Japan. Future work using attenuation and scatter corrected SPECT imaging may provide insight into the contribution of patient size to these population differences in apparent cardiac MIBG uptake.

Limitations

This study was retrospective in nature, utilizing control and patient data recruited to our center for wider Lewy body disease studies. The inclusion of older adults with a range of cardiac risk factors and without detailed cardiac assessment to detect undiagnosed cardiovascular disease makes it difficult to determine the impact of cardiac disease on HMR. Very large samples of older adults with and without risk factors for cardiovascular disease could help to determine this, but as seen here, older people often have cardiac risk factors, making it impractical to recruit true healthy controls. As obesity is both a risk factor for cardiovascular disease and a potential cause of artifactually reduced HMR, this would also need to be accounted for. If it were feasible, a large prospective study in multiple countries recruiting cognitively normal older adults typical of the local population and scanning them using the same protocol could determine whether HMR is population specific in a more rigorous manner.

New Knowledge Gained

We determined cardiac 123I-MIBG heart-to-mediastinum ratio cut-off values derived from a group of older adults in average health without evidence of cognitive decline for the first time. After accounting for camera differences, these values were lower than the values used in Japan, higher than in a large US study but similar to those found in another UK center.

Conclusion

Our results suggest that normal cardiac 123I-MIBG heart-to-mediastinum ratio values may be population specific, although processing differences cannot be excluded. Further work is required to determine the reasons for these differences, which could include contributions from heart disease, obesity, and other co-morbid conditions. This work highlights the importance of deriving HMR cut-offs from controls that are representative of the population being studied.