Introduction

Free fatty acid is a main myocardial energy source under normal conditions, but under ischemic conditions, myocardial metabolism preferentially shifts from aerobic β oxidization of free fatty acid in mitochondria to anaerobic glycolysis in cytosol.1,2 Under a condition called post-ischemic metabolic stunning, reduced uptake of 123I-β-methyl-p-iodophenlypentadecanoic acid (BMIPP), which reflects impaired myocardial β oxidization is prolonged despite restoration of perfusion.3,4 Such a perfusion-metabolism mismatch, where the deficit of free fatty acid metabolism is discordantly more severe than that of perfusion, is known to be predictive of myocardial viability and prognosis in chronic ischemia5,6 and in the (sub)acute phase of myocardial infarction,7,8 and can be evaluated by dual-isotope nuclear imaging. 99mTc-labelled isotopes (t1/2= 6 hours) are theoretically preferable perfusion radiotracers to 201Tl-based ones (t1/2 = 73 hours), as they have a shorter physical half time and higher photon-energy, leading to less attenuation artifact and radiation exposure.9 However, as the photo-peaks of 99mTc (140 keV) and 123I (159 keV) are close, quantitative assessment using simultaneous acquisition of 99mTc-/123I-labelled dual tracers is challenging.10,11 Recently, semiconductor cadmium-zinc-telluride (CZT) detectors have been launched in clinical settings, and they have provided shorter scanning times while improving energy and spatial resolution in comparison with conventional NaI scintillation detectors.10,12,13,14,15 More recently, a novel SPECT scanner with CZT detectors, the Discovery NM/CT 670 CZT (GE Healthcare, Milwaukee, Wisconsin), was developed for whole-body imaging including cardiac SPECT imaging. This novel scanner is equipped with 130 pixelated detectors with registered collimation to improve the overall system resolution, as well as the energy resolution; although evidence of the benefits of these features from clinical settings is currently lacking, these are promising properties for multi-tracer imaging.

In our current study, we sought to evaluate the utility of a Discovery NM/CT 670 CZT for simultaneous 99mTc-tetrofosmin (TF)/123I-BMIPP dual-isotope acquisitions by comparing dual-tracer imaging with single-tracer imaging, in patients having undergone primary percutaneous coronary intervention (PCI) for acute myocardial infarction (AMI).

Methods

Study Design and Patient Selection

Patients aged over 20 years who underwent PCI for AMI from December 2017 to June 2018 at Saitama Medical University, International Medical Center, were prospectively recruited for this study (UMIN ID: 00029768). Patients with an artificial pacemaker, potential pregnancy, potential inability to complete the examination due to mental or physical reasons, or recent myocardial perfusion/metabolism imaging, were excluded. This study was approved by the Institutional Review Board of Saitama International Medical Center (reference number: 17-126) and Saitama Medical University (reference number: 17-094). All patients signed written informed consent.

Clinical Data Collection

Baseline demographics, electrocardiographic findings, and data regarding disease severity and in-hospital course were recorded.

Examination Protocol

After completion of in-hospital cardiac rehabilitation, patients underwent a 1-day 99mTc-TF (Nihon Medi-physics Co., Ltd., Tokyo, Japan) and 123I-BMIPP (Nihon Medi-Physics Co., Ltd.) dual-isotope imaging examination at Saitama Medical University Hospital. All examinations were performed under electrocardiographic gating, as recommended for quantification.16 Patients were assigned to either a 99mTc-TF or 123I-BMIPP first protocol (Figure 1). In the 99mTc-TF first protocol, single-tracer 99mTc-TF gated-SPECT/CT imaging was first acquired followed by a dual-tracer gated-SPECT/CT acquisition. In the 123I-BMIPP first protocol, single-tracer 123I-BMIPP gated-SPECT/CT was first acquired followed by a dual-tracer gated-SPECT/CT acquisition.

Figure 1
figure 1

Schema of the examination protocols. Patients were assigned to either a 99mTc-TF or 123I-BMIPP first protocol. All acquisition times were set for 16 minutes. 99mTc-TF and 123I-BMIPP were injected 30 and 15 minutes prior to initiation of acquisition, respectively. The injected isotope doses were fixed at 600 and 111 MBq for 99mTc-TF and 123I-BMIPP, respectively

SPECT Data Acquisition and Post-processing of Images

The Discovery NM/CT 670 CZT has 130 CZT modules and wide-energy high-resolution collimators. Each detector contained 32 × 32 pixelated 2.46 × 2.46 mm CZT elements. Using the Lister tool (GE Healthcare) on a Xeleris 4.0 workstation (GE Healthcare), images with 30 views/scan and 65 seconds/view were reframed. The peak energy and window widths were set to 140 ± 7.5% and 159 ± 7.5% KeV for 99mTc-TF and 123I-BMIPP, respectively, as recommended by GE Healthcare. Using the Myocardial Evolution software for 99mTc and 123I (GE Healthcare) supplied on the same workstation, images for quantitative gated-SPECT (QGS) and quantitative perfusion SPECT (QPS) were reconstructed to obtain views in the standard axes (short axis, vertical long axis and horizontal axis) and polar maps of the left ventricle. The iterative algorithm with integrated collimator geometry modeling used maximum penalized likelihood iterative reconstruction.17 A Butterworth post-processing filter (critical frequency, 0.4 cycle/cm; order, 10) was applied to the reconstructed slices.17 Attenuation correction was not applied because it would have resulted in low uniformity.10 Manual processing was applied as necessary to adjust the left-ventricular axis, center, and endocardial/epicardial contour with masking/constraint of extracardiac accumulation.

Image Analysis

The left-ventricular myocardium was divided in to 17 segments,18 and the vascular territories corresponding to the three major coronary arteries were defined, with the anterior, septal and apical regions being assigned to the left anterior descending artery (LAD), the inferior region to the right coronary artery (RCA), and the lateral region to the left circumflex artery (LCX).18 For the QGS with 99mTc-TF, the following parameters were assessed. Left-ventricular wall motion (mm) and wall thickening (%) in each segment, and the summed values per coronary territory were recorded. Semi-quantitative scales for the wall motion and the wall thickening were automatically assigned to each segment to calculate the summed motion scores (SMS) and summed thickening scores (STS) of all segments.19 Left-ventricular end-diastolic volume (EDV, ml), end-systolic volume (ESV, ml), and ejection fraction (EF, %) were recorded. For the QPS with 99mTc-TF and 123I-BMIPP, the following parameters were assessed. Semi-quantitative scales for uptake of radioactive counts was automatically assigned to each segment to calculate the summed rest scores (SRS) of all segments, and the summed values per coronary territory were recorded.20 The mismatch scores were then calculated as SRS for BMIPP-TF. The perfusion extent (%) and the total perfusion deficit (TPD, %) were recorded. A summary and details of the quantitative and semi-quantitative analysis are shown in Supplementary Table 1 and Figure 1.

For the visual analysis, two experienced nuclear radiologists (I.M. and K.F.) interpreted the dual-tracer imaging for the presence or absence of infarction and perfusion-metabolism mismatch (stunning/hibernation) in the global left ventricle (i.e., per patient basis), and the LAD, RCA and LCX coronary territories.

Statistical Analysis

Continuous variables are expressed as the mean ± standard deviation or median [first-third quartile], and categorical variables as number (%). The Shapiro–Wilk test was performed to test for normal distributions.

The correlation between the quantitative values obtained from dual- and single-tracer imaging was evaluated using intra-class correlation (ICC; two-way mixed-effects model, for consistency) and Bland–Altman analysis. Repeatability and reproducibility of the quantitative analysis with respect to intra-(S.N.) and inter-(S.N. and I.M) observer variability were evaluated using ICC (two-way mixed-effects model for absolute agreement) and ICC (two-way random-effects model, for absolute agreement), respectively, with the reframed dual-tracer images of ten patients being randomly assigned to each analysis.

The concordance in the visual analysis was evaluated using unweighted κ statistics between the interpretations of the two radiologists.

Statistical analyses were performed using JMP Pro 11.2.0 (SAS Institute Inc., Cary, NC, USA) or SPSS statistics software (version 25, IBM Corp., Armonk, NY, USA), as appropriate.

Results

Clinical Characteristics of the Studied Patients

From December 2017 to June 2018, 110 consecutive patients who underwent primary PCI for AMI at Saitama Medical University, International Medical Center, were screened. After application of the exclusion criteria, 31 patients participated in the study, with one patient being further excluded from the final analysis because of a suspected CD36 deficiency (Supplementary Figure 2). The characteristics of the remaining 30 patients (99mTc-TF first protocol: n = 14, 123I-BMIPP first protocol: n = 16) are shown in Table 1.

Table 1 Patient characteristics

Concordance in the Visual Analysis of Dual-Tracer Imaging

The Cohen’s κ coefficients between the two radiologists for the presence of infarction and perfusion-metabolism mismatch were 0.85 (P < 0.001) and 1.00 (P < 0.001), respectively for the global left ventricle, and 0.94 (P < 0.001) and 0.89 (P < 0.001) for the coronary territory-based regions of the left ventricle (a total of 90 territories).

Quantification and Semi-quantification of Dual- and Single-Tracer Imaging

The ICC coefficient with 95% confidence interval and the Bland–Altman mean of difference with 95% limits of agreement between the values obtained from dual- and single-tracer imaging are shown in Figures 2, 3, and 4. The ICC coefficients between dual- and single-tracer imaging were high (> 0.91) in all the data. The correlations were high in the QGS for the global left ventricle (Figure 2), except for apparent outlying EDV (Figure 2C) and ESV (Figure 2D) values in a patient with a severely dilated left ventricle and prominent extra-cardiac accumulation. No obvious tracer-specific trend was observed in the QPS results from the global left ventricle (Figure 3), and no obvious coronary territory-related heterogeneity was observed in the QGS and QPS regional left ventricle data (Figures 4, 5; Supplementary Figure 3).

Figure 2
figure 2

Quantitative Gated SPECT (QGS) for the global left ventricle: dual- vs single-tracer imaging. Upper graphs: Intra-class correlation (ICC) between the QGS data for the global left ventricle obtained from dual- (x axis) and single- (y axis) tracer imaging. Lower graphs: Bland-Altman analysis showing mean (x axis) and difference between (y axis) the corresponding two values. CI, confidence interval; LOA, limit of agreement

Figure 3
figure 3

Quantitative Perfusion SPECT (QPS) for the global left ventricle: dual- vs single-tracer imaging. Upper graphs: Intra-class correlation (ICC) between dual- and single-tracer QPS data for the global left ventricle, Lower graphs: Bland-Altman analysis. CI, confidence interval; LOA, limit of agreement. Purple circle: 99mTc-TF, Green diamond: 123I-BMIPP

Figure 4
figure 4

Quantitative Gated/Perfusion SPECT (QGS/QPS) per coronary territory: dual- vs single-tracer imaging. Upper graphs: Intra-class correlations (ICC) between dual- and single-tracer QGS/QPS data per coronary territory. CI, confidence interval; LOA, limit of agreement. White circle: Left anterior descending artery, red square: right coronary artery, blue triangle: left circumflex

Figure 5
figure 5

Representation of dual- and single-tracer Quantitative Perfusion SPECT (QPS) imaging. Quantitative Perfusion SPECT images for 99mTc-TF obtained from single-(99mTc-TF alone) and dual-(99mTc-TF and 123I-BMIPP) tracer. (A) a 77-year-old man with right coronary artery occlusion showing a matched defect pattern (mismatch score: 1). (B) a 57-year-old man with left anterior descending artery occlusion showing a mismatched defect pattern (mismatch score: 22)

Repeatability and Reproducibility

The ICC coefficients for the repeatability and reproducibility analysis were high (> 0.93) in all the data; the intra- and inter-observer ICC coefficients were 1.00 and 0.99 for SMS, 1.00 and 0.99 for STS, 1.00 and 1.00 for EF, 0.98 and 1.00 for SRS on 99mTc-TF, and 0.99 and 0.98 for SRS on 123I-BMIPP. The details are shown in Supplementary Table 2.

Discussion

Main Findings

Our current study first demonstrated that a Discovery NM/CT 670 CZT scanner could be used for simultaneous 99mTc-TF /123I-BMIPP dual-tracer acquisitions, affording clinically available and reproducible images in patients with AMI having undergone primary PCI. The quantitative and semi-quantitative values for global and regional left-ventricular function, perfusion, and fatty acid metabolism obtained from the simultaneous 99mTc-TF /123I-BMIPP dual-tracer isotope acquisitions were closely comparable to those obtained in the single-tracer mode, possibly owing to the properties of the novel scanner.

Utility of CZT Detectors for Myocardial Perfusion and Cardiac Function in Coronary Artery Disease

Currently, there are two semiconductor CZT camera systems available commercially, the Discovery NM/CT series (530c and 670 CZT) and the D-SPECT (Spectrum Dynamics, Biosensors, Caesarea, Israel). Although the collimation and reconstruction algorithms are different, the heart-centric techniques of the Discovery NM/CT 530c and D-SPECT, and the specific reconstruction algorithms facilitate enhanced count sensitivity, improved energy, and higher spatial resolution than conventional NaI crystal detectors.13,21,22,23,24 Consequently, the advent of these CZT detectors has led to improved image quality, and a reduction in acquisition time and/or radiation exposure.17,25,26

In patients with known or suspected coronary artery disease, fast myocardial perfusion imaging using CZT-based detectors has provided better diagnostic accuracy with improved regional sensitivity compared with conventional cameras.27,28 Left-ventricular global and regional functions have also been assessed using CZT-based scanners, with quantification of EF and the regional extent of scarring being similar to measurements made on magnetic resonance imaging,29 and the EF demonstrating higher reproducibility than achieved with a conventional NaI camera.30 Our study demonstrated that the quantitative and semi-quantitative parameters obtained for myocardial perfusion and cardiac function were highly reproducible, both for global left ventricle and regional measures (Supplementary Table 1).

Simultaneous 99mTc-TF /123I-BMIPP Dual-Isotope Acquisition Using CZT Detectors

Areas showing a perfusion-metabolism mismatch may correspond to jeopardized but viable myocardium as functional improvement in such areas has been observed after revascularization.7,8,31,32,33123I-BMIPP is a commonly used tracer for evaluation of myocardial-free fatty acid metabolism in clinical studies.3,5,6,7,8,34 On the other hand, for myocardial perfusion, 201Tl, which has been recognized as providing a diagnostic accuracy similar to the 99mTc-family of tracers in most coronary artery disease35, may frequently be used in simultaneous dual-isotope imaging in combination with 123I-BMIPP, as the distant photo peaks of the two tracers facilitate enhanced energy resolution.36,3799mTc-TF may be a theoretically preferable perfusion tracer in respect to better image quality, as it has a shorter half time and administration of higher dose is possible. However, when used in combination, the close photo-peaks of the 99mTc- and 123I-isotopes require a special attention to discriminate their energy properties while maintaining count sensitivity. To ameliorate potentially interfering factors in dual-tracer imaging using conventional NaI scanners such as scatter contamination of isotopes, cross-talk from the primary photons of each radionuclide, attenuation and distance-dependent collimator responses, techniques using constrained spectral factor analysis, artificial neural networks and Monte Carlo-based joint ordered-subset expectation maximization iterative reconstruction have been applied, although all required complicated data processing22,38,39. Instead, using CZT scanners, excellent concordance between 99mTc-/123I-labeled dual- and single-images has been demonstrated in both phantom and animal studies,10,40 with the excellent spatial and energy resolution of the CZT scanners allowing the count sensitivity to be maintained. A previous phantom study using similar CZT-scanner systems showed improved energy resolution for dual-tracer imaging (300 MBq of 99mTc-TF and 100 MBq of 123I-BMIPP), with only 4.0% downscatter contamination from 123I-BMIPP in the energy window of 99mTc-TF and almost no upscatter contamination from 99mTc-TF in that of 123I-BMIPP.10 A clinical study on simultaneous 99mTc-MIBI/123I-BMIPP dual-tracer imaging using the D-SPECT system showed the correct detection (in comparison with angiographic findings) of infarcted and stunned areas in patients who underwent revascularization for AMI.11 In our current clinical study, we demonstrated images, and quantitative and semi-quantitative values obtained from 99mTc-/123I-labeled dual-tracer imaging on a Discovery NM/CT 670 CZT that were analogous to those obtained from single-tracer imaging with fixed window settings in most of the studied patients. Exceptionally, one patient with a severely dilated left ventricle with low LVEF (LVEDV/LVESV = 178/145 ml, LVEF 19% in dual-mode) revealed a prominent discrepancy between the single- and dual-tracer data. In this patient, enhanced extra-cardiac accumulation and weak cardiac uptake might have led to inaccurate QGS analysis. Furthermore, another outlier that revealed discrepant results in the QPS for BMIPP demonstrated a failure in the automatic and manual QPS processing to adjust the endocardial/epicardial contour. The Discovery NM/CT 670 CZT has each collimator registered with a single semi-conductor detector pixel, which affords high resolution (6.4 mm at center) and image contrast. Together with the thick detector (7.5 mm in thickness) designed for improved count sensitivity, these properties of the scanner are beneficial for multi-tracer imaging. Our current study demonstrated remarkable interrater agreement in visual interpretation of infarction and perfusion-metabolism mismatch in the global and regional left ventricle, suggesting promising clinical feasibility of the Discovery NM/CT 670 CZT for simultaneous 99mTc-TF /123I-BMIPP dual-tracer imaging. The coronary territories defined by the radiologists were also compatible with angiographic findings. Disagreement on the infarcted area was observed in patients with modest infarction (n = 2), and on the perfusion-metabolism mismatched area in patients with mildly impaired free fatty acid metabolism (n = 1) and concomitant hibernation (n = 1). The patients with a visually determined mismatched defect (n = 17) tended to show a higher mismatch score (SRS for BMIPP-TF) than those without such a defect (n = 13) (6 [2,3,4,5,6,7,8,9,10,11,12]) vs 2 [2,3,4], P = 0.095, Supplementary Figure 4).

Limitations

This study has some limitations. First, it is a single-center, observational study with a relatively small number of patients. Using R version 3.5.1, package ICC.Sample.Size, the p0 (ICC coefficient for null hypothesis) was 0.60 to achieve a power of 0.80, when p (ICC coefficient obtained in the study) = 0.90, α = 0.05, and N = 14. However, we suggest that the correlation between the values obtained from dual- and single-tracer imaging, as well as intra- and inter-observer variability, are sufficient to propose the clinical feasibility of the technique. Further confirmative studies on larger numbers of patients are needed to ensure that our findings apply to heterogenous clinical cases, particularly those showing low LVEF. Second, the time elapsed from AMI onset to examination varied, with a median value of 25 days.16,17,18,19,20,21,22,23,24,25,26 Therefore, post-ischemic functional recovery owing to revascularization may not have been appropriately predicted, as metabolic alterations may have been resolved in some patients with delayed examination.41

New Knowledge Gained

We demonstrated the clinical feasibility of the Discovery NM/CT 670 CZT with a novel CZT detector for simultaneous 99mTc-TF /123I-BMIPP dual-tracer isotope acquisitions in patients with AMI having undergone primary PCI. The quantitative and semi-quantitative values obtained from the dual-tracer mode for global and regional left-ventricular function, perfusion, and fatty acid metabolism were closely comparable to those obtained in the single-tracer mode, possibly owing to the improved energy and spatial resolution as well as count sensitivity of the new scanner.