1 Introduction

Amyloid positron emission tomography (PET) imaging can reveal in vivo brain amyloid-β plaques, a pathological hallmark of Alzheimer’s disease (AD). The 11C-labeled Pittsburgh Compound B (11C-PiB) is a widely used amyloid imaging agent in research [1]. Dynamic 11C-PiB PET imaging with kinetic analysis has been performed for the accurate quantification of amyloid deposition [2].

On a related note, PET/computed tomography (CT) cameras with three-dimensional (3D) acquisition are commonly used in oncological whole-body (WB) PET imaging due to higher sensitivity and shorter scanning time than cameras with two-dimensional (2D) acquisition. However, 3D acquisition may be influenced by singles and random coincidence, which diminish image quality and quantitative capability [3,4,5,6,7,8]. In the brain 11C-PiB dynamic scan using 3D acquisition, the injected activity is mostly distributed throughout the body, particularly in the early phase post-injection. Such activity outside the field of view (FOV) can lead to errors in the quantification of brain PET images within the FOV [9,10,11,12].

Despite this, the effect of body activity on brain 11C-PiB dynamic scans has not been evaluated, because no studies have gathered detailed data on 11C-PiB biodistribution in the neck, chest, and WB, with the exception of one study addressing radiation dosimetry [13].

In the present study, we acquired the early phase biodistribution data of 11C-PiB in various axial body areas in subjects who had been enrolled in various brain research protocols and had undergone a brain PET scan in the later phase. We then performed phantom experiments based on the actual human biodistribution data using a number of PET cameras in 3D mode. Finally, based on these phantom experiments, the effects of body activity on the quantitative capability of brain 11C-PiB PET and on kinetic analysis were estimated.

2 Materials and methods

2.1 Subjects

In the present study, a total of 10 subjects were enrolled in various research protocols, each focusing primarily on the acquisition of late-phase brain 11C-PiB PET images. In addition, we took advantage of the accumulation phase and set up an add-on protocol to measure sequential activity over time in each axial body segment. Approval was obtained from the ethics committee and informed consent was obtained from each subject for the primary study, the add-on protocol, and the subsequent data analysis.

The demographic profiles of the subjects and scans are summarized in Table 1 and described below. Six subjects with mild cognitive impairment (MCI) or suspected MCI (mean age 72.2 years, range 65–82 years) underwent a 2D WB or neck-to-chest dynamic scan with 11C-PiB. Two healthy subjects and one patient with AD (mean age 59.9 years, range 51–64 years) underwent a 2D brain dynamic scan with 11C-PiB. A tenth subject, a 63-year-old patient with AD, underwent a brain 3D dynamic 11C-PiB scan using another PET camera. None of the subjects had any cardiovascular, respiratory, or renal diseases that could have influenced WB biodistribution or pharmacokinetics.

Table 1 Subject data and scan information

2.2 PET scanner and scan mode

We used an ECAT EXACT HR + scanner (Siemens/CTI, Knoxville, TN) in 2D mode to acquire whole-body 11C-PiB biodistribution [14]. To allow quantitative measurement of each axial body area, dynamic WB, neck-to-chest, and brain PET scans were performed in 2D acquisition mode.

In the subsequent phantom study, we used six PET camera models and performed phantom scans in 3D mode. These six PET cameras were considered to be used for brain 3D dynamic 11C-PiB PET scans in the Japanese Alzheimer’s disease neuroimaging initiative (J-ADNI) project [15]. These 3D scans were performed in accordance with the standard PET imaging protocol detailed in the J-ADNI. Table 2 shows the PET camera specifications and scan parameters that were used in the phantom study.

Table 2 PET camera specifications and scan parameters in the phantom study

The tenth subject (J) underwent a 11C-PiB brain 3D dynamic scan using a Shimadzu SET-3000G/X (Table 2) to evaluate the effect of body activity on the brain PET acquired and analyzed in the standard J-ADNI protocol.

2.3 Dynamic whole-body and neck-to-chest PET scans to measure 11C-PiB biodistribution

Dynamic WB PET scans were performed on three subjects (A, B, and C) to measure the biodistribution of 11C-PiB from the top of the head to the mid-thigh. Dynamic neck-to-chest scans were performed on another three subjects (D, E, and F) to measure the 11C-PiB biodistribution from the neck to the chest area in more detail.

A transmission scan was performed using a 68Ge/68Ga rod source before 11C-PiB injection. The transmission scan protocols were as follows: 3 min/bed × 8 beds for the WB scan and 3 min/bed × 2 beds for the neck-to-chest scan. A dynamic emission scan was performed for a total period of about 40 min starting immediately after injection of 11C-PiB. The WB emission scan protocol was as follows: 30 s/bed × 8 beds, repeated six times, starting at approximately 0, 7, 14, 21, 27, and 35 min after injection. The neck-to-chest emission scan protocol was as follows: 30 s/bed × 2 beds, five repeats, 45 s/bed × 2 beds, two repeats, 60 s/bed × 2 beds, two repeats, and 90 s/bed × 2 beds, one scan; therefore, the protocol consisted of a total of 10 repeats (starting at approximately 0, 2.25, 4.5, 6.75, 9, 12, 15.5, 19, 25, and 35 min after injection). Subsequently, a 20-min (4 frames × 300 s) emission scan of the brain was performed starting 50 min after injection to evaluate brain 11C-PiB uptake. Finally, a transmission scan of the head was performed for 6 min.

The acquired WB dynamic data were reconstructed using Fourier rebinning and a filtered back projection (FBP) algorithm with a Gaussian post-filter [8-mm full-width at half-maximum (FWHM)]. In addition, to locate the regions of interest (ROIs), PET data were reconstructed using the ordered-subset expectation–maximization (OSEM) algorithm (4 iterations, 8 subsets) with a Gaussian post-filter (8-mm FWHM). Late-phase brain data (50–70 min after injection) were also reconstructed using the OSEM algorithm (4 iterations and 16 subsets) without a post-smoothing filter [12]. The brain images of 50–70 min after injection were visually interpreted and classified as either positive or negative by an expert 11C-PiB PET physician in accordance with the J-ADNI criteria [16].

2.4 Dynamic brain PET scans to measure brain 11C-PiB activity

The brain 2D dynamic PET scans were performed on three subjects (G, H, and I) to measure the time course of brain 11C-PiB for a 90-min period after injection of 11C-PiB. The 2D emission scan protocol was as follows: 15 s × 4, 30 s × 8, 60 s × 9, 180 s × 2, 300 s × 8, and 600 s × 3 for a total of 34 frames. A 15-min transmission followed. PET data were reconstructed using filtered back projection with a Gaussian post-filter (8-mm FWHM). Images of 50–70 min after injection were generated and visually interpreted in terms of brain 11C-PiB uptake.

Another subject (J) underwent a brain 3D dynamic emission scan for 70 min starting immediately after injection based on the standard J-ADNI protocol (10 s × 6, 20 s × 3, 60 s × 2, 180 s × 2, and 300 s × 12 for a total of 25 frames). A 6-min transmission scan followed. PET data were reconstructed using the OSEM (4 iterations and 16 subsets) without a post-smoothing filter. Images of 50–70 min after injection were generated and visually interpreted in terms of brain 11C-PiB uptake.

2.5 Phantom experiments

In the phantom experiments, we assumed a 11C-PiB injection dose of 555 MBq and a 70-min dynamic scan. As shown in Fig. 1, we designed a set of phantoms simulating the 11C-PiB biodistribution in the human body; these consisted of a cylindrical phantom (16-cm diameter, 15-cm length; Itoi Factory Inc.) for the head, three saline bags (500 mL; Otsuka Pharmaceutical Co., Ltd.) for the neck, a National Electrical Manufacturers Association/International Electrotechnical Commission body phantom (without lung insert and spheres; Data Spectrum Corp.) for the chest, and a cylindrical phantom (15-cm diameter and 20-cm length; Itoi Factory Inc.) for the upper abdomen. These phantom sets entirely consisted of water; thus, other structures (e.g., bone) were not taken into account.

Fig. 1
figure 1

Photograph and diagram of the phantom alignment. The head phantom was placed in the center of the axial field of view. The abdominal phantom was placed in the liver position off the midline. For the neck phantom, of the three bags, the one that was placed in the middle was filled with radioactivity

Table 3 shows the target radioactivity of 18F that was injected into each part of the phantom set, simulating the 11C-PiB biodistribution at the following three timepoints: post-injection 1–2 min (early phase; early1–2), 10 min (early phase; early10), and 50 min (late phase; late50). The radioactivity in each phase was determined based on the human WB, neck-to-chest, and brain dynamic scan data (subjects A–I). The activity of the neck and chest phantoms was determined based on the mean %injected dose (%ID) values in the neck and chest using the dynamic WB and neck-to-chest data. The %ID was calculated as follows:

$$ {\text{\% ID}}\;{ = }\; \frac{\text{Tissue activity (MBq)}}{\text{Total injected dose (MBq)}}. $$
Table 3 Target activity injected into each phantom

The activity of the abdominal phantom was determined based on the sum of mean %ID values in the liver, spleen, and kidney using the dynamic WB data. Data from the first WB and neck-to-chest scans were used as the early1–2 data, and those from the second and third WB scans, and from the fifth and sixth neck-to-chest scans, were used as the early10 data. The final WB and neck-to-chest scan data were corrected for decay and were used as the late50 data, because WB and neck-to-chest data were only available up to about 35 min after injection.

At first, a transmission or CT scan was performed with all the phantoms aligned, so that the head phantom was located in the center of the FOV (Fig. 1); an emission scan followed. Next, after removing the neck, chest, and upper abdomen phantoms, we performed emission scans of the head phantom without body activity. We performed these scans in three phantom conditions, simulating the activity at early1–2, early10, and late50, respectively. The emission data simulating early1–2 and early10 were acquired in 5 min × 1 frame, and the late50 data were acquired in 5 min × 4 frames.

2.6 Data analysis

2.6.1 Dynamic whole-body, neck-to-chest, and brain PET data

Image analysis was performed using the PMOD software ver.3.17 (PMOD Technologies Ltd., Switzerland). To analyze the dynamic WB and neck-to-chest PET data (subjects A–F), the ROIs were manually placed on the sum images of all frames over the brain, thyroid, lungs, heart, liver/gallbladder, spleen, kidneys, bladder, and intestine, as well as over the entire neck area (from the lower border of cerebellum to the lower border of the thyroid) and the entire chest area (from the lower border of thyroid to the upper border of the liver). A time-activity curve (TAC) was obtained and the %ID was calculated based on the radioactivity distribution in each organ and the injected dose.

To analyze the 2D dynamic brain PET data (subjects G–I), a whole-brain ROI was placed on the sum images of all frames. The whole-brain ROI value was then measured for each frame of the dynamic data and the %ID was calculated.

To analyze subject J, TACs were obtained using manual ROIs placed over the frontal cortex, parietal cortex, and cerebellum.

2.6.2 Phantom data with three types of activity

A 10-cm-diameter circular ROI was placed in the center of the reconstructed images of the head phantom in the early1–2, early10, and late50 image data. The ROI mean value for each slice with and without body activity was obtained and compared to evaluate the effect of body activity.

2.6.3 Effect of body activity on quantitative values

Axial slices of the head phantom images were divided into three parts (caudal, middle, and cranial), each covering an axial length of about 3 cm (excluding the slices at both ends). The distance from the caudal area to the middle was about 3 cm. These three parts corresponded to the axial level of the cerebellum, frontal cortex, and parietal cortex, respectively. The estimated error (EE) was defined as the percentage difference between the head phantom ROI mean value with and without phantoms simulating body activity as follows:

$${\text{Estimated error (EE)}}\;{ = }\; \frac{{{\text{Mean activity with phantoms (Bq/mL) }} - {\text{mean activity without phantoms (Bq/mL)}}}}{\text{Mean activity without phantoms (Bq/mL)}}.$$

The EE was calculated in each slice of the brain phantom and averaged to evaluate the effect of body activity on the brain. All scan frames belonging to the late50 phase of the phantom experiments were averaged to reduce the effect of image noise. The EE at any timepoint was estimated by linear interpolation of the EE at the measured three timepoints.

TACs of the cerebellum, frontal cortex, and parietal cortex were obtained from the brain dynamic PiB-positive PET data [subject J; measured TAC (TACmeas)]. Next, we corrected for the effect of body activity using the EE of the TACmeas at each timepoint, as determined using the SET-3000G/X camera. We then generated corrected TAC data (TACcorr). We performed kinetic analysis using both TACmeas and TACcorr, and then compared the distribution volume ratios (DVRs). The reference region was placed over the cerebellar cortex. Multi-linear reference tissue model 2 (MRTM2) [17] and Logan graphical analysis with reference tissue (LGAR) [18] were used in the kinetic analysis. We used the TAC data of 0–70 min to obtain the operational equation model. Subsequently, the DVRs were estimated using linear or multi-linear regression for t > t*, where t* is the equilibrium time for the regression. The LGAR requires a fit to the latter linear portion of the plot only. In contrast, the MRTM2 can include more data points, since it uses a multi-linear approach [19]. Based on these, we set t* equal to 1650 s for LGAR and 0 s for MRTM2 (all data were used).

3 Results

3.1 Whole-body 11C-PiB biodistribution

Figure 2 shows the sequential WB and neck-to-chest maximum intensity projection (MIP) images after injection of 11C-PiB. Immediately after 11C-PiB injection, images reflecting blood perfusion were obtained. The organ uptake was rapidly washed out, except for that in the brain, liver, intestine, and urinary tract. The biliary tracts and intestine gradually became intensely visible over time. Figure 3 shows the TACs for the whole brain, thyroid, neck, chest, liver/gallbladder, spleen, kidneys, bladder, and intestine that were obtained from the dynamic WB, neck-to-chest, and brain data. Whole-brain uptake of 11C-PiB reached a maximum about 2 min after injection. Uptake by the thyroid, chest, spleen, and kidneys was at its highest in the first frame of the dynamic data. Conversely, uptake by the liver, bladder, and intestine gradually increased over time up to 35 min after injection.

Fig. 2
figure 2

Serial whole-body (WB) maximum intensity projection (MIP) images after 11C-PiB injection. A WB MIP image from the head to femoral region in subject B (a). Neck-to-chest MIP image in subject E (b)

Fig. 3
figure 3

Time-activity curves (TACs) for each organ (whole-brain, thyroid, neck, chest, liver/gallbladder, spleen, kidneys, bladder, and intestine). Each line indicates single-subject data. The whole-brain TAC was calculated from the dynamic brain scan. The other TACs were calculated from the dynamic whole-body and neck-to-chest scans

3.2 Phantom experiments

Table 3 shows the radioactivity injected into each phantom, determined based on the dynamic human PET scan data. Figures 4, 5, 6 show the relative ROI value at each slice level of the head phantom image with and without the phantoms simulating body activity for the three phases of post-injection. The EE calculated from the phantom experiments is also shown in Table 4. The ROI value at each slice level was generally constant without the body activity for using each PET camera and during all phases. However, using the ECAT ACCEL, an underestimation of 3.7% was observed at the caudal level during the early1–2 phase. Conversely, using the SET-3000G/X, there was a maximum 6.0% overestimation at the cranial level. The differences in the values measured in ROIs with or without body activity were relatively small during the late50 phase (Fig. 6).

Fig. 4
figure 4

Plots of the head phantom region of interest values in the head at each slice level; the values without body activity are used as a baseline. The graphs show changes with and without body activity at simulated 1–2 min after injection (early1–2). The PET camera models used were the SET-3000G/X (a), Discovery ST Elite (b), Advance (c), Biograph Hi-Rez (d), Biograph TruePoint TrueV (e), and ECAT ACCEL (f). In the caudal, middle, and cranial areas, the EE was calculated

Fig. 5
figure 5

Plots of the head phantom relative values measured in regions of interest at each slice position with or without body activity at simulated 10 min after injection (early10). The PET camera models used were the SET-3000G/X (a), Discovery ST Elite (b), Advance (c), Biograph Hi-Rez (d), Biograph TruePoint TrueV (e), and ECAT ACCEL (f)

Fig. 6
figure 6

Plots of the head phantom relative values measured in regions of interest at each slice position with or without body activity at simulated 50 min after injection (late50). The PET camera models used were the SET-3000G/X (a), the Discovery ST Elite (b), Advance (c), Biograph Hi-Rez (d), Biograph TruePoint TrueV (e), and ECAT ACCEL (f)

Table 4 Estimated error (EE) calculated from the phantom data

3.3 Effect of body activity on kinetic analysis

Figure 7 shows TACmeas and TACcorr in the cerebellum, frontal cortex, and parietal cortex of subject J. The TACcorr values were lower than the TACmeas values. The DVR calculated using TACcorr in the frontal and parietal cortexes was slightly lower than that calculated using TACmeas. The percentage differences in the frontal and parietal cortexes calculated using MRTM2 were −0.9 and 0.0%, respectively, and those calculated using LGAR were 0.0 and −1.2%, respectively.

Fig. 7
figure 7

Time-activity curves (TACs) for the frontal cortex, parietal cortex, and cerebellum, obtained using brain a three-dimensional dynamic scan with an SET-3000G/X camera in a patient with Alzheimer’s disease. The vertical axis indicates activity (Bq/mL) normalized by the injected activity (MBq). The broken lines are the measured TACs (TACmeas). The solid lines are the TACs corrected for the effect of body activity using EE obtained from the phantom experiments (TACcorr)

4 Discussion

In the present study, we measured the sequential whole-body biodistribution of 11C-PiB over time. We also estimated the effect of body activity on the quantified values of the brain 11C-PiB PET based on self-designed phantom experiments using a number of PET camera models. We found that body activity had only a small effect on measured brain activity, with exceptions in the cases of a few PET cameras in the early post-injection phase. We also found that the error of the estimated DVR value was relatively small (≤1.2%) when using one such camera (SET-3000G/X scanner) based on the kinetic methods tested in this study (MRTM2 and LGAR).

Sequential WB images of 11C-PiB distribution showed activity in the heart and lungs, in a blood pool in the neck and chest area, and in the kidney and intestine, for about 10 min after injection (Fig. 2). There was high consistency among subjects, irrespective of whether they were positive or negative for late-phase brain amyloid deposition. In the early phase, uptake by the lungs, liver, and blood in the neck and chest was much higher than in the brain. Uptake in the neck and chest decreased by 20 min after injection and fell further in the later phase. In contrast, uptake in the bladder and intestine gradually increased. These results were consistent with a previous dosimetry study [13]. Because activity in the neck-to-chest area is considered the main source of error in brain quantitation, late-phase images, which are routinely used for visual interpretation and standardized uptake value ratio (SUVR) measurement, are minimally affected by the body activity.

Using self-designed phantom experiments that were based on the measured human whole-body distribution of 11C-PiB, we estimated the error caused by the body activity at each axial level of the brain using various PET cameras. We also evaluated the effect of such activity on kinetic analysis using the SET-3000G/X camera. The results showed that body activity caused non-uniform errors that depended on the axial localization within the brain and on the PET camera model. This result corroborated the previous studies [10, 12]. When radioactivity exists outside the FOV, single gamma rays from outside the FOV can be detected, especially at the far end of the axial FOV. This radiation increases singles and random coincidence, and they might influence brain PET camera uniformity, as shown in Fig. 4 [5, 10, 12]. It should be noted that an end shield may be available that can be attached to the gantry [9]. This would be effective in blocking those gamma rays from the body. However, such a shield was not used in the present study.

Furthermore, in the present study, the quantitative value of DVR was scarcely affected by the body activity. Specifically, the effect of body activity on the DVR estimate was small, because the kinetic analyses performed in this study did not rely heavily on early phase data, which were more influenced by body activity in a non-uniform way. If we use the smaller t* timepoint for the LGAR model, or if we use other kinetic models [2, 19,20,21], body activity may influence the quantified estimates derived by the kinetic analysis. Conversely, simple ratio quantification, such as SUVR, is considered robust for body activity, because it does not use the early phase data. However, the activity outside the FOV may influence image noise level, although we did not evaluate the effect of body activity on the image noise. Therefore, a further study is required to examine the effect of activity outside the FOV on the voxel-based analysis, because it is sensitive to the image noise level.

In the results of the present study, the influence of body activity on the quantitative accuracy of 11C-PiB PET was generally small when we did not use much early phase data. Conversely, several previous studies have reported that an external shield is useful in eliminating the effect of body activity [9, 10, 12, 22,23,24]. However, this carries an additional cost and may potentially limit the flexibility of subject positioning.

5 Conclusion

We measured the serial, whole-body biodistribution of 11C-PiB in detail. In addition, we evaluated the effect of body activity on brain 11C-PiB PET quantification and kinetic analysis. Our results indicate that the effect of body activity on the quantitative capabilities of brain 11C-PiB PET is generally small and that it depends on the method of quantitative analysis, the location of the axial FOV, and the PET camera model. Therefore, when using early phase PET data from the brain, it may be prudent to evaluate the effect of activity outside the FOV using appropriate phantoms beforehand.