Abstract
A loop-mediated isothermal amplification (LAMP)-based method has been developed for the analysis of animal feed for the presence of Salmonella spp. The method is compatible with the standard culture-based method ISO 6579:2002. Samples comprising 100 g feed are pre-enriched in 900 ml buffered peptone water, and then 2× 100-μl aliquots are taken, one for secondary enrichment in a LAMP-friendly broth and the other for secondary enrichment according to ISO 6579:2002. A 3-μl aliquot of the LAMP-friendly culture is added directly into a LAMP reaction. LAMP-positive signals are considered presumptive of the presence of Salmonella spp.; confirmation is obtained by continuation of ISO 6579:2002. Negative LAMP reactions can be considered as signifying that the sample does not contain viable Salmonella. The LAMP assay incorporates an internal amplification control (IAC); the target and IAC signals are distinguished by the difference in annealing temperatures of the amplicons. The method provides simple and reliable screening of feedstuffs for Salmonella spp., with uncontaminated samples being identified in 24 h less than using the current ISO standard method alone. The method was used to analyse 79 samples of animal feed and identified 13 Salmonella-positive samples.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
It has long been recognised that a link exists between contamination of animal feeding stuffs with Salmonella and human disease (Jones 2011). Surveys of animal feed have indicated that the presence of this pathogenic zoonotic bacterium is a significant issue for the feed industry (Foley et al. 2008; Papadopoulou et al. 2009; Li et al. 2012). Animal feeds are routinely analysed for the presence of this agent, with the analysis being generally based on the use of conventional culturing of Salmonella. However, nucleic acid amplification (NAA)-based methods can offer increased diagnostic capabilities, such as the ability to rapidly identify uncontaminated samples.
An NAA technique which is becoming increasingly reported in the scientific literature is loop-mediated isothermal amplification (LAMP), and several LAMP methods for detection of Salmonella have been reported (Kokkinos et al. 2014). Amplification of DNA does not per se provide an indication of bacterial viability, and methods based solely on NAA may not necessarily be completely acceptable to industry (D’Agostino and Rodriguez-Lazaro 2009), which very often prefers to rely on established international standard culture-based methods. Any target signal obtained using NAA is considered to only represent the presumptive presence of the target. There is, nonetheless, an approach which embodies the benefits of identification of viable contaminants by conventional culturing and rapid screening by NAA (D’Agostino and Rodriguez-Lazaro 2009). This entails using the initial stages of a standard culture-based method to enrich a sample and then dividing the enrichment into two subsamples: one of which is taken for analysis by NAA and the other by continuation of the standard culture-based method. NAA is so sensitive and specific that if no target signal is obtained, the target can be considered to be absent. Thus, the analysis of the subsample by NAA screening for target-negative samples and the analysis of the sample using the culture-based method confirm NAA presumptive target-positive samples. This approach ensures that screening and confirmation are performed on the same sample.
We report a novel rapid method for analysing animal feed for the presence of Salmonella, which is based on the screening for Salmonella-negative samples using a LAMP assay, with confirmation of presumptively positive signals using a standard culture-based method (International Organisation for Standardization 2002).
Materials and Methods
Salmonella Strains
Salmonella enterica subsp. enterica serovar Typhimurium ATCC 14028 and S. enterica subsp. enterica serovar Cerro NCTC 5801 were used for the development of the method. For artificial contamination of the test samples with Salmonella Typhimurium, one Lenticule® disc (Public Health England) containing a mean of 1–5 colony-forming units (cfu) was used per test sample.
Animal Feed
Soya meal was used as the matrix during the development of the method; it was kindly supplied by John Thompson and Sons Ltd., Belfast. One hundred gram soya meal was used as test sample size, and 100-g soya meal samples without artificial contamination were also tested.
Culture-Based Confirmation of Salmonella spp. Detection
The method used was based on that described in “ISO 6579:2002 - Microbiology of food and animal feeding stuffs—horizontal method for the detection of Salmonella spp.” (2002) with the following exceptions:
-
1.
For primary enrichment of the sample, 100 g soya meal was added to 900 ml buffered peptone water (BPW; Oxoid, Basingstoke, UK).
-
2.
Secondary enrichment was performed using only Rappaport-Vassiliadis medium with soya (RVS broth; Oxoid, Basingstoke, UK).
-
3.
Plating out was performed using only xylose lysine desoxycholate agar (XLD; Oxoid, Basingstoke, UK).
Biochemical characterisation of presumptively positive colonies was performed using the API 20E kit (bioMérieux, UK).
LAMP-Friendly Secondary Enrichment
One hundred microlitres of the primary enrichment was added into 100 ml of Yersinia-PCR-compatible enrichment (YPCE) broth (Knutsson et al. 2002) and incubated at 37 °C 24 h. Three microlitres was used directly in the LAMP reaction.
Purification of Salmonella Genomic DNA
Nucleic acid was purified from 100 ml culture using a commercial DNA extraction kit (Qiagen, Manchester, UK). Three microlitres was used as a template in the LAMP assay. The concentration and purity of the DNA were determined using a NanoDrop instrument (Thermo Scientific, UK). The number of genome copies in the extract was determined by dividing the amount of DNA present by the weight of one genome copy.
LAMP Assay
The LAMP assay was adapted from that of Hara-Kudo et al. (2005) which targets a region of the invA gene (accession number embl|U43237). The reaction contained 1× Tin isothermal master mix, 1 μM each of primers invASalm FIP and invASalm BIP, 500 nM each of primers invASalmF-loop and invASalmB-loop, 200 nM each of primers invASalmF3 and invASalmB3 and 1 μl of internal amplification control (IAC) in a 22 μl volume; 3 μl of the template was added. The thermal parameters of the reaction were 95 °C for 2 min (cell lysis), followed by 65 °C for 50 min (nucleic acid amplification). Amplicon annealing profiling was performed by heating to 98 °C and then cooling to 80 °C at a rate of 0.05 °C s−1. The samples were analysed using a Genie II instrument (OptiGene Ltd., Horsham, UK).
IAC Design
The IAC was designed to be amplified by the same primer set used for target amplification, by incorporating the target sequences into the IAC sequence. To ensure that the annealing temperature of the IAC amplicon was higher than the target amplicon, the sequence between the FIP and BIP primer regions was modified from the original ttatcttgattgaagccgatg to gcatctggagcgcagccgatgcgcaggcctc to increase the G:C content of the IAC amplicon to 50 % (target amplicon 46 %), thereby increasing the annealing temperature by 1.6 °C according to the Oligo Calc: Oligonucleotide Properties Calculator (http://www.basic.northwestern.edu/biotools/OligoCalc.html).
IAC Construction
The IAC was supplied by Eurofins MWG Operon (Ebersberg, Germany). The IAC sequence was cloned into plasmid pEX-A, which was linearised using ScaI.
IAC Optimisation and Demonstration
The optimal amount of IAC per reaction, i.e. which will consistently produce a signal in the absence of target nucleic acid, was determined according to the procedure of Cook et al. (2012). To demonstrate that the IAC will not produce a signal in the presence of substances in the feed sample which can be inhibitory to the LAMP reaction and is thus able to identify a failed reaction, a flask containing 900 ml BPW plus 100 g soya meal was inoculated with 1 ml of a mid-log phase culture of Salmonella Typhimurium. The culture was incubated at 37 °C for 24 h. Three microlitres of the culture was added directly as a template to a LAMP reaction mix, and the reaction performed as described above. To demonstrate that the IAC will produce a signal when the reaction has performed correctly, 100 g soya meal was added to 900 ml BPW and incubated at 37 °C for 24 h. One hundred microlitres of the primary enrichment was added into 100 ml of YPCE broth and incubated at 37 °C 24 h. Three microlitres was used directly in the LAMP reaction.
Detection Limit of the LAMP Assay
Purified Salmonella Cerro DNA was diluted to give concentrations of 3.3 × 104, 3.3 × 103, 3.3 × 102 and 3.3 × 101 genome copies per μl. Three microlitres of each dilution was then used as a template in the LAMP assay. Triplicate reactions were performed for each dilution.
Survey of Animal Feeds
Seventy-nine samples of animal feed were analysed. The types of feedstuff analysed included rapemeal, maize, corn, soya bean, wheat, sunflower pellets, distiller grains, whole rape, microalgae, palm kernel expeller and other meals for poultry, fish, etc. All the samples were screened using the LAMP-based method with confirmation of positive samples by following ISO 6579:2002.
Results
A Loop-Mediated Amplification-Based Method for Analysing Animal Feed for the Presence of Salmonella
One hundred grams of each of the animal feed samples were added to 900 ml BPW. These were incubated at 37 °C for 16–20 h and tested using the LAMP-based method. One hundred microlitres of primary enrichment culture was added to 100 ml of YPCE broth pre-warmed to 37 °C, and simultaneously, 0.1 ml of primary enrichment culture was added to 10 ml Rappaport-Vassiliadis (RV) medium. The LAMP-based method and the standard method were then both followed (see Fig. 1). A positive signal obtained using the LAMP-based method was considered as presumptively positive, and the ISO standard method was continued to be confirmed. A negative target signal and a positive IAC signal obtained using the LAMP-based method were considered to demonstrate that the sample was uncontaminated by Salmonella spp. The method could identify Salmonella-negative test samples of soya meal in less than 48 h.
Annealing Temperatures of IAC and Target
The amplicon annealing temperatures observed after the amplification process were sufficiently different to distinguish between target and IAC signals. Figure 2 shows the signal produced using 104 genome copies of Salmonella Typhimurium DNA.
IAC Optimisation
The optimal amount of IAC per reaction was determined at 40 copies, as this quantity (and higher quantities; not shown) yielded amplicon signals at a similar time (Fig. 3a), whereas the signals obtained from 20 copies were obtained at differing times, indicating inconsistent amplification (Fig. 3b).
Demonstration of Correct and Failed Assays
Neither target nor IAC signal was obtained when neat primary enrichment culture was added directly to the LAMP reaction (Fig. 4a, b); this assay had therefore failed.
When the culture was diluted at 1:10, the target signal was restored (Fig. 4c, d).
Detection Limit of the LAMP Assay
Amplicon peaks for 3.3 × 104 genome copies per reaction were consistently clearly visible. For lower numbers of genome copies per reaction, it was not always possible to see a clearly defined peak.
Survey of Animal Feeds for Salmonella spp. Contamination
Thirteen out of 79 samples were presumptively positive for Salmonella spp. using the LAMP method, and all was confirmed as positive by ISO 6579:2002. Table 1 lists the positive samples and the strain of Salmonella where it was able to be identified.
Discussion
The LAMP-based method for analysis of animal feeding stuffs allows, through screening, the identification of uncontaminated samples in approximately 24 h less than the standard culture-based method alone. If the product is required to be certified as Salmonella free before it is released to the market, this would provide a significant cost saving. If a positive LAMP signal is obtained, the sample can be considered as presumptively positive for Salmonella contamination. Confirmation by following the standard method will satisfy any requirement for further identification of the Salmonella contaminant.
This is the first report of incorporation of a competitive IAC (using the same primers which amplify the target sequences) into a LAMP assay. The LAMP assay for Salmonella (Hara-Kudo et al. 2005) could detect ∼30 genome copies per reaction when no IAC was present. Incorporation of an IAC resulted in a reduction in sensitivity; however, as the LAMP is preceded by sample enrichment, it is not essential that the assay is highly sensitive.
The two-step enrichment procedure can be relied on to indicate the viability of a presumptive positive. It increases the number of viable cells originally in the sample to detectable amounts and will dilute out dead cells originally present. The likelihood that a LAMP signal will be derived from viable Salmonella can be demonstrated arithmetically. Approximately 104 genome copies and, hence, cells per LAMP are necessary to obtain a target signal. This means that 400 cells must be present in the 3 μl used as a template in the reaction. Consequently, it is necessary to have 1.33 × 107 viable cells in the 100 ml YPCE after 24 h of incubation. Assuming a doubling time of 30 min, to reach this number the LAMP-friendly enrichment would require only one viable cell at 0 h. Thus, the 1 l primary-enriched culture needs to contain at least 1 × 104 viable cells after 16 h of incubation in order to contain one viable cell in the 100 μl that is put into the 100 ml YPCE. One viable Salmonella cell in the original 100-g sample will reach this number in 16–20 h, even with a lag phase of ∼8 h. The likelihood that a LAMP signal will not be derived from dead Salmonella can also be demonstrated arithmetically. The original sample is added to BPW to make the primary culture and then one out of 10,000th of that is added to the YPCE. Finally, less than one out of 30,000th of the secondary culture is incorporated into the LAMP. The LAMP assay can detect ∼400 cells per reaction. Thus, at least 1 × 1011 dead Salmonella cells must be present in the original 100 g of animal feed to obtain a positive LAMP signal. Confirmation of the viability of presumptive positives is anyway conferred using the standard culture-based method.
If monitoring of animal feed for pathogens is to be effectively performed, then it is vitally necessary that the reliability of the analytical results can be verified (Rodríguez-Lázaro et al. 2007). In NAA-based diagnostics, IACs are essential to identify failed reactions, through e.g. inhibition or operator error (Rodríguez-Lázaro et al. 2007; D’Agostino et al. 2011). To date, however, no LAMP assay has been published which contains an IAC which uses the same primers as for the amplification of the target to ensure complete control of the assay. LAMP assays for plant-related diagnostics have been published which have contained controls based on amplification of non-target genes found within the plant material, but this approach is not universally valid. We modified the inter-primer region to ensure that the IAC amplicons can be distinguished from the target amplicons by a difference in melting and annealing temperature (see e.g. Fig. 1).Footnote 1 Optimization of IAC concentration ensures consistent detection of IAC in the absence of a target (Rodríguez-Lázaro et al. 2004) (see Fig. 2). An example of a failed LAMP assay is shown in Fig. 3a, b. Upon performing a 1:10 dilution of the sample, the target signal has been restored (Fig. 3c, d), demonstrating that the false-negative result was due to the presence of inhibitors in the culture. Another interesting aspect of the LAMP assay used here is the use of the thermostable enzyme Thermodesulfotator indicus (Tin) DNA polymerase enzyme. This allows the direct addition of whole bacterial cells to the LAMP reaction as when the temperature rises past 90 °C, they are lysed and release their nucleic acids directly into the reaction. This removes the necessity for prior nucleic acid extraction and is the most useful feature. YPCE secondary enrichment medium has been shown to be effective in a polymerase chain reaction-based method for the detection of Listeria monocytogenes (D’Agostino et al. 2004); in the current study, it was equally effective and, therefore, no other secondary enrichment broths were evaluated.
The utility of the LAMP-based method was demonstrated by its use in the analysis of a variety of actual feed samples. Several naturally Salmonella-contaminated feeding stuffs were identified using the LAMP-based method. The method has been verified as equivalent to ISO 6579:2002 through intra- and inter-laboratory validation, as will be described in a future publication.
Notes
This IAC approach will not be applicable if using turbidity as an end-point detection system (Chen et al. 2011; Soli et al. 2013). However, users of turbidity-based detection should recognise its limitations and the potential ambiguity of its results; in that, false-negative result interpretation cannot be totally avoided.
References
Chen S, Wang F, Beaulieu JC, Stein RE, Ge B (2011) Rapid detection of viable salmonellae in produce by coupling propidium monoazide with loop-mediated isothermal amplification. Appl Environ Microbiol 77:4008–4016
Cook N, de Ridder GA, D’Agostino M, Taylor MB (2012) Internal amplification controls in real-time polymerase chain reaction-based methods for pathogen detection. In: Rodriguez-Lazaro, D. (ed.) Real-time PCR in food science: current technology and applications. Caister Academic, pp. 35–41
D’Agostino M, Rodriguez-Lazaro D (2009) Harmonisation and validation of methods in food safety—“FOOD-PCR”, a case study. In: G. Barbosa-Canovas, G. Mortimer A, Colonna P, Lineback D, Spiess W, Buckle K (eds.) Global issues in food safety and technology. Academic, pp. 199–209
D’Agostino M, Wagner M, Vazquez-Boland JA, Kuchta T, Karpiskova R, Hoorfar J, Novella S, Scortti M, Ellison J, Murray A, Fernandes I, Kuhn M, Pazlarova J, Heuvelink A, Cook N (2004) A validated PCR-Based method to detect Listeria monocytogenes using raw milk as a food model—towards an international standard. J Food Prot 67:1646–55
D’Agostino M, Cook N, Rodriguez-Lazaro D, Rutjes S (2011) Nucleic acid amplification-based methods for detection of enteric viruses: definition of controls and interpretation of results. Food Environ Virol 3(2):55–60
Foley SL, Lynne AM, Nayak R (2008) Salmonella challenges: prevalence in swine and poultry and potential pathogenicity of such isolates. J Anim Sci 86(14 Suppl):E149–62
Hara-Kudo Y, Yoshino M, Kojima T, Ikedo M (2005) Loop-mediated isothermal amplification for the rapid detection of Salmonella. FEMS Microbiol Lett 253:155–161
International Organisation for Standardization (2002) ISO 6579:2002—microbiology of food and animal feeding stuffs—horizontal method for the detection of Salmonella spp.
Jones FT (2011) A review of practical Salmonella control measures in animal feed. J Appl Poult Res 20:102–113
Knutsson R, Fontanesi M, Grage H, Radstrom P (2002) Development of a PCR-compatible enrichment medium for Yersinia enterocolitica: amplification precision and dynamic detection range during cultivation. Int J Food Microbiol 72:185–201
Kokkinos PA, Ziros PG, Bellou M, Vantarakis A (2014) Loop-mediated isothermal amplification (LAMP) for the detection of Salmonella in food. Food Anal Methods 7:512–526
Li X, Bethune LA, Jia Y, Lovell RA, Proescholdt TA, Benz SA, Schell TC, Kaplan G, McChesney DG (2012) Surveillance of Salmonella prevalence in animal feeds and characterization of the Salmonella isolates by serotyping and antimicrobial susceptibility. Foodborne Pathog Dis 9:692–8
Papadopoulou C, Carrique-Mas JJ, Davies RH, Sayers AR (2009) Retrospective analysis of Salmonella isolates recovered from animal feed in Great Britain. Vet Rec 165:681–8
Rodríguez-Lázaro D, D’Agostino M, Pla M, Cook N (2004) Construction strategy for an internal amplification control for real-time diagnostic assays using nucleic acid sequence-based amplification: development and clinical application. J Clin Microbiol 42:5832–6
Rodríguez-Lázaro D, Lombard B, Smith H, Rzezutka A, D’Agostino M, Helmuth R, Schroeter A, Malorney B, Miko A, Guerra B, Davison J, Koblinsky A, Hernández M, Bertheau Y, Cook N (2007) Trends in analytical methodology in food safety and quality: monitoring microorganisms and genetically modified organisms. Trends Food Sci Technol 18:306–319
Soli KW, Kas M, Maure T, Umezaki M, Morita A, Siba PM, Greenhill AR, Horwood PF (2013) Evaluation of colorimetric detection methods for Shigella, Salmonella, and Vibrio cholerae by loop-mediated isothermal amplification. Diagn Microbiol Infect Dis 77:321–3
Acknowledgments
The research leading to these results has received funding from the European Union Seventh Framework Programme (FP7/2007‐2013) under grant agreement no. 265702. MD-V was supported by a grant from the Spanish Ministry. MD and NC were supported by a grant Q-SAFFE. We thank Q-SAFFE project partners, particularly John Thompson and Sons Ltd., for the supply of feed samples. We thank Almudena Perez and Zohartze Monteagudo for their technical assistance.
Conflict of Interest
Marta Diez-Valcarce has no conflict of interest. Nigel Cook has no conflict of interest. Martin D’Agostino has no conflict of interest. Susana Robles has no conflict of interest. Begoña Losilla-Garcia has no conflict of interest. This article does not contain any studies with human or animal subjects.