Introduction

Measurement of cerebral blood flow (CBF), cerebral blood volume (CBV), cerebral oxygen extraction fraction (OEF), and cerebral metabolic rate of oxygen (CMRO2) by positron emission tomography (PET) with 15O-labeled carbon dioxide (C15O2), 15O-labeled carbon monoxide (C15O), or 15O-labeled oxygen (15O2) is widely used for investigation into the pathophysiology of several brain diseases, especially occlusive cerebrovascular disease [1,2,3,4,5,6,7]. Decreased cerebral perfusion pressure due to major cerebral arterial occlusive disease causes cerebral autoregulatory vasodilatation for maintenance of CBF (stage I hemodynamic change). Decreased cerebral perfusion pressure below the lower limit of cerebral autoregulation causes a decrease in CBF with an increase in OEF, for maintenance of CMRO2 (stage II hemodynamic change). For the assessment of stage I hemodynamic compromise, measurements of CBF at rest and during acetazolamide stress by PET or single photon emission tomography (SPECT) are widely performed [8, 9]. Stage II hemodynamic compromise, however, can be assessed by measuring OEF, which can be done only by PET [10].

The integrated design of the positron emission tomography/magnetic resonance imaging (PET/MRI) scanner system allows simultaneous data acquisition for PET and MRI [11], and therefore, it must have potential for clinical and research settings of neurological and psychiatric diseases including cerebrovascular diseases [12, 13]. Such PET/MRI scanner system can be applied for correction of partial volume effects, motion correction, and determination of image-derived input function of brain PET [12].

In the present study, CBF, CBV, OEF and CMRO2 were measured using the integrated design of the PET/MRI scanner system and 15O-labeled gases with two kinds of attenuation corrections for PET by MRI. A real-time head motion correction of PET images using MR images simultaneously measured with PET was also applied. Values of CBF, CBV, OEF, and CMRO2 with two kinds of attenuation correction for PET with and without a real-time motion correction were compared. This is a first attempt to measure cerebral perfusion and oxygen metabolism using PET/MRI with 15O-labeled gases.

Materials and methods

Subjects

Nine healthy men (20–27 years old) were recruited. No subject had notable organic lesions in the brain according to MRI. The subjects were free of somatic, neurological and psychiatric disorders on the basis of their medical history. They were free from medications having central nervous action. The study was approved by the Institutional Review Board of the Fukushima Medical University, Fukushima, Japan. Written informed consent was obtained from all subjects.

PET/MRI experimental procedure

All PET studies were performed with a Siemens mMR PET/MRI scanner, which provides 127 sections with an axial field of view of 25.8 cm [14]. The intrinsic spatial resolution was 4.3 mm in-plane and 4.3 mm full-width at half maximum (FWHM) axially. Data were acquired in three-dimensional mode. Scatter was corrected [15]. PET measurements with the steady-state method of 15O-labeled gases, C15O, 15O2, and C15O2 were performed on all subjects [16, 17]. Static PET scanning was started 3 min after 1 min of continuous inhalation of C15O gas (a total of approximately 3 GBq supplied by mouth). Scanning time was 4 min. Then, static PET scanning was performed during inhalation of 15O2 gas after equilibrium had been reached and confirmed by the head radioactivity curve (a total of approximately 8.4 GBq supplied by mouth). Scanning time was 10 min, and the time from the beginning of inhalation to the beginning of scanning was 10 min. Static PET scanning was performed during inhalation of C15O2 gas using the same protocol as with 15O2 gas (a total of approximately 2.8 GBq supplied by mouth). During each PET scanning, arterial blood sampling was performed to measure the radioactivity concentration in the blood and plasma. Arterial blood gases were also measured. For the arterial blood sampling, non-magnetic medical devices, such as a plastic needle and plastic syringe were used. Total oxygen content in arterial blood was estimated from PaO2, pH and hemoglobin concentration (Hb) [17].

All MRI studies were performed with a Siemens mMR PET/MRI scanner, equipped with a 3.0-T MR scanner, during and between PET scanning. Dixon sequence (DIXON) (3D-VIBE (volumetric interpolated breath-hold examination), TR: 3.56 ms, TE: 1.23 ms and 2.46 ms; field of view: 500 mm, slice thickness: 3.12 mm, resolution: 2.6 × 2.6 × 3.1 mm, 1 slab: 128 slices) and Dixon sequence with model-based bone segmentation (DIXONbone) were performed for attenuation correction of PET [18, 19]. The MR images with gradient echo-type echo planar imaging (2D GR-EPI) sequence (TR: 1000 ms, TE: 30 ms, flip angle 75°; field of view: 240 mm, acquisition matrix size: 64 × 64, slice thickness: 3.0 mm, slice gap: 1.5 mm, 15 slices) to detect head motion were simultaneously measured with PET scanning for a real-time motion correction of PET images [20, 21]. Three-dimensional volumetric T1-weighted images (T1WI) and T2-weighted images (T2WI), diffusion-tensor images, arterial spin labelling images, and MR angiography were also acquired. In the present study, T1WI (three-dimensional MPRAGE magnetization prepared rapid acquisition with gradient echo sequence, TR: 1800 ms, TE: 1.98 ms, flip angle 9°; field of view: 250 mm, acquisition matrix size: 256 × 256, slice thickness: 1.0 mm) were used for analyses.

Image reconstruction of PET was carried out by the ordered-subset expectation maximization (OSEM) algorithm (iterations: 3, subsets: 21) with postreconstruction Gaussion filter of 5 mm FWHM. Two attenuation corrections for PET using DIXON and DIXONbone were applied. A real-time motion correction of PET images was also applied using simultaneously measured MR images with GR-EPI sequence.

Calculation of parametric images

Using reconstructed PET images, the radioactivity concentration in arterial blood and plasma, and arterial blood gases data, the parametric images of CBF, CBV, OEF and CMRO2 were calculated [16]. The parametric images were calculated for attenuation correction by DIXON and DIXONbone with and without real-time motion correction.

Data analysis

Regions of interest (ROIs) were drawn on all PET images, referring to T1WI. Circular ROIs (10 mm in diameter) were defined for the cerebellar cortex, pons, thalamus, putamen, centrum semiovale, and four neocortical regions representing the base side of the frontal cortex, lateral side of the temporal cortex, inferior parietal lobule cortex, and cuneus of the occipital cortex. The average of values of four neocortical regions were calculated as the cerebral cortices.

Results

Figure 1 shows typical images of attenuation coefficient by DIXON and DIXONbone. In the image of attenuation coefficient by DIXONbone, the skull was depicted.

Fig. 1
figure 1

Typical images of attenuation coefficient using the Dixon sequence (DIXON) and Dixon sequence with model-based bone segmentation (DIXONbone). T1WI for the same slices of attenuation images are also shown

Table 1 gives mean and SD values of CBF, CBV, OEF, and CMRO2 with attenuation correction by DIXON and DIXONbone with and without real-time motion correction. Values of CBF and CMRO2 measured with attenuation correction by DIXON were significantly lower than those measured with attenuation correction by DIXONbone in all brain regions for no real-time motion correction. Values of CBV measured with attenuation correction by DIXON were also significantly lower than those measured with attenuation correction by DIXONbone in all brain regions without the cerebellum and frontal cortex. The OEF values were not affected by attenuation correction methods.

Table 1 Values of cerebral blood flow (CBF), cerebral blood volume (CBV), cerebral oxygen extraction fraction (OEF) and cerebral metabolic rate of oxygen (CMRO2)

Values of CBF and CMRO2 with real-time motion correction were slightly but significantly higher than those without real-time motion correction in all brain regions for both DIXON and DIXONbone. No changes in CBV or OEF were observed between with and without real-time motion correction.

The coefficients of variation of CBF, CBV, OEF, and CMRO2 in the cerebral cortices were 13%, 9%, 17%, and 12%, and 13%, 9%, 17%, and 12% for attenuation correction by DIXON with and without real-time motion correction, respectively. The coefficients of variation of CBF, CBV, OEF, and CMRO2 in the cerebral cortices were 14%, 9%, 17%, and 12%, and 14%, 8%, 17%, and 13% for attenuation correction by DIXONbone with and without real-time motion correction, respectively.

Figure 2 shows the relations of CBF, CBV, OEF and CMRO2 in the cerebral cortices of each subject between values with attenuation correction by DIXON without real-time motion correction and values with attenuation correction by DIXON with real-time motion correction, or values with attenuation correction by DIXONbone with or without real-time motion correction. Significant correlations were observed in all relations.

Fig. 2
figure 2

The relations of cerebral blood flow (CBF) (a), cerebral blood volume (CBV) (b), cerebral oxygen extraction fraction (OEF) (c) and cerebral metabolic rate of oxygen (CMRO2) (d) in the cerebral cortices of each subject between the values with attenuation correction by Dixon sequence (DIXON) without real-time motion correction and values with attenuation correction by DIXON with real-time motion correction, or values with attenuation correction by Dixon sequence with model-based bone segmentation (DIXONbone) with or without real-time motion correction. Significant correlations are observed in all relations

Typical PET images of CBF, CBV, OEF, and CMRO2 with attenuation correction by DIXON and DIXONbone with and without real-time motion correction are shown in Fig. 3.

Fig. 3
figure 3

Typical PET images of cerebral blood flow (CBF), cerebral blood volume (CBV), cerebral oxygen extraction fraction (OEF) and cerebral metabolic rate of oxygen (CMRO2). a Attenuation correction is performed using Dixon sequence (DIXON). No real-time motion correction is performed. b Attenuation correction is performed using DIXON. A real-time motion correction is performed. c Attenuation correction is performed using Dixon sequence with model-based bone segmentation (DIXONbone). No real-time motion correction is performed. D: Attenuation correction is performed using DIXONbone. A real-time motion correction is performed. Scale maximum values are 70 mL/100 mL/min, 16 mL/100 mL, 0.8, and 5 mL/100 mL/min for CBF, CBV, OEF, and CMRO2 images, respectively. The image slices are the same as Fig. 1

Discussion

Using the PET/MRI scanner system with the steady-state method of 15O-labeled gases, CBF, CBV, OEF, and CMRO2 could be measured. Because of the PET measurement in the static magnetic field, non-magnetic medical devices were used for arterial blood sampling. Similar regional distributions of CBF, CBV, OEF, and CMRO2 were obtained using a PET/MRI scanner as compared with those by PET alone scanner system reported previously [22].

The values of CBF, CBV, and CMRO2 measured with attenuation correction by DIXON were significantly lower than those measured with correction by DIXONbone. One of the reasons for this is that attenuation correction of DIXON does not take into consideration of the photon absorption by bone [18, 23]. Since the values of CBF, CBV, and CMRO2 includes scale factors to radioactivity concentration in brain, these values might be affected by the attenuation correction methods. On the other hand, the values of OEF are relative values corresponding to ratios of CMRO2 to CBF. Thus, OEF values were not affected by the attenuation correction methods.

The values of CBF and CMRO2 with a real-time motion correction were slightly higher than those without correction. It has been reported that real-time motion corrected images with the present method showed about 10% large FWHM as compared with static images in phantom studies [21]. However, improving tissue mixture between gray and white matter and cerebrospinal fluid caused by head movement using a real-time motion correction might introduce increasing values of CBF and CMRO2. Although head fixation during PET measurement often causes physical stress for patients, a real-time motion correction technique might realize the PET measurement without head fixation.

The coefficients of variation of CBF, CBV, OEF, and CMRO2 in the cerebral cortices were almost the same among attenuation corrections by DIXON and DIXONbone with and without real-time motion correction. These coefficients of variation were almost the same as those previously reported, indicating the validity of measurement of CBF, CBV, OEF, and CMRO2 using the PET/MRI scanner system [17, 22].

Using the PET/MRI scanner system with adequate corrections, similar values of CBF, CBV, OEF, and CMRO2 with the PET alone scanner system reported previously were obtained [17, 22]. Measurement of CBF, CBV, OEF, and CMRO2 using 15O-labeled gases is useful for investigation of the pathophysiology of occlusive cerebrovascular disease, especially for assessment of the stage II hemodynamic compromise [1,2,3,4,5,6,7]. The PET/MRI scanner system can also be used for assessment of the stage II hemodynamic compromise.

The diffusion-weighted image (DWI) is a better imaging method to detect early ischemic lesions in stroke patients, and is widely used for clinical diagnosis of acute brain ischemia [24]. It has been reported that the area of cerebral infarction was slightly smaller than that of abnormality demonstrated by the DWI in the acute phase of brain ischemia in animal and human [25, 26]. Although the pathophysiology in such area showing discrepancy between the final area of cerebral infarction and the area of lesion detected by DWI is unknown, it can be investigated in relation to regional CMRO2 using the PET/MRI scanner system with 15O-labeled gases.

In conclusion, using the PET/MRI scanner system with the steady-state method of 15O-labeled gases, CBF, CBV, OEF, and CMRO2 could be measured. Values of CBF, CBV, and CMRO2 measured with attenuation correction by DIXON were significantly lower than those measured with correction by DIXONbone. Values of CBF and CMRO2 with real-time motion correction were slightly higher than those without correction. Using the PET/MRI scanner system with adequate corrections, similar values of CBF, CBV, OEF, and CMRO2 with PET alone scanner system reported previously were obtained. PET/MRI scanner system can be used for assessment of the stage II hemodynamic compromise.