Abstract
A novel cloud point extraction was developed for the pre-concentration and determination of Cu(II), Zn(II) and Fe(III) in bottled mineral water samples by flame atomic absorption spectrometry. The procedure was based on the formation of a charge transfer ion-pairing agent between the 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol (5-Br-PADAP) and tetraphenylborate (TPB) in the presence of 0.1 mol L−1 HCl, and the selective cloud point extraction was achieved after its complexation with target metals. Under the optimum conditions, good linearity was observed in the concentration ranges of 10–160 μg L−1 for Cu(II), 2–30 μg L−1 for Zn(II) and 20–300 μg L−1 for Fe(III). In a 50 mL of sample, concentration factors of 38.4, 39.0 and 29.9 and limits of detection of 0.44, 0.14 and 0.78 μg L−1 for Cu(II), Zn(II) and Fe(III) were obtained, respectively. Compared with single complexing reagent, the ion pair (5-Br-PADAP-TPB) significantly enhanced the enrichment factor of each analyte and tolerance levels of interference ions and obtained the lower detection limit. The proposed method was validated via the analysis of certified reference materials and successfully used for the determination of Cu(II), Zn(II) and Fe(III) in ten brands of bottled mineral water samples. All samples contained lower levels of Cu(II), Fe(III) and Zn(II) according to the maximum allowable concentration recommended by Chinese legislation. And recoveries for Cu(II), Zn(II) and Fe(III) were found to be between 95.0 and 105%; the relative standard deviation of this method was 0.620−1.67%
Graphic abstract
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Trace metals refer that their amount are less than 0.01% of the total human body weight, and they are involved in nearly all types of significant life processes, such as circulation and optimal growth of the body. Iron (Fe(III)), copper(Cu(II)) and zinc(Zn(II)) as essential trace metals play a key role in cellular metabolism. The Fe(III) is an essential component of oxygen-binding molecules (hemoglobin and myoglobin) and many enzymes involved in biological oxidation (Maru 2013). The Cu(II) is necessary for the growth and formation of bone and myelin sheaths in the nervous system. The Zn(II) functions as a cofactor and is associated with more than 200 enzymatic systems (Maru 2013). These metals are critical for life processes and sustainability but are only needed at trace and the excess intake of them can also lead to adverse health effects (Liu et al. 2013; Williams 2005; Zoni and Lucchini 2013). All trace metals must have an intake range that is safe in relation to toxicity but adequate to meet nutrient requirements (Fairweather-Tait et al. 2011). Therefore, the potential human health risk from excess trace metals cannot be ignored. Various metals occur naturally in many parts of the world, and the people are unknowingly exposed to them such as via water. For a variety of reasons, including the perception that mineral water is better than drinking water (Arve et al. 1999) or fear of contamination, many consumers now choose to drink bottled mineral water (Chiarenzelli and Pominville 2008). However, the technological processes employed during the purification of mineral water cannot completely remove inorganic contaminants in certain cases. In addition, a lower or higher mineral content has a negative impact on human health. Hence, the determination, nutrition and risk assessment of these trace metals in mineral water are necessary.
Different approaches have been reported for the analysis of Cu(II), Zn(II) and Fe(III) using highly sensitive analytical instruments including cathodic striping voltammetry (CSV) (Ghoneim 2010), inductively coupled plasma-optical emission spectrometry (Dogan et al. 2014; Javed et al. 2018), inductively coupled plasma-mass spectrometry (Tamburo et al. 2015), graphite furnace atomic absorption spectrometry (Ogunbileje et al. 2013) and flame atomic absorption spectrometry (Dasbasi et al. 2016; Sorouraddin et al. 2020). The flame atomic absorption spectrometry, a comparatively simple method, is cheap and available in many research laboratories. However, due to the low concentrations of target metals in environmental samples and the complexity of the matrix, analysis requires separation and pre-concentration methods for detection of metals (Lemos and David 2010). Numerous separation/pre-concentration methods such as solid phase extraction (Chatterjee et al. 2016; Diaz-de Alba et al. 2012; Xie et al. 2008), dispersive liquid–liquid microextraction (El-Shahawi and Al-Saidi 2013), homogeneous liquid–liquid extraction, immersed single-drop microextraction (Pena et al. 2008) and cloud point extraction(CPE) have been developed (Pytlakowska et al. 2013). Among these, CPE based on nonionic surfactants aggregation via an increase in temperature is considered to be an efficient and environmental benign extraction technique for the extraction of hydrophobic species (Manzoori and Bavili-Tabri 2002; Rusinek et al. 2015). Until to present, the majority applications of CPE to pre-concentrate metal ions are based on the formation of hydrophobic metal complexes that are extractable by choosing effective chelating agents.
5-Br-PADAP, a ligand capable of forming complexes with various metal ions, has high sensitivity and stability. It is usually used as a chromogenic reagent for the determination of metals. Presently, the application of 5-Br-PADAP as a chelating agent in CPE is receiving more attention (Unsal et al. 2015). There are some reports of 5-Br-PADAP utilized in CPE for the determination of metal ions, including iron, cobalt, nickel, copper, vanadium, cadmium, chromium, lead, mercury and manganese with atomic spectrometry and UV–Vis spectrophotometry (Aranda et al. 2008; Bezerra et al. 2006; Borkowska-Burnecka et al. 2010; de Oliveira et al. 2018; Deng et al. 2014). The uses of 5-Br-PADAP in CPE benefit from its broad-spectrum properties, which can chelate with a variety of metal ions to form a hydrophobic compound at an appropriate pH value so that the hydrophilic metal ions in the water sample can be extracted. However, it should be noted that 5-Br-PADAP method has several drawbacks such as relatively lower extraction efficiencies and weaker tolerability of interfering ions (Sabermahani and Taher 2014). If increasing the hydrophobicity of 5-Br-PADAP and metal complex, this method could further improve the selectivity and enrichment factor of target metals. Moreover, the detection limit could be concomitantly lowered.
The addition of the lipophilic counter anions can affect the overall hydrophobicity of the metal complex significantly, as well as their lipid/water partition coefficients. (Pytlakowska et al. 2013; Wang et al. 2017) Tetraphenylborate (TPB), which has lipophilic properties, consists of a boron anion and four phenyl groups (Fig. 1D). 5-Br-PADAP has three dissociation steps, and it can be protonated to form ion-pairing agent with TPB anion in acidic condition. The 5-Br-PADAP reacts with metal ion to form an amphiprotic complex cation, but in the presence of TPB anion it forms a water-insoluble complex. Here, TPB− has participated in ion-pairing complex formation as an auxiliary ligand that achieves higher hydrophobicity. Due to a more stable ternary complex formation of a metal ion with 5-Br-PADAP-TPB, the proposed method is promising for the extraction of metal ions. However, to date, there is no report on the use of 5-Br-PADAP-TPB as a chelating agent for extraction in CPE.
In this study, a CPE method using the 5-Br-PADAP-TPB ion-pairing agent was proposed for the pre-concentration of Cu(II), Zn(II) and Fe(III) prior to their determination by FAAS. The issues involved in the optimization of CPE conditions; a comparison of characteristics and results with 5-Br-PADAP were described. Ten brands of bottled mineral water samples were analyzed for these metals by the proposed method.
Materials and methods
Apparatus
An AA-6200 flame atomic absorption spectrometer (Shimadzu, Japan) with a deuterium background corrector was used in this study. All measurements were taken in an air/acetylene flame. The operating conditions were: analytical line, 324.7 nm for Cu, 248.3 nm for Fe, and 213.9 nm for Zn; slit width, 0.7 nm for Cu, 0.2 nm for Fe, 0.7 nm for Zn; lamp current, 6 mA for Cu, 12 mA for Fe, and 8 mA for Zn; air flow rate, 8 L min−1; fuel gas (acetylene) flow rate, 1.8 L min−1 for Cu, 2.2 L min−1 for Fe, and 2 L min−1 for Zn. A DK-600 thermostated bath (Shanghai Precision Experimental Equipment Co. Ltd., Shanghai, China) was used to maintain the desired temperature, and phase separation was assisted using a centrifuge (TG12Y; Hunan Xiangli Scientific Instrumentation Co. Ltd., Hunan, China). The pH measurements were taken with a pH meter (Shanghai Hongyi Instrumentation Co. Ltd., Shanghai, China).
Reagents and solution
All chemicals and reagents used in this study were of analytical reagent grade or higher purity. Ultrapure water was prepared using a SZ-97 automatic triple water distiller (Shanghai Yarong Biochemical Instrument Factory, Shanghai, China). The working solutions of Cu(II), Zn(II) and Fe(III) were prepared by serial dilution with ultrapure water from stock solutions (Cu(II), 1000 mg L−1; Zn(II), 1000 mg L−1; and Fe(III), 1000 mg L−1, China National Center for Quality Supervision and Testing of Iron and Steel, China Iron and Steel Research Institute Group). The following solutions were prepared: 5-Br-PADAP (1 × 10–3 mol L−1, Adamas Reagent Co. Ltd., Switzerland) in ethanol, a 1 × 10–3 mol L−1 solution of sodium tetraphenylborate (Tianjin Yongda Chemical Reagent Company Limited, Tianjin, China) in ultrapure water, Triton X-114 (5.0%, w/v) in ultrapure water and HNO3 (10%, v/v) in methanol. Buffer solutions were prepared from acetic acid and sodium acetate for the pH range between 3 and 6, from sodium phosphate monobasic dihydrate and disodium hydrogen phosphate dodecahydrate for pH 7–8 and disodium tetraborate decahydrate and sodium hydroxide for pH 9–11.
All laboratory glassware used for trace analysis was kept in 5% (v/v) nitric acid for at least 24 h and then washed several times with ultrapure water before use.
Preparation of ion-pairing agent (5-Br-PADAP-TPB)
A 1 × 10–3 mol L−1 5-Br-PADAP solution was prepared by dissolving 0.0349 g of 5-Br-PADAP in 3.5 mL of 1 mol L−1 HCl and then diluting 100 mL with ethanol. A TPB solution, 1 × 10–3 mol L−1, was prepared in ultrapure water and then slowly mixed with the 5-Br-PADAP solution in equal volume ratio. The mixed solution was stirred for at least 30 min until it reached room temperature. The prepared 5 × 10–4 mol L−1 ion-pairing of 5-Br-PADAP and TPB solution was stored in a brown bottle.
CPE procedure
An aliquot of a solution containing Cu(II) (10–160 μg L−1), Zn(II) (2–30 μg L−1), or Fe(III) (20–300 μg L−1) was taken in a 50-mL centrifuge tube. Adjust the pH value of the solution to 7 when extracting Cu(II) and Zn(II) with a suitable buffer, and adjust it to 4 when extracting Fe(III). 0.8 mL of 5% (m/v) Triton X-114 and 0.8 mL of 5 × 10–4 mol L−1 5-Br-PADAP-TPB were consecutively added to the tube. The solution was kept for 15 min in a thermostatic bath maintained at 40 °C. When the aqueous solution turned turbid, phase separation was achieved by centrifugation at 3500 rpm for 10 min and then the surfactant-rich phase was cooled in an ice-bath for 10 min to increase the viscosity. The supernatant aqueous phase was carefully discarded with a pipette. To decrease the viscosity and facilitate sample handling before the determination by the FAAS, the remaining surfactant-rich phase was diluted to 1.0 mL with 10% (v/v) HNO3 in methanol. Then, the sample solution was injected into FAAS for analysis.
Real samples
Ten brands of bottled mineral water samples were purchased from supermarkets, such as Trust-Mart, Carrefour and Parkn Shop. We chose the brands of mineral water at random, mainly from Guangdong, which people regularly purchase. All the samples were filtered by 0.45-µm-pore-size membrane filters and stored at 4 °C. Then, 30 mL of each sample was subjected to CPE.
Results and discussion
Characteristics of the ion-pairing agent
Reacting TPB with 5-Br-PADAP in the presence of hydrochloric in ethanol produced dark red precipitates at saturated condition. The ion-pair agent was soluble in CCl4. The proton nuclear magnetic resonance (1H NMR) was carried out to determine the structures of the complexes. Before forming ion pair, it showed two signals at 3.45 (2H) and 1.26 (3H) ppm corresponding to the methylene (CH2) and protons of methyl (CH3) groups, respectively. The spectrum displayed six signals at 5.88 (1H) to 8.50 (1H) ppm attributed to the aromatic cycloparaffin at positions, respectively. Moreover, singlet signal was observed at 15.8 ppm, as denoted by OH group (Fig. 1A). After forming ion pair, larger 1H NMR line shifts of OH group, CH2 group and CH3 group to lower δ values, reflecting slightly different chemical environments around this hydrogen species. Compared with 5-Br-PADAP, the NH+ group in 5-Br-PADAP-TPB, has electron receptor effect can increase shielding effect to H in CH2 and CH3 group because of a high charged density. This phenomenon is explained by the TPB of the B− ligands to the 5-Br-PADAP protonated in the H+, and the effect of electronic is decreased (Fig. 1B, C). The 1H NMR spectra for compounds were considered to be consistent with the proposed structures (Fig. 1D). Table 1 shows the 1H NMR spectra for 5-Br-PADAP before and after forming ion pairs with TPB.
Effect of pH
The metal complex formation and its chemical stability are two significant parameters in CPE. The pH plays a distinct role in the metal complex formation and subsequent extraction. The effect of pH ranging from 3 to 10 on the extraction efficiency of Cu(II), Zn(II) and Fe(III) was separately analyzed, and the results are shown in Fig. 2. The maximum absorbance of Fe(III) was around at pH 4, while the absorbances of Cu(II) and Zn(II) increased with increasing pH and reached plateau values in the pH range of 6–8. At pH > 5, a low extraction efficiency of Fe(III) was observed, which was attributed to the formation of hydroxide precipitate. Consequently, pH 4 was used for as the optimum pH values for maximum extraction of Fe(III), and pH 7 was selected for Cu(II) and Zn(II).
In addition, the influence of buffer volume on the absorbance was assayed at the optimum pH 4 and 7. The results showed that the maximum absorbance of metal ions was achieved at 2 mL of buffer solution. Therefore, 2 mL of the buffer solution was used in subsequent experiments.
Effect of Triton X-114 concentration
Surfactant concentration is an important parameter for effective extraction and should be sufficient to quantitatively determine the target analyses. Triton X-114 is a nonionic surfactant extensively used in CPE. Triton X-114 was chosen because of its low cloud point temperature, which is close to the ambient temperature (22-25 °C), and its high density, which can enhance the phase separation. The effect of Triton X-114 concentration on extraction efficiency was investigated in a volume range of Triton X-114 (5% w/v) from 0.4 to 1.2 mL (Fig. 3). Triton X-114 was found to be quantitatively extracted the target metal complexes from aqueous sample at surfactant volume above 0.8 mL. Beyond this optimal range, lower concentration of Triton X-114 led to the decrease of extraction efficiency. This is because the surfactant was not sufficient to entrap the metal complexes quantitatively. However, the consequence employing more surfactant volume (higher than 1.0 mL) was that the absorbance signals of these metal ions decreased. This can be interpreted that the volume and viscosity of the surfactant-rich phase increased in this situation/condition. Therefore, more solvent was needed to dilute the surfactant-rich phase, thus resulting in a loss of sensitivity. The optimal volume of 5% (w/v) Triton X-114, 0.8 mL, was chosen in the following experiments.
Effect of 5-Br-PADAP-TPB concentration
The CPE can be used for the pre-concentration of metal ions after the formation of hydrophobic complexes. The extraction efficiency strongly depends on the hydrophobicity of the chelating agent and the formed metal ion complex. The variations of the analytical signal as a function of the added 5-Br-PADAP-TPB (5 × 10−4 mol L−1) volume between 0.05 and 1.0 mL were optimized. As illustrated in Fig. 4, the analytical signals of Cu(II), Zn(II) and Fe(III) had positive relationships with the 5-Br-PADAP-TPB quantity. With an increasing volume of 5 × 10−4 mol L−1 5-Br-PADAP-TPB, the absorbance signals of these metal ions increased initially, reached the maximums at 0.4 mL and then remained constant until the highest volume tested. To ensure complete chelation of target metal ions, 0.8 mL of 5 × 10−4 mol L−1 5-Br-PADAP-TPB was selected for subsequent experiments.
Effect of equilibration temperature and time
The optimal equilibration temperature and the incubation time are necessary for complete reactions. The desirable condition was the most suitable equilibration temperature and shortest time to achieve efficient pre-concentration and complete phase separation. The equilibration temperature ranging from 30 to 50 °C was investigated. The higher extraction efficiency was observed at 40 °C for each analyte that remained stable at 50 °C. The incubation time was studied during a time interval of 5–30 min. The maximum absorbance signals for all metal ions were obtained after 15 min. Hence, temperature of 40 °C and 15 min time were selected for further experiments.
Effect of interference ions
In consideration of the high selectivity of FAAS, the interferences may come from the extraction step. The foreign ions in mineral water samples may react with chelating agent and decrease the extraction efficiency of target analytes. The effects of common foreign ions were investigated using 5-Br-PADAP and 5-Br-PADAP-TPB as the chelating agent, respectively. In this analysis, aqueous solutions containing either 50 μg L−1 of Cu(II), 10 μg L−1 of Zn(II) or 100 μg L−1 of Fe(III) as well as foreign ions were analyzed according to the proposed procedure. An ion was considered to interfere when its presence altered the absorbance signals of Cu(II), Zn(II) or Fe(III) by more than ± 5.0%. The effects of various ions are presented in Table 2.
As summarized in Table 2, both of 5-Br-PADAP and 5-Br-PADAP-TPB can form stable complexes with various metal ions, and most of the cations examined do not interfere at the mg level. Nevertheless, some of them may lead to the decrease of extraction efficiency because of the competitive complexing between the coexisting ions and target ions. However, this effect can be eliminated by masking agent. Hg(II), Mn(II) and Fe(II) were tolerated in all cases by using triethanolamine as a masking agent in acidic/weakly acidic conditions. NaF was used to mask Al(III), and sodium thiosulfate masked Cd(II) and Cu(II). 2,3-Dimercaptopropanol was used for the chemical masking of Zn(II) in acidic/weakly acidic conditions. Furthermore, it can be also observed that 5-Br-PADAP-TPB has higher selectivity for target analytes than 5-Br-PADAP, especially for Zn(II).
Analytical characteristics
To evaluate the ion-pairing agent, 5-Br-PADAP-TPB influence on extraction efficiency, the extraction of the analytes was performed in the presence of 5-Br-PADAP-TPB and 5-Br-PADAP, respectively. The analytical characteristics of the two methods are summarized in Table 3, which displayed both of two methods had good analytical characteristics. Moreover, 5-Br-PADAP-TPB method was superior to 5-Br-PADAP method for limit of detection (LOD) and enrichment factor. LODs of Cu(II), Zn(II) and Fe(III) for 5-Br-PADAP-TPB were 0.44, 0.14 and 0.78 μg L−1 in sequence, whereas LODs of target metal ions for 5-Br-PADAP were greater than the former. Enrichment factors for 5-Br-PADAP-TPB were 38.4, 39.0 and 29.9 for Cu(II), Zn(II) and Fe(III), all of which were enhanced in comparison with 5-Br-PADAP. These indicated that 5-Br-PADAP-TPB method had higher selectivity and sensitivity for Cu(II), Zn(II) and Fe(III) than 5-Br-PADAP method as a result of an increase of the overall hydrophobicity of metal complex with 5-Br-PADAP-TPB.
Table 4 compares the characteristic data of the proposed method with those reported in literatures. The proposed method had the lower determination limits and the higher enrichment factors than other reported methods. This method is more suitable for detecting trace metals in environmental samples.
Analysis of real samples
The proposed method was employed for the determination of Cu(II), Zn(II) and Fe(III) in mineral water samples, and the results are presented in Table 5. To verify the accuracy of the proposed method, two environmental water certified reference materials (GBW08608 for Cu(II), Zn(II) and GBW(E)080,407 for Fe(III)) were analyzed. The obtained values were 49 ± 0.5 μg L−1, 91 ± 2.4 μg L−1, 9.9 ± 0.11 mg L−1 (n = 6) for Cu(II), Zn(II) and Fe(III), respectively. These results were in good agreement with the certified values of 50 ± 1.0 μg L−1, 90 ± 3.6 μg L−1, and 10 ± 0.20 mg L−1. Recoveries for Cu(II), Zn(II) and Fe(III) were found to be between 95.0% and 105%, thus confirming the application of the method to water samples without matrix effects.
All samples contained lower levels of Cu(II), Fe(III) and Zn(II) according to the maximum allowable concentration recommended by Chinese legislation (GB8537-20082008). In fact, every sample contained detectable quantities of Zn(II) and Fe(III), but Cu(II) was not detected in any of the bottled water samples. A comparison of these mineral water results and the current standards for drinking water showed that all samples would pass all drinking mineral water standards for the recommended allowable concentration of trace metals (Cu < 1 mg L−1). On the basis of Baba's study (Baba et al. 2008), Cu(II) content was also below the detection level (< 1 μg L−1); for Zn(II) and Fe(III), most values were found to be below 1 and 8 μg L−1, respectively.
Types and concentrations of metals in mineral water are associated with a geological background. Guangdong province supply was dominated by metasilicate, a function of the geology of the groundwater. Hence, Cu, Zn and Fe concentrations are proportionally low.
Drinking water is an important source for the daily intake of many of these essential and nonessential metals. Most metals with a physiological function have a proper dose, and intake above or below this level will result in health problems. The average adult intake is 2 L water per day. Compared to the tolerable upper intake level of Cu (8 mg day−1), Zn (40 mg day−1) and Fe (40 mg day−1), the intakes of Cu, Zn and Fe for adults via consumption of mineral water were in safe limits. However, for the recommend intake amounts (Cu, 0.8 mg day−1; Zn, 12.5 mg day−1 for man and 7.5 mg day−1 for woman; Fe, 12 mg day−1 for man and 20 mg day−1 for woman) in Chinese dietary reference intakes, the intakes of these trace metals only relying on drinking mineral water are not enough. Thus, the intake of sufficient trace metals needs to focus on ingesting various types of foods which are rich in these metals, especially the special groups needing for nutrition.
Conclusions
A novel CPE was developed for the determination of Cu(II), Zn(II) and Fe(III) in mineral water samples by FAAS. To date, 5-Br-PADAP-TPB ion-pairing agent was first used as chelating agent in CPE. The 5-Br-PADAP-TPB method had better sensitivity and selectivity than 5-Br-PADAP method, which was attributed to an increase of the hydrophobicity of metal complex. Besides that, the method had good detection limits and enrichment factors for the Cu(II), Zn(II) and Fe(III). As a result, the proposed method offered a simple, rapid, robust and sensitive alternative to other methods.
Furthermore, 10 bottle mineral water samples from Guangdong province were analyzed by the proposed method. The Cu(II), Zn(II) and Fe(III) concentrations in all bottled mineral water samples were below the maximum allowable concentration recommended by Chinese legislation, while the intakes via water were not enough to meet the daily nutritional needs. This method could be used to determine trace metals in bottled mineral water to serve as a guide to consumers with particular health needs, and enhance susceptibility, and/or preferences in drinking water selection.
References
Aranda PR, Gil RA, Moyano S, De Vito IE, Martinez LD (2008) Cloud point extraction of mercury with PONPE 7.5 prior to its determination in biological samples by ETAAS. Talanta 75:307–311. https://doi.org/10.1016/j.talanta.2007.11.012
Arve M, Bjørn F, Ulrich S, Clemens R (1999) Variation of 66 elements in European bottled mineralwaters. Science Total Environ. https://doi.org/10.1016/S0048-9697(99)00307-1
Baba A, Erees FS, Hicsonmez U, Cam S, Ozdilek HG (2008) An assessment of the quality of various bottled mineral water marketed in Turkey. Environ Monit Assess 139:277–285. https://doi.org/10.1007/s10661-007-9833-9
Bezerra MA, Bruns RE, Ferreira SLC (2006) Statistical design-principal component analysis optimization of a multiple response procedure using cloud point extraction and simultaneous determination of metals by ICP OES. Anal Chim Acta 580:251–257. https://doi.org/10.1016/j.aca.2006.07.056
Borkowska-Burnecka J, Szymczycha-Madeja A, Zyrnicki W (2010) Determination of toxic and other trace elements in calcium-rich materials using cloud point extraction and inductively coupled plasma emission spectrometry. J Hazard Mater 182:477–483. https://doi.org/10.1016/j.jhazmat.2010.06.057
Chatterjee M, Srivastava B, Barman MK, Mandal B (2016) Combined cation-exchange and solid phase extraction for the selective separation and preconcentration of zinc, copper, cadmium, mercury and cobalt among others using azo-dye functionalized resin. J Chromatogr A 1440:1–14. https://doi.org/10.1016/j.chroma.2016.02.030
Chiarenzelli J, Pominville C (2008) Bottled water selection and health considerations from multi-element analysis of products sold in New York state. J Water Health 6:505–512. https://doi.org/10.2166/wh.2008.064
Dasbasi T, Sacmaci S, Ulgen A, Kartal S (2016) Determination of some metal ions in various meat and baby food samples by atomic spectrometry. Food Chem 197:107–113. https://doi.org/10.1016/j.foodchem.2015.10.093
de Oliveira ESAF, de Castro WV, de Andrade FP (2018) Development of spectrophotometric method for iron determination in fortified wheat and maize flours. Food Chem 242:205–210. https://doi.org/10.1016/j.foodchem.2017.08.110
Deng J, Huang K, An L, Hu C, Ju S, Xiao N (2014) Cloud point extraction and simultaneous spectrophotometric determination of V(V), Co(II) and Cu(II) ions in water samples by 5-Br-PADAP using partial least squares regression. J Radioanal Nucl Chem 300:835–842. https://doi.org/10.1007/s10967-014-3062-9
Diaz-de Alba M, Galindo-Riano MD, Garcia-Vargas M (2012) Solid phase extraction of copper traces using poly(styrene-divinylbenzene) membrane disks modified with pyridoxal salicyloylhydrazone in water samples. Talanta 100:432–438. https://doi.org/10.1016/j.talanta.2012.08.014
Dogan Y, Unver MC, Ugulu I, Calis M, Durkan N (2014) Heavy metal accumulation in the bark and leaves of Juglans regia planted in Artvin City, Turkey. Biotechnol Biotechnol Equip 28:643–649. https://doi.org/10.1080/13102818.2014.947076
El-Shahawi MS, Al-Saidi HM (2013) Dispersive liquid-liquid microextraction for chemical speciation and determination of ultra-trace concentrations of metal ions. TrAC Trends Anal Chem 44:12–24. https://doi.org/10.1016/j.trac.2012.10.011
Fairweather-Tait SJ, Harvey LJ, Collings R (2011) Risk-benefit analysis of mineral intakes: case studies on copper and iron. Proc Nutr Soc 70:1–9. https://doi.org/10.1017/S0029665110003873
Filik H, Giray D (2012) Cloud point extraction for speciation of iron in beer samples by spectrophotometry. Food Chem 130:209–213. https://doi.org/10.1016/j.foodchem.2011.07.008
Galbeiro R, Garcia S, Gaubeur I (2014) A green and efficient procedure for the preconcentration and determination of cadmium, nickel and zinc from freshwater, hemodialysis solutions and tuna fish samples by cloud point extraction and flame atomic absorption spectrometry. J Trace Elements Med Biol Organ Soc Minerals Trace Elements 28:160–165. https://doi.org/10.1016/j.jtemb.2013.12.004
GB8537-2008 (2008) Drinking natural mineral water. Ministry of Health of the People’s Republic of China, Beijing
Ghoneim EM (2010) Simultaneous determination of Mn(II), Cu(II) and Fe(III) as 2-(5’-bromo-2’-pyridylazo)-5-diethylaminophenol complexes by adsorptive cathodic stripping voltammetry at a carbon paste electrode. Talanta 82:646–652. https://doi.org/10.1016/j.talanta.2010.05.025
Manzoori JL, Bavili-Tabri A(2002) Cloud point preconcentration and flame atomic absorptionspectrometric determination of Cd and Pb in human hair. Anal Chim Acta 470:6. https://doi.org/10.1016/S0003-2670(02)00765-1
Javed H, Islam A, Chauhan A, Kumar S, Kumar S (2018) Efficacy of engineered GO Amberlite XAD-16 picolylamine sorbent for the trace determination of Pb (II) and Cu (II) in fishes by solid phase extraction column coupled with inductively coupled plasma optical emission spectrometry. Sci Rep 8:17560. https://doi.org/10.1038/s41598-018-35656-1
Kolachi NF et al (2011) Multivariate optimization of cloud point extraction procedure for zinc determination in aqueous extracts of medicinal plants by flame atomic absorption spectrometry. Food Chem Toxicol Int J Pub Br Ind Biol Res Assoc 49:2548–2556. https://doi.org/10.1016/j.fct.2011.06.065
Lemos VA, David GT (2010) An on-line cloud point extraction system for flame atomic absorption spectrometric determination of trace manganese in food samples. Microchem J 94:42–47. https://doi.org/10.1016/j.microc.2009.08.008
Liu XM, Song QJ, Tang Y, Li WL, Xu JM, Wu JJ, Wang F, Charles PB (2013) Human health risk assessment of heavy metals in soil-vegetable system: a multi-medium analysis. Sci Total Environ 463–464:530–540. https://doi.org/10.1016/j.scitotenv.2013.06.064
Maru MBT, Tessema DA (2013) Calcium, Magnesium, Iron, Zinc and Copper, Compositions of Human Milk from Populations with Cereal and ‘Enset’ Based Diets. Ethiop J Health Sci 23:8. https://doi.org/10.1002/gps.3950
Meeravali NM, Reddy MA, Kumar SJ (2007) Cloud point extraction of trace metals from seawater and determination by electrothermal atomic absorption spectrometry with iridium permanent modifier. Anal Sci 23:6. https://doi.org/10.2116/analsci.23.351
Mortada WI, Kenawy IMM, Abdel-Rhman MH, El-Gamal GG, Moalla SMN (2017) A new thiourea derivative [2-(3-ethylthioureido)benzoic acid] for cloud point extraction of some trace metals in water, biological and food samples. J Trace Elements Med Biol Organ Soc Minerals Trace Elements 44:266–273. https://doi.org/10.1016/j.jtemb.2017.09.003
Ogunbileje JO, Sadagoparamanujam VM, Anetor JI, Farombi EO, Akinosun OM, Okorodudu AO (2013) Lead, mercury, cadmium, chromium, nickel, copper, zinc, calcium, iron, manganese and chromium (VI) levels in Nigeria and United States of America cement dust. Chemosphere 90:2743–2749. https://doi.org/10.1016/j.chemosphere.2012.11.058
Pena F, Lavilla I, Bendicho C (2008) Immersed single-drop microextraction interfaced with sequential injection analysis for determination of Cr(VI) in natural waters by electrothermal-atomic absorption spectrometry. Spectrochim Acta Part B 63:6. https://doi.org/10.1016/j.sab.2008.01.005
Pytlakowska K, Kozik V, Dabioch M (2013) Complex-forming organic ligands in cloud-point extraction of metal ions: A review. Talanta 110:202–228. https://doi.org/10.1016/j.talanta.2013.02.037
Rusinek CA, Bange A, Papautsky I, Heineman WR (2015) Cloud point extraction for electroanalysis: anodic stripping voltammetry of cadmium. Anal Chem 87:6133–6140. https://doi.org/10.1021/acs.analchem.5b00701
Sabermahani F, Taher MA (2014) Determination of ultra trace amounts of copper by a multi-injection technique of electrothermal atomic absorption spectrometry after using solid-phase extraction. J AOAC Int 97:1713–1718. https://doi.org/10.5740/jaoacint.12-168
Shoaee H, Roshdi M, Khanlarzadeh N, Beiraghi A (2012) Simultaneous preconcentration of copper and mercury in water samples by cloud point extraction and their determination by inductively coupled plasma atomic emission spectrometry. Spectrochim Acta Part A Mol Biomol Spectrosc 98:70–75. https://doi.org/10.1016/j.saa.2012.08.027
Shokrollahi A, Ghaedi M, Hossaini O, Khanjari N, Soylak M (2008) Cloud point extraction and flame atomic absorption spectrometry combination for copper(II) ion in environmental and biological samples. J Hazard Mater 160:435–440. https://doi.org/10.1016/j.jhazmat.2008.03.016
Silva EL, Roldan PS, Gine MF (2009) Simultaneous preconcentration of copper, zinc, cadmium, and nickel in water samples by cloud point extraction using 4-(2-pyridylazo)-resorcinol and their determination by inductively coupled plasma optic emission spectrometry. J Hazard Mater 171:1133–1138. https://doi.org/10.1016/j.jhazmat.2009.06.127
Sorouraddin SM, Farajzadeh MA, Okhravi T (2020) Application of deep eutectic solvent as a disperser in reversed-phase dispersive liquid-liquid microextraction for the extraction of Cd(II) and Zn(II) ions from oil samples. J Food Compos Anal 93:103590. https://doi.org/10.1016/j.jfca.2020.103590
Tamburo E, Varrica D, Dongarra G, Grimaldi LM (2015) Trace elements in scalp hair samples from patients with relapsing-remitting multiple sclerosis. PLoS ONE 10:e0122142. https://doi.org/10.1371/journal.pone.0122142
Unsal YE, Soylak M, Tuzen M, Hazer B (2015) Determination of lead, copper, and iron in cosmetics, water, soil, and food using polyhydroxybutyrate-b-polydimethyl siloxane preconcentration and flame atomic absorption spectrometry. Anal Lett 48:1163–1179. https://doi.org/10.1080/00032719.2014.971365
Wang CY, Chang DA, Shen Y, Sun YC, Wu CH (2017) Vortex-assisted liquid-liquid microextraction of strontium from water samples using 4 ’,4 "(5 ")-di-(tert-butylcyclohexano)-18-crown-6 and tetraphenylborate. J Sep Sci 40:3866–3872. https://doi.org/10.1002/jssc.201700205
Williams MH (2005) Dietary Supplements and Sports Performance Minerals. J Int Soc Sports Nutr 2:7. https://doi.org/10.1186/1550-2783-2-1-43
Xie F, Lin X, Wu X, Xie Z (2008) Solid phase extraction of lead (II), copper (II), cadmium (II) and nickel (II) using gallic acid-modified silica gel prior to determination by flame atomic absorption spectrometry. Talanta 74:836–843. https://doi.org/10.1016/j.talanta.2007.07.018
Zoni S, Lucchini RG (2013) Manganese exposure: cognitive, motor and behavioral effects on children: a review of recent findings. Curr Opin Pediatr 25:255–260. https://doi.org/10.1097/MOP.0b013e32835e906b
Acknowledgements
This work was supported by Guangzhou Science and Technology Project (202002030409, 202008040002), the Student Innovation Training Program of China of Guangdong Province, China (grant numbers 201710573037), Innovative and Strong School Project of Guangdong Provincial Department of Education (2020KZDZX1122), Guangdong Science and Technology Project (2020B111106001) and Foshan Science and Technology Innovation Project (2020001000416).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Xie, M., Liu, W., Hao, X. et al. Cloud point extraction combined with flame atomic absorption spectrometry for the determination of copper, iron and zinc using 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol-tetraphenylborate ion-pairing agent. Chem. Pap. 75, 5849–5859 (2021). https://doi.org/10.1007/s11696-021-01765-x