Introduction

Retinal vein occlusion

Fig. 1
figure 1

An annotated overview of key elements during RVO [10]

Patients with retinal vein occlusion (RVO) suffer from an obstructed blood flow in the retina [21]. This can lead to severe vision impairing damage due to neovascularization, ischaemia, and oedema [15]. RVO is the second most common retinal vascular disorder after diabetic retinal disease and affects 16.4 million people worldwide [24]. Currently, no curative treatment is clinically available. A promising but very challenging experimental treatment is retinal vein cannulation (RVC) [29]. During such an intervention, the goal is to cannulate the clotted retinal vein, followed by the injection of a clot-dissolving drug to restore blood circulation. This is done using a cannulation instrument containing a microneedle, which is inserted into the eye via a small incision in the sclera (Fig. 1). During the procedure, the surgeon has visual feedback through a stereomicroscope. The illumination of the surgical scene is performed internally, by introducing a light probe, or a pair of optical fibres (i.e. chandeliers) into the eye. Given the minute scale of retinal veins, between 30 and \(400~\upmu \hbox {m}\), this is an extremely challenging task. Three key challenges can be associated with RVC: (1) guiding the needle to the target vein with an accuracy of \(10~\upmu \hbox {m}\) [23], (2) stopping at the proper depth when puncturing the vein and (3) maintaining the needle position during the injection [12]. Successfully performing these feats relies strongly on the surgeon’s capacity to manage undesired hand motion—typically in the order of 300–400 \(~\upmu \hbox {m}\) [20]—as well as on the degree of local awareness of the target anatomy. Due to these challenges, RVC remains controversial and is not performed in clinical practice today.

Robotic assistance

Several robotic systems for eye surgery have been reported to address the key surgical challenges related to retinal surgery [19]. A distinction between three categories of architectures can be made: telemanipulation, comanipulation and handheld. Physiological hand tremor is typically managed by means of motion or force scaling. Except for handheld architectures, a shared key functionality is typically the limiting of instrument motion to a single entry point in the sclera [Remote Center of Motion (RCM)] (Fig. 1). Despite the advantages offered by surgical robotics, local estimation of tool-to-surface distance and force remain unresolved problems, resulting in increased surgical risk. Depth estimation is only approximated through the observation of the shadows and the focus of the image on the instrument of the stereomicroscope [7] and is impaired at higher magnifications [6]. Furthermore, instrument-to-surface interaction forces have been confirmed to be below the threshold of human perception during retinal surgery [8, 11]. It is clear that an inaccurate perception of the depth and an absence of the force estimate create high surgical risks and is able to induce severe complications [25].

Measuring functions for surgical instruments

A precise local estimation of the interaction between the needle tip and the retina could provide a solution. For this purpose, two measurement methods are envisaged: a proximity sensor and a contact sensor, both located near the end of the cannulation instrument.

Proximity sensing

A significant amount of research has been conducted to allow intraocular detection or distance sensing. Optical coherence tomography (OCT) forms a promising approach, visualizing the anatomical structure along a single ray (A-scan), swept over a line (B-scan) or across a surface (C-scan). OCT provides non-invasive (no contact) information of the retina with micrometre resolution, at rate of few milliseconds for A-scans [7]. Previous research has typically been done relying on two different probe designs.

A lateral-imaging probe measures the tissue at the side of the probe [2]. A recent in-human pilot study including three patients has been reported. For this configuration, the feasibility of defining a retinal structure with acceptable resolution during a peel procedure of the epiretinal membrane has been demonstrated.

A forward-imaging probe measures anatomical structures directly in front of the probe tip. Research dedicated to this type of imaging probe is further developed, as indicated by the amount of the literature on the topic [3, 28]. Different applications show the potential for intraoperative use and control including compensation of physiological motions and retina tracking [28]. Balicki et al. [3] demonstrated on a synthetic phantom how OCT-based distance can be used to set up a virtual wall as a positional safety boundary.

Contact sensing

Instrument to retina interaction forces have been reported to range between 0.6 and 17.5 mN [10] in typical VRS with a possible resolution below 0.25 mN [13]. Previous research on interaction force estimation for retinal surgery has been aimed at preventing the application of excessive force with the instrument, during specific surgical tasks such as RVC and epiretinal membrane peeling [9, 14, 17]. In addition to surgical safety, intraoperative force data could contribute in identifying mechanical properties of retinal tissue. This information would be useful to build up a more realistic virtual training environments [11].

Different technologies have been proposed to measure the tool-tissue force in vitreoretinal: semiconductor strain gauges [18], microelectromechanical systems-based diffractive optical encoders [30] and intensity-modulated fibre optic sensing [22]. Nevertheless , these methods experience problems in distinguishing contributions from force exerted on the instrument tip and forces that are exchanged at the interface with the sclera. A proposed solution is to integrate the measurement functionality at the tip of the instrument. This imposes additional challenges in terms of biocompatibility, sterilizability and safety.

Fibre Bragg grating (FBG) based on optical fibre provides a solution to these issues. FBG can detect fine changes in strain and thus when paired with a known flexure. FBGs are adequately small sized (diameter of 60–200 \(\upmu \hbox {m}\)), inherently safe, biocompatible, sterilizable, relatively inexpensive and highly sensitive. Their output is insulated from electrostatic and electromagnetic noise [1]. Several works in the literature describe how FBGs have been embedded in ophthalmological instruments [9, 11, 13, 17, 26].

Objectives

A first effort to combine two sensing modalities to assist RVC was reported in [27]. Hereby, it was shown that relevant distance and force data can be acquired on ex vivo enucleated porcine eyes. However, this conceptual design did not integrate the two sensors within a single instrument. A second version of the instrument was designed and characterized in [26]. This second version combines both sensing modalities in a single instrument and was validated in vitro and ex vivo. This paper provides some insights on the data (force and distance) collected with the instrument for in vivo animal trial. The experiments were carried out to evaluate the previously developed sensorized instruments. Depth and contact were the major issues, different groups have developed sensors in the past, but there have been very little validated in a realistic in vivo context. With this experiment, we investigate practical issues that may occur. Understanding of the force and damage that occurs may allow developing improved robotic assistance schemes that improve safety and reliability of interactions with the retina. This paper progresses our prior work [26, 27] and introduces two prototypes in “Design of a cannulation needle with combined distance and force” section. “Animal trial” section depicts a first animal trial to evaluate the combined sensor in real operation. The performance of the combined OCT distance and FBG force is reported. Finally, conclusion and future perspectives are given in “Conclusion and future work” section.

Design of a cannulation needle with combined distance and force

This section describes the design of a combined force and distance sensing cannulation instruments based on FBG and OCT A-scan technology. Both sensing modalities are briefly described, after which the conceptual designs of two prototypes are presented.

Force sensing

In order to integrate force sensing modality into a cannulation needle, FBG fibres were chosen. An initial design of a 2-DOF force-sensitive needle serves as a basis [9]. In addition to the FBG fibres, a Micron Optics sm130 interrogator was used, which offers a wavelength detection resolution and scan frequency of, respectively, 2 pm and 500 Hz.

The resulting prototype offers a thermally compensated 1-DOF force measurement, showing negligible drift, and a precision and measurement range of 0.20 and 20 mN, respectively. A detailed overview of the force sensor characterization can be found in [26].

Distance sensing

A custom-made long-range optical distance sensing fibre providing OCT A-scan data, with an outer diameter of \(125~\upmu \hbox {m}\), was made by Medical Laser Center Lübeck (MLL). The OCT A-scan fibre was connected via an optical switch to an available medical grade iOCT scanner (OptoMedical Technologies (OPMed)). The OCT A-scan fibre was aligned to target a sample with approximately 1–3 mm distal from the needle tip reference. The OCT A-scan data were collected at a sample rate of 200 Hz, at a pixel resolution of \(3.8~\upmu \hbox {m}\). Precision and accuracy of the distance measurement were characterized at 0.064 and 0.010 mm, respectively, for in vitro samples, at a maximum response time of 0.1 s. A detailed overview of the distance sensor characterization can be found in [26].

Proposed prototypes

Two different configurations have been proposed to evaluate this combination.

A hollow shaft with four parallel grooves acts as a flexure, with the FBG fibres glued into two opposing grooves, and the OCT fibre in one of the other remaining grooves (Fig. 2). These FBG fibres provide strain measurements due to bending in the y direction. Each opposing strain measurement provides an estimate of a radial force component (\(F_{p}\)) along the instrument’s y direction applied at the needle tip. The difference between strains provides an estimate that is free of thermal effects.

Fig. 2
figure 2

Conceptual model of the force in the cannulation needle

Shared tube (ST) prototype

STP (Fig. 3) combines all fibres on a single grooved hollow shaft of 0.55 mm outer diameter. The fibres are routed near the tip of the instrument. The main advantages of this design are: ease manufacturing, simplicity and small outer diameter. A limitation is that the OCT fibre is not oriented towards the tip of the needle, but measures the distance at a certain offset, in x and y direction. Therefore, the needle tip is not in the field of view of the OCT A-scan. This requires a pre-calibration of the needle tip location with respect to the OCT fibre. Furthermore, the sensorized tube is directly exposed to the sclera entry point.

Fig. 3
figure 3

Conceptual sketch of STP: OCT and FBG fibres are aligned along three grooves in the same shaft

Outer sleeve (OS) prototype

OSP (Fig. 4) is composed of an internal part for force measurements and an external part made of an outer sleeve for distance measurements. The assembly is more complex but allows to position the needle tip within the OCT view. This prototype allows collocated force and distance measurement. The outer sleeve shields the force measuring part, protecting the measurement segment from large sclera interaction forces. However, the resulting diameter of the outer sleeve that is to be inserted inside the eye is 2.3 mm with a 0.6 mm protuberance.

Fig. 4
figure 4

Conceptual sketch of OSP: OCT and housing surrounds a force sensing cannulation needle

Animal trial

Fig. 5
figure 5

Operating theatre of animal trial

Fig. 6
figure 6

Distance measurement based on OCT fibre

The performed animal study was approved by the Ethical Committee for Animal Research at Medanex Clinic. In total, experiments were performed on 3 pigs. Four surgeons from three different institutes participated, taking turns throughout 2 days to perform the experiments. A protocol for the experiments was set up and adhered to. All required equipments were transported to the facility and field-tested beforehand.

A Haag-Streit surgical stereomicroscope (Hi-R NEO 900A NIR) was used, paired with a dedicated iOCT system (OPMedt GmbH). In addition, a fibre OCT A-scan probe from MLL embedded within the cannulation needle was plugged into the same OCT system via a purposely developed fibre switch. Force was measured using a Micron Optics sm130 interrogator connected to a real-time PXI industrial computer from national instruments. Two cannulation needle prototypes were fixed to a previously developed comanipulation surgical robot [10]. For research purposes, surgical robot control was implemented on a real time (Xenomai 6). The hardware of the robot (motors, encoders and two foot pedals) was connected to a national instrument NI9144 EtherCAT slave with embedded FPGA used for data acquisition. The control programme was running on OROCOS a middle-ware for real-time robot control [5].

A general view of the operating theatre at Medanex Clinic facility is provided in Fig. 5. During the animal trial, both prototype needles were tried out for evaluation. Even though suitable for laboratory work, the OS prototype was found infeasible to safely introduce into the eye due to its large outer diameter (2.3 mm). In the following, only the ST prototype was used and its data depicted.

OCT distance versus insertion depth

Fig. 7
figure 7

Evaluation of the OCT distance versus the robot insertion depth motion

Fig. 8
figure 8

Evaluation of the force vs. the robot insertion depth motion

The first experiment was conducted to check the quality of the distance measurements with the OCT fibre. The estimation of the distance within the 4 mm possible by the OCT is shown in Fig. 6. The extraction of the distance from the A-scan was based on the work presented in [4] for further details.

Figure 7 depicts the resulting curves of the robot insertion depth motion (translation along the needle axis) and the OCT distance estimation within 4 mm range of the OCT sensor. The OCT distance shows a similarity with the robot insertion depth but not correspondence. This is most likely due to slight bending of the instrument caused by sclera interaction forces, resulting in a measurement mismatch between the robot workspace and local measurement data. In conclusion, the OCT A-scan signal can get a real-time estimate of the distance to the retina surface, but care must be taken to filter and interpret its data. Furthermore, it highlights the challenge of combining both robotic and local measurement data.

Force versus insertion depth

Fig. 9
figure 9

Evaluation of the force vs. the OCT distance for a short time contact

A second experiment was conducted to evaluate the behaviour of the force sensor inside a closed porcine eye. This was the first time—to the knowledge of the authors—that such a test with a FBG sensor was conducted on a closed eye. The sensor allows to measure the force in two directions such as if the force is applied in opposition to the needle tip, it will decrease and conversely. Continuous contact with the needle on the retina was tested. The curve of the experiment is depicted in Fig. 8. The robot is locked between 5 and 18 s where the insertion depth is constant. The force shows an increase of 2.5 mN in the force measurements. It was observed that this force is caused by eye motion of the sedated animal, assumed to be caused by breathing. This motion is not noticeable from the robot encoders but can be seen on the stereomicroscope. Note when retracting the needle, between 20 and 26 s, noticeable increase of the force can be seen. These forces were observer to be caused by sclera interaction forces during instrument extraction.

Multiple approaches to the retina were made to understand the behaviour of the force sensor in the closed eye. Data of the force and OCT distance were gathered simultaneously during this multiple contact approach. Figure 9 depicts the evolution of the two signals. Between 155 and 165 s the distance does not change, yet a force peak is seen. This variation of the force is related to the physiological motion that will be demonstrated in the next subsection. Between 165 and 175 s, the surgeon is retracting the robot and a force peak in the same direction is measured.

STP was able to measure forces inside the eye. The values of the force were between \(+\,20\) and \(-\,20\) mN for the different experiments. The continuous contact shows the possible use of the force to estimate the force at the needle tip even when such is not easily noticed with the distance measurement. However, even though not directly measured, a clear correlation between robot motion and force measurements can be observed. Furthermore, when purposefully misaligning the surgical robot with the sclera entry point, this correlation was observed to scale up. It is concluded that this correlation is caused by the direct interaction between the fibres in the grooved tube and the trocar at the sclera entry. This highlights a key design concern for the usage of FBG fibre-based through a trocar, as this requires some form of mechanical isolation for the bare, non-measuring section of the fibres. This could be achieved as proposed in the OS prototype, using a stiff outer tube concept.

Acquisition of physiological motion

Fig. 10
figure 10

Evaluation of force vs. robot insertion depth motion for the estimation of the physiological motion

Physiological motion caused by breathing and the heartbeat could be observed by using the OCT distance estimation. The breathing motion depicted in Fig. 10 approximates a sinusoidal trajectory. The breathing frequency was approximately 11 breaths per minute. Moreover, the effect of the heartbeat was visible as a small variation superpositioned on the sinusoidal motion at approximately 120 heartbeats per minute. Peak-to-peak position amplitudes are observed to be approximately 0.1  and 0.3 mm for motions caused by heartbeat and breathing, respectively.

The force sensor was used to measure the physiological motion. Figure 10 shows the evolution of the OCT distance measurement for a constant insertion depth and for different positions on the retina. From the graph we see that also when in contact the distance measurement works as well. The corresponding force signal is depicted in Fig. 10. Even though slightly scaled down, the force sensor continues to clearly depict the physiological forces when not in direct contact with the retina. This confirms the earlier made conclusion of force interference due to sclera interaction forces.

Conclusion and future work

This paper presented two designs of sensorized cannulation needles enabling combined force and distance sensing. The first results of in vivo porcine animal experiments are reported. For both prototypes, the force measurement relies on two FBG fibres that are glued on a custom 0.6 mm-diameter hollow flexure. The distance measurement that relies on an integrated OCT fibre is used. Two different layouts have been investigated here. In STP, the same shaft of the cannulation needle is employed, leading to a small outer diameter of 0.6 mm, providing non-collocated force and distance sensing. In OSP, OCT fibre is embedded in a separate construction leading to a collocated measurement, but resulting in a large outer diameter of 2.3 mm. A characterization and calibration of the two prototypes was conducted. In order to validate the feasibility of using such sensorized instrument in an in vivo setting, an animal trial campaign was conducted at Medanex Clinic facility. During these experiments, surgeons validated the OCT A-scan probe distance estimation technology, in order to enable future robot-assisted procedures within the context of an initial in vivo human study. At the same time, these experiments were conducted to further validate the functioning of the combined OCT and force sensing cannulation needles. Within this effort, a unique data set was gathered of in vivo OCT and force sensing.

Whereas both instruments (ST and OS ) worked well in a laboratory setting, it was found that the larger outer diameter OS concept was not feasible in an in vivo setting. Surgeons managed to successfully use the ST prototype throughout the preconceived experimental protocol. A key observation was that a clear coupling between force measurements and sclera interaction forces was observed. This highlights the need for a reliable mechanical decoupling of the present fibres from the sclera entry. Despite this setback, valuable force and OCT data were able to be measured and is still being processed. Intraoperative retinal motion due to heartbeat and breathing motion was found to be in the order of 0.1 and 0.3 mm, respectively. To the knowledge of the authors , this has been the first time that such a combined force/OCT cannulation needle was tested in vivo. Valuable insights have been obtained, which will form the foundation for future work on the development of improved smart surgical instrumentation.