1 Introduction

Atherosclerosis is a major vascular disease and remains the leading cause of world mortality [9, 14]. It has been widely accepted that cardiovascular diseases are associated with hemodynamic conditions [20]. Hemodynamic parameters contribute to the initiation, localization and development of stenosis as a major health risk [1, 17, 38]. Some of the mechanical parameters that have been verified to be associated with development of arterial stenosis are as follows: hypertensive pulsatile blood pressure and high cyclic circumferential stress (CS) on arterial wall [6], disturbed flow patterns [15], low, excessive and highly oscillating wall shear stress (WSS) [2, 4, 37], temporal variation of the shear stress [26] and spatial wall shear stress gradient [18].

Among the abovementioned parameters, mean wall shear stress (time-averaged WSS) and oscillatory shear index (OSI) are two critical parameters, which play important roles in mechanics of atherosclerosis [28, 30, 34]. From the pathological point of view, in some regions such as bifurcations and stenotic sites, OSI has been shown to be relatively high and mean WSS relatively low compared to nearby sites [34]. In addition to the direct influence of wall mechanics and fluid dynamics, mechanical stresses are influenced by wall–flow interactions. Because of the flexibility of arteries, the arterial wall displacement, which is caused by the pressure pulse, affects the blood flow pattern. Theoretical considerations of interactions between elastic wall and blood flow significantly affect characteristics of flow field and wall domain and therefore influence hemodynamic parameters, particularly pulsatile wall shear stress, and pressure gradient waves in both healthy and stenotic arteries [27].

A considerable number of studies, both experimental and computational, have investigated effects of hemodynamic parameters on blood flow and development of arterial diseases [15, 2729, 33, 36, 41]. Although scholars have attempted to conduct experiments on cardiovascular diseases, they performed under simplifying assumptions overlooking some parameters such as physiological flow condition, wall elasticity and fluid–wall interaction. Such experiments focused on the flow downstream of the stenosis, and particularly on the disturbances induced by the stenosis; however, they failed to consider the effects of wall elasticity [23]. However, most studies focused on the conditions which may be related to artery collapse and plaque cap rupture such as considerable compressive stress in the tube wall and critical flow conditions like negative pressure, high shear stress and flow separation [36].

Fluid–structure interaction (FSI) modeling and computational methods have been extensively employed in the study of cardiovascular function and pathology, and also blood flow characteristics through stenotic arteries have been investigated under different assumptions [15, 2729, 36]. In most studies blood flow has been restricted by some assumptions such as steady flow [3, 36]; models of stenotic arteries as rigid-walled tubes [3, 11] compliant wall under static no-flow conditions [13, 22]; or disregarding effects of arterial wall–blood interactions [3, 20]. The scope of the aforesaid investigations was not extended into two parameters: (1) the relation between stenosis severity and shear stress pulse and (2) the angular phase difference between wall shear stress and circumferential stress waves due to elastic wall–fluid coupling. These two parameters seem to be determinants of endothelial cell pathology in predication of potential site of stenotic plaque growth [8].

The main purpose of this study is to provide a more accurate understanding of hemodynamic parameters of blood flow in models of stenotic arteries, considering nearly realistic conditions of physiological blood flow. An experimental setup was designed and developed to produce original pulsatile arterial flow. This setup is capable of monitoring and measuring blood pressure waves for carrying out experiments in different scales of symmetric stenosis using a compliant tube similar to an elastic artery. Moreover, a numerical model considering effects of FSI was developed to investigate wall shear stress, circumferential stress waves and the phase shift between them, based on the output of experimental setup. Given the concept of stress phase angle, the wall–blood coupling was studied through interaction of pressure flow pulses, and the consequent cyclic shear and circumferential stress values were evaluated. The results can provide a useful means to extend scope of effects of stenosis on hemodynamic parameters and mechanisms involved in plaque growth.

2 Methods

2.1 Experimental setup

The setup is shown in Fig. 1, and it is capable of producing a flow wave similar to that of an original physiological blood flow in desired arteries and also detecting and measuring the consecutive blood pressure and pressure difference waves in real time (Fig. 1), given the flow wave pattern as the input. For the current study, a model of the brachial artery was chosen due to the availability of its physiological flow wave and pressure pulse data, together with its mechanical and geometrical parameters. In addition, the brachial artery is highly compliant and the effects of elasticity of the wall on flow patterns are prominent. Within the human body most of the arteries are less distensible; however, even in those cases it is important to know how the compliance of that artery might contribute to the flow regime specifically when luminal obstruction is involved. Hence, although brachial arteries are less prone to atherosclerotic plaque formation, for a typical distensible artery the influence of wall elasticity on the flow can be studied when the artery is obstructed with different scales of stenosis. See Appendix for more details about the experimental setup.

Fig. 1
figure 1

Schematic representation of the experimental setup, which is included a closed loop and Windkessel model (elastic and resistant elements, pulsatile pump and reservoir)

2.2 Experimental protocol

Before running each experiment, the pump was discharged of any remaining bubbles using discharge valves arrayed on both transducers. Calibration of the pump was then accompanied to minimize probable errors in producing the original flow and in measurement system using Chart for Windows™ v5.0.1 software, which the output signals of the transducers were adjusted to the real numerical values.

The initial experiment simulated the flow in a healthy brachial artery. The blood flow wave of brachial artery was set to the pump to produce an original arterial pulsatile flow [24]. Mean flow rate, \( {\bar{\dot{\text{Q}}}} \), was equal to 3.66 (ml s-1), and the frequency of the wave, ω, was 1.16 (Hz) (70 heart BPM). The fluid (a mixture of water, 65 % by weight; glycerin, 35 % by weight) was chosen to be a Newtonian fluid with a viscosity of μ = 3.5 (cP) and density of ρ = 1050 (kg m-3), to mimic the density and viscous properties of blood in large arteries [12]. The dimensionless parameters of Reynolds (Re) and Womersley (Nw) numbers for flow in this experiment were calculated as 297 and 1.4, respectively. The values comply with published data for physiological flow conditions of the same level arteries and are in close analogy.35 The inlet, P1, and outlet, P2, pressure waves were detected by means of the data acquisition and processing system, and consequently the pressure difference wave was calculated.

The main experiment was designed and performed with the aim to study the effect of different scales of symmetric stenosis on pressure and pressure difference waves with assumption of constant flow to distal part. The characteristics of the flow wave and the setup were maintained consistent with the previous experiment, except different scales of symmetric stenosis were set to the elastic tube. The stenosis profiles were generated by external constraints around the wall of elastic tube as described above and shown in Fig. 2. The degree of stenosis severity was defined as the degree of reduction in cross-sectional area of lumen, varied as 5, 10, 20, 30, 50, 60, 70 and 80 % to investigate effects of stenosis severity on hemodynamic parameters. To avoid the circumferential effects on deformation within the stenotic site and altered velocity pattern, boundary conditions were allocated far enough.

Fig. 2
figure 2

a Transverse section of the meshed model, which was used in numerical modeling, the wall section has 8-node hexahedral mesh and the fluid field has 4-node tetrahedral mesh (see appendix). b Geometry of the total model, including the stenotic area. Z1 and Z2 are longitudinal distances of start and end of stenosis from the tip of arterial model, respectively, and the length of stenosis is shown in Detail A. r and t are internal radius and thickness of the tube, respectively. c External housing and bolts for creation of the stenosis

2.3 Numerical analysis

The governing equations in fluid domain (continuity and momentum conservation) for non-Newtonian modified Casson model, in solid wall (elastodynamics equations), and also fluid–solid coupling are considered in this study, which are all discussed in detail in the appendix. Also the numerical modeling and boundary conditions are discussed in the appendix.

To verify accuracy of numerical simulation, results for effective stresses at stenotic throat in different stenosis severities (20, 50 and 70 %) were compared to the published numerical results of Valencia and Baeza’s (2009) [39] using ADINA software. All computational models including elastic modulus, lengths, density, profile of stenosis and viscosity of fluid were considered similar to those used in Valencia and Baeza’s. The given results from our simulation describe good agreements as shown in Fig. 3a. Moreover, minimum and maximum effective stresses at pre-stenotic zone for the stenosis severity of 50 % were calculated to be 52.3 (kPa) and 88.1 (kPa), respectively, comparable to the reported values of 55 (kPa) and 84.9 (kPa) [39]. Additionally, acceptable compatibility was achieved between results of WSS in this study compared to those of Kaazempur-Mofrad et al. [16] (Fig. 3b).

Fig. 3
figure 3

Numerical validation: a comparison of time variation of effective stress for three different stenosis severities at stenotic throat with Valencia et al. study, b comparison of WSS along the axisymmetric stenotic artery (z) with S = 56 % and Reynolds number = 250 with Kaazempour-Mofrad et al. study

3 Results

3.1 Experimental results for models of stenosed arteries with different severities

In accordance with experimental conditions, the results for inlet and outlet pressure waves (P1 and P2) for normal and stenosed models with degrees of severities from 5 to 80 % are presented in Fig. 4a, b. Although luminal obstruction causes change in the inlet and outlet pressure waves, the elevation of P1 is dominant and changes in P2 are less noticeable.

Fig. 4
figure 4

a Inlet (P1), which is elevated by increasing stenosis severity, b outlet (P2) pressure waves, which remained almost unchanged by increasing stenosis severity, c pressure difference waves obtained in experiments in different scales of stenosis and d inlet, outlet and pressure difference waves with error bars for a typical model (S = 60 %)

In addition to the inlet and outlet pressure waves, luminal obstruction influences the pressure difference wave. The pressure difference wave was evaluated by subtracting the outlet (P2) and inlet (P1) pressure waves in real time, measured by pressure transducers simultaneously. The resultant pressure difference waves are shown in Fig. 4c. The pressure drop is more intensified when the severity is elevated. Such effect is highly prominent in higher grades of stenosis. The increased pressure difference is due to increase in the resistance against the flow by stenosis. Additionally, increased severity results in rightward shift of peak of pressure difference as demonstrated in Fig. 4c. The resultant pressure waves have been obtained from different operating cycles of the pump and hence are presented in averaged values. Due to accumulation of results figures are not provided with error bars. To demonstrate deviation of results, Fig. 4d shows pressure waves for a typical model with stenosis severity of 60 % with error bars. The rest of results indicate deviations within almost the same range.

The mean values of inlet pressures are calculated and presented in Fig. 4 for different degrees of stenosis. In lower scales of stenosis (lower than 60 %), alterations in the magnitude of mean inlet pressure are small, beyond that changes are considerable. Hence, to maintain a certain flow due to metabolic requirements of distal tissues, the pressure difference and inlet pressure have to be elevated. The significant rise in the inlet pressure is highly prominent after specific degree of stenosis, in current study about 60 % (Fig. 5). Such degree of stenosis describes a critical condition, beyond which the mean inlet pressure has to be elevated significantly to maintain the flow.

Fig. 5
figure 5

Effect of elevation of stenosis severity on mean P1 value. At 50 % stenosis the mean upstream blood pressure increased by 10 mm Hg, which it can be considered as a critical percentage of stenosis severity

3.2 Numerical results for FSI models of stenosed arteries with different severities

3.2.1 Model verification

Resultant P1 values obtained from numerical modeling with boundary conditions of P2 and flow waves as experimental parameters are presented in Fig. 6 for stenosis severities of 20, 50 and 80 %. The P1 waves resulted from FSI modeling are compared to those obtained from experiments.

Fig. 6
figure 6

Comparison of experimental and numerical inlet pressure waves for 20, 50 and 80 % stenosis severity. All the comparisons show a good agreement between numerical and experimental approaches

Results indicate average values of deviation of 1.1, 1.1 and 1.07 % for models with 20, 50 and 80 % severities, respectively. For maximum values of deviation these values correspond to 4.7, 4.8 and 4 %, respectively. The results of numerical modeling for P1 values are in reasonable agreement with experimental results, as the waves only differ by the criteria of normalized root mean square deviation of NRMSD = 0.01. The NRMSD is calculated by Eq. 1:

$$ {\text{NRMSE}} = \frac{{\sqrt {\frac{{\mathop \sum \nolimits_{i = 1}^{n} \left( {{\text{P}}_{{{ \exp },i}} - {\text{P}}_{{{\text{num}},i}} } \right)^{2} }}{\text{n}}} }}{{{\text{P}}_{ \hbox{max} } - {\text{P}}_{ \hbox{min} } }} $$
(1)

where n indicates total time steps in each cycle and Pexp and Pnum correspond to experimental resulted and numerical calculated inlet pressures, respectively. Such accordance enables calculation of shear and circumferential stress values based on experimental conditions. To study details of such stress waves, results are presented and discussed in three separate regions: pre-stenotic, stenotic and post-stenotic. Pre- and post-stenotic regions are defined as 1.5D (1.5 times of lumen diameter) before and after the stenotic region, respectively.

3.3 Wall circumferential stress

Figure 7 illustrates the circumferential stress waves for models with three degrees of stenosis at three sites.

Fig. 7
figure 7

Comparison of circumferential stress pulses for S = 20, 50 and 80 %. By increasing stenosis severity, the CS at stenotic throat decreased severely, but remained constant at post-stenotic region

3.3.1 Wall shear stress (WSS)

In contrast to circumferential stress, wall shear stress (WSS) increases significantly at the stenotic throat (Fig. 8).

Fig. 8
figure 8

Comparison of wall shear stress pulses for stenosis severities of 20, 50 and 80 % (all diagrams are normalized to the peak WSS). By increasing stenosis severity, the WSS at stenotic throat increased significantly and became more negative at post-stenotic region

The conduct of the WSS along the axis of the artery is demonstrated in Fig. 9 through mean shear stress (the average value of shear stress during arterial pulse) and oscillatory shear index (OSI) parameters.

Fig. 9
figure 9

Mean wall shear stress and oscillatory shear index along the artery axis in the model with different stenosis severities. There is a peak value of mean WSS at stenotic region, and it became more severe by increasing stenosis. Also, the OSI became wider in 80 % stenosis compared to lower stenosis

3.4 Angular phase shift

It has been shown that angular phase difference between wall shear stress and circumferential stress waves (stress phase angle, SPA), which an endothelial cell in arterial wall experiences, is a key parameter in endothelial cell pathology [28, 29]. The higher absolute values of SPA describe more critical conditions [28, 29].

Figure 10 indicates magnitudes of SPA based on the experimental results and FSI modeling. The calculation of angular phase difference is based on discrete Fourier transform for the first eight harmonic frequencies, at the stenotic throat for severities of 20, 50 and 80 %.

Fig. 10
figure 10

Angular phase difference between wall shear stress and circumferential stress waves, for first eight harmonic frequencies at post-stenotic region. SPA values became more negative in more severe stenosis

4 Discussion

Endothelial injury is the major determinant of atherosclerotic plaque formation and progression [8, 42]. Local and systemic hemodynamic factors such as WSS, CS and the phase difference between them contribute to such phenomenon. In this study, the calculated stress values through experimental setup and computational models were in accordance with published data. The range of WSS for specific arteries with 50–56 % stenosis severity has been shown to be −0.2 to 6.5 (Pa) reported by Kaazempour-Mofrad et al. [16] and −0.6 to 5.7 (Pa) reported by Lee and Xu [19]. In this study, such values were within the range of −0.5 to 7 (Pa). The peak WSS value for the healthy common carotid and brachial artery has been proposed to be 3.4 ± 0.8 (Pa) and 3.9 ± 0.8 (Pa) and for the mean WSS value to be 1.15 ± 0.21 (Pa) and 0.48 ± 0.15 (Pa), respectively [7].

Considering flow in cylindrical tubes, pressure gradient is mostly sensitive to the tube radius. Local decrease in luminal radius due to elevation of severity leads to the increase of pressure gradient. In current study, flow regime was set to resemble physiological conditions where the pressure in arterial tree is regulated to maintain blood flow through stenotic arteries for metabolic requirements. If the flow rate has to be constant due to downstream metabolic requirements, the pre-stenotic pressure and pressure difference rise (Fig. 4b, c). To keep flow constant, as described by numerical modeling, the inlet pressure has to be elevated. The increase in the upstream pressure, to compensate resistance of stenosis, leads to elevation of pressure of parallel non-stenotic arteries. If the rise of pressure is significant due to the critical severity, a systemic hypertension is manifest in real biological conditions. By assumption of constant flow rate, if the inlet pressure is elevated excessively, it is assumed that such trend cannot be maintained due to effects on systemic pressure. Hence, to avoid such pressure rise, flow rate may decrease. This means that by increase of stenosis severity, until the inlet pressure reaches the critical point, the flow is set almost constant. However, further increase in severity leads to blood deficiency to avoid systemic hypertension.

The degree of critical severity might vary among different arterial sites. Such degree depends on local geometry and hemodynamic parameters. In this study, the critical severity occurred almost at 50 %, which cased to 10 mm Hg elevation in upstream blood pressure. Based on previous studies, 10 mm Hg pressure increase may result in endothelial cells damage [10]. For severities smaller than 50 %, to maintain flow rate, the inlet pressure increase was less than 3 %. However, a further elevation of severity from 50 to 70 % caused an extra increase of 14 % in inlet pressure. The value of critical severity changes when the radius of artery and the value of blood flow are altered. In other words, the critical degree of severity depends on arterial site. Study of such value is of great importance in major arteries at risk, such as coronary and carotid arteries.

While in general, by elevation of stenosis severity, the maximum and minimum values of circumferential stress waves (31.5–43 (kPa)) remained constant in pre-stenotic site, such values were reduced in stenotic regions (from the range of 31.5–43 (kPa) non-stenotic model to 13–17.5 (kPa) for 80 % stenosis severity) due to increased plaque thickness which makes the wall displacement smaller (from 1.1 × 10−6 (m) to 3.9 × 10−7 (m) at the peak value) and consequently the stress level smaller.

The circumferential stress pulse amplitude at the stenotic site decreased significantly as the stenosis was elevated by the amounts of 8.3, 6.6 and 4.7 (kPa) for S = 20, 50 and 80 %, respectively, but it almost remained constant (11.5 (kPa)) at pre-stenotic region. Hence, for higher values of luminal obstruction, the variation of circumferential stress along the lumen axis is more prominent.

In post-stenotic section, small reductions in circumferential stress are observed. This is due to sharp pressure drop in more severe stenotic region, which leads to lower maximum post-stenotic pressure and consequently lower circumferential stress. In critical conditions with significant pressure drop the post-stenotic neighborhood undergoes collapse [43]. In current study, for the model with 80 % severity, the minimum value of post-stenotic CS was as low as 30 kPa, while the values of CS at pre-stenotic region remained almost constant. For idealistic models with high values of pressure gradient, significant collapse has been shown to occur in post-stenotic region [36].

In contrast to circumferential stress, wall shear stress (WSS) increased significantly at the stenotic throat (Fig. 8). Such increase was markedly intensified by elevated severity. Both maximum shear stress value and shear stress pulse amplitude were elevated by increased severity. The magnitude of maximum shear stress increased 171 % from 20 to 50 % severity and 262 % from 50 to 80 % severity, and shear stress pulse amplitude increased 150 % from 20 to 50 % and 253 % from 50 to 80 % severity. Additionally, by elevation of stenosis severity, a more negative WSS was observed at post-stenotic site. As can be inferred by negative values of WSS in Fig. 8, increased severity induced more back flow and vortexes in post-stenotic region. For the severity of 80 %, the peak value of WSS dropped to −3.7 Pa which might contribute to endothelial injury and plaque progression in axial direction. For higher values of luminal obstruction WSS fluctuated mostly with negative values, describing potential pathological outcome such as endothelial injury and post-stenotic axial plaque growth [34].

The post-stenotic high variation of mechanical stresses induced by stenosis severity manipulates the phase difference between shear and circumferential stresses. Since WSS was reduced and the circumferential stress was elevated at the post-stenotic site, as the stenosis of artery became more severe, the stress gradient was intensified along the axis of tube with higher severities resulting in altered phase difference and possible pathological consequences. Models with high grade of stenosis showed higher elevation of WSS value at stenosis and higher reduction of WSS distal to plaque.

Alteration in WSS pattern caused by disturbed blood flow is a major determinant of endothelial damage along the arterial tree through altered permeability to LDL and other large molecules, platelet activation and aggregation [31]. Changes in WSS modulate diameter adaptive responses, intimal thickening and thrombosis [31]. Maximum WSS occurred at the throat of the stenosis. Very high shear stress values may contribute to endothelial injury and be related to platelet activation and aggregation [4, 31]. Results showed that in higher stages of stenosis severity, WSS values (both maximum value and stress wave amplitude) were significantly elevated. Figure 8 shows WSS wave at stenotic throat for different scales of stenosis. Unlike circumferential stress, shear stress pulse is increased. For a highly obstructed model (severity of 80 %) the value of WSS is within the range of published data for endothelial damage.

Low and in some cases negative shear stress trigger plaque formation and further growth of initiated plaques [21]. In the post-stenotic region, WSS is reduced leading to a negative value for high grades of stenosis (Fig. 8). Models with high grade of stenosis result in sharper reduction of WSS distal to plaque and more intense fluctuation during pressure pulse. Recirculation zones as a result of shear stress patterns downstream of the tube intensify possibility of transferring large molecules within the wall, including lipids, which often are found in plaque composition [21]. Patterns of WSS in post-stenotic region shown in Fig. 8 suggest that the vulnerable site of further plaque formation and progression is the post-stenotic neighborhood. This is in agreement with findings that the post-stenotic segment is of more pathological importance [26] and higher possibility of endothelial injury and plaque growth to even more luminal obstruction before plaque rupture [4, 31].

At the systolic peak for the stenotic arteries, WSS at the inner wall peaks at the neck of the stenosis reaches the minimum value in the post-stenosis section and then gradually recovers downstream until it levels off. The more severe stenosis is the further down the location of minimum WSS along the inner wall of the artery. The WSS at the inner wall is significantly influenced by the presence and severity of the stenosis (Fig. 9), especially in the post-stenosis site. Additionally, the magnitude of OSI in the post-stenosis region is noticeable especially for the higher degrees of severity as it increases by elevation of luminal obstruction. The high magnitude of OSI and low and even negative values of shear stress after luminal obstruction suggest the possible post-stenotic endothelial injury and progression of the stenosis downward. It has been shown that OSI values near 0.5 describe the situation in which WSS exceedingly fluctuates with higher possibility of endothelial damage [35]. In this study, severities of 50 to 80 % describe such situation.

As the stenosis becomes more severe, the angular phase difference at first harmonic frequency tends to become more negative at the post-stenotic site. Such trend at higher scales of stenosis leads to critical environment on endothelium, which suggests potential post-stenotic endothelial injury and axial plaque growth [28]. As SPA values become lower than −180 degree, increased LDL and possible endothelial injuries have been proposed [25, 35]. In this study, as described in Figs. 8, 9 and 10, the negative values of SPA are closely correlated with negative values of WSS.

5 Conclusion

By conducting experiments, we found that increased severity of luminal obstruction led to elevated inlet pressure. In lower scales of stenosis, the pressure rise was minor, but as the cross section of the lumen was reduced to a critical level, the pressure rose significantly. In this study we found that the critical severity was evaluated to be nearly 50 %. The negative values of shear stress, high fluctuations of shear stress, together with more negative angular phase difference in the first harmonic frequency, might be key parameters in predicting potential sites for endothelial injury and subsequently plaque growth. The negative value of WSS will cause flow separation in post-stenotic region and may result in further growth of plaque [32]. The angular phase difference between wall shear stress and circumferential stress waves became more negative at post-stenotic region as the degree of stenosis severity was elevated. The shear stress also became more negative with higher fluctuations in post-stenotic section as the scale of stenosis was raised. The trend of alteration in parameters was intensified for higher degrees of stenosis. Such parameters seem to be determinants of endothelial cell pathology and thrombus formation and predication of potential progression sites of the plaque. In this study, post-stenotic neighborhood was shown to be the location in which all three major determinants (negative WSS, high OSI and negative SPA) reach their critical values. This is in agreement with physiological findings, as it has been reported that the plaques tend to grow in directions distal to stenotic region [26, 37]. The results can be applied in the study of plaque maturity and growth, leading to clinical consequences of atherosclerosis.